K Number
K051660
Date Cleared
2005-08-04

(43 days)

Product Code
Regulation Number
880.6870
Panel
HO
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DDS 7000 is a countertop convective dry heat sterilizer designed for use in Healthcare facilities for the sterilization of un-bagged dental instruments that can withstand exposure temperatures up to 216°C (420° F). There is a single software-controlled cycle that includes a warm-up and cool down phase in addition to the 3-minute sterilization (exposure) phase. Sterilize (exposure) starts when the RTD/control temperature reaches the operating temperature set point of 190°C (374° F). Complete cycle times average 34-44 minutes and are load dependent. The DDS 7000 achieves optimum results by uniform load distribution utilizing a dedicated rack and tray system as detailed below.

  1. Maximum recommended loads validated for the DDS 7000-115 sterilizer: -
    a) Maximum load for a four (4)-rack configuration is four (4) plier racks each holding nine (9) orthodontic pliers each for a total of thirty-six (36) pliers. Other maximum load configuration options are: 32 pliers on eight mini vertical racks; or 20 pliers and hand instruments on four Combo Racks -(OR)-
    b) Maximum load for a full tray configuration is one (1) Horizontal Tray each holding up to six (6) orthodontic pliers or up to (12) single-handle hand instruments. Other maximum load combinations include: (5) orthodontic pliers with up to (2) single-handle hand instruments; or (4) orthodontic pliers with up to (4) single-handle hand instruments; or (3) orthodontic pliers with up to (6) single-handle hand instruments, or (2) Orthodontic pliers with up to (8) single-handle hand instruments; or (1) Orthodontic pliers with up to (10) single-handle hand instruments. - (OR)-
    c) Maximum load for the one (1) half tray and two (2) racks is one (1) Half Tray holding three (3) Orthodontic pliers or six (6) single-handle hand instruments in combination with two (2) plier racks capable of holding nine (9) pliers each for a total of eighteen (18) pliers on plier racks. Other maximum load combinations include: (2) orthodontic pliers with up to (2) single-handle hand instruments in combination with two (2) plier racks capable of holding nine (9) pliers each for a total of eighteen (18) pliers on plier racks; or (1) orthodontic pliers with up to (4) singlehandle hand instruments in combination with two (2) plier racks capable of holding nine (9) pliers each for a total of eighteen (18) pliers on plier racks.
Device Description

The DDS 7000, like the DDS 5000 is a convective dry heat batch sterilizer capable of sterilizing un-bagged dental instruments and cooling them for immediate use. It is designed for use with a dedicated rack system that provides for uniform loading to achieve optimum load distribution.
The DDS 7000 is equipped with a replaceable HEPA filter to comply with ANSI/AAMI ST50: 2004 cooling requirements. The HEPA filter has a filtration efficiency of 99.7% for 0.3-micron particles.
The DDS 7000 is a software-controlled device that allows for the user to preprogram options such as display units (Celsius or Fahrenheit), audible alarm enabled/disabled and safety interlock enabled/disabled The DDS 7000 monitors the entire cycle and will provide diagnostic error codes in the event of any failure or disruption to the sterilization cycle. A printable log is available for the most recently performed cycle that includes pass/fail notification and error codes if applicable

AI/ML Overview

The Dentronix "DDS 7000" Rapid Dry Heat Sterilizer Sterilization System is designed for the dry heat sterilization of un-bagged dental instruments that can withstand exposure temperatures up to 216℃ (420° F).

Here's an analysis of its acceptance criteria and the study proving it:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Implied)Reported Device Performance
Achieve Sterility Assurance Level (SAL) of 10^-6Half cycle kills result in an overall 12-log reduction of spores, thereby producing a 10^-6 SAL.
Sterilize un-bagged dental instrumentsDemonstrated to sterilize un-bagged dental instruments.
Withstand exposure temperatures up to 216°C (420°F)Instruments capable of withstanding these temperatures are intended for use. The sterilizer operates at 190°C (374°F).
Complete sterilization cycle (warm-up, 3-minute exposure, cool down)Complete cycle times average 34-44 minutes, load dependent, including warm-up, 3-minute sterilization at 190°C (374°F), and cool down.
Uniform load distributionAchieved through a dedicated rack and tray system.
Effective across various validated load configurationsTested and validated with specified maximum load configurations for pliers and hand instruments.
Temperature monitoring and error detectionYes (LED Display, software monitors cycle parameters, provides diagnostic error codes and audible alarm).

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: The document mentions "Performance testing of the DDS 7000 was performed using Bacillus atrophaeus spore strips and inoculated tools." While the specific number of spore strips or inoculated tools used is not quantified, it implies a sufficient sample was used to demonstrate a 12-log reduction of spores. Full, partial, and empty chambers were also included in temperature studies.
  • Data Provenance: The study was conducted by Coltene/Whaledent Incorporated for the purpose of a 510(k) premarket notification to the FDA. The data provenance is internal testing performed by the manufacturer in the USA (Cuyahoga Falls, Ohio). It is a prospective study as it involves specific performance testing of the device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This type of device (sterilizer) relies on established microbiological and engineering principles rather than clinical expert consensus for its "ground truth."

  • Experts: Not applicable in the sense of medical image interpretation or diagnosis. The "ground truth" for sterilization efficacy is typically defined by microbiologists who assess the viability of biological indicators (spore strips) after exposure to the sterilization cycle.
  • Qualifications: While not explicitly stated, it is assumed that personnel with expertise in microbiology, sterilization science, and engineering were involved in designing, executing, and analyzing the performance tests.

4. Adjudication Method for the Test Set:

Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving human interpretation (e.g., radiology reads where multiple experts evaluate cases and discrepancies are resolved). For sterilizer performance, the assessment is objective, based on laboratory results (e.g., no growth of Bacillus atrophaeus spores).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where the performance of human readers, with and without AI assistance, is compared. The DDS 7000 is a sterilization device, not a diagnostic imaging device.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance):

Yes, a standalone performance study was done. The performance testing described (using Bacillus atrophaeus spore strips and inoculated tools) is a direct evaluation of the device's ability to sterilize, independent of human interaction during the sterilization cycle itself (though human operation is required to initiate the cycle). The study directly assesses the sterilizer's efficacy.

7. Type of Ground Truth Used:

The ground truth used is microbiological evidence. Specifically, the absence or reduction of viable microorganisms (spores of Bacillus atrophaeus) on biological indicators and inoculated tools after the sterilization cycle. This is an objective measure of sterilization efficacy.

8. Sample Size for the Training Set:

Not applicable. The DDS 7000 is a hardware device (dry heat sterilizer) with software controls, not a machine learning or AI algorithm that requires a "training set" in the conventional sense. The "software validation" mentioned refers to ensuring the software correctly controls the device and functions as intended, not to training a predictive model.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no "training set" for an AI or machine learning model in this context. The "software" validation principles refer to ensuring appropriate design, development, and testing of the control software to meet functional and safety requirements, as per FDA guidance documents.

§ 880.6870 Dry-heat sterilizer.

(a)
Identification. A dry-heat sterilizer is a device that is intended for use by a health care provider to sterilize medical products by means of dry heat.(b)
Classification. Class II (performance standards).