AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VNUS RFS and RFSFlex devices are intended for use in vessel and tissue coagulation including: Treatment of incompetent (i.e., refluxing) perforator and tributary veins.

Device Description

The VNUS RFS family of devices is a bipolar, high frequency electrosurgical system designed for use in general surgical procedures where blood vessel and tissue coagulation is desired. These devices are compatible with separately cleared Radiofrequency (RF) Generator, and the Instrument Cable manufactured by VNUS Medical Technologies. The VNUS RFS family of devices are sterile disposable devices intended for a singleuse only. The device's function is to deliver bipolar RF energy to the desired treatment site and relay temperature and other feedback to the RF Generator. The disposable device is available in 2 bipolar configurations and multiple lengths for selection by the physician based on preference for method of vessel access, location and length of the vessel to be treated. This submission reflects a name change, minor modification of the indication statement (more specific inclusion of perforator and tributary veins), and minor design changes.

AI/ML Overview

The provided text indicates that the VNUS® RFS and RFSFlex Devices were evaluated for substantial equivalence to predicate devices (VNUS® Closure® System and VNUS® Vessel and Tissue Coagulation System) rather than undergoing a de novo clinical study with specific acceptance criteria and performance metrics. Therefore, many of the requested details about acceptance criteria, study design, and ground truth establishment are not present in the given document.

Here's a breakdown of the information that is available based on the provided text, and where gaps exist:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Not explicitly defined as numerical acceptance criteria for a new device study. The evaluation was for "substantial equivalence" to predicate devices.Results of in-vitro testing demonstrate that the VNUS RFS and RFSFlex devices are safe and effective for their intended function. Materials are biocompatible.
Implicit Criteria for Substantial Equivalence:
- Similar intended useIntended for "vessel and tissue coagulation including: Treatment of incompetent (i.e., refluxing) perforator and tributary veins." This is comparable to predicate device indications ("coagulation of blood vessels in patients with superficial vein reflux" and "vessel and tissue coagulation").
- Similar technological characteristics and principles of operationRelies on delivery of RF energy, similar to predicate devices. Acknowledged potential advantages of bipolar RF over monopolar (no grounding pads, minimized damage to adjacent tissue, lower voltage/power).
- Safety and Efficacy"Results of in-vitro testing demonstrate that the VNUS RFS and RFSFlex devices are safe and effective for their intended function." "The RF ablation/coagulation of blood vessels (i.e., VNUS Closure) has a well-established history of safe an efficacious use in over 100,000 procedures."
- Biocompatibility"The materials used in the VNUS RFS and RFSFlex Devices have been shown to be biocompatible."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size for test set: Not applicable. The submission relies on "in-vitro testing" and comparison to the "well-established history" of predicate devices, rather than a clinical test set with a defined sample size for subjects.
  • Data provenance: Not explicitly stated for specific test data. The submission refers to "in-vitro testing." The predicate device's history (over 100,000 procedures) is cited as evidence of efficacy, but this is a general statement about the technology, not specific data from the device under review.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. There is no mention of an expert panel establishing ground truth for a clinical test set in the context of this 510(k) submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. No clinical test set requiring adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a surgical instrument (bipolar electrosurgical system), not an AI diagnostic tool that would involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a surgical instrument, not an algorithm. Its "standalone" performance would relate to its physical and energetic output, as confirmed by in-vitro testing.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the "in-vitro testing," the ground truth would likely be defined by engineering specifications, physical/chemical measurements, and potentially histological analysis of treated tissue in laboratory settings. The document does not specify the exact nature of this "ground truth."
  • For the claim of "well-established history of safe an efficacious use" for RF ablation/coagulation, the ground truth would be historical clinical outcomes data, morbidity, and mortality statistics associated with the predicate technology over thousands of procedures.

8. The sample size for the training set

  • Not applicable. This is not an AI/machine learning device that would require a "training set."

9. How the ground truth for the training set was established

  • Not applicable. There is no training set for this type of device.

In summary, this document is a 510(k) submission for "substantial equivalence," not a report of a new clinical efficacy study. Therefore, the details requested about acceptance criteria, clinical test sets, expert ground truth, and AI performance are largely absent because they are not typically part of a substantial equivalence determination for this class of medical device. The safety and effectiveness are primarily supported by in-vitro testing and the established history of the predicate devices.

{0}------------------------------------------------

VNUS® RFS and RFSFlex Devices

K052003

Summary of Safety and Effectiveness

  • Determination of Substantial Equivalence A. VNUS RFS and RFSFlex devices

B. Common Name

Bipolar Electrosurgical Instrument

C. Predicate Device(s)

VNUS® Closure® System (K982816, K003092, and K030557); and VNUS® Vessel and Tissue Coagulation System (proprietary name: TBD) (K033547)

D. Device Description

The VNUS RFS family of devices is a bipolar, high frequency electrosurgical system designed for use in general surgical procedures where blood vessel and tissue coagulation is desired. These devices are compatible with separately cleared Radiofrequency (RF) Generator, and the Instrument Cable manufactured by VNUS Medical Technologies.

The VNUS RFS family of devices are sterile disposable devices intended for a singleuse only. The device's function is to deliver bipolar RF energy to the desired treatment site and relay temperature and other feedback to the RF Generator. The disposable device is available in 2 bipolar configurations and multiple lengths for selection by the physician based on preference for method of vessel access, location and length of the vessel to be treated. This submission reflects a name change, minor modification of the indication statement (more specific inclusion of perforator and tributary veins), and minor design changes.

ய் Intended Use

The VNUS RFS and RFSFlex Devices are intended for use in vessel and tissue coagulation including:

  • Treatment of incompetent (i.e., refluxing) perforator and tributary veins .

Intended Use of Predicate Devices F.

The specified predicate devices are indicated for "coagulation of blood vessels in patients with superficial vein reflux" (VNUS Closure System) and "vessel and tissue coagulation" (VNUS Vessel and Tissue Coagulation System).

G. Technological Comparison

RF energy has been widely used in electrosurgical equipment for many years. The safety and efficacy of such devices has been well established for a variety of intended uses. The use of bipolar RF energy delivery has potential advantages over monopolar systems. No grounding pads are required, and the potential for damage to adjacent tissue is minimized, as the patient is no longer the return path for electrical current. The efficiency of bioolar RF energy delivery allows systems such as the VNUS RF Generator to be used at lower voltage and power settings as compared with monopolar systems.

{1}------------------------------------------------

Page 2 of (2)

The technological characteristics and principals of operation of the VNUS RFS family of devices are substantially equivalent to the noted predicate devices relv on the delivery of RF energy to achieve their intended use.

Discussion of Clinical/Non-Clinical Tests and Conclusions H.

Performance

Results of in-vitro testing demonstrate that the VNUS RFS and RFSFlex devices are safe and effective for their intended function.

Biocompatibility

The materials used in the VNUS RFS and RFSFlex Devices have been shown to be biocompatible.

l. Summary of Safety and Effectiveness

Based upon the intended use, design, materials, function, comparison with currently marketed devices and the non-clinical testing it is concluded that the VNUS RFS and RFSFlex Devices are substantially equivalent to the noted predicate devices. The RF ablation/coagulation of blood vessels (i.e., VNUS Closure) has a well-established history of safe an efficacious use in over 100,000 procedures.

Signature

Brady Esch . Directory, R&D VNUS Medical Technologies, Inc.

{2}------------------------------------------------

Public Health Service

0C1 1 1 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

VNUS Medical Technologies, Inc. c/o Sam Nanavati 750 Chesapeake Drive Redwood City, California 94063

Re: K052003

Trade/Device Name: VNUS® RFS and RFS Flex Devices Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: July 22, 2005 Received: July 25, 2005

Dear Mr. Nanavati:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2- Sam Nanavati

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Harlow Annett
to!

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Device Name: Derices VNUS® RFS and RFSFlex

510(k) Number (if known):

): K052003

Indications for Use:

The VNUS RFS and RFSFlex devices are intended for use in vessel and tissue coagulation including:

Treatment of incompetent (i.e., refluxing) perforator and tributary veins .

Over the Counter Use: (Per 21 CFR 801.109)

or

Prescription Use:

(Please do not write below this line - continue on another Page if needed)

CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

Caitano Pereira

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K052005

. - -

Premarket Notification [510(K)] Submission

Page 42

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.