K Number
K053100

Validate with FDA (Live)

Date Cleared
2005-12-19

(46 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ESTECH Cobra Bipolar System is intended for the coagulation of soft tissue during general surgery. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

Device Description

The Estech Cobra Bipolar System is comprised of two components: a reusable surgical clamp an associated coagulating device known as the Cobra Bipolar Insert. The Bipolar Insert is a sterile, single use device that transmits radiofrequency (RF) energy to coagulate soft tissues and fits in a jaw of the clamp. The Bipolar System will be used with the Boston Scientific Cobra Electrosurgical Unit (ESU). The ESU delivers 460 kHz of RF energy to selected Bipolar Insert electrodes, modulates the RF energy to keep all selected Bipolar Insert electrodes' temperatures essentially the same, and adjusts the power output to maintain the maximum temperature of all selected electrodes close to the set point.

AI/ML Overview

This document is a 510(k) Premarket Notification for the ESTECH Cobra Bipolar System, which is essentially a re-filing of an existing device (Boston Scientific Cobra Bipolar System, K023288) due to a transfer of sales and responsibility. Therefore, the information provided focuses on demonstrating substantial equivalence to the predicate device rather than presenting new performance data from a dedicated study.

Based on the provided text, the acceptance criteria and study information are as follows:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Intended UseIdentical to predicate device (Boston Scientific Cobra Bipolar System)
TechnologyIdentical to predicate device
DesignIdentical to predicate device
MaterialsIdentical to predicate device
ManufactureIdentical to predicate device
PackagingIdentical to predicate device
SterilizationIdentical to predicate device
Safety & EffectivenessSubstantially equivalent to predicate device

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states that "Supporting data is not necessary to support this submission since the Estech Cobra Bipolar System is identical to the predicate device, the Boston Scientific Cobra Bipolar System (K023288)."

Therefore, there was no dedicated test set or new performance data generated for this specific 510(k) submission. The data provenance relies on the prior approval of the predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Since no new test set was used, no experts were specifically used to establish ground truth for this submission. The assumption is that the predicate device's original approval relied on appropriate expert evaluations as part of its regulatory process.

4. Adjudication Method for the Test Set

As there was no new test set, there was no adjudication method employed for this submission.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No MRMC comparative effectiveness study was done for this submission. The device is claimed to be identical to the predicate, implying similar performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This device is an electrosurgical probe, not a software algorithm. Therefore, a standalone algorithm performance study is not applicable and was not performed.

7. The Type of Ground Truth Used

The ground truth for this submission is based on the established performance and safety of the predicate device (Boston Scientific Cobra Bipolar System, K023288), which was previously cleared by the FDA. The submission asserts that the new device is identical in all relevant aspects, hence its functionality and safety are equivalent to the pre-existing device.

8. The Sample Size for the Training Set

Since this submission relies on demonstrating substantial equivalence to an identical predicate device and no new performance studies were conducted, there was no new training set used or described for this particular submission.

9. How the Ground Truth for the Training Set Was Established

As no training set was used for this submission, there is no description of how its ground truth was established. The regulatory burden was met by declaring the new device identical to the predicate.

{0}------------------------------------------------

· ESTECH Cobra Bipolar System Premarket Notification

Section 10: Executive Summary
K053100
Device Information:
CategoryComments
Sponsor:Estech4135 Blackhawk Plaza Circle.Suite 150Danville, CA 94506Tel: 925-648-3500
Correspondent ContactInformation:Craig CoombsCoombs Medical Device Consulting1193 Sherman StreetAlameda, CA 94501Tel: 510-337-0140Fax: 510-337-0416
Device Common Name:Electrosurgical Probe
Device Classification & Code:Class II, GEI (21 CFR 870.4420)
Device Classification Name:Electrosurgical cutting and coagulation device andaccessories
Device Proprietary Name:Estech Cobra Bipolar System

Predicate Device Information:

Predicate Devices:Cobra Bipolar System (K023288)
Predicate Device Manufacturers:Boston Scientific
Predicate Device Common Name:Electrosurgical Probe
Predicate Device Classification:21 CFR 878.4400
Predicate Device Classification & Code:Class II, GEI

b. Date Summary Prepared

October 31, 2005

c. Description of Device

The Estech Cobra Bipolar System is identical to the Boston Scientific Cobra Bipolar System.

Both Bipolar Systems are comprised of two components: a reusable surgical clamp an associated coagulating device known as the Cobra Bipolar Insert. The Bipolar Insert is a sterile, single use device that transmits radiofrequency (RF) energy to coagulate soft tissues and fits in a jaw of the clamp. The Bipolar System will be used with the Boston Scientific Cobra Electrosurgical Unit (ESU). The ESU delivers 460 kHz of RF energy to selected Bipolar Insert electrodes, modulates the RF energy to keep all selected Bipolar Insert electrodes' temperatures essentially the same, and adjusts the power output to maintain the maximum temperature of all selected electrodes close to the set point.

{1}------------------------------------------------

C 503100

Endoscopic Technologies, Inc.

Image /page/1/Picture/2 description: The image shows the text "Page 2 of 2 ESTECH Cobra Bipolar System Premarket Notification". The word "Page" is written in cursive. The number "2" appears three times in the image, twice written in cursive and once circled.

d. Intended Use

The ESTECH Cohra Bipolar System is intended for the coagulation of soft tissue during general surgery. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

e. Comparison to Predicate Device

The Estech Cobra Bipolar System is identical in intended use, technology, design, materials, manufacture, packaging and sterilization to the predicate Boston Scientific Cobra Bipolar System.

This 510(k) is being filed to transfer sales and responsibility of this portion of Boston Scientifics' product line over to Estech. For the time being Boston Scientific will continue to manufacture the device. Estech will add additional contract manufacturers as needed.

Estech concludes that the Estech Cobra Bipolar System is substantially equivalent to the Boston Scientific Cobra Bipolar System.

f. Summary of Supporting Data

Supporting data is not necessary to support this submission since the Estech Cobra Bipolar System is identical to the predicate device, the Boston Scientific Cobra Bipolar System (K023288).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized image of an eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 9 2005

Endoscopic Technologies, Inc. % Mr. Craig Coombs President Coombs Medical Device Consulting 1193 Sherman Street Alameda, California 94501

Re: K053100

Trade/Device Name: Estech Cobra Bipolar System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and Coagulation device and accessories. Regulatory Class: II Product Code: GEI Dated: November 2, 2005 Received: November 8, 2005

Dear Mr. Coombs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Chateaumichin)

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Section 4: Indications for Use Statement

KO53100 510(k) Number (if known):

Device Name: ESTECH Cobra Bipolar System

Indications For Use:

The ESTECH Cobra Bipolar System is intended for the coagulation of soft tissue during general surgery. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

Prescription UseX
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER
PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)to MXM
Division of General, Restorative,
and Neurological Devices
510(k) NumberK053100
Page 1 of

CONFIDENTIAL

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.