K Number
K052519
Date Cleared
2006-11-30

(442 days)

Product Code
Regulation Number
866.3940
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Spectral West Nile virus IgM STATus™ test is a rapid immunochromatographic lateral flow assay that utilizes recombinant West Nile virus (WNV) antigen (E glycoprotein) for the qualitative detection of IgM antibodies to WNV in human serum or plasma (sodium heparin or sodium citrate). This test is for the presumptive laboratory diagnosis of West Nile virus infection in patients having signs and symptoms of meningoencephalitis. Positive results must be confirmed by the PRNT (Plaque Neutralization Reduction Test), or by using the current CDC guidelines for diagnosis of this disease. The Spectral test is not intended for point of care testing, home use or in screening blood donor samples.

Device Description

A Simple and Rapid Immunoassay for the Qualitative Detection of West Nile Virus IgM Antibodies in Human Serum or Plasma. The Spectral WNV IgM STATus™ test employs solid-phase immunochromatographic assay technology to qualitatively detect the presence of WNV IgM antibodies in serum or plasma. When the specimen to be tested is dispensed into the sample well of the Spectral device, anti-WNV IgM in the sample will bind to the recombinant WNV antigen (envelop glycoprotein (E) of West Nile virus, NY99 strain) to form a tertiary complex with gold-labeled monoclonal murine antibody against flavivirus family glycoprotein E. This tertiary complex will migrate through reaction strip and be captured by goat anti-human IgM antibodies at the Test area. Excess, unreacted gold complex detector is captured by immobilized anti-mouse IgG antibodies at the Control area. A visible pinkish-purple horizontal band will appear in the Test area within 15 minutes following the addition of a sample if the level of the WNV IgM antibodies in the human serum sample is above the cut-off level. A pinkish-purple band in the Control area indicates that the test is working properly and such a band must always appear, irrespective of the WNV IgM levels, in order for the test to be valid.

AI/ML Overview

Acceptance Criteria and Device Performance Study for Spectral West Nile Virus IgM STATus™ Test

This document outlines the acceptance criteria and the studies that demonstrate the Spectral West Nile Virus IgM STATus™ Test meets these criteria, based on the provided 510(k) summary (K052519).

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria with specific numerical targets in a formal table. However, based on the performance data presented, the implicit acceptance criteria appear to be high agreement (sensitivity and specificity/negative agreement) with a legally marketed predicate device (Focus West Nile Virus IgM Capture ELISA) and reference methods like PRNT, along with robust reproducibility.

Performance MetricImplicit Acceptance Criteria (Derived from Predicate)Reported Device Performance
Negative AgreementHigh agreement with predicate device in non-flavivirus IgM samples.Overall: 98.8% (95% CI: 97.1-99.7%) (n=346)
Site 1: 99.4% (95% CI: 96.5-99.9%) (n=158 out of 160)
Site 2: 98.2% (95% CI: 93.8-99.8%) (n=114)
Site 3: 98.6% (95% CI: 92.4-99.9%) (n=71 out of 72)
Serological Sensitivity (Acute)High agreement with PRNT and CDC WNV IgM & IgG ELISA positive (acute).Site 4: 95% (95% CI: 83.1%-99.4%) (n=40)
Site 5: 100% (95% CI: 88.3%-100%) (n=24)
Serological Sensitivity (Late)High agreement with PRNT and CDC WNV IgM & IgG ELISA positive (late).Site 5: 80% (95% CI: 59.3%-93.2%) (n=25)
Negative Agreement (CDC WNV IgM ELISA Negative)High agreement with CDC WNV IgM ELISA negative.Site 4: 96% (95% CI: 86.3%-99.5%) (n=50)
Site 5: 97% (95% CI: 82.8%-99.9%) (n=30)
ReproducibilityConsistent results across different sites, lots, and operators.All sites and operators produced expected results for all 15 panel members on every day of testing.
Interfering SubstancesNo interference from common blood constituents at specified concentrations.No interference observed up to the listed concentrations of human serum albumin, bilirubin, hemoglobin, triglycerides, and human IgM.
IgM SpecificityConfirmation of IgM class antibodies in positive samples.All 14 DTT-treated samples produced negative results, while untreated paired samples remained positive.
Cross-ReactivityAcceptable levels of cross-reactivity with other pathogens.Varied results, specific percentages reported for each agent (e.g., Herpes Simplex: 10%, ANA: 39%, Rheumatoid Factor: 30%, Dengue Virus: 10%, Yellow Fever: 20%).

2. Sample Size Used for the Test Set and Data Provenance

The test set consisted of several distinct groups of samples tested at different study sites.

  • Non-flavivirus IgM samples (for Negative Agreement):
    • Total Sample Size: 346 specimens.
    • Provenance: Prospectively collected from patients with a variety of non-flavivirus ailments (e.g., rash, febrile, bronchitis, diarrhea, drug, neuropathy, acute coronary syndrome) and from endemic normal populations.
    • Country of Origin:
      • Site 1: South-Western State of America (endemic region).
      • Sites 2 & 3: South Eastern Provinces of Canada (endemic regions).
  • WNV PRNT Confirmed / CDC WNV IgM ELISA Negative specimens (Study Site 4):
    • Total Sample Size: 90 specimens (40 WNV PRNT confirmed; 50 CDC WNV IgM ELISA negative).
    • Provenance: Randomized retrospective patient serum specimens from a provincial health laboratory, sent with suspected WNV infection.
    • Country of Origin: South Eastern Province of Canada.
  • Banked Panel of Clinical Serum Specimens (Study Site 5):
    • Total Sample Size: 79 specimens.
    • Provenance: Banked clinical serum specimens from patients with clinical symptomology consistent with WNV infection (including symptomatic acutely infected, neuro-invasive disease, and asymptomatic for previous exposures).
    • Country of Origin: Prairie Province at Central Canada.
  • Reproducibility Panel: 15 clinical specimens.
  • Interfering Substances Panel: Positive and negative serum specimens spiked with specific substances.
  • IgM Specificity Panel: 14 paired positive samples.
  • Cross-Reactivity Panel: Varied number of samples for each potentially cross-reactive agent (e.g., 10 for HSV, 28 for ANA, 33 for Rheumatoid Factor).
    • Provenance: Sera sero-positive to other potentially cross-reactive pathogens.
    • Country of Origin: North Eastern USA (site 1), Mid-West Canada (site 3 for Dengue), in-house (site 2 for ANA/RF), CDC laboratory at Mid-West USA (site 4 for EEE).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not specify the number or qualifications of experts used to establish the ground truth for the test set. Instead, it relies on established reference methods and the comparator predicate device.

4. Adjudication Method for the Test Set

The document does not mention an explicit adjudication method for the test set (e.g., 2+1, 3+1). The "ground truth" was established by reference assays or clinical characterization of samples.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no indication of a multi-reader multi-case (MRMC) comparative effectiveness study or any assessment of human readers improving with AI vs. without AI assistance. This device is a diagnostic assay (rapid immunochromatographic test) designed for laboratory use, not an AI-assisted diagnostic imaging or interpretation tool for human readers.

6. Standalone Performance Study

Yes, a standalone performance study was conducted. The "Method Comparison" section explicitly tests the Spectral device against a comparator device (Focus West Nile Virus IgM Capture ELISA) and other reference methods like PRNT and CDC WNV IgM ELISA. The reported performance metrics (e.g., negative agreement, serological sensitivity) are measures of the algorithm's (device's) performance without human-in-the-loop intervention beyond reading the visible bands.

7. Type of Ground Truth Used

The ground truth for the different test sets was established using a combination of the following:

  • Reference Devices/Assays:
    • Comparator device: Focus West Nile Virus IgM Capture ELISA (for negative agreement studies).
    • PRNT (Plaque Neutralization Reduction Test): Considered a gold standard for confirming WNV infection. Used for characterizing positive WNV samples at Study Site 4 and 5.
    • CDC WNV IgM ELISA and IgG ELISA: Used as reference assays for characterizing WNV positive and negative samples at Study Site 4 and 5.
  • Clinical Characterization: Samples at Study Site 4 were from patients with "suspected WNV infection based on clinical signs and symptoms." Samples at Study Site 5 were from patients with "clinical symptomology consistent with the West Nile virus infection."
  • Pathogen Sero-positivity: For cross-reactivity studies, samples were characterized as sero-positive to other potentially cross-reactive pathogens (e.g., Herpes Simplex Virus, Cytomegalovirus, Syphilis, Epstein Barr Virus, ANA, Rheumatoid Factor, HIV, certain Encephalitis viruses, Dengue Virus, Hepatitis B/C, Legionella, Yellow Fever, E. coli infection).

8. Sample Size for the Training Set

The document does not provide any information regarding a training set or its sample size. This is a rapid diagnostic test, and the reported studies focus on validation and performance characteristics as opposed to machine learning model development which would typically involve distinct training and test sets.

9. How the Ground Truth for the Training Set Was Established

Since no information on a training set is provided, the method for establishing its ground truth is also not available in the document.

{0}------------------------------------------------

K052519

510(k) Summary of Safety and Effectiveness Spectral West Nile Virus IgM STATus 1M Test Prepared November 29, 2006 Page 1 of 8

ApplicantSpectral Diagnostics Inc.135-2 The West MallToronto, Ontario, CanadaM9C 1C2NOV 30 2006
Establishment Registration No.9617857
Contact PersonDr. Nisar A. ShaikhTel (416) 626-3233 ex 2008Fax (416) 626-9738nshaikh@spectraldx.com
Summary DateNovember 29, 2006
Propriety NameSpectral West Nile Virus IgM STATusTM Test
Generic NameRapid 1 WNV IgM Test
ClassificationWest Nile Virus Serological Reagents21 CFR § 866.3940Class II
Predicate DeviceFocus West Nile Virus IgM Capture ELISA (K031952)

Device Description

A Simple and Rapid Immunoassay for the Qualitative Detection of West Nile Virus IgM Antibodies in Human Serum or Plasma

Intended Use

STATus™ virus The Spectral West Nile test is a rapid immunochromatographic lateral flow assay that utilizes recombinant West Nile virus (WNV) antigen (E protein) for the qualitative detection of IgM antibodies to WNV in human serum or plasma (sodium heparin or sodium citrate). This test is for the presumptive laboratory diagnosis of West Nile virus infection in patients having signs and symptoms of meningoencephalitis. Positive results must be confirmed by the PRNT (Plaque Neutralization Reduction Test), or by using the current CDC guidelines for diagnosis of this disease. The Spectral test is not intended for point of care testing, home use or in screening blood donor samples.

{1}------------------------------------------------

Test Principle:

The Spectral WNV IgM STATus™ test (henceforth referred to as the Spectral device) employs solid-phase immunochromatographic assay technology to qualitatively detect the presence of WNV IgM antibodies in serum or plasma. When the specimen to be tested is dispensed into the sample well of the Spectral device, anti-WNV IgM in the sample will bind to the recombinant WNV antigen (envelop glycoprotein (E) of West Nile virus, NY99 strain) to form a tertiary complex with gold-labeled monoclonal murine antibody against flavivirus family glycoprotein E. This tertiary complex will migrate through reaction strip and be captured by goat anti-human IgM antibodies at the Test area. Excess, unreacted gold complex detector is captured by immobilized anti-mouse IgG antibodies at the Control area.

A visible pinkish-purple horizontal band will appear in the Test area within 15 minutes following the addition of a sample if the level of the WNV IgM antibodies in the human serum sample is above the cut-off level. A pinkish-purple band in the Control area indicates that the test is working properly and such a band must always appear, irrespective of the WNV IgM levels, in order for the test to be valid.

Test bandControl bandWNV IgM STATus Result
Band absentBand presentNegative
Band presentBand presentPositive*
Band presentBand absentInvalid Test**
Band absentBand absentInvalid Test**

Result Interpretation: Spectral WNV IgM STATus™ device

  • Consider West Nile Virus infection: WNV IgM levels are at or above the cut-off level.

** Invalid test. Test must be repeated with a new Spectral WNV IgM STATus™ device.

Expected Results:

Serum specimens from a total of 346 presumably healthy individuals (n=67) or non-WNV ailments (n=279) including 61 febrile patients were tested with the Spectral WNV IgM STATus™ test (henceforth referred to as the Spectral device) in a general population from endemic regions of USA and Canada. The samples were about equally distributed between male (47.1%) and female (52.9%) populations. Positive results were observed and determined to be positive due to heterophile antibodies interferences.

{2}------------------------------------------------

Age (Years)NegativePositive% Positive
0 to 9000.0% (0/0)
10 to 19800.0% (0/8)
20 to 293912.5% (1/40)
30 to 395611.7% (1/57)
40 to 495311.8% (1/54)
50 to 595600.0% (0/55)
60 to 694912.0% (1/50)
70 to 793912.5% (1/40)
80+4100.0% (0/41)
Overall34151.4% (5/346)

Prevalence of WNV IgM and Efficacy of Spectral Device in Endemic Population

Performance Characteristics:

Reproducibility:

Reproducibility studies were performed using 15 patient serum samples at 3 study sites over a three day period. Each day a new lot was used and a different operator was employed. The 15 member reproducibility panel consisted of 6 clinical specimens with a mean index value 25% above the cutoff, 6 clinical specimens with a mean index value 15% below the cutoff, and 3 clinical specimens with a mean index value 6 times the cutoff level of a commercially available WNV IgM Capture ELISA device (comparator device). All sites and operators produced the expected result for all panel members on every day of testing.

Method Comparison

Study sites 1-3: Negative agreement of Spectral device with comparator in non-flavivirus IgM samples:

Studies were performed at three sites on clinical serum specimens collected prospectively from patients with a variety of non-flavivirus ailments (rash, febrile, bronchitis, diarrhea, drug, neuropathy, acute coronary syndrome etc) and from endemic normal populations either at emergency departments of hospitals or at inhouse collection. Site 1 represents an endemic region at the South-Western State of America; sites 2 and 3 represent endemic regions of South Eastern Provinces of Canada. These specimens were tested at individual sites concomitantly with the Spectral and a commercially available comparator device. The results are tabulated as "Pos" (positive), or "Neg" (negative). Indeterminate results produced by the comparator device are listed as equivocal ("Eqv")

{3}------------------------------------------------

Study Site 1: Negative agreement of Spectral device with comparator in non-flavivirus IgM samples (n=160).

Comparator device
PosEqvNegTotal
Spectral DevicePos001**1
Neg02*157159*
Total02158160

Negative agreement with the comparator device = 157/158* 99.4% 96.5-99.9% (95% CD)

*Note: Two specimens that produced equivocal results with the comparator device were determined to have interference from crossreacting antibodies.

**One Spectral false positive result was observed and determined to be due to interference from crossreacting antibodies

Study Site 2: Negative agreement of Spectral device with comparator in non-flavivirus
IgM samples (n=114).
Comparator device
PosEqvNegTotal
Spectral DevicePos002*2
Neg0-112112
Total00114114

Negative agreement with the comparator device = = 112/114 98.2% 93.8-99.8% (95% CI)

*Two Spectral false positive results were observed and determined to be due to interference from crossreacting antibodies

igm samples (n=72).
Comparator device
PosEqvNegTotal
Spectral DevicePos1*012**
Neg007070
Total107172

Study Site 3: Negative agreement of Spectral device with comparator in non-flavivirus IgM samples (n=72)

Negative agreement with the comparator device = 70/71 98.6% 92.4-99.9% (95% CI)

*False positive due to cross reacting antibodies, as the sample remained positive with both devices when antigen was removed.

**Two Spectral false positive results were observed and determined to be due to interference from crossreacting antibodies.

{4}------------------------------------------------

All Sites (1-3): Negative agreement of Spectral device with comparator in non-flavivirus IgM samples (n= 346)

Comparator device
Spectral DevicePosEqvNegTotal
Pos1*045***
Neg02**339341*
Total12343346

Negative agreement with the comparator device = 339/343 98.8% 97.1-99.7% (95% CI)

*False positive due to cross reacting antibodies (e.g. RF & heterophiles), as the sample remained positive with both devices when antigen was removed.

** Two specimens that produced equivocal results with the comparator device were determined to have interference from crossreacting antibodies.

***Five Spectral false positive results were observed and determined to be due to interference from crossreacting antibodies

Study Site 4: A provincial health laboratory located at South Eastern Province of Canada assessed the performance of the Spectral device with randomized retrospective patient serum specimens (n=90) that were sent to the laboratory with suspected WNV infection based on clinical signs and symptoms. This batch of samples constituted 40 WNV PRNT confirmed and 50 CDC WNV IgM ELISA negative specimens. All specimens were randomized and tested blinded with the Spectral device.

Results Study Site 4: Spectral WNV IgM STATus™ test reactivity with WNV PRNT Positive/CDC WNV IgM ELISA positive and CDC WNV IgM ELISA negative specimens

Specimen Characterized by ReferenceAssaysPosNegTotalPos or NegAgreement95%CI
Scrological sensitivity (CDC WNV IgM and IgGELISA Positive and WNV PRNT Positive), n=403824095%83.1%-99.4%
Negative Agreement with presumptive CDCWNV IgM ELISA Negative, n=502485096%86.3%-99.5%

Study Site 5:

Spectral device reactivity was assessed at a public health laboratory located in a prairie Province at Central Canada using a banked panel of clinical serum specimens that were drawn from patients (n=79) with clinical symptomlogy consistent with the West Nile virus infection. These samples also included symptomatic acutely infected patients (PRNT confirmed, n= 24) exhibiting either febrile illness or neuro-invasive disease such as meningoencephalitis (n=5) and, others specimens from asymptomatic patients (n=25) that were sent to the

{5}------------------------------------------------

laboratory to detect previous West Nile virus exposures. All these samples were then randomized, blinded and tested with the Spectral device.

Results Study Site 5: Spectral WNV IgM STATus™ test reactivity with WNV PRNT
Positive/CDC WNV IgM Elisa positive and CDC WNV IgM Elisa negative specimens
Specimen Characterized by ReferenceAssaysSpectral WNV IgM STATus™ ResultsPos or NegAgreement95%CI
PosNegTotal
Serological sensitivity for acute PRNT Pos, CDCWNV IgM & IgG Elisa Pos (n=24)24024100%88.3%-100%
Serological sensitivity for late PRNT Pos, CDCWNV IgM & IgG Elisa Pos (n=25)2052580%59.3%-93.2%
Negative agreement with presumptive CDC WNVIgM Elisa Neg (n=30)1293097%82.8%-99.9%

Interfering Substances

The following constituents of human blood were tested to ensure that they did not interfere with the performance of the Spectral device. The concentration of the serum proteins or lipids shown in Table below were added to positive and negative serum specimens and tested with the Spectral device. No interference up to the concentrations listed below was observed.

Potential Interferents and Concentrations

Potential Interfering materialConcentration
Human serum albumin16 g/dL
Bilirubin (unconj.)60 mg/dL
Hemoglobin25 g/dL
Triglycerides1.3 g/dL
Human IgM300 mg/dL

IgM Specificity:

The IgM specificity of the Spectral WNV IgM test was assessed in 14 paired positive samples by treating with 5mM dithiothreitol (DTT) solution for one hr at room temperature (22-24℃). All 14 treated samples produced negative results following the treatment with DTT and their respective untreated paired samples remained positive, indicating the presence of IgM class antibodies in the positive samples (9).

{6}------------------------------------------------

Cross Reactivity:

Cross- reactivity studies with sera that were sero-positive to other potentially cross-reactive pathogens were conducted at a Public Health Laboratory in the North Eastern USA (site 1). Three out of seven Dengue post infection samples were tested at Public Health Laboratory at Mid-West Canada (site 3); 29 out of 33 RF positive sera and 23 out of 28 ANA positive sera were tested in-house (site 2). Cross reactivity with Eastern Equine Encephalitis specimens was studied at a CDC laboratory at Mid-West USA (site 4). Results of these studies are tabulated below.

Potentially Cross-Reactive AgentsSpectral WNV IgM STATUST™ test Results
SiteNumberofSamplesNegativePositive% Cross-ReactiveSamples
Herpes Simplex Virus 1and/or 21109110%
CytomegalovirusPolyvalent Test, allreactive15500%
Syphilis (trep Reactive)1101000%
Epstein Barr Virus, IgGand IgM reactive15500%
ANA (anti-nuclearantibody)1,228171139%
Rheumatoid Factor1,233231030%
HIV(humanimmunodeficiency virus)154120%
Japanese EncephalitisVirus (vaccinerecipients)15500%
Saint Louis EncephalitisVirus. SLE IgM (4/5)and IgG (5/5) reactive,CDC15500%
Dengue Virus (5 recentand 2 post infection)-DEN IgM & IgG1,3109110%

Results: Cross Reactivity Results

K052519

Page 7 of 8

{7}------------------------------------------------

K052518

510(k) Summary of Safety and Effectiveness Spectral West Nile Virus IgM STATus™ Test Prepared November 29, 2006 Page 8 of 8

reactive, CDC
Hepatitis B Virus(HbsAg)15500%
Hepatitis C Virus(HCAb)15500%
California EncephalitisPRNT reactive as JC or LAX16600%
Legionella, Polyvalent Test, all reactive15500%
Yellow Fever (vaccine recipients)154120%
E. coli infection (culture confirmed)24400%
Eastern Equine Encephalitis42200%

Summary

The existing WNV IgM tests marketed in North America are all ELISA based tests (e.g. Focus Diagnostics West Nile Virus IgM Capture ELISA, InBios' West Nile Detect™ IgM Capture ELISA, Panbio WNV IgM capture ELISA, etc). These assays are multi-step procedures that require instrumentation to read the tests, need subsequent calculation for the interpretation of the test results and take more nece bussequent carbandi.or fertral WNV IgM STATus™ test, described here is a rapid format test that produces the results visibly in 15 minutes and is shown to be substantially equivalent to the commercially available device. Data presented above describes the assay performance characteristics and, safety and effectiveness of the Spectral rapid format device.

Conclusion:

The Spectral WNV IgM STATus™ test is a safe and effective device for the qualitative determination of WNV IgM antibodies in human serum and plasma as proven through performance characteristics and methods comparison with the predicate device and confirmed through the testing of both normal and clinical samples.

{8}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, with three horizontal lines representing the feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Dr. Nisar A. Shaikh Senior VP, Corporate Quality & Scientific Affairs Spectral Diagnostics, Inc. 135 The West Mall Toronto, Ontario Canada M9C 1C2

NOV 3 U 2006

Re: K052519

Trade/Device Name: Spectral West Nile Virus IgM STATus™ Test Regulation Number: 21 CFR 866.3940 Regulation Name: West Nile Virus Serological Reagents Regulatory Class: Class II Product Code: NOP Dated: November 10. 2006 Received: November 14, 2006

Dear Dr. Shaikh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{9}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to 10gally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll frem and (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sally, anton

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indication for Use

510(k) Number (if known): K052519

Device Name: __ Spectral WNV IgM STATus™ test

Indications for Use:

The Spectral West Nile virus IgM STATus™ test is a rapid immunochromatographic lateral flow assay that utilizes recombinant West Nile virus (WNV) antigen (E glycoprotein) for the qualitative detection of IgM antibodies to WNV in human serum or plasma (sodium heparin or sodium citrate). This test is for the presumptive laboratory diagnosis of West Nile virus infection in patients having signs and symptoms of meningoencephalitis. Positive results must be confirmed by the PRNT (Plaque Neutralization Reduction Test), or by using the current CDC guidelines for diagnosis of this disease. The Spectral test is not intended for point of care testing, home use or in screening blood donor samples.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off
---------------------

Office of In Vitro Diagnostic Device
Evaluation and Safety

510(k)K052519
-----------------

§ 866.3940 West Nile virus serological reagents.

(a)
Identification. West Nile virus serological reagents are devices that consist of antigens and antisera for the detection of anti-West Nile virus IgM antibodies, in human serum, from individuals who have signs and symptoms consistent with viral meningitis/encephalitis. The detection aids in the clinical laboratory diagnosis of viral meningitis/encephalitis caused by West Nile virus.(b)
Classification. Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus.” See § 866.1(e) for the availability of this guidance document.