(83 days)
The VPAP ADAPT is intended to provide non-invasive ventilatory support to treat adult patients with OSA and Respiratory Insufficiency caused by central and/or mixed apneas and periodic breathing.
The VPAP ADAPT is a non-invasive flow generator device designed to provide adaptive servo-ventilation therapy (ASV) to stabilize a patient's ventifation. The device continually measures the patient's instantaneous ventilation, and calculates a target ventilation equal to 90% of the patient's recent average ventilation (time constant 100 seconds). It then adjusts the degree of support to servo-control the patient's ventilation to at least equal the target ventilation.
The VPAP ADAPT is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed) whereas others are clinically selectable. Also, a graphic user interface displays clinical data and enables the user (patient/clinician) to set and adjust certain parameters.
Here's a breakdown of the acceptance criteria and study information for the VPAP ADAPT device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Bench Testing | |
| Minute Ventilation | Behaved similar to predicate device (VPAP III ST-A) |
| Tidal Volume | Behaved similar to predicate device (VPAP III ST-A) |
| Clinical Study | |
| Apnea/Hypopnea Index (AHI) | Clinically equivalent to predicate device (VPAP III ST-A) |
| Respiratory-related Arousal Index (RAI) | Clinically equivalent to predicate device (VPAP III ST-A) |
| Effectiveness for central, mixed apnea, or periodic breathing | Demonstrated effective |
| Superiority over predicate device | VPAP ADAPT device was statistically significantly superior (p=0.001) to the VPAP III ST-A for providing effective therapy (based on non-equivalent AHI results) |
| All secondary endpoints | Provided further evidence of equivalency of devices |
2. Sample Size Used for the Test Set and Data Provenance
The text explicitly mentions "Predicate bench testing and clinical studies were used to show substantial equivalence." However, it does not specify the sample size for either the bench test or the clinical test set.
Regarding data provenance:
- The text does not explicitly state the country of origin.
- The text refers to the study as "clinical studies" and "equivalence study," which are typically prospective in nature.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the clinical study. It refers to "clinical studies" and "findings," implying expert interpretation of data like AHI and RAI, but no specifics are given.
4. Adjudication Method for the Test Set
The document does not specify any adjudication method used for the test set in the clinical study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted in the typical sense for algorithm evaluation. Instead, a comparative clinical study was performed between the new device (VPAP ADAPT) and a predicate device (VPAP III ST-A) in human patients.
- Effect size of how much human readers improve with AI vs without AI assistance: This metric is not applicable as the study compares device performance on patients, not human reader performance with or without AI assistance. The study found that the VPAP ADAPT device itself was, in one aspect (non-equivalent AHI results), superior (statistically significant p=0.001) in providing effective therapy compared to the predicate device.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
The study described is a standalone device performance study, not an algorithm-only study. The device, VPAP ADAPT, is a physical medical device that provides "adaptive servo-ventilation therapy" and includes "processing electronics." The performance data relates to the device's ability to treat patients, rather than the isolated performance of an algorithm within it. The "servo-control" mechanism is essentially an algorithm embedded in the device.
7. Type of Ground Truth Used
The ground truth for the clinical study was based on clinical parameters such as:
- Apnea/Hypopnea Index (AHI)
- Respiratory-related Arousal Index (RAI)
These are standard metrics used in sleep medicine to assess the severity of breathing disorders and the effectiveness of therapies. While not explicitly stated, these metrics are typically derived from polysomnography studies, which involve expert interpretation.
8. Sample Size for the Training Set
The document does not mention a separate training set or its sample size. This suggests that the device's "adaptive servo-ventilation therapy" mechanism might have been developed and refined using internal methods, but no external "training set" with ground truth is referenced in the context of this 510(k) submission. The device continually measures ventilation and calculates a target ventilation, which is inherent to its real-time operation, not a pre-trained model in the sense of modern AI.
9. How the Ground Truth for the Training Set Was Established
As no separate training set is mentioned as part of the submission documentation, there is no information on how its ground truth would have been established. The device design and verification activities were performed based on risk analysis and product requirements, implying internal R&D processes rather than training on a specific external dataset with pre-established ground truth.
{0}------------------------------------------------
AUG 16 2005
Sponsor
- 11
ResMed Ltd
Traditional 510(k) Premarket Notification
VPAP ADAPT
510(k) Summary – AutoSet CS2 [As required by §807.92 (c) ]
| Submitter Name: | ResMed Corp. |
|---|---|
| Submitter Address: | 14040 Danielson Street, Poway CA 92064-6857, USA |
| Contact Person: | David D'Cruz, VP Requlatory & Clinical Affairs US |
| Phone Number: | (858) 746 2238 |
| Fax Number: | (858) 746 2915 |
| Date Prepared: | May 19, 2005 |
| Device Trade Name | VPAP ADAPT |
| Device Common Name/Classification Reference | Bi level Positive Pressure Ventilator/ ContinuousVentilator, Passive Exhalation Port, Non-Critical Care.21 CFR868.5895 |
| Product Code | 73MNS |
| Predicate Devices: | VPAP III ST-A (K033276) |
| Reason for submission | New device |
| Intended Use: | The VPAP ADAPT is intended to provide non-invasiveventilatory support to treat adult patients with OSA andRespiratory Insufficiency caused by central and/or mixedapneas and periodic breathing. |
{1}------------------------------------------------
Device Description
The VPAP ADAPT is a non-invasive flow generator device designed to provide adaptive servo-ventilation therapy (ASV) to stabilize a patient's ventifation. The device continually measures the patient's instantaneous ventilation, and calculates a target ventilation equal to 90% of the patient's recent average ventilation (time constant 100 seconds). It then adjusts the degree of support to servo-control the patient's ventilation to at least equal the target ventilation.
The VPAP ADAPT is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed) whereas others are clinically selectable. Also, a graphic user interface displays clinical data and enables the user (patient/clinician) to set and adjust certain parameters.
Performance Data
Design and verification activities were performed on the VPAP ADAPT as a result of the risk analysis and product requirements. All tests confirmed that the VPAP ADAPT met the acceptance criteria. The VPAP ADAPT complies with the applicable standards and requirements referenced in the following quidance documents:
- FDA Draft Reviewer Guidance for Ventilators (July 1995)
- FDA Reviewer Guidance for Premarket Notification Submissions (November ﺎ 1993)
- FDA Reviewer's and Industry, Guidance for the content of premarket submissions for software contained in medical devices, May 1998
Substantial Equivalence
The VPAP ADAPT (new device) and the VPAP III ST-A (predicate device) are both positive pressure flow generators comprising of a blower (motor/fan assembly), flow and pressure sensors, and processing electronics. The blower supplies pressurized air to the patient via a mask and air tubing. The VPAP ADAPT has the same intended use as the VPAP III ST-A.
Predicate bench testing and clinical studies were used to show substantial equivalence between the VPAP ADAPT and VPAP III ST-A.
In the predicate bench testing, the minute ventilation and tidal volume produced by VPAP ADAPT and the VPAP 111 ST-A were measured under various patient breathing efforts. The results show the VPAP ADAPT behaves similar to the predicate device, the VPAP III ST-A.
The findings of the VPAP ADAPT equivalence study demonstrated that the VPAP ADAPT is effective for use in patients with central, mixed apnea or periodic breathing. The primary endpoints were achieved, showing clinically equivalent ApnealHypopnea (AHI) and Respiratory-related Arousal Index (RAI) sleep and respiratory parameters between the VPAP ADAPT and VPAP III ST-A devices. Furthermore, the non-equivalent AHI results indicated that the VPAP ADAPT device is superior (statistically significant p=0.001) to the VPAP III ST-A device for providing effective therapy. Additionally, all secondary endpoints provided further evidence of the equivalency of the devices.
Conclusion
The VPAP ADAPT is substantially equivalent to the VPAP III ST-A.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle faces to the right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
AUG 16 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David D'Cruz Vice President, Regulatory & Clinical Affairs ResMed Limited 14040 Danielson Street Poway, California 92064-6857
Re: K051364
Trade/Device Name: VPAP Adapt Regulation Number: 21 CFR 868.5895 Regulation Name: Bi-Level positive Pressure Ventilator/Continuous Ventilator Regulatory Class: II Product Code: MNS Dated: May 20, 2005 Received: May 31, 2005
Dear D'Cruz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. D'Cruz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known):
VPAP ADAPT Device Name: ____________________
Indications For Use:
The VPAP ADAPT is intended to provide non-invasive ventilatory support to treat adult patients with OSA and Respiratory Insufficiency caused by central and/or mixed apneas and periodic breathing.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aun Sufeom
(Division Sign-Off) Division of Anesthasiology, General Hospital, Infection Control, Dental Devices
510(k) Number:
.
Page 1 of
§ 868.5895 Continuous ventilator.
(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).