K Number
K053255

Validate with FDA (Live)

Date Cleared
2006-03-14

(113 days)

Product Code
Regulation Number
878.4300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nexus Ligating Clip is indicated for use in surgical procedures on vessels or other tubular structures where a metal ligating clip is required.

Device Description

The Nexus Ligating Clip is a titanium clip used during surgical procedures. The clips are supplied sterile in four sizes: small, medium-large, and large; six clips per cartridge. The size of clip should be chosen based on experience, judgment, and needs, ensuring that the tissue to be occluded fits completely within the clip. The devices are heart-shaped to provide a firm grip on the structure to be ligated, and there are transverse grooves on the inner surface of each clip to resist slippage.

AI/ML Overview

This 510(k) summary (K053255) describes the Nexus Ligating Clip, a titanium clip used in surgical procedures. The submission focuses on demonstrating substantial equivalence to a predicate device, the Horizon Ligation Clips (K982313) from Weck Closure Systems, rather than establishing de novo acceptance criteria through clinical studies.

As such, the document does not contain information on dedicated clinical studies to proactively establish and meet acceptance criteria for a novel device. Instead, it relies on demonstrating similar characteristics and performance to an already legally marketed device. Therefore, many of the requested sections (sample size, expert qualifications, ground truth, MRMC study, training set details) are not applicable in the context of this 510(k) submission.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Sterilization Cycle Validation In accordance with accepted international standards."The sterilization cycle was validated in accordance with accepted international standards."
Shelf Life Support a labeled shelf life period of five years."Shelf life testing was performed to support the labeled shelf life period of five years. The results of all testing were within acceptable limits."
Technological Characteristics"The indications for use, principles of operation, and device design of the Nexus Ligating Clip are virtually identical to those of the predicate device."
Performance Specifications"There are no significant differences in either technology nor performance specifications [compared to the predicate device]."

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided because the submission relies on demonstrating substantial equivalence through comparison of technical characteristics and existing validated processes (sterilization, shelf life) rather than a clinical test set. No specific patient data or test set data provenance is mentioned.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable. No "ground truth" for a clinical test set was established in this 510(k) submission.

4. Adjudication Method for the Test Set

This information is not applicable. No clinical test set requiring adjudication was conducted or described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No MRMC comparative effectiveness study was mentioned or performed as part of this 510(k) submission. The focus is on device characteristics, not human reader performance with or without AI.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

This is not applicable, as the device is a titanium ligating clip, not an AI algorithm.

7. The Type of Ground Truth Used

The ground truth concept, as typically applied in AI/clinical studies (e.g., pathology, outcomes data), is not relevant here. The "ground truth" for this submission revolves around:

  • Adherence to international sterilization standards.
  • Successful completion of shelf-life testing.
  • Demonstrated similarity in design, materials (titanium), indications for use, and principle of operation to the predicate device.

8. The Sample Size for the Training Set

This information is not applicable, as this device does not involve a "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable.

{0}------------------------------------------------

510(k) SUMMARY

K053255

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

Submitter's Name: MicroSurgical Laboratories, dba Wexler Surgical Supplies 11333 Chimney Rock Road, Suite 110 Houston, TX 77035 Telephone: (713) 723-6900 Fax: (713) 723-6906 Contact person: Mr. Danny Fishman, COO

Date of Summary: February 17, 2006

Device Names:

Trade Name: Nexus Ligating Clips Common Name: Titanium Hemostatic Clip Classification Name: Implantable Clip (21 CFR 878.4300, Product Code FZP)

Legally Marketed Device to which Equivalence is Claimed: The legally marketed predicate devices are the Horizon Ligation Clips (K982313) manufactured by Weck Closure Systems, determined to be substantially equivalent to a legally marketed (preAmentment) device on August 10, 1998.

Device Description: The Nexus Ligating Clip is a titanium clip used during surgical procedures. The clips are supplied sterile in four sizes: small, medium-large, and large; six clips per cartridge. The size of clip should be chosen based on experience, judgment, and needs, ensuring that the tissue to be occluded fits completely within the clip. The devices are heart-shaped to provide a firm grip on the structure to be ligated, and there are transverse grooves on the inner surface of each clip to resist slippage.

Intended Use: The Nexus Ligating Clip is indicated for use in surgical procedures on vessels or other tubular structures where a metal ligating clip is required.

Descriptive Summary of Technological Characteristics and Those of Predicate Device: The indications for use, principles of operation, and device design of the Nexus Ligating Clip are virtually identical to those of the predicate device. The Nexus devices are essentially a subset of the Horizon System, in that the Nexus Clip is offered in fewer sizes, only one cartridge capacity, and are made with titanium only, rather than titanium or tantalum. There are no significant differences in either technology nor performance specifications.

Performance Data: The sterilization cycle was validated in accordance with accepted international standards. Shelf life testing was performed to support the laboled shelf life period of five years. The results of all testing were within acceptable limits.

Conclusion: The information and data provided in this 510(k) Notification establish that the Nexus Ligating Clip is substantially equivalent to the legally marketed predicated device.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings and two lines representing its legs.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MicroSurgical Laboratories, dba Wexler Surgical

MAR 1 4 2006

Supplies c/o Ms. Lisa S. Jones Regulatory Affairs Consultant Devices for the Future 540 College Street Bellaire, Texas 77401

Re: K053255

Trade/Device Name: Nexus™ Ligating Clip Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: II Product Code: FZP Dated: January 23, 2006 Received: January 24, 2006

Dear Ms. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and lisuing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Jones

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/indcx.html.

Sincerely yours,
elmee

Mark N. Melkerson, M.S. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K053255

Image /page/3/Picture/1 description: The image shows a logo for "YECLER SURGICAL SUPPLIES". The logo features a stylized abstract design on the left, followed by the company name in a bold, sans-serif font. The words "SURGICAL SUPPLIES" are printed in a smaller font size below the main name.

Indications for Use

510(k) Number: K053255

Device Name: Nexus™ Ligating Clip

Indications for Use: The Nexus Ligating Clip is indicated for use in surgical procedures on vessels or other tubular structures where a metal ligating clip is required.

Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)

OR Over-The-Counter
(21 CFR 80

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CD ffice of Device Evaluation (ODE)

(Division Sign-Off

Division of General, Restorative, and Neurological Devices

Page 1 of 1

510(k) Number_K053255

11333 Chimacy Rock, Suite 110 - Houston, Texas 77935 - Phone: 713.723.6900 - Toll Free: 800,414.1476 F-mail: sales@wexlersurgical.com - 17ax: 713.723.6906 - www.weslersurgical.com

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.