K Number
K052682
Device Name
ALIGNRT
Manufacturer
Date Cleared
2006-01-06

(100 days)

Product Code
Regulation Number
892.5050
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AlignRT system is used to position patients at the isocentre of the linear accelerator for radiation therapy procedures. Patient contour data can be extracted and exported from the acquired data for the purpose of treatment planning. AlignRT is used by radiotherapy professionals in appropriate hospital environments. It can be used with all types of radiotherapy patients and can image any visible anatomical region. The system is completely non-invasive and does not require the application of any external markers to the patient.

Device Description

AlignRT is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. The system consists of advanced software, a computer workstation, two 3D camera units, cables, and templates that are used for camera calibration. The system is non-invasive, does not require the use of body markers and produces no irradiation during the imaging process. AlignRT first generates a reference surface of the optimum treatment position determined during treatment simulation. This reference image is generated by either recording the surface of a patient placed in a conventional radiotherapy simulator in which the system is installed or by importing skin contours from CT (X-Ray Computer Tomography) volumetric data generated via third party treatment planning software. Prior to each treatment session the patient's position is imaged and compared to the reference image by the system's surface matching software. Where movement from the reference position is detected the software calculates new co-ordinates to adjust the treatment couch for optimal positioning of the patient. The system is designed to interface directly with certain third party couch control systems. In order to minimise the setup errors that can result from respiratory motion, an option is available to gate the image capture of the system so that all data is acquired at a repeatable point within the breathing cycle such as the end of exhalation. Real-time imaging (1 frame/sec) of the patient is possible during the actual treatment and software verification tools are provided in order to determine any patient movement. Patient contour data may be extracted from surface data acquired by the system. In addition, lung contour data may be estimated by utilising AlignRT 3D surface data and 2D X-ray data acquired from either a conventional simulator or portal imager. These data may be exported for the purpose of treatment planning by radiotherapy professionals.

AI/ML Overview

Based on the provided 510(k) summary (K052682) for the AlignRT system, the following information regarding acceptance criteria and study details can be extracted:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document (K052682) is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed performance study with explicit acceptance criteria and corresponding performance metrics for a novel technology. Therefore, the "acceptance criteria" listed below are inferred from the device's stated purpose and comparison to predicates, and the "reported device performance" is a general statement from the summary. Specific quantitative metrics are not detailed in this document.

Acceptance Criteria (Inferred from Device Purpose and Comparison)Reported Device Performance (General Statement in Summary)
Ability to image the skin surface of a patient in 3D before and during radiotherapy treatment."AlignRT is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment."
Capability to generate a reference surface of the optimum treatment position."AlignRT first generates a reference surface of the optimum treatment position determined during treatment simulation."
Comparison of patient's current position to the reference image."Prior to each treatment session the patient's position is imaged and compared to the reference image by the system's surface matching software."
Calculation of new co-ordinates to adjust the treatment couch for optimal patient positioning where movement is detected."Where movement from the reference position is detected the software calculates new co-ordinates to adjust the treatment couch for optimal positioning of the patient."
Real-time imaging capability during treatment to detect patient movement."Real-time imaging (1 frame/sec) of the patient is possible during the actual treatment and software verification tools are provided in order to determine any patient movement."
Ability to extract patient contour data."Patient contour data may be extracted from surface data acquired by the system."
Ability to estimate lung contour data by utilizing 3D surface data and 2D X-ray data."In addition, lung contour data may be estimated by utilising AlignRT 3D surface data and 2D X-ray data acquired from either a conventional simulator or portal imager."
No new efficacy or safety concerns compared to predicate devices despite technological differences."performance data has been submitted to show that AlignRT achieves its intended use and that these technological differences raise no new efficacy or safety concerns."

2. Sample Size Used for the Test Set and Data Provenance

The 510(k) summary does not explicitly state a sample size for a test set or any details about data provenance (e.g., country of origin, retrospective or prospective data) for a specific clinical study. General "performance data" is mentioned as being submitted, but the specifics are not included in this publicly available summary.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The 510(k) summary does not provide information regarding the number of experts, their qualifications, or how ground truth was established for any specific test set. The device's function involves comparing to a "reference surface" determined during "treatment simulation" or from "CT volumetric data," which implies clinical input from radiotherapy professionals, but specific details about establishing ground truth for validation are absent.

4. Adjudication Method for the Test Set

The 510(k) summary does not describe any adjudication method used for a test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The 510(k) summary makes no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, nor does it quantify any effect size of human readers improving with or without AI assistance. The AlignRT system is presented as a positioning and monitoring tool for radiotherapy, not an AI-assisted diagnostic or interpretive tool for human readers in the traditional MRMC sense.

6. Standalone (Algorithm Only) Performance Study

The document states that "performance data has been submitted to show that AlignRT achieves its intended use." This implies that a standalone performance evaluation of the system's capabilities (e.g., accuracy of 3D surface imaging, precision of position detection, correlation with actual patient movement) was likely performed and submitted to the FDA. However, the details of such a standalone study, including specific metrics and methodologies, are not present in this summary. The core of the device is its algorithm for surface matching and coordinate calculation.

7. Type of Ground Truth Used

The ground truth for the core function of the AlignRT system is implicitly related to:

  • The "optimum treatment position determined during treatment simulation." This would be established by radiotherapy professionals.
  • "Skin contours from CT (X-Ray Computer Tomography) volumetric data" which are considered a gold standard for anatomical representation in radiotherapy planning.
  • "2D X-ray data acquired from either a conventional simulator or portal imager" when estimating lung contour data.

Therefore, the ground truth leans towards expert consensus (for optimal treatment position) and imaging modalities considered high-fidelity anatomical references (CT, X-ray). Pathology or outcomes data are not mentioned as direct ground truth for this positioning device.

8. Sample Size for the Training Set

The 510(k) summary does not provide any information regarding the sample size of a training set. Given the described functionality (3D surface imaging, matching, coordinate calculation), the system likely relies on image processing algorithms that might be trained or calibrated, but details are not disclosed.

9. How the Ground Truth for the Training Set Was Established

Since no information on a training set is provided, there is correspondingly no information on how its ground truth was established.

{0}------------------------------------------------

510(k) Summary

K052682

As required by 21 CFR 807.92

1 Owner

Vision RT Ltd Daws House 33-35 Daws Lane London NW7 4SD United Kingdom

Principle Contact:

Dr Norman Smith Chief Executive Officer Tel: +44 (0)20 8906 6620 Fax:+44 (0)20 8906 6621 Email: nsmith@visionrt.com

2 510(k) Summary Preparation Date

July 19, 2005

3 Device Details

Name AlignRT

Common Name

Radiotherapy positioning system

Classification Name System, radiation therapy, charged-particle Product code: LHN Regulation: 21 CFR 892.5050

4 Predicate Devices

Exactrac system (K983660) and, Osiris system (K981153).

5 Device Description

AlignRT is a video-based three-dimensional (3D) surface imaging system, which is used to image the skin surface of a patient in 3D before and

Page 1 of 5

{1}------------------------------------------------

during radiotherapy treatment. The system consists of advanced software, a computer workstation, two 3D camera units, cables, and templates that are used for camera calibration. The system is non-invasive, does not require the use of body markers and produces no irradiation during the imaging process.

AlignRT first generates a reference surface of the optimum treatment position determined during treatment simulation. This reference image is generated by either recording the surface of a patient placed in a conventional radiotherapy simulator in which the system is installed or bv importing skin contours from CT (X-Ray Computer Tomography) volumetric data generated via third party treatment planning software. Prior to each treatment session the patient's position is imaged and compared to the reference image by the system's surface matching software. Where movement from the reference position is detected the software calculates new co-ordinates to adjust the treatment couch for optimal positioning of the patient. The system is designed to interface directly with certain third party couch control systems.

In order to minimise the setup errors that can result from respiratory motion, an option is available to gate the image capture of the system so that all data is acquired at a repeatable point within the breathing cycle such as the end of exhalation.

Real-time imaging (1 frame/sec) of the patient is possible during the actual treatment and software verification tools are provided in order to determine any patient movement.

Patient contour data may be extracted from surface data acquired by the system. In addition, lung contour data may be estimated by utilising AlignRT 3D surface data and 2D X-ray data acquired from either a conventional simulator or portal imager. These data may be exported for the purpose of treatment planning by radiotherapy professionals.

5 Indications for Use

The AlignRT system is used to position patients at the isocentre of the linear accelerator for radiation therapy procedures. Patient contour data can be extracted and exported from the acquired data for the purpose of treatment planning.

AlignRT is used by radiotherapy professionals in appropriate hospital environments. It can be used with all types of radiotherapy patients and can image any visible anatomical region. The system is completely non-Can Irrage any Visible anatomical region. The system is completely now
Previous and the comment - www.comment.com - Page 2 of 5

Page 2 of 5

{2}------------------------------------------------

invasive and does not require the application of any external markers to the patient.

Caution: Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.

Comparison of Technological Characteristics 6

The AlignRT system is substantially equivalent to the predicate devices in terms of their intended use and technological characteristics. There are differences between the Align RT and predicate devices in terms of their principles of operation, materials, performance, human factors and energy delivered by the system. However, performance data has been submitted to show that AlignRT achieves its intended use and that these technological differences raise no new efficacy or safety concerns.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three wing-like shapes, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 6 2006

Vision RT Ltd. %Mr. Ned Devine Responsible Third Party Official Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719

Re: K052682 Trade/Device Name: AlignRT Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: December 20, 2005 Received: December 22, 2005

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

. Nancy C. Bigdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use
------------------------------

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

The AlignKT system is used to position patients undergoing external beam radiation therapy. Prior to each treatment session the patient is registered to the position which best reproduces the patient's position during treatment simulation. The system and also monitor the patient's position during treatment so that movements from the optimum position can be detected. Patient contour data can be extracted and exported from the acquired data for the purpose of treatment planning by radiotherapy professional .

AlignRT is used by radiotherapy professionals in appropriate hospital environments. It can be used with all types of radiotherapy patients and can image any visible anatomical region.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Larson

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices-[K052682](https://510k.innolitics.com/search/K052682)
510(k) Number and Radiological Devices-`K052682

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.