K Number
K051509

Validate with FDA (Live)

Manufacturer
Date Cleared
2005-07-27

(50 days)

Product Code
Regulation Number
878.4350
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ERBE ERBOKRYO® CA Cryosurgical Unit and accessories are intended for devitalization (destruction) of tissue during surgical procedures by the application of extreme cold and for removal of foreign bodies, mucous plugs, necrotic tissue, and tissue biopsy by cryoadhesion.

Clinical Indications for Cryosurgery

GynecologyCervical Erosions, Cervical Polyps, Condylomas, Chronic Cervicitis, VulvaCarcinoma (palliative), Neoplasia
DermatologyLeukoplakia, Fibroma, Condylomas, Basal Cell Carcinoma, Skin Tumor(palliative), Warts, Naephus
OphthalmologyAblatio Retinae, Glaucoma, Lid Tumor
ENTLeukoplakia, Inoperable Tumor (palliative), Laryngeal Papilloma, Fibroma,Angioma, Haemangioma
Thoracic SurgeryPost-Operative
UrologyProstate Tumor (palliative), Condylomas, Penile Tumor (palliative)
PhlebologyVaricose Veins of the Lower Limbs (Cryo Stripping)
ProctologyHemorrhoids (1st and 2nd Degree), Pari-Anal Condylomas, Anal Tumor(palliative), Rectal Tumor (palliative), Acute Anal Fissures
PulmonologyTumors, Granulomatous Tissue, Malignant Lesions (palliative)
PneumologyTracheobronchial Stenoses (Cryorecanalization)
Device Description

The ERBE USA, Inc. ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories is used to apply extreme cold to tissue during surgical procedures. It uses the Jouleused to apply extreme oold to toose aumy and Thompson principle where probe tip and the surrounding tissue. a fapid drop in temperature and frous Oxide) or CO2 (Carbon Dioxide) gas. The The unit is ascu with older 1120 (11) in the cryo probes are reusable and urnt and accessones are provided new surgical probes are provided in the Notes for Use.

AI/ML Overview

The provided 510(k) summary (K051509) for the ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories does not contain specific acceptance criteria or a detailed clinical study report with performance metrics in the format requested.

The document discusses the substantial equivalence of the modified device to its predicate (K934261), highlighting similarities in basic technology and indications for use, and differences related to materials and an added internal filter. The key part relevant to your request is the expansion of indications for use to include "removal of foreign bodies, mucous plugs, necrotic tissue, and tissue biopsy by cryoadhesion" and "Tracheobronchial Stenoses (Cryorecanalization) in Pneumology."

The 510(k) process primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than requiring new, full-scale clinical trials with pre-defined acceptance criteria and performance metrics typically seen in a PMA or de novo submission.

However, I can extract the information regarding how the expanded indications were supported:

1. Table of Acceptance Criteria and Reported Device Performance:

  • No specific quantitative acceptance criteria or device performance metrics (e.g., sensitivity, specificity, accuracy, success rates in a clinical trial) are defined or reported in the provided text for the expanded indications.
  • The document implies that the device "can be used safely and effectively" for the new indications based on existing studies involving "cryosurgical units" and physician consensus.

2. Sample Size Used for the Test Set and Data Provenance:

  • Not explicitly stated in terms of a formal "test set" for performance evaluation.
  • The document refers to "all available/known studies involving 'cryosurgical units' in conjunction with the expanded indications for use." These studies provided "documentation provided demonstrates that the ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories can be used safely and effectively to remove foreign bodies... and tissue biopsy by cryoadhesion."
  • For Tracheobronchial Stenoses, it states, "The documentation also shows that physicians consider cryotherapy to be a safe procedure for the treatment of bronchial obstruction in Pneumology. Pneumologists in a number of change cases, and overall with patients with malignant tumors that cryotherapy is a good palliative treatment in these cases, and is suitable for patients with obstruction >50% (or not)."
  • Data Provenance: Implied to be from existing scientific literature and clinical experience ("available/known studies," "physicians consider," "Pneumologists in a number of change cases"). No specific country of origin is mentioned, nor is it explicitly labeled as retrospective or prospective for a single study, but rather a review of existing evidence.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

  • Not specified as a formal "ground truth" establishment process for a specific test set.
  • The text mentions "physicians" and "Pneumologists" without specifying their number, specific qualifications (e.g., years of experience, board certification), or their role in establishing a ground truth for a test set. This likely refers to their general clinical consensus and published findings.

4. Adjudication Method for the Test Set:

  • Not applicable / Not specified. There is no mention of a formal adjudication method (like 2+1, 3+1) for a test set, as no dedicated clinical test set with independent expert adjudication is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study is not mentioned. The document does not describe any study comparing human readers with and without AI assistance for this cryosurgical unit.

6. Standalone (Algorithm Only) Performance Study:

  • No. This device is a cryosurgical unit, not an AI algorithm. Therefore, "standalone (algorithm only)" performance is not applicable. The device's performance is inherently linked to its use by a human operator.

7. Type of Ground Truth Used:

  • For the expanded indications, the "ground truth" (or supporting evidence) appears to be based on:
    • Expert consensus/Clinical Opinion: "physicians consider cryotherapy to be a safe procedure," "Pneumologists in a number of change cases."
    • Outcomes Data (implied): "Patients tolerate the procedure well and show improvement in symptoms," "good palliative treatment."
    • Review of existing literature/studies: "All available/known studies involving 'cryosurgical units' in conjunction with the expanded indications for use have been provided."

8. Sample Size for the Training Set:

  • Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device undergoing such a development and validation process. The supporting evidence for the expanded indications comes from broader existing clinical knowledge and studies.

9. How the Ground Truth for the Training Set was Established:

  • Not applicable. See point 8.

{0}------------------------------------------------

JUL 2 7 2005 510K SUMMARY

ERBE USA, Inc. Submitted By: 2225 Northwest Parkway Marietta, GA 30067

Tel: 770-955-4400 Fax: 770-955-2577

  • Julie Stephens, President/Consultant Contact Person: Regulatory Resources Group, Inc.
    May 31, 2005 Date Prepared:

Cryosurgical Unit and Accessories Common Name:

ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories Trade/Proprietary Name:

Unit, Cryosurgical, Accessories (21 CFR 878.4350) Classification Name:

GEH. FAZ Product Code:

Legally Marketed Predicate Device: ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories - 510(k) Number: K934261

Device Description:

The ERBE USA, Inc. ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories is used to apply extreme cold to tissue during surgical procedures. It uses the Jouleused to apply extreme oold to toose aumy and Thompson principle where probe tip and the surrounding tissue. a fapid drop in temperature and frous Oxide) or CO2 (Carbon Dioxide) gas. The The unit is ascu with older 1120 (11) in the cryo probes are reusable and urnt and accessones are provided new surgical probes are provided in the Notes for Use.

Intended Use:

The ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories are intended for The ENDE ENDE ENDE of tissue during surgical procedures by the application of devitalization (dootraction) of about bodies, mucous plugs, necrotic tissue, and tissue biopsy by cryoadhesion. Also see Clinical Indications.

{1}------------------------------------------------

510K SUMMARY

Similarities and Differences of the Modified Device to the Current Device (Predicate Comparison/Substantial Equivalence):

Similarities

The ERBE USA, Inc. ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories The ERBE USA, Inc. ENDE ENDORATE is to induce extreme cold to tissue as has the same basic technology oncreens and the indications for use are the same except for the additions noted below under differences.

The ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories has the same The ERBE ERBORINTO "OA Oryosary.car only submission # K934261 which are as follows:

"For use with nitrous oxide (N₂O) or carbon dioxide (CO₂) and cryoprobes for cryosurgery."

"Clinical Indications:

Call indications.
Gynecology - Cervical erosions, Cervical polyps, Condylomas, Chronic Cervicitis, Vulva Carcinoma (palliative), Neoplasia

Dermatology - Leukopłakia, Fibroma, Condylomas, Basal Cell Carcinoma, Skin Tumor (palliative), Warts, Naephus

Ablatio Retinae, Glaucoma, Lid Tumor Ophthalmology

ENT - Leukoplakia, Inoperable Tumor (palliative), Laryngeal Papilloma, Fibroma, Angioma, Haemangioma

Thoracic Surgery - Post-Operative

Urology - Prostate Tumor (palliative), Condylomas, Penile Tumor (palliative)

Phiebology - Varicose Veins of the Lower Limbs (Cryo Stripping)

Proctology - Hemorrhoids (1st and 2nd Degree), Pari-Anal Condylomas, Anal Tumor (palliative), Rectal Tumor (palliative), Acute Anal Fissures

Pulmonology - Tumors, Granulomatous Tissue, Malignant Lesions (palliative)

Differences

The ERBE USA, Inc. ERBE ÉRBOKRYO® CA Cryosurgical probe (cryo probe) utilizes different materials from the predicate; however, the materials are biocompatible [see Section III, Declaration of Conformance with Consensus Standards] The ERBE USA, Inc. ERBE ERBOKRYO® CA Cryosvrgcal unit has an added internal filter at the gas outlet in order to comply with the ASTM standard.

{2}------------------------------------------------

510K SUMMARY

The ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories have and The ERBE ERBORNTO - OA STYSTYSTYCE
expanded the indications for use. The difference to the indication is:

    1. The addition of "for removal of foreign bodies, mucous plugs, necrotic tissue, and tissue biopsy by cryoadhesion."
  • Tracheobronchial Stenoses of addition 2. The (Cryorecanalization)"

All the ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories modifications were verified or validated in design control.

Conclusion:

The modifications to the ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories The modifications to the ERDC LRDORN of St. or set and were "himmal" and ormance with Consensus Standards".

All available/known studies involving "cryosurgical units" in conjunction with the All available/Known studies involving "crysuright" atthis submission. The Clinical expanded indications for use have been provided in espany of consisted documentation provided demonstrates than the Erively to remove foreign bodies (i.e., Unit and Accessones can be used surely and charges plugs, etc.) and tissue biopsy food such as peanuts, chicken bones, count, phile, has at the system can be used in by cryoadhesion. The documentation and Showsicians consider cryotherapy to be Pheumology for Tracheobronial Occhoooe. The physicalism at the end a sate procedure for the treather. Of brononial in symptoms at the end of
Patients tolerate the procedure well and show improvement in sympthassy, in a Pattents tolerate the procedure with malignant tumors that crycherapy is a the procedure. They also determinou with Thange Cases, and overally and over Mach palliative treathern in these cases, and of obstruction >50% or not). Most of and 80% measuring performance onceral (romover cryotherapy has a delayed effect in the studies did add the warning that because organ will be used as emergency treatment in acute respiratory distress.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 7 2005

ERBE USA, Inc. c/o Ms. Julie Stephens Regulatory Resources Group, Inc. 111 Laurel Ridge Drive Alpharetta, Georgia 30004

Re: K051509

KUJ1509
Trade/Device Name: ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: II Product Code: GEH Dated: June 2, 2005 Received: June 9, 2005

Dear Ms. Stephens:

We have reviewed your Section 510(k) premarket notification of intent to market the indication We nave reviewed your becaon 910(t) premained is substantially equivalent (for the indications referenced above and nave uctchimed the aerrooms and devices marketed in interstate for use stated in the encrosuly manatinent date of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 28, 1770, the enaement and of the Federal Food. Drug.
devices that have been reclassified in accordance with the provisions of the UMAA devices that have been recialsmed in accordance while approval approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval of the and Cosment AC. (Act) that do not require appro the general controls provisions of the Act. The You may, merelore, market the devices, becycer of the set include registration, listing of general controls provisions of the Fist labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it II your device is classified (Sec above) me existing major regulations affecting your device can
may be subject to such additional controls. Existing major and contribution may be subject to such additional controller Entisting may in 1998. In addition, FDA may be found in the Code of I subtal suggerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean Please be advised that 1 Dr. 5 issualite of a device complies with other requirements of the Act that I DA has made a determination administered by other Federal agencies. You must or any rederal statutes and regulations daminding, but not limited to: registration and listing (21 Comply with an the Act 31equirements, 01 ); good manufacturing practice requirements as set and CFK Fart 807), labeling (21 CFR Part 820), good if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 5 0 locs) This letter will anow you to begin marketing your antial equivalence of your device to a legally premitsel notheadon: The PDA miang of backation for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your do not the car as a same and the regulation entitled, and Colliact the Office of Compilance w (21 trip Part 807.97). You may obtain Misoranding by releveloc to promantonitiities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,
Mark N Millan

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): KOS/509

Device Name: ERBE ERBOKRYO® CA Cryosurgical Unit and Accessories

Indications For Use:

The ERBE ERBOKRYO® CA Cryosurgical Unit and accessories are intended for The ERBE ERBOKKYO CA Clydsuring surgical procedures by the application of extreme
devitalization (destnuction) of tissue during surgical procedures by devitalization (destruction) of tissue during surgical procedules of the ppp
cold and for removal of foreign bodies, mucous plugs, necrotic tissue, and tissue biopsy by cryoadhesion.

Clinical Indications for Cryosurgery

GynecologyCervical Erosions, Cervical Polyps, Condylomas, Chronic Cervicitis, VulvaCarcinoma (palliative), Neoplasia
DermatologyLeukoplakia, Fibroma, Condylomas, Basal Cell Carcinoma, Skin Tumor(palliative), Warts, Naephus
OphthalmologyAblatio Retinae, Glaucoma, Lid Tumor
ENTLeukoplakia, Inoperable Tumor (palliative), Laryngeal Papilloma, Fibroma,Angioma, Haemangioma
Thoracic SurgeryPost-Operative
UrologyProstate Tumor (palliative), Condylomas, Penile Tumor (palliative)
PhlebologyVaricose Veins of the Lower Limbs (Cryo Stripping)
ProctologyHemorrhoids (1st and 2nd Degree), Pari-Anal Condylomas, Anal Tumor(palliative), Rectal Tumor (palliative), Acute Anal Fissures
PulmonologyTumors, Granulomatous Tissue, Malignant Lesions (palliative)
PneumologyTracheobronchial Stenoses (Cryorecanalization)

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark H. Millman

Page 1 of 1

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K05

§ 878.4350 Cryosurgical unit and accessories.

(a)
Identification —(1)Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold.(2)
Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold.(3)
Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.(b)
Classification. Class II.