K Number
K052219

Validate with FDA (Live)

Device Name
CRONO GO.
Manufacturer
Date Cleared
2005-09-13

(29 days)

Product Code
Regulation Number
880.5725
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

the Crono go has been designed for use in subcutaneous infusion of prescribed liquid medicines.

Device Description

Crono go is an ambulatory syringe infusion pump intended for the controlled administration of liquids into the patients. This means that the pump can be worn during the infusion. Crono go combines high technology with innovative design. Thanks to the small size of the pump Crono go can administer the drug any time during the day without interrupting daily life or leisure activities. Crono go pump uses dedicated syringes which are available in two sizes: 10 and 20 ml. Crono go has a particular mechanism which pushes directly on the rubber syringe piston, which guarantees a significant thrust force and high accuracy of administration. Additional bolus doses can be easily administered by pressing a dedicated button, and the administered bolus dose is shown on the display screen. In case of occlusion, an innovative infusion control system makes it possible to proceed with the infusion automatically and, after the occlusion is eliminated, to complete it. Crono go is fitted with a liquid crystal display which shows the time it takes to complete the delivery, the syringe size and the battery charge status.

AI/ML Overview

Here's an analysis of the provided text to extract the requested information about device acceptance criteria and study details.

Important Note: The provided text is a 510(k) summary and FDA clearance letter for a medical device (Crono go ambulatory infusion pump). This type of document is heavily focused on demonstrating substantial equivalence to a predicate device, rather than presenting a novel clinical study with a defined set of acceptance criteria in the traditional academic sense. Therefore, some of the requested categories (e.g., ground truth for training data, MRMC studies, number of experts for ground truth) are not applicable (N/A) or cannot be explicitly found in this type of regulatory submission. The "acceptance criteria" here are aligned with the performance specifications demonstrated to show equivalence.


Acceptance Criteria and Device Performance Study

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Predicate Device K013840)Crono go (New Device) Reported Performance
Intended UseSubcutaneous infusion of prescribed liquid medicines.SAME
Size3" x 1.85" x 1.14" (77 x 48 x 29 mm)SAME
Weight4.0 oz (115 g) (battery included)SAME
BatteryPower Source Lithium battery (3V) of the 123 A type.SAME
Infusion per impulse22 μlSAME
Flow rate accuracy+/- 2%+/- 2% (Reported as "SAME" as predicate)
Max. Occlusion pressure4.5 bar +/-1 bar4.5 bar +/-1 bar (Reported as "SAME" as predicate)
Capacity10 or 20 mlSAME

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document refers to "bench, EMC, and user testing" but does not give specific numbers of units tested or duration of tests.
  • Data Provenance: The device manufacturer, Canè S.r.l., is based in Rivoli-Cascine Vica (Torino), Italy. It can be inferred that testing was conducted by or under the supervision of the manufacturer, likely in Italy. The study is retrospective in the sense that it's comparing a new device against an existing, cleared predicate device's specifications. It is, however, prospective testing of the "Crono go" device itself.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • N/A. This information is not relevant or typically provided in a 510(k) submission for an infusion pump. The "ground truth" for performance is based on engineering specifications and direct measurement, not expert interpretation.

4. Adjudication Method for the Test Set

  • N/A. Not applicable, as expert adjudication is not used for mechanical device performance testing. Performance is measured against predefined engineering specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This is not an AI-assisted diagnostic device, but rather a standalone medical pump. Therefore, MRMC studies and AI assistance are not applicable.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, implicitly. The device itself is a standalone infusion pump. The performance characteristics (flow rate accuracy, occlusion pressure) are measured independently of human interaction during the actual infusion process, reflecting the device's inherent functional capabilities. The user testing mentioned would pertain to usability and safety aspects of human interaction with the device, not its core mechanical performance.

7. The Type of Ground Truth Used

  • Engineering Specifications and Direct Measurement: The ground truth for the device's performance characteristics (e.g., flow rate accuracy, occlusion pressure) is based on established engineering principles, metrology, and direct measurement against predetermined acceptable ranges. For comparison, the "ground truth" is also the established performance of the predicate device.

8. The Sample Size for the Training Set

  • N/A. This device does not involve a "training set" in the context of machine learning or AI models. It's a mechanical device whose performance is engineered and then verified.

9. How the Ground Truth for the Training Set was Established

  • N/A. As above, no training set is applicable.

{0}------------------------------------------------

SEP 1 3 2005

Image /page/0/Picture/1 description: The image shows the word "CAINE" in bold, black letters. To the left of the word is a symbol that resembles a plus sign, but with a more geometric and stylized design. The letters are slightly tilted to the right, giving the impression of movement or a dynamic feel.

Canė S.r.l. società unipersonale

P.IVA - Reg. Impr. TO: 04384410017

R.Ε.Α. ΤΟ Ν. 629783

Cap. Soc. Euro 10.400 i.v.

Image /page/0/Picture/4 description: The image shows a logo with the text "CERMET" in bold letters. Below that, there is a smaller box with the text "SISTEMA DI GESTIONE CERTIFICATO". Further down, the text "REG. N. 3506 UNI EN ISO 9001:200 ISO 13485:2002" is displayed, indicating a certification or registration number along with ISO standards.

EXHIBIT 2

CANÈ S.r.I. Via Pavia, 105/1 10090 Rivoli-Cascine Vica (Torino) Italy Tel.: ++39-011-957.48.72 Fax ++39-011-959.88.80 Contact: Mario Cané, President July 4, 2005

510(k) Summary

1. Identification of the Device:

Proprietary - Trade Name: Crono go. Classification Name: 80 FRN. Common/Usual Name: Ambulatory Infusion Pump.

2. Equivalent legally marketed devices

2: Equivalient regulty marriction and design to the Crono cleared under 510(k) number K041414.

3. Indications for Use (intended use)

or manuallence on ogo infusion devices have been designed for use in subcutaneous infusion of prescribed liquid medicines.

4. Description of the Device

Cane s.r.l., a company that specializes in the production of ambulatory pumps, has now produced Same of miration of compact pumps: Crono, a perfect combination of high technology and innovative design.

Crono go is an ambulatory syringe infusion pump intended for the controlled administration of liquids into the patients. This means that the pump can be worn during the infusion.

Crono go combines high technology with innovative design. Thanks to the small size of the pump Orono go sombinister the drug any time during the day without interrupting daily life or leisure activities.

Crono go pump uses dedicated syringes which are available in two sizes: 10 and 20 ml.

Crono go has a particular mechanism which pushes directly on the rubber syringe piston, which Orono go has a particular mestignificant thrust force and high accuracy of administration.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the word "CAINE" in a stylized font. The letters are bold and have a unique design, with horizontal lines running through them. To the left of the word is a symbol that resembles a plus sign or a cross, also in a bold and stylized design. The overall image has a modern and geometric aesthetic.

Canė S.r.l. società unipersonale
P.IVA - Reg. Impr. TO: 04384410017
R.E.A. ΤΟ Ν. 629783
Cap. Soc. Euro 10.400 i.v.

Image /page/1/Picture/3 description: The image shows a logo with the word "SINCERT" at the top. Below that is a logo with the word "CERMET" inside of it. Underneath the logo, the words "SISTEMA DI GESTIONE CERTIFICATO" are written. The text "REG. N. 3506, UNI EN ISO 9001:2000, ISO 13485:2002" is at the bottom of the image.

Additional bolus doses can be easily administered by pressing a dedicated button, and the administered bolus dose is shown on the display screen.

In case of occlusion, an innovative infusion control system makes it possible to proceed with the infusion automatically and, after the occlusion is eliminated, to complete it.

Crono go is fitted with a liquid crisfal display which shows the time it takes to complete the delivery, the syringe size and the battery charge status.

5. Safety and Effectiveness, comparison to predicate device.

The results of bench, EMC, and user testing indicates that the new device is as safe and effective as the predicate device.

CharacteristicCrono go - K013840Crono go
Intended UseSubcutaneous infusion of prescribed liquid medicines.SAME
Physical characteristicsCrono go - K013840Crono go
Size3" x 1.85" x 1.14" (77 x 48 x 29 mm).SAME
Weight4.0 oz (115 g) (battery included).SAME
BatteryPower Source Lithium battery (3V) of the 123 A type.SAME
Infusion per impulse22 μl.SAME
Flow rate accuracy+/-2%.SAME
Max. Occlusion pressure4.5 bar +/-1 bar.SAME
Capacity10 or 20 ml.SAME

6. Substantial Equivalence Chart

7. Conclusion

After analyzing both bench and user testing data, it is the conclusion of CANÈ S.r.l. that Crono pumps are as safe and effective as the predicate device, has few software differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the caduceus. The seal is black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 1 3 2005

Mr. Carlo Musso Quality Manager Cane S.R.L. Via Pavia, 105/I 10090 Rivoli-Cascine Vica (Torino) ITALY

Re: K052219

Trade/Device Name: Crono Go Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: July 15, 2005 Received: August 15, 2005

Dear Mr. Musso:

We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your bocaon by (1) person the device is substantially equivalent (for the referenced above and have determined the arrest and the marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mersuale collimence prior to ritay 20, 1970, accordance with the provisions of Amendinents, on to devroes that have of the Act (Act) that do not require approval of a prematic the Fedelar Pood, Drug, und Conners , therefore, market the device, subject to the general approval application (1 Mir). The general controls provisions of the Act include controls provisions of the Fea. "The garries, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III II your device is classified (see above) into exontrols. Existing major regulations affecting (FMA), it may be subject to back addinf Federal Regulations, Title 21, Parts 800 to 898. In your device can oc found in the Seas nouncements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Musso

Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA S Issualled of a baseaux and as the other requirements
mean that FDA has made a determination that your device the Federal agencies mean that FDA has made a decemmance and regulations administered by other Federal agencies.
of the Act or any Federal statutes and regulations administered by seciences and of the Act of any Federal statutes and regulaents, including, but not limited to: registration
You must comply with all the Act's requirements, and manufacturing protice You must comply with an the Fee s requirements and 801); good manufacturing practices
and listing (21 CFR Part 807); labeling (21 CFR Part 801); gED Box 800); and i and listing (21 CFR I art 607), lacemig (21 CFR Part 820); and if requirements as set form in the quality by control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 5 10(k) I his letter will anow you to begin maticemly your substantial equivalence of your device to a premarket nothleation. The I Dr Imaling of basemans of your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Patt 801), If you desire specific auvice for your device at (240) 276-0115. Also, please note the regulation please comact the Other or Comphanes are (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may outain other general miormational and Consumer Assistance at its toll-free Division of Division of (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with a cross symbol on the left and the word "CAINE" on the right. The cross symbol is made up of thick black lines, and the word "CAINE" is written in bold, sans-serif letters. The logo appears to be for a company or organization with the name "CAINE".

Cané S.r.l. società unipersonale
P.IVA - Reg. Impr. TO: 04384410017
R.Ε.Α. ΤΟ Ν. 629783
Cap. Soc. Euro 10.400 ί.ν.

Image /page/4/Picture/2 description: The image shows a logo with the text "CERMET" in bold letters. Below that, it says "SISTEMA DI GESTIONE CERTIFICATO". Further down, it reads "REG. N. 3506" followed by "UNI EN ISO 9001:2000" and "ISO 13485:2002". The logo includes a stylized symbol at the top.

SINGERT

j) Indications for Use

Kos 2219 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Crono go.

Indications For Use: the Crono go has been designed for use in subcutaneous infusion of prescribed liquid medicines.

Prescription Use X . (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rihd C Chapman for ACw

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Denta

510(k) Number:_

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).