K Number
K051460
Device Name
C.CAM-AC
Date Cleared
2005-06-15

(12 days)

Product Code
Regulation Number
892.1200
Panel
RA
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Attenuation correction is a method where the emission images acquired during SPECT acquisitions are compensated for attenuation within the patient. The purpose of attenuation correction is to compensate for varying attenuation of gamma rays in the body.

The intended use of the c.cam Attenuation Correction option is to provide the customer with means to measure the attenuation of gamma rays within the body of a patient, when using a c.cam gamma camera system for SPECT studies, and to map the measured attenuation into images that can be used to compensate emission images for the attenuation. The c.cam-AC includes additional acquisition of data for scatter correction of these images.

Device Description

The c.cam-AC is an attenuation correction device, which will be marketed as an optional device for the c.cam gamma camera.

C.cam gamma camera is a dual detector system and the attenuation correction device (two units) will be mounted opposite of the two detectors, one device on each detector. The attenuation correction device is formed as two housings each containing 14 line sources in fixed positions, shutter and all shutter control.

The following described functions are performed on both detectors simultaneously.
When the scanning is initiated the emission image and the transmission images are acquired sequentially. When the emission image has been acquired, a shutter, common for all sources, located in the device housing will open for exposure from 14 line sources (max. 20 mCi each 153 Gd with a total of max. 96 mCi). A collimated beam of gamma ray photons is focused on the opposite detector field of view to create the transmission images of the patient placed in the field of view of the detector. Having acquired both emission and transmission images, the data for emission, transmission, emission scatter and transmission scatter are stored in one dataset. A scaled version of the blank transmission scan is stored in another dataset. The complete study can be export via DICOM to the OEM customer provided processing station for later processing and reviewing.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the c.cam-AC device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text for K051460 describes a device that is essentially equivalent to existing devices and focuses on its design and radioactive source rather than specific performance metrics tied to acceptance criteria. The only performance-related data explicitly stated are for Radiation Leakage.

Acceptance Criteria (What the device aims to achieve)Reported Device Performance
Radiation Leakage:
Exposure from surface of housing, shutter closed

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.