(69 days)
Lakeshore Technologies Inc., formally Syracuse Scientific Inc., intends to use its SS960 CCD TV camera in combination with its Image Star II digital image processor for the presentation of Fluoroscopic or Radiographic TV images as presented by the image intensifier of the X-ray system.
A) The images may also be used to assist in operative procedures.
B) This device is to be used in diagnostic radiology or cardiology.
C) The clinical settings for this device are hospitals and for medical centers.
D) This device is to be used by a qualified radiologist or cardiologist.
E) This device is substantial equivalent to our Fluorecord Imaging System (510k K952493) in terms of the operation, design and intended use.
The Image Star II digital image processor in combination with the SS960 CCD TV camera makes up a simple, low cost, high performance imaging system. This system is designed to be used in place of the conventional 100mm or 105mm spot film cameras for any standard RF procedure. This digital imaging system will be used instead of, or in addition the standard spot film device. This system is designed to interface easily, in a generion no, ner, to any x-ray system that is designed to accommodate a spot rapid and /or cinc camers. The SS960 CCD TV camera will present a flickerless image when interfaced to anex. Flux generator performing pulsed fluoroscopy and/or cine radiography.
Image Star II is a complete digital imaging system and is comprised of two main components; a main acquisition control unit or Work Station and a Review Station. Images are acquired, processed, displayed and reviewed to allow high quality image analysis and diagnosis with the following special features: Image Analysis, Image Processing, Exams storage, CD Recording, Intranet Server and Medical Report Editing. This system performs functions such as edge enhancement, noise reduction, multi frame loop, and digital subtraction. This makes it useful for diagnostic radiology. cardiology. or mobile C-arm use.
The Work Station is interfaced to the x-ray generator and TV camera and is primarily used for acquisition. The Review Station is an optional independent Station connected yo the Work Station through a high speed network and is primarily used for storage and review. Recorded images can be reviewed on both the Work Station and the Review Slation. Studies can be opened, edited, analyzed. saved to a medical report and printed. Current or past patient studies can be viewed. Recorded images or image sequences can be zoomed, the window and level adjusted, annotation added and many other features. Patient data can be added, edited or deleted prior to the actual date and time of the exam. Patient data and information from a recorded study can be easily viewed. A full data redundancy system is used to store and maintain study data to eliminate loss or damage. Interfacing to the Work Station and Review Station is accomplished with the use of a standard text keyboard and mouse. An optional Touch Screen can also be used to interface with the Graphies User Interface.
The camera is a two piece configuration including a camera head and a camera control unit (CCU). The camera head is a single board design utilizing a single flexible intercombetting cable for power and low voltage digital signal (I.VDS) output to the CCU. The CCU has both standard video outputs such as 1049/60 Hz interlaced or XVGA and also an LVDS digital output to the Image Star image processor.
This system is DICOM 3.0 compliant for interfacing to a DICOM network and can be connected to paper printers for generating medical report documents.
Here's an analysis of the provided 510(k) summary for the Lakeshore Technologies Inc. "Image Star II" device, focusing on acceptance criteria and study information.
Missing Information: It's important to note that the provided 510(k) summary does not contain information about specific acceptance criteria or a study designed to prove the device meets those criteria. 510(k) submissions for devices like the Image Star II often rely on demonstrating substantial equivalence to a predicate device rather than presenting formal performance studies with defined acceptance criteria in the manner one might see for novel diagnostic AI.
The document primarily focuses on describing the device, its intended use, and comparing its specifications to a predicate device to claim substantial equivalence.
Based on the provided text, here's what can be extracted and what remains unknown regarding acceptance criteria and performance studies:
Acceptance Criteria and Device Performance Study Analysis: Lakeshore Technologies Inc. "Image Star II" (K052610)
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (If explicitly stated) | Reported Device Performance |
|---|---|
| Not explicitly stated within the provided document. The submission focuses on demonstrating substantial equivalence to a predicate device (Fluorecord Imaging System, K952493) through a comparison of specifications and intended use. The performance criteria are implicitly related to meeting or exceeding the predicate device's capabilities or function for equivalent diagnostic image processing. | The device is described as a "simple, low cost, high performance imaging system" aiming to be used "in place of the conventional 100mm or 105mm spot film cameras for any standard RF procedure." It offers features like "edge enhancement, noise reduction, multi frame loop, and digital subtraction." Concrete, quantified performance metrics against specific acceptance thresholds are not provided. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not specified. No formal "test set" in the context of a performance study with quantitative metrics is mentioned. The equivalence claim appears to be based on design, functionality, and specification comparison.
- Data Provenance: Not specified. Given the nature of a 510(k) for an image processor, the "data" would typically be the images it processes. There's no information on a specific dataset or its origin used for evaluation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Not applicable. The document does not describe a study involving expert-established ground truth for a test set.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. No formal test set or ground truth establishment described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study Done: No. The provided information does not describe any MRMC study. The submission focuses on the device itself as an image processor, not on its impact on human reader performance in a controlled study scenario.
- Effect Size of Human Reader Improvement with AI vs. without AI: Not applicable, as no MRMC study was conducted or reported.
6. Standalone (Algorithm Only) Performance Study
- Standalone Study Done: No. While the device is an "Image Star II digital image processor," the 510(k) does not present a standalone, quantitative performance study of its processing capabilities against defined metrics. Its performance is implied to be equivalent to or improved over the predicate based on its features and technical specifications (e.g., CCD camera, digital processing capabilities).
7. Type of Ground Truth Used
- Type of Ground Truth: Not applicable. No ground truth is described as having been established for evaluating the device's performance in a study. The "ground truth" for a device like this is inherently linked to its ability to present diagnostic images clearly, which is assessed through equivalence to existing, accepted systems in diagnostic radiology/cardiology.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This device is an image processor, not an AI/ML algorithm that typically requires a distinct training set in the modern sense. Its functionality is based on established digital image processing techniques (edge enhancement, noise reduction, etc.) rather than learning from a large dataset.
9. How the Ground Truth for the Training Set was Established
- Ground Truth Establishment for Training Set: Not applicable.
{0}------------------------------------------------
WD52611
SUMMARY
NOV 3 0 2005
510k Summary for "K052610" premarket notification Submission for a TV camera and digital imaging processor
o o1...
Lakeshore Technologies Inc., formally Syracuse Scientific Inc., is requesting approval to market its SS960 CCD TV camera in combination with its Image Star II digital image processor. We manufacture this product at 7536 Murray Dr. Cicero, NY 13039.
| A) Classification Name: | Image Star System |
|---|---|
| Common / usual name: | Digital Image Processor System |
| Proprietary Name: | Image Star II |
- B) Establishment Registration number: 9020592
- C) Classification: Since our system becomes a component of a class two devise, we have considered it as a class two devise, but we are unaware of it's actual classification. This system is not subject to any performance standards. We do not manufacture the TV monitor, which is a class two devise, and we do not manufacture the x-ray machine to which our imaging system is connected. Our Fluorecord system which we are comparing our new system to has 510k number K952493. is a class 2 devise and the product code is JAA.
- D) Performance Standards: Not applicable
- I:) Label / Labeling / Advertisements: Copics of the package labeling, operation manual. technical and installation instructions. and schematics are enclosed.
- F) Intended Use: For the presentation of a Fluoroscopic or Radiographic TV image as presented by the image intensifier of the X-ray system. This is to be used in diagnostic radiology or cardiology. These images may also be used to assist in operative procedures.
- (i) Substantial Equivalence: This devise is similar to our Fluorecord Imaging System (510k) # K952493 in terms of the operation, design and intended use. A side by side comparison of specifications is enclosed.
{1}------------------------------------------------
SECTION: 1.1
General Description of the Image Star System
The Image Star II digital image processor in combination with the SS960 CCD TV camera makes up a simple, low cost, high performance imaging system. This system is designed to be used in place of the conventional 100mm or 105mm spot film cameras for any standard RF procedure. This digital imaging system will be used instead of, or in addition the standard spot film device. This system is designed to interface easily, in a generion no, ner, to any x-ray system that is designed to accommodate a spot rapid and /or cinc camers. The SS960 CCD TV camera will present a flickerless image when interfaced to anex. Flux generator performing pulsed fluoroscopy and/or cine radiography.
Image Star II is a complete digital imaging system and is comprised of two main components; a main acquisition control unit or Work Station and a Review Station. Images are acquired, processed, displayed and reviewed to allow high quality image analysis and diagnosis with the following special features: Image Analysis, Image Processing, Exams storage, CD Recording, Intranet Server and Medical Report Editing. This system performs functions such as edge enhancement, noise reduction, multi frame loop, and digital subtraction. This makes it useful for diagnostic radiology. cardiology. or mobile C-arm use.
The Work Station is interfaced to the x-ray generator and TV camera and is primarily used for acquisition. The Review Station is an optional independent Station connected yo the Work Station through a high speed network and is primarily used for storage and review. Recorded images can be reviewed on both the Work Station and the Review Slation. Studies can be opened, edited, analyzed. saved to a medical report and printed. Current or past patient studies can be viewed. Recorded images or image sequences can be zoomed, the window and level adjusted, annotation added and many other features. Patient data can be added, edited or deleted prior to the actual date and time of the exam. Patient data and information from a recorded study can be easily viewed. A full data redundancy system is used to store and maintain study data to eliminate loss or damage. Interfacing to the Work Station and Review Station is accomplished with the use of a standard text keyboard and mouse. An optional Touch Screen can also be used to interface with the Graphies User Interface.
The camera is a two piece configuration including a camera head and a camera control unit (CCU). The camera head is a single board design utilizing a single flexible intercombetting cable for power and low voltage digital signal (I.VDS) output to the CCU. The CCU has both standard video outputs such as 1049/60 Hz interlaced or XVGA and also an LVDS digital output to the Image Star image processor.
This system is DICOM 3.0 compliant for interfacing to a DICOM network and can be connected to paper printers for generating medical report documents.
{2}------------------------------------------------
SECTION: 2.2
Brochures on Predicate Device
The following pages are brochures of the Fluorecord imaging system.
| 510(k) Number | K952493 |
|---|---|
| Devise Name | FLUORECORD MODEL 762 |
| Product Code | JAA |
| Decision Date | 06/22/1995 |
| Decision | Substantially Equivalent (SE) |
{3}------------------------------------------------
SECTION: 4.1
Software Considerations
Our system is mainly intended for presenting a high quality TV image for the purpose of diagnoses or to assist in operative procedures. These TV images are stored to the hard disks redundantly by both of the system computers. The TV images may then be transferred to film by a video camera or laser imager or saved to an archival system via the DICOM network. We do have software in this system for performing measurements and quantitative analysis and the method used to derive these measurements are based an established techniques. (Refer to Section 4.4 for the methods used for these measurements),
The software in the system is in the form of executable files that run on the Windows XP* operating system platform. This software is used for the following purposes.
- Control of entering patient identification and hospital identification and 1) displaying this information on the TV monitor.
- Control of transferring image data to and from the disk storage system and to a 2) laser imager, video camera device or DICOM network.
- Control of image processing circuitry for enhancement of image quality. Such 3) as, edge enhancement, noise reduction etc.
- Control for acting upon operator input from the keyboard or mouse. 4)
The software control in the system has a minor level of concern in terms of patient safety, It is extremely unlikely for a failure to occur having an effect on the TV image that would produce an artifact that could be mistaken for a patient anomaly. All conccivable failures in our validation procedures would produce an obvious (problem) image.
We have displayed the patient name on the TV monitor in order to protect from images being stored with the incorrect patient identification. Our philosophy is that we cannot protect from an operator error in terms of entering the wrong information. Therefore, the information is displayed so that the entire staff in the procedure room can see the patient identification on the monitor to allow someone to point out an error.
We have taken additional precaution to protect the patient identification and image data that represents the patient s medical document files in two ways. First, all TV images are stored to hard disk redundantly on both the work station and the review station computers. Secondly, in many systems when the hard disks are full, the oldest exams are deleted or overwritten automatically to make room for the newer exams. In this system, the oldest TV images can only be deleted after being saved on the DICOM network or CD / DVD media and only with direct intervention by the user.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure with outstretched arms.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Jack Race Lakeshore Technologies, Inc. Formerly Syracuse Scientific. Inc. 7536 Murray Drive CICERO NY 13039
AUG 2 1 2013
Re: K052610 Trade/Device Name: Image Star II Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: November 8, 2005 Received: November 10, 2005
Dear Mr. Race:
This letter corrects our substantially equivalent letter of November 30, 2005.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known): __ k052610_
Device Name: _________________________________________________________________________________________
Indications for Use:
Lakeshore Technologies Inc., formally Syracuse Scientific Inc., intends to use its SS960 CCD TV camera in combination with its Image Star II digital image processor for the presentation of Fluoroscopic or Radiographic TV images as presented by the image intensifier of the X-ray system.
- A) The images may also be used to assist in operative procedures.
- B) This device is to be used in diagnostic radiology or cardiology.
- C) The clinical settings for this device are hospitals and for medical centers.
- D) This device is to be used by a qualified radiologist or cardiologist.
- E) This device is substantial equivalent to our Fluorecord Imaging System (510k K952493) in terms of the operation, design and intended use.
Daniel A. Leggerson
(Division Sian-Off Division of Reproductive. Abde and Radiological Devices 510(k) Number
Prescription Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
ANDIOR
Over-The-Counter Use ___________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
§ 892.1650 Image-intensified fluoroscopic x-ray system.
(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.