K Number
K060174

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-02-17

(25 days)

Product Code
Regulation Number
880.5200
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indicated for use in vascular access to intravenously administer nutrients, fluids, chemotherapeutic agents, and other drugs for therapy, blood sampling, blood delivery, and venous pressure monitoring.

Device Description

The Spectrum Five Lumen Central Venous Catheter Set is a central venous catheter designed to intravenously administer nutrients, fluids, chemotherapeutic agents and other drugs for therapy, blood sampling, blood delivery, and venous pressure monitoring through the vasculature. The Spectrum Five Lumen Central Venous Catheter will be 10 Fr. available in lengths of 15, 20, 25, and 30 cm. The Spectrum Five Lumen Catheter will be impregnated with an average of 683.3 ug/cm minocycline and 564.1 ug/cm rifampin. This device will be supplied sterile and is intended for one-time use.

AI/ML Overview

Here's an analysis of the provided text regarding the Cook Incorporated Spectrum® Five Lumen Central Venous Catheter, formatted to answer your questions where information is available.

It's important to note that this document is a 510(k) summary for a medical device and, as such, focuses on demonstrating substantial equivalence to existing predicate devices rather than providing a detailed clinical study report typically found for a novel AI/software device. Therefore, many of your questions related to AI performance metrics, ground truth, and expert evaluation will not be applicable.


Acceptance Criteria and Reported Device Performance

1. A table of acceptance criteria and the reported device performance

TestAcceptance CriteriaReported Device Performance
Tensile Testing(Explicit criteria not stated in the summary, but implied to meet industry standards for catheter integrity under tension.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Shaft Flexibility Testing(Explicit criteria not stated in the summary, but implied to meet industry standards for catheter flexibility.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Shaft Flexural Fatigue Tolerance Testing(Explicit criteria not stated in the summary, but implied to meet industry standards for catheter durability under repeated flexing.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Gravity Flow Rate Testing(Explicit criteria not stated in the summary, but implied to meet clinically relevant flow rates for intravenous administration.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Air Leakage Testing(Explicit criteria not stated in the summary, but implied to demonstrate absence of air leaks.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Collapsibility under Vacuum Testing(Explicit criteria not stated in the summary, but implied to demonstrate resistance to collapse under vacuum conditions.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Liquid Leakage Testing(Explicit criteria not stated in the summary, but implied to demonstrate absence of liquid leaks.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Burst Pressure Testing(Explicit criteria not stated in the summary, but implied to demonstrate integrity under high-pressure conditions.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Biocompatibility Testing(Implied to meet ISO 10993 standards and other relevant biocompatibility requirements.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Tensile Testing after Aging(Implied to meet tensile strength requirements after simulated aging to ensure long-term integrity.)The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.
Amount of Minocycline and Rifampin Impregnated(Explicitly states an average concentration of 683.3 ug/cm minocycline and 564.1 ug/cm rifampin to indicate consistency of impregnation.)Average of 683.3 ug/cm minocycline and 564.1 ug/cm rifampin.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • The document does not specify sample sizes for these engineering and material tests, nor does it detail data provenance. These are typically bench and lab tests, not clinical studies with human subjects.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • This question is not applicable. The device validation involves engineering and material testing, not interpretation of data requiring expert consensus or ground truth in the way an AI/diagnostic device would.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This question is not applicable. Adjudication methods are relevant for diagnostic interpretations, not for physical material and performance tests of a catheter.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This question is not applicable. The device is a physical medical device (central venous catheter), not an AI/software diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This question is not applicable. This is not an AI/software device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • This question is not applicable in the context of diagnostic "ground truth." The "ground truth" for this device's performance would be the physical properties and functional specifications of the catheter itself, verified through standardized testing methods, not a clinical diagnosis.

8. The sample size for the training set

  • This question is not applicable. This is not an AI/machine learning device that uses training sets.

9. How the ground truth for the training set was established

  • This question is not applicable. This is not an AI/machine learning device.

Summary of the Study (Device Verification/Validation):

The "study" described in this 510(k) summary is a series of in-vitro, bench-top engineering, and material tests designed to assure the reliable design and performance of the Cook Incorporated Spectrum® Five Lumen Central Venous Catheter. These tests are standard for evaluating the physical, chemical, and mechanical properties of medical devices like catheters. The purpose is to demonstrate that the device meets predetermined specifications for safety and effectiveness, and is substantially equivalent to legally marketed predicate devices. The document explicitly states that "The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter."

{0}------------------------------------------------

Ko60174

33

510(k) Summary

Submitted By:

FEB 1 7 2006

Evangeline D. Loh, Ph.D., RAC Regulatory Affairs Scientist Cook Incorporated 750 Daniels Way (PO Box 489) Bloomington, IN 47404 (47402) 812-339-2235 January 23, 2006

Device:

Trade Name:Cook Incorporated Spectrum Five Lumen Central VenousCatheter Set
Proposed Classification:Intravascular CatheterFOZ (21 CFR§880.5200)

Indications for Use

Indicated for use in vascular access to intravenously administer nutrients, fluids, chemotherapeutic agents, and other drugs for therapy, blood sampling, blood delivery, and venous pressure monitoring.

Predicate Devices:

The Spectrum Five Lumen Central Venous Catheter Set is similar in terms of intended use, principles of operation, materials of construction and technological characteristics to predicate devices reviewed as devices for venous access to intravenously administer solutions needed in medical treatment.

Device Description:

The Spectrum Five Lumen Central Venous Catheter Set is a central venous catheter designed to intravenously administer nutrients, fluids, chemotherapeutic agents and other drugs for therapy, blood sampling, blood delivery, and venous pressure monitoring through the vasculature. The Spectrum Five Lumen Central Venous Catheter will be 10 Fr. available in lengths of 15, 20, 25, and 30 cm. The Spectrum Five Lumen Catheter will be impregnated with an average of 683.3 ug/cm minocycline and 564.1 ug/cm rifampin. This device will be supplied sterile and is intended for one-time use.

{1}------------------------------------------------

K060174 34

Substantial Equivalence:

The Spectrum Five Lumen Central Venous Catheter is similar to many devices in commercial distribution for venous access. These devices include the Cook Incorporated Five Lumen Central Venous Catheter Set (D.C. # K032274), Cook Incorporated Spectrum and Spectrum Glide Central Venous Catheter Set (D.C. #K033843), the Arrow International Arrowgard Blue Quad-Lumen Central Venous Catheter (D.C. #K962577), and the Medical Components Medcomp Quad Lumen (D.C.#K010021).

The similar indications for use, principles of operation, technological characteristics, and performance testing results of the Spectrum Five Lumen Central Venous Catheter as compared to the predicate devices support a determination of substantial equivalence.

Test Data:

The Spectrum Five Lumen Central Venous Catheter was subjected to the following tests to assure reliable design and performance under the specified testing parameters.

  • 1 . Tensile Testing
  • Shaft Flexibility Testing in
  • Shaft Flexural Fatigue Tolerance Testing ﻢ ﺍﻟﻤﺴﺘﻘﺒﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ
    1. Gravity Flow Rate Testing
  • న్, Air Leakage Testing
  • Collapsibility under Vacuum Testing ().
  • Liquid Leakage Testing 7.
  • Burst Pressure Testing oc
  • Biocompatibility Testing ().
  • Tensile Testing after Aging 】().
  • Amount of Minocycline and Rifampin Impregnated | | .

The results of these tests provide reasonable assurance that the device conforms to the requirements necessary for its use as a central venous catheter.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement. The logo is black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 7 2006

Evangeline D. Loh, Ph.D., RAC Regulatory Affairs Scientist Cook, Incorporated 750 Daniels Way Bloomington, Indian 47404

Re: K060174

K000174
Trade/Device Name: Cook Incorporated Spectrum® Five Lumen Central Venous Catheter Regulation Number: 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: January 23, 2006 Received: January 24, 2006

Dear Dr. Loh:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rowed your sected by the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvial appressions of the Act. The general controls provisions of the Act include controls provisions of annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 WA), it har ob subject in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device to rear roublish further announcements concerning your device in the Ecderal Register.

{3}------------------------------------------------

Page 2 - Dr. Loh

Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA s Issualice of a substantials with other requirements
mean that FDA has made a determination that your device Fodard assessios mean that FDA has made a decemination until Josefinistered by other Federal agencies.
of the Act or any Federal statutes and regulations administered by posignity of the Act of any Federal statutes and registerents, including, but not limited to: registration You must comply with an the Act 3 requirement 801); good manufacturing practice
and listing (21 CFR Part 807); labeling (21 CFR Part 801); gEP Rat 800); and i and listing (21 CFR Fall 807), labeling (21 CFR Pat 820), and if and 820); and if requirements as set form in the quality systems (Sections 531-542 of the Act);
applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 5 10(k) I his letter will anow you to begin marketing your alence of your device to a premarket notification. The PDA miding of babbaan of a
legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), a If you desire specific advice for your device of your device Also, please note the regulation please contact the Other or Comphanes and in (21CFR Part 807.97). You entitled, "Misoranding by reserved to problems on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free Division of Binas 2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely wours,

Susan Yunae

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

K060174

Device Name:

Cook Incorporated Spectrum™ Five Lumen Central Venous Catheter Set

The Spectrum™ Five Lumen Central Venous Catheter Set is Indications for Use: indicated for the intravenous administration of nutrient fluids, chemotherapeutic agents, and other drugs for therapy, blood sampling, blood delivery, and venous pressure monitoring.

Prescription Use X (Per 21 CFR 801 Subpart D) ાર

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anton Viny

They, General Hospital,

K060174

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).