K Number
K052716
Date Cleared
2006-11-16

(413 days)

Product Code
Regulation Number
866.3110
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RAPIRUN H. pylori Antibody Detection Kit is a rapid immunochromatographic assay intended for the qualitative detection of antibodies against Helicobacter pylori (H. pylori) in urine to aid in the diagnosis of H. pylori infection. The RAPIRUN Kit is suitable for use in both point-of-care and clinical laboratory settings.

Device Description

The RAPIRUN H. pylori Antibody Detection Kit is an immunochromatographic method for the qualitative detection of anti-H. pylori antibodies in urine using a nitrocellulose membrane immobilized with proteins extracted from H. pylori as an antigen. The test device includes a sample window in which the test specimen is placed in a dropwise manner and an evaluation window in which a test line and a control line are visible. Colloidal gold-conjugated anti-human IgG (Fc) polyclonal antibody (goat) is present between the sample window and the evaluation window. The test line and the control line in the evaluation window are immobilized with H. pylori-extracted protein and with anti-human IgG polyclonal antibodies (goat), respectively.

When a urine specimen diluted with the sample diluent is added dropwise to the sample window, the IgG in the diluted sample reacts with the conjugated antibodies. Immunocomplexes are formed and flow into the nitrocellulose membrane. If anti-H. pylori IgG antibody complexes are present in these immunocomplexes, they react with the H. pylori antigen immobilizing on the test line and are captured producing a red band. On the other hand, immunocomplexes without anti-H. pylori IgG antibodies pass the test line and are captured by the anti-human IgG polyclonal antibodies (goat) immobilizing on the control line producing a red band. A test specimen is judged to be positive for anti-H. pylori antibodies if red colored bands appear in both the test zone and the control zone, or is judged to be negative for anti-H. pylori antibodies if a red colored band only appears in the control zone.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the RAPIRUN H. pylori Antibody Detection Kit, based on the provided text:

Acceptance Criteria and Device Performance

  • Note: The document does not explicitly state pre-defined "acceptance criteria" in terms of specific performance thresholds that were required for clearance. Instead, it presents the results of the clinical study, which are then used to demonstrate substantial equivalence to predicate devices. The implicit acceptance criteria would be for the device's performance to be comparable to, or "substantially equivalent" to, the predicate devices.
MetricPredicate Device 1 (HM-CAP™ EIA) ComparisonPredicate Device 2 (UBiT-IR300) Comparison
% Overall Agreement87.23% (95% CI: 81.97, 91.39)93.09% (95% CI: 88.51, 96.27)
% Positive Agreement84.71% (95% CI: 75.82, 91.30)86.46% (95% CI: 78.53, 92.40)
% Negative Agreement89.32% (95% CI: 81.96, 94.31)100.00% (95% CI: 96.17, 100.00)

Study Information

1. Sample Size Used for the Test Set and Data Provenance:
* Sample Size: 188 evaluable subjects.
* Data Provenance: Prospective, multi-center study conducted at four Physician Office Laboratory (POL)/Point-of-Care (POC) settings. The country of origin is not explicitly stated, but the sponsor is Japanese and the contact information includes a US contact, suggesting the study was likely conducted in the US (given the FDA submission context).

2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
* The study did not use human experts to establish "ground truth" for the test set in the traditional sense of diagnostic interpretation. Instead, it used predicate medical devices (HM-CAP™ H. pylori Enzyme Immunoassay Test and UBiT-IR300 Infrared Spectrophotometer) as the reference standards for comparison. Therefore, no information is provided about the number or qualifications of experts for ground truth establishment.

3. Adjudication Method for the Test Set:
* Not applicable. The "ground truth" was established by the results of predicate devices, not through expert consensus requiring adjudication.

4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
* No, an MRMC comparative effectiveness study was not done. This was a direct comparison of the new device to existing predicate devices, not a study evaluating human reader performance with and without AI assistance.

5. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
* Yes, this was a standalone performance study. The RAPIRUN H. pylori Antibody Detection Kit is a rapid immunochromatographic assay with visual interpretation of test and control lines. Its performance was evaluated purely based on its own output compared to the predicate devices, without human-in-the-loop assistance in the sense of an algorithm supporting a human reader.

6. The Type of Ground Truth Used:
* The "ground truth" for evaluating the RAPIRUN Kit was established by predicate medical devices:
* HM-CAP™ H. pylori Enzyme Immunoassay Test (serum)
* UBiT-IR300 Infrared Spectrophotometer method (breath)
* These predicate devices themselves are considered established methods for diagnosing H. pylori infection.

7. Sample Size for the Training Set:
* The document does not mention a "training set" or "training data" for the device itself. Immunochromatographic assays like the RAPIRUN Kit are typically developed and optimized through laboratory studies and then directly validated in clinical performance studies. They are not machine learning algorithms that require a distinct training phase with annotated data in the same way.

8. How the Ground Truth for the Training Set Was Established:
* Not applicable, as there is no mention of a distinct "training set" or "ground truth for the training set" in the context of this device's development as described.

{0}------------------------------------------------

K052716

NOV 16 2006

510(k) Summary for RAPIRUN H. pylori Antibody Detection Kit

1. SPONSOR

Otsuka Pharmaceutical Co., Ltd. 2-9 Kanda Tsukasa-cho, Chiyoda-ku Tokyo 101-8535 Japan

Contact Person:

Japan:Kuniaki Natsume
Telephone:81-3-3257-0539

U.S.: Kyoko Tsuchiya Telephone: 720-479-6449

Date Prepared: September 28, 2005

2. DEVICE NAME

Proprietary Name:RAPIRUN H. pylori Antibody Detection Kit
Common/Usual Name:Qualitative Immunoassay for H. pylori
Classification Name:Campylobacter pylori

3. PREDICATE DEVICES

  • HM-CAP™ H. pylori Enzyme Immunoassay Test . Enteric Products, Inc. K984544, K955085, K944159
  • . UBiT-IR300 Infrared Spectrophotometer Otsuka Pharmaceutical Co., Ltd. K013371
  • . Instant-View H. pylori Rapid Test Alfa Scientific Designs, Inc. K024360

{1}------------------------------------------------

DEVICE DESCRIPTION 4.

The RAPIRUN H. pylori Antibody Detection Kit is an immunochromatographic method for the qualitative detection of anti-H. pylori antibodies in urine using a nitrocellulose membrane immobilized with proteins extracted from H. pylori as an antigen. The test device includes a sample window in which the test specimen is placed in a dropwise manner and an evaluation window in which a test line and a control line are visible. Colloidal gold-conjugated anti-human IgG (Fc) polyclonal antibody (goat) is present between the sample window and the evaluation window. The test line and the control line in the evaluation window are immobilized with H. pylori-extracted protein and with anti-human IgG polyclonal antibodies (goat), respectively.

When a urine specimen diluted with the sample diluent is added dropwise to the sample window, the IgG in the diluted sample reacts with the conjugated antibodies. Immunocomplexes are formed and flow into the nitrocellulose membrane. If anti-H. pylori IgG antibody complexes are present in these immunocomplexes, they react with the H. pylori antigen immobilizing on the test line and are captured producing a red band. On the other hand, immunocomplexes without anti-H. pylori IgG antibodies pass the test line and are captured by the anti-human IgG polyclonal antibodies (goat) immobilizing on the control line producing a red band. A test specimen is judged to be positive for anti-H. pylori antibodies if red colored bands appear in both the test zone and the control zone, or is judged to be negative for anti-H. pylori antibodies if a red colored band only appears in the control zone.

5. INTENDED USE

The RAPIRUN H. pylori Antibody Detection Kit is a rapid immunochromatographic assay intended for the qualitative detection of antibodies against Helicobacter pylori (H. pylori) in urine to aid in the diagnosis of H. pylori infection. The RAPIRUN Kit is suitable for use in both point-of-care and clinical laboratory settings.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.

The RAPIRUN Kit and the predicate devices cited above are all qualitative tests intended for use in the diagnosis of H. pylori infection. Like the RAPIRUN Kit, both the HM-CAP™ Test and the Instant-View Test detect IgG antibodies to H. pylori in the test sample. The UBiT-IR300 method measures changes in 1302 content in breath CO2 gas. The test samples vary among the devices. The RAPIRUN Kit is

{2}------------------------------------------------

used with urine samples, while the HM-CAP™ Test is used with serum or plasma, the UBiT-IR300 with breath, and the Instant-View Test with whole blood or serum.

Both the RAPIRUN Kit and the Instant-View Test are laminar flow chromatographic immunoassays with detection by visual inspection. The HM-CAP™ Test is an enzyme immunoassay with absorbance measured at 450 nm and the UBiT-IR300 is an infrared spectrophotometric method with detection by infrared absorption. The RAPIRUN Kit, UBiT-IR300, and Instant-View Test are primarily intended for use in point-of-care settings while the HM-CAPTM Test is for use in clinical laboratory settings.

The differences between the RAPIRUN Kit and the predicate devices do not impact the safety or effectiveness of the proposed RAPIRUN Kit for its intended use.

7. PERFORMANCE TESTING

7.1 Nonclinical Testing

A series of nonclinical studies was conducted to assess the performance of the RAPIRUN H. pylori Antibody Detection Kit. These studies evaluated reproducibility, interfering substances, cross-reactivity, high-dose hook effect, sample treatment and handling, and kit stability. The results of all studies demonstrated that the RAPIRUN Kit performed according to its specifications.

7.2 Clinical Testing

A clinical study was conducted to evaluate the performance of the RAPIRUN H. pylori Antibody Detection Kit for the qualitative detection of antibodies to H. pylori in urine. The multi-center, prospective study was designed to compare the RAPIRUN H. pylori Antibody Detection Kit for the qualitative detection of antibodies to H. pylori in urine to the HM-CAP™ H. pylori Enzyme Immunoassay Test for the detection of antibodies to H. pylori in serum and to the UBiT-IR300 Infrared Spectrophotometer method for measuring 13CO2 enrichment in breath. Subjects were recruited at four Physician Office Laboratory (POL)/Point-of-Care (POC) settings. The total number of evaluable subjects was 188 across all participating sites.

{3}------------------------------------------------

The primary endpoint was the percent agreement of the RAPIRUN Kit results as compared to the HM-CAP™ EIA Test results. Both of these methods are tests for the qualitative detection of antibodies to H. pylori. The percent agreement for all subjects is as follows:

% Overall Agreement:87.23%[95% CI: (81.97, 91.39)]
% Positive Agreement:84.71%[95% CI: (75.82, 91.30)]
% Negative Agreement:89.32%[95% CI:(81.96, 94.31)]

A secondary endpoint was the percent agrecment of the RAPIRUN H. pylori Antibody Detection Kit as compared to the UBiT-IR300 Infrared Spectrophotometer method for measurement of '300_enrichment in breath. The percent agreement for all subjects is as follows:

% Overall Agreement:93.09% [95% CI: (88.51, 96.27)]
% Positive Agreement:86.46% [95% CI: (78.53, 92.40)]
% Negative Agreement:100.00% [95% CI:(96.17, 100.00)]

For reference purposes, the results from the HM-CAP™ EIA Test also were compared to the results from the UBiT-IR300 Breath Test. The percent agreement for all subjects is as follows:

% Overall Agreement:88.83% [95% CI: (83.69, 92.65)]
% Positive Agreement:83.33% [95% CI: (75.02, 90.16)]
% Negative Agreement:94.57% [95% CI:(87.97, 97.83)]

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Otsuka Pharmaceutical Co., Ltd. Cynthia A. Sinclair, RAC Principal Consultant, Regulatory Services Medical Device Consultants, Inc. 49 Plain Street North Attleboro, MA 02760

NOV 16 2006

K052716 Re:

Trade/Device Name: RAPIRUN H. pylori Antibody Detection Kit Regulation Number: 21 CFR 866.3110 Regulation Name: Campylobacter Fetus Serological Reagents Regulatory Class: Class I Product Code: LYR Dated: October 3, 2006 Received: October 5, 2006

Dear Ms. Sinclair:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sally antrom

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: __ RAPIRUN H. pylori Antibody Detection Kit

Indications for Use:

The RAPIRUN H. pylori Antibody Detection Kit is a rapid immunochromatographic assay. intended for the qualitative detection of antibodies against Helicobacter pylori (H. pylori) in urine to aid in the diagnosis of H. pylori infection. The RAPIRUN Kit is suitable for use in both point-of-care and clinical laboratory settings.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Rucell. Park.
Division Sign-Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K052716

Otsuka Pharmaceutical Co., Ltd. 510(k) RAPIRUN H. pylori Antibody Detection Kit

September 28, 2005

Page vii

§ 866.3110

Campylobacter fetus serological reagents.(a)
Identification. Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identifyCampylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases.Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.(b)
Classification. Class I (general controls).