K Number
K053640
Device Name
INTEGRA
Date Cleared
2006-03-01

(61 days)

Product Code
Regulation Number
876.5990
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INTEGRA is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and upper ureter.

Device Description

INTEGRA is a transportable Electromagnetic (EM) Extracorporeal Shock Wave Lithotripter (ESWL) used for urinary stones treatment. The device is the result of integrating the cleared fluoroscopy imaging device - 3Dscope and lithotripter. INTEGRA also contains the necessary interfaces for ultrasonic imaging devices.

INTEGRA includes the following features:

  • EM Shockwave generator with parabolic reflector for focusing mechanism A
  • High Level of integration: the shock wave generator is fully integrated with A the fluoroscopic system and stationary patient table.
  • In lieu of the patient table movement the reflector is moved in three A directions
  • In-line X-Ray localization system which is executed with lateral movement

  • Off-line movable Ultrasound transducer

AI/ML Overview

The provided text describes the Integra Extracorporeal Shock Wave Lithotripter (ESWL) and its FDA 510(k) submission. However, it does not contain specific acceptance criteria or reported device performance data in a quantitative format related to clinical outcomes or device efficacy for stone fragmentation.

The document primarily focuses on:

  • Device Description and Intended Use: Fragmenting urinary stones in the kidney and upper ureter.
  • Predicate Devices: Storz Modulith SL-20/(SLX), Storz Modulith Lithotripter Model SLK, Direx Tripter X - 1 Compact Duet, Direx 3Dscope.
  • Performance Testing (Engineering Standards): Compliance with various IEC, FDA CDRH, ANSI/AAMI/ISO, and CISPR standards related to electrical safety, EMC, and radiation control.These are technical safety and performance standards for the device itself, not clinical efficacy metrics.
  • Clinical Tests: It states that a "Confirmatory study was conducted according to 'Guidance for the Content of Premarket Notifications (510(k)s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi; August 9, 2000, Chapter 8 subchapter D: (Clinical Performance Testing, confirmatory clinical study)'." However, the results or specific acceptance criteria from this clinical study are not provided in this document excerpt.
  • Substantial Equivalence Argument: Comparing Integra's features (shockwave generation, focusing, X-ray and Ultrasound localization) to its predicate devices.

Therefore, I cannot fulfill the request for a table of acceptance criteria and reported device performance from the provided text. The text indicates that a clinical study was performed, but the details of that study, including the acceptance criteria and its results, are missing.

Here's a breakdown of what can be extracted or inferred, and what information is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Not specified in the provided text for clinical efficacy.Not specified in the provided text for clinical efficacy. The document states that a "Confirmatory study was conducted" following FDA guidance, but the outcomes or specific metrics (e.g., stone-free rate, fragmentation success rate) are not detailed.
Compliance with various safety and performance standards (e.g., IEC 60601-1, FDA CDRH 21CFR 1020.30, etc.)The device was tested according to these standards, implying compliance, but specific results (e.g., measured EMI levels, radiation leakage) are not listed.

2. Sample size used for the test set and the data provenance

  • Sample Size (Test Set): Not specified in the provided text.
  • Data Provenance: The document states a "Confirmatory study was conducted," implying clinical data, but the country of origin, retrospective/prospective nature, and specific details are not available in this excerpt.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Adjudication Method: Not specified.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: Not mentioned. This device is an ESWL lithotripter, a therapeutic device, not an AI-powered diagnostic imaging tool. Therefore, an MRMC study related to human readers improving with AI assistance would not be applicable or expected for this type of device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable. This is a hardware therapeutic device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: For the "Confirmatory study," it would likely involve outcomes data directly related to stone fragmentation and removal (e.g., post-procedure imaging for stone-free status or residual stone size). However, this is inferred, not explicitly stated, and specific details are not available.

8. The sample size for the training set

  • Sample Size (Training Set): Not applicable, as this is not an AI/ML algorithm requiring a training set.

9. How the ground truth for the training set was established

  • Ground Truth (Training Set): Not applicable, as this is not an AI/ML algorithm requiring a training set.

Summary of Missing Information: The provided 510(k) summary explicitly refers to a "Confirmatory clinical study" but does not include the details of its methodology (sample size, ground truth establishment, expert involvement, etc.) or its results (acceptance criteria, device performance). The document primarily addresses engineering performance standards and argues for substantial equivalence based on device features rather than clinical efficacy data within this excerpt.

{0}------------------------------------------------

SECTION 5 - 510(k) SUMMARY ___________________________________________________________________________________________________________________________________________________

K053640

[Submitted pursuant to 21 CFR 807.92(a). All data included in this document are accurate and complete to the best of DSC's knowledge.]

1. Submitter Information

Submitter:Direx Systems Corporation437 Turnpike StreetCanton, MA 02021
Telephone:Fax:Contact Person(339) 502-6013(339) 502-6018Larisa GershteinQA Manager
Contact Person e-mail address:lgershtein@direxusa.com
2. Device
Trade/Proprietary Name:Integra
Common/Usual Name:Extracorporeal Shock Wave Lithotripter (ESWL)
Regulation Number:21 CFR 876.5990
Requlatory Class:Class II (special controls)
Product code:78 LNS
Panel:Gastroenterology and Urology

3. Predicate Devices

Storz Modulith® SL-20/ (SLX) (P920051) Storz Modulith® Lithotripter Model SLK (K010340) Direx Tripter X - 1 Compact Duet (K041582) Direx 3Dscope (K041213)

{1}------------------------------------------------

4. Intended Use:

The INTEGRA is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and upper ureter.

5. Description

INTEGRA is a transportable Electromagnetic (EM) Extracorporeal Shock Wave Lithotripter (ESWL) used for urinary stones treatment. The device is the result of integrating the cleared fluoroscopy imaging device - 3Dscope and lithotripter. INTEGRA also contains the necessary interfaces for ultrasonic imaging devices.

INTEGRA includes the following features:

  • EM Shockwave generator with parabolic reflector for focusing mechanism A
  • High Level of integration: the shock wave generator is fully integrated with A the fluoroscopic system and stationary patient table.
  • In lieu of the patient table movement the reflector is moved in three A directions
  • In-line X-Ray localization system which is executed with lateral movement

  • Off-line movable Ultrasound transducer

6. Performance Testing

The Integra Lithotripter was tested according to the following standards:

  • IEC 60601-1 (1988) +A1 (1991) +A2 (1995) .
  • IEC 60601-1-1 (2000) .
  • IEC 60601-1-2 (2001) .
  • IEC 60601-1-3 (1994) .
  • IEC 60601-2-7 (1998) .
  • . FDA CDRH 21CFR 1020.30
  • FDA CDRH 21CFR 1020.32 .
  • IEC 606001-2-36 (1997) .
  • IEC 61846 (1998) .
  • IEC 60601-2-38 (1996) .

{2}------------------------------------------------

  • ANSI/ AAMI/ ISO 10993– 1 (1998) .
  • CISPR 11 (1997) + A1 (1999) class B �
  • IEC 60601-1-4 (2000) .
  • Guidance for the Content of Premarket Notifications (510(k)s) for . Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi; August 9, 2000

7. Clinical Tests

The Confirmatory study was conducted according to "Guidance for the Content of Premarket Notifications (510(k)s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi; August 9, 2000, Chapter 8 subchapter D: (Clinical Performance Testing, confirmatory clinical study ).

8. Substantial Equivalence

INTEGRA is substantially equivalent to the predicate devices, since the basic features, design and intended uses are the same or similar.

Specifically, The devices have the same intended uses (the Storz devices and Compact Duet).

In terms of generation and focusing of pressure waves Integra is similar to the Storz predicates. (both use parabolic reflectors that converge pressure waves to a focal point). Both Integra and Storz SL-20/ (SLX) have also the same shockwave orientation (the angle of reflector is 90°).

Similarly to the Storz devices, Integra contains an in-line X-ray localization system.

The Ultrasound localization is similar to that of Compact Duet in the sense that the ultrasound probe of both devices is located outside the reflector.

Most of the X-ray parameters are the same as the ones of 3Dscope.

The minor differences in design, dimensions and features between Integra and its predicates raise no new issues of safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the edge of the circle. Inside the circle is a stylized image of an eagle with its wings spread. The eagle is facing to the left. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 2006

Ms. Larisa Gershtein OA Manager DiREX Systems Corporation 437 Turnpike Street CANTON MA 02021

Re: K053640

Trade/Device Name: Integra Regulation Number: 21 CFR 8876.5990 Regulation Name: Extracorporeal shock wave lithotripter Regulatory Class: II Product Code: LNS Dated: December 27, 2005 Received: December 30, 2005

Dear Ms. Gershtein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I ms lotter with and on. The FDA finding of substantial equivalence of your device to a legally prematics notification: "Thesults in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you doors affect of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Anot production other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for DiREX Systems Corp. The logo is in black and white, with the word "DiREX" in large, bold letters. Below the word "DiREX" is a horizontal line, and below that is the text "SYSTEMS CORP." in smaller letters. The logo is simple and professional-looking.

Section 4: Indications for Use Statement

Indications for Use STATEMENT

510(k) Number (if known): K05 3690

Device Name:

Integra

Indications for Use:

The INTEGRA is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and upper ureter

Prescription Use X
(Per 21 CFR § 801.109)

OR

Over the Counter Use_

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Daniel A. Segarra

(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number

437 Turnpike Street ♦ Canton. MA 02021 Phone (339) 502-6013 + Fax (339) 502-6018 e - mail promo (a), dire x u sa. com

§ 876.5990 Extracorporeal shock wave lithotripter.

(a)
Identification. An extracorporeal shock wave lithotripter is a device that focuses ultrasonic shock waves into the body to noninvasively fragment urinary calculi within the kidney or ureter. The primary components of the device are a shock wave generator, high voltage generator, control console, imaging/localization system, and patient table. Prior to treatment, the urinary stone is targeted using either an integral or stand-alone localization/imaging system. Shock waves are typically generated using electrostatic spark discharge (spark gap), electromagnetically repelled membranes, or piezoelectric crystal arrays, and focused onto the stone with either a specially designed reflector, dish, or acoustic lens. The shock waves are created under water within the shock wave generator, and are transferred to the patient's body using an appropriate acoustic interface. After the stone has been fragmented by the focused shock waves, the fragments pass out of the body with the patient's urine.(b)
Classification. Class II (special controls) (FDA guidance document: “Guidance for the Content of Premarket Notifications (510(k)'s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi.”)