(175 days)
No
The device description and performance studies focus on the physical properties and sterilization capabilities of a sealed container, with no mention of AI or ML.
No
The device is a sterilization container for holding medical devices and surgical supplies, not for treating a disease or condition.
No
The device, ONE TRAY, is described as a sealed sterilization container used to hold medical devices for sterilization and transport. Its function is to facilitate the sterilization process and ensure the safe delivery of instruments, not to diagnose medical conditions or provide medical information about a patient.
No
The device description clearly outlines a physical, rigid container with filters and covers, indicating it is a hardware device for sterilization and transport. There is no mention of software as the primary or sole component.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states that the device is used to hold and sterilize medical devices and supplies for immediate use. It is a container for sterilization and transport.
- Device Description: The description details the physical components of a sterilization container, focusing on its ability to allow steam penetration and act as a barrier.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body to provide information for diagnostic, monitoring, or compatibility purposes.
IVD devices are specifically designed to perform tests on biological samples (like blood, urine, tissue) to diagnose conditions, monitor health, or determine compatibility. This device's function is entirely related to the sterilization and handling of medical instruments.
N/A
Intended Use / Indications for Use
The ONE TRAY® Sealed Sterilization Containers are intended to be used for rapid sterilization of instruments or instrument sets in flash cycles. The containers are intended to be used to hold medical devices during steam sterilization. The complete line of ONE TRAY® Sealed Sterilization Containers can be processed in both flash steam pre-vacuum and gravity cycles.
After sterilization, ONE TRAY® provides for the safe transport and assured delivery of the enclosed devices in a sealed container with tamper evident protection according to AAMI and AORN guidelines.
ONE TRAY® Sealed Sterilization Containers are intended to be used to hold temperature tolerant medical devices, surgical supplies, single instruments or an instruments or an instrument set for immediate use following flash sterilization. This includes sterilization of lumens 3 mm in diameter or larger with lengths of up to 400 mm.
After sterilization, ONE TRAY provides for the safe transport and assured delivery for immediate use of the enclosed devices in a sealed container with tamper evident security and load record documentation according to AAMI and AORN guidelines.
The performance and intended use of ONE TRAY Sealed Sterilization Containers should comply at all times with the methods of use and flash sterilization guidelines as recommended by the manufacturer of the devices being sterilized, AAMI (Association for the Advance of Medical Instrumentation), AORN (Association of periOperative Nurses), ASHCSP (American Society of Healthcare Central Service Professionals), NIHSPO (National Institute for the Certification of Healthcare Sterile Processing and Distribution Personnel) and IAHCSMM (International Association of Healthcare Central Service Materiel Management)
Product codes (comma separated list FDA assigned to the subject device)
80FRG, KCT
Device Description
ONE TRAY® is a sealed rigid container with a rectangular patterned group of perforations forming vented areas in the lid and base. Dispossable hydrophobic SMS filters cover each vented area and are held firmly in place by a perforated stainless steel filters cover. This assembly permits the penetration of steam during the sterilization process and serves as a bacterial and fluid barrier at the conclusion of the sterilization cycle.
The ONE TRAY® design is rectangular in shape and has vents in the lid and floor. The ONE TRAY® design/performance concept takes advantage of the thermodynamic behavior of steam to facilitate the expeditious and complete introduction of the sterilant throughout the container. This is accomplished by incorporating a specific number of proportionate size vents in strategic locations in both the lid and base. As the sterilizing media (steam) fills up the chamber of the sterilizer, it is introduced into the container through the dedicated entry vent in the lid of ONE TRAY®. This dedicated entry port provides for displacement of the atmosphere within the container from the top to the bottom.
The vents located in the extreme lateral portions of the floor provide two dedicated exit ports that offer twice the vapor exchange capacity of the single entry vent in the lid. This relationship facilitates the complete, consistent and expeditious displacement of the atmosphere by the sterilant throughout the ONE TRAY® container,
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Testing was performed in order to determine the functional equivalency between the ONE TRAY® and the predicate device. Performance testing of the ONE TRAY® included, steam penetration, half cycle validation, package integrity, shelf life, material compatibility and biocompatibility studies. Based on the results of laboratory tests; the ONE TRAY® is substantially equivalent to the predicate container.
The ONE TRAY® components listed in the preceding table have been validated in all of the various size configurations listed in the following table to process a twenty five pound (25 lb) gross weight load (single container plus contents) in a steam pre-vacuum cycle at 132°C for 4 minutes exposure time; or in a steam gravity cycle at 132°C for 34 minutes exposure time.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 880.6850 Sterilization wrap.
(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).
0
510(k) Summary of
Safety and Effectiveness
Submitter: Sterile Containment Technology, LLC 1301 Quarry Court Suit 204 Point Richmond, CA 94801 510-412-6383 (phone) 510-588-4699 (fax)
◆ Establishment FDA Registration No .: 3003793526
. Date Summary was prepared August 30th, 2005
Percival Banks Printed name of person submitting for 510(k)
Signature of person submitting for 510(k)
President Title of person submitting for 510(k)
Device Name and Classification
Trade Name: | ONE TRAY® Sealed Sterilization Container |
---|---|
Classification Name: | Sterilization Wrap |
Common Name: | Sterilization, Rigid Reusable Case |
Device Classification: | General Hospital, Class II, Regulation No. 880.6850 |
Product Code: | 80FRG |
Predicate Device
.
Case Medical SteriTite® K960738 and K022978
Additional Predicate Devices
Medin K833343 and FlashPak K871202
Device Description
ONE TRAY® is a sealed rigid container with a rectangular patterned group of perforations forming vented areas in the lid and base. Dispossable hydrophobic SMS filters cover each vented area and are held firmly in place by a perforated stainless steel filters cover. This assembly permits the penetration of steam during the sterilization process and serves as a bacterial and fluid barrier at the conclusion of the sterilization cycle.
1
Intended Use
The ONE TRAY® Sealed Sterilization Containers are intended to be used for rapid sterilization of instruments or instrument sets in flash cycles. The containers are intended to be used to hold medical devices during steam sterilization. The complete line of ONE TRAY® Sealed Sterilization Containers can be processed in both flash steam pre-vacuum and gravity cycles.
After sterilization, ONE TRAY® provides for the safe transport and assured delivery of the enclosed devices in a sealed container with tamper evident protection according to AAMI and AORN guidelines.
Technical Characteristics
The ONE TRAY® design is rectangular in shape and has vents in the lid and floor. The ONE TRAY® design/performance concept takes advantage of the thermodynamic behavior of steam to facilitate the expeditious and complete introduction of the sterilant throughout the container. This is accomplished by incorporating a specific number of proportionate size vents in strategic locations in both the lid and base. As the sterilizing media (steam) fills up the chamber of the sterilizer, it is introduced into the container through the dedicated entry vent in the lid of ONE TRAY®. This dedicated entry port provides for displacement of the atmosphere within the container from the top to the bottom.
The vents located in the extreme lateral portions of the floor provide two dedicated exit ports that offer twice the vapor exchange capacity of the single entry vent in the lid. This relationship facilitates the complete, consistent and expeditious displacement of the atmosphere by the sterilant throughout the ONE TRAY® container,
Performance Testing
Testing was performed in order to determine the functional equivalency between the ONE TRAY® and the predicate device. Performance testing of the ONE TRAY® included, steam penetration, half cycle validation, package integrity, shelf life, material compatibility and biocompatibility studies. Based on the results of laboratory tests; the ONE TRAY® is substantially equivalent to the predicate container.
Conclusion
Supportive data has demonstrated that the ONE TRAY® is substantially equivalent to the predicate device in that they have the same intended uses and the performance attributes are the same. The use of the ONE TRAY® raises no issues related to its safety or effectiveness and therefore the ONE TRAY® should be allowed for market in the United States.
2
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three heads, representing health, human services, and well-being. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.
MAR 1 3 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Gage Garman Sterile Containment Technology, Limited Liability Company 1301 Quarry Court, Suite 204 Point Richmond, California 94801
Re: K052567
Trade/Device Name: ONE TRAY® Sealed Sterilization Container Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: February 11, 2006 Received: February 16, 2006
Dear Mr. Garman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 - Mr. Gage Garman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Saith y. Michael Oond.
Chíu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
INDICATIONS for USE
510(k) Number: K052567
Device Name: ONE TRAY® Sealed Sterilization Container
Indications For Use:
ONE TRAY® Sealed Sterilization Containers are intended to be used to hold temperature tolerant medical devices, surgical supplies, single instruments or an instruments or an instrument set for immediate use following flash sterilization. This includes sterilization of lumens 3 mm in diameter or larger with lengths of up to 400 mm.
ONE TRAY® Sealed Sterilization Containers are available in the following sizes:
ONE TRAY System Components | ||
---|---|---|
Product No. | Product Name | Description |
ONE 20 Model | ||
OTS-MB2000 | ONE 20ML Base | ONE 20ML Base - 20" Low Profile Base (Metal, Blue, 11.5"x20.5"x3") |
OTS-MB2001 | ONE 20MH Base | ONE 20MH Base - 20" High Profile Base (Metal, Blue, 11.5"x20.5"x5") |
OTS-MB2010 | ONE 20ML Cover | ONE 20ML Cover - 20" Low Profile Cover (Metal, Blue, 11.5"x20.5"x1") |
OTS-MB2011 | ONE 20MH Cover | ONE 20MH Cover - 20" High Profile Cover (Metal, Blue, 11.5"x20.5"x2.5") |
OTS-MB2020 | ONE 20M Deck | ONE 20M Deck -20" Deck Plate (Metal, Blue, 11.5"x20.5") |
ONE 23 Model | ||
OTS-MB2300 | ONE 23ML Base | ONE 23ML Base - 23" Low Profile Base (Metal, Blue, 11.5"x23.5"x3") |
OTS-MB2301 | ONE 23MH Base | ONE 23MH Base - 23" High Profile Base (Metal, Blue, 11.5"x23.5"x5") |
OTS-MB2310 | ONE 23ML Cover | ONE 23ML Cover - 23" Low Profile Cover (Metal, Blue, 11.5"x23.5"x1") |
OTS-MB2311 | ONE 23MH Cover | ONE 23MH Cover - 23" High Profile Cover (Metal, Blue, 11.5"x23.5"x2.5") |
OTS-MB2320 | ONE 23M Deck | ONE 23M Deck -23" Deck Plate (Metal, Blue, 11.5"x23.5) |
The ONE TRAY® components listed in the preceding table have been validated in all of the various size configurations listed in the following table to process a twenty five pound (25 lb) gross weight load (single container plus contents) in a steam pre-vacuum cycle at 132°C for 4 minutes exposure time; or in a steam gravity cycle at 132°C for 34 minutes exposure time.
Shule A. Murgheigh, 13/4/06
Page 1 of 2
ing, Genere
Lakes
Koszta
5
ONE 20 Model | ONE 23 Model | ||||
---|---|---|---|---|---|
Name | Description | Product No. | Name | Description | Product No. |
ONE 20MH | High Profile Base | OTS-MB2001 | ONE 23MH | High Profile Base | OTS-MB2301 |
High Profile Cover | OTS-MB2011 | High Profile Cover | OTS-MB2311 | ||
Deck Plate | OTS-MB2020 | Deck Plate | OTS-MB2320 | ||
ONE 20MHL | High Profile Base | OTS-MB2001 | ONE 23MHL | High Profile Base | OTS-MB2301 |
Low Cover | OTS-MB2010 | Low Cover | OTS-MB2310 | ||
Deck Plate | OTS-MB2020 | Deck Plate | OTS-MB2320 | ||
ONE 20MLH | Low Profile Base | OTS-MB2000 | ONE 23MLH | Low Profile Base | OTS-MB2300 |
High Profile Cover | OTS-MB2011 | High Profile Cover | OTS-MB2311 | ||
Deck Plate | OTS-MB2020 | Deck Plate | OTS-MB2320 | ||
ONE 20ML | Low Profile Base | OTS-MB2000 | ONE 23ML | Low Profile Base | OTS-MB2300 |
Low Profile Cover | OTS-MB2010 | Low Profile Cover | OTS-MB2310 | ||
Deck Plate | OTS-MB2020 | Deck Plate | OTS-MB2320 |
After sterilization, ONE TRAY provides for the safe transport and assured delivery for immediate use of the enclosed devices in a sealed container with tamper evident security and load record documentation according to AAMI and AORN guidelines.
The performance and intended use of ONE TRAY Sealed Sterilization Containers should comply at all times with the methods of use and flash sterilization guidelines as recommended by the manufacturer of the devices being sterilized, AAMI (Association for the Advance of Medical Instrumentation), AORN (Association of periOperative Nurses), ASHCSP (American Society of Healthcare Central Service Professionals), NIHSPO (National Institute for the Certification of Healthcare Sterile Processing and Distribution Personnel) and IAHCSMM (International Association of Healthcare Central Service Materiel Management)
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use ___X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 2
Shiela P. Murphy, A 3/9/04
General Hos,
Dental Devices