(171 days)
The noninvasive INVOS 5100B is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain of an individual. It also is intended for use as an adjunct trend monitor of hemoglobin oxygen saturation of blood in a region of skeletal muscle tissue beneath the sensor in infants, children, or adults at risk for reduced-flow or no-flow ischemic states. The prospective clinical value of data from the INVOS System has not been demonstrated in disease states. The INVOS System should not be used as the sole basis for diagnosis or therapy.
The principles of operation of the cerebral oximeter system are based on the assumption that hemoglobin exists in two principal forms in the blood: oxygenated hemoglobin (HbO2) and reduced hemoglobin (Hb). Functional oxygen saturation (SO2) is defined as the ratio of oxyhemoglobin (HbO2) to total hemoglobin (HbO2 + Hb) and is commonly presented as a percentage.
SO2 = (HbO2 / (HbO2 + Hb)) * 100%
Since oxygenated and reduced hemoglobin are different colors and absorb light as a known function of wavelength, selected wavelengths of light can be used to assess the relative percentage of these two constituents. This fundamental approach of assessing the color of blood using various wavelengths of light to measure hemoglobin oxygen saturation trends is used in all currently marketed oximetry systems.
For cerebral monitoring, a disposable sensor of medical grade materials is applied to the patient's forehead (Figure 1). The sensor incorporates a light source and two return signal detectors at different predetermined distances from the light source. The signal detector nearest the light source (3 cm) is considered the "shallow detector" and the further detector from the light source (4 cm) the "deep detector."
While the light reaching the deep detector has sampled about the same amount of skin, scalp, and skull as the light reaching the shallow detector, it has sampled more brain tissue. This difference is used to help separate out the brain signal and suppress anatomical differences in patients. The additional information unique to the deep signal return is predominately from brain tissue blood which is composed mostly of venous blood. The information contained in the shallow and deep signal returns is processed by an algorithm to measure changes in hemoglobin oxygen saturation in a small region of tissue beneath the sensor, predominately in the brain.
For non-cerebral monitoring, a somatic sensor which is similar to the cerebral sensor is placed on skeletal muscle in an area free of large fat deposits, bony protuberances and hair. Since the tissue being interrogated is relatively homogeneous, the resultant rSQ2 index is an average of the hemoglobin oxygen saturation of blood in the region of tissue below the sensor.
The SomaSensor is connected to a preamplifier (1.4 x 7.65 x 3.75 in.) which is placed close to the patient and amplifies the rSO2 signal. The signal is then carried to a display unit (8.4 x 9.6 x 8.5 in.) where the values and trends are displayed on the screen. The display unit controls all functions of the system with selections made by keys with onscreen labels. The system will operate for up to 120 minutes on battery, enabling patient transport without loss of data.
The Somanetics INVOS 5100B Oximeter System, including the SSFB Somatic SomaSensor, was granted substantial equivalence based on prior clinical studies and nonclinical testing that demonstrated its performance. The provided document does not explicitly present a table of acceptance criteria with numerical targets and direct comparisons to device performance for this specific 510(k) submission. Instead, it relies heavily on the substantial equivalence to previously cleared devices (INVOS 5100, Spectros Corporation T-Stat™ 303, and Hutchinson InSpectra) and results from previous studies.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not provide a formal table with acceptance criteria outlined for this specific submission. However, it references performance characteristics from a previous animal study for skeletal muscle tissue monitoring.
| Acceptance Criteria (Implied from previous study) | Reported Device Performance (from animal study) |
|---|---|
| Ability to track changes in blood oxygen saturation in skeletal muscle tissue over a range of 5% to 95% with an unspecified bias and standard deviation. | Tracked changes in blood oxygen saturation in skeletal muscle tissue over a range of 5% to 95% with a bias of 1 and standard deviation of 7. |
| Ability to accurately track changes in cerebral blood oxygen saturation over a wide range of arterial saturation compared to co-oximetry. | Demonstrated accurate tracking of changes in cerebral blood oxygen saturation compared to co-oximetry in adult hypoxia study and pediatric cath lab study. |
2. Sample Sizes and Data Provenance:
- Test Set (for the mentioned animal study related to somatic sensor): The document states an "animal study using isolated perfused porcine limbs" was conducted. The specific number of limbs or animals used is not provided.
- Data Provenance:
- Adult Hypoxia Study: Not explicitly stated, but typically these are conducted in controlled clinical environments, likely in the US given the submission to the FDA. Retrospective/Prospective is not specified.
- Pediatric Cath Lab Study: Not explicitly stated, but likely clinical data collected in a hospital setting, probably in the US. Retrospective/Prospective is not specified.
- Animal Study (Somatic Sensor): Porcine limbs, prospective in nature. Country of origin not specified.
- Multiple Peer-Reviewed Studies: Referencing existing literature, data provenance would vary by study.
- Two Published Studies (Exercise): Referencing existing literature, data provenance would vary by study.
- The 510(k) submission itself primarily relies on substantial equivalence and previously established performance, rather than new, large-scale clinical trials for the specific 5100B device.
3. Number of Experts and Qualifications for Ground Truth:
- The document does not specify the number of experts or their qualifications for establishing ground truth for any of the studies mentioned.
- For the cerebral studies, the "ground truth" was established by co-oximetry analysis of blood samples, which is considered a gold standard for blood oxygen saturation measurement. This is an objective measurement rather than an expert interpretation.
- For the animal study, the "ground truth" was also a direct measurement, implied to be derived from the isolated perfused porcine limbs' blood oxygen saturation.
4. Adjudication Method:
- Not applicable/Not specified. Since objective measurements like co-oximetry were used as ground truth, expert adjudication methods (like 2+1 or 3+1 for imaging consensus) were not relevant or mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No. An MRMC comparative effectiveness study was not done or reported in this document. The INVOS 5100B is a trend monitor, not an interpretive diagnostic device that relies on human reader input in the same way imaging AI does. Therefore, a study to assess how human readers improve with AI assistance is not relevant to this type of device.
6. Standalone Performance Study:
- Yes, in essence. The reported performance from the adult hypoxia study, pediatric cath lab study, and the animal study on porcine limbs demonstrate the standalone performance of the INVOS System in tracking oxygen saturation against a gold standard (co-oximetry or direct measurement). The device is intended to provide measurements, and its accuracy in doing so is what these studies assessed.
7. Type of Ground Truth Used:
- Objective Measurement/Physiological Gold Standard:
- For cerebral oxygen saturation, the ground truth was established using co-oximetry analysis of blood samples.
- For skeletal muscle tissue, the ground truth was also a direct measurement of blood oxygen saturation in the isolated perfused porcine limbs.
8. Sample Size for the Training Set:
- The document does not provide information on the sample size used for training any algorithms within the INVOS 5100B system. This product pre-dates the widespread use of deep learning and large-scale training sets as discussed in modern AI/ML medical devices. The device's operation is based on established principles of near-infrared spectroscopy and algorithmic processing of light absorption, likely with parameters developed through engineering and calibration, rather than extensive "training sets" in the contemporary AI sense.
9. How Ground Truth for Training Set was Established:
- As the document does not refer to a distinct "training set" in the context of modern AI/ML, it does not describe how ground truth for such a set was established. The device's underlying principles and algorithms would have been developed and validated through scientific understanding of optical properties of blood and tissue, and engineering calibration, possibly against reference measurements in laboratory settings or smaller physiological studies.
{0}------------------------------------------------
| NOV - 4 2005 | K051274 | |
|---|---|---|
| Appendix A10 | 510(k) Summary | |
| Date Prepared: | 10/6/2005 | |
| Submitter/Contact: | Ronald A. Widman | |
| Vice President, Medical Affairs | ||
| Somanetics Corporation | ||
| 1653 East Maple Road | ||
| Troy, MI 48083 | ||
| Phone: (248) 689-3050Fax: (248) 689-4272 | ||
| Trade Name: | Somanetics INVOS® 5100B and SSFB Somatic SomaSensor®Accessories: | |
| SAFB | Adult Cerebral SomaSensor (>40 kg) | |
| SAFB-SM | Small Adult Cerebral SomaSensor (>40 kg) | |
| SPFB | Pediatric Cerebral SomaSensor (<40 kg) | |
| SSFB | Somatic SomaSensor | |
| BRCB | Bilateral Reusable Sensor Cable | |
| 5100B-W | One-year Extension of Warranty | |
| 313137A | 5100B System Operations Manual | |
| 5100B-FTD | Field Test Device | |
| 5100B-STD | Portable Mobile Stand | |
| 5100B-TC | Travel Case | |
| DB9DB9 | Computer Connection Serial Cable | |
| 5100B-DD | 3.5" Floppy Disk Drive | |
| Classification Name: | Tissue Oximeter | |
| Common Name: | Oximeter/Cerebral Oximeter/NIRS | |
| Regulatory Class: | Oximeters have been classified in Class II by theCardiovascular device panel, see 21 CFR 870.2700, productcode MUD. | |
| Performance Stds: | FDA has not developed performance standards for this device.Also, oximeters have not been assigned any special controls. |
Indications for Use: The noninvasive INVOS 5100B is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain of an individual. It also is intended for use as an adjunct trend monitor of hemoglobin oxygen saturation of blood in a region of skeletal muscle tissue beneath the sensor in infants, children, or adults at risk for reduced-flow or no-flow ischemic states. The prospective clinical value of data from the INVOS System has not been demonstrated in disease states. The INVOS System should not be used as the sole basis for diagnosis or therapy.
Device Description:
The principles of operation of the cerebral oximeter system are based on the assumption that hemoglobin exists in two principal forms in the blood: oxygenated hemoglobin (HbO2) and reduced hemoglobin (Hb). Functional oxygen saturation (SO2) is defined as the ratio of oxyhemoglobin (HbO2) to total hemoglobin (HbO2 + Hb) and is commonly presented as a percentage.
$$SO_1 = \frac{HbO_2}{HbO_2 + Hb} \times 100%$$
{1}------------------------------------------------
Since oxygenated and reduced hemoglobin are different colors and absorb light as a known function of wavelength, selected wavelengths of light can be used to assess the relative percentage of these two constituents. This fundamental approach of assessing the color of blood using various wavelengths of light to measure hemoglobin oxygen saturation trends is used in all currently marketed oximetry systems.
Image /page/1/Figure/2 description: This image is a cross-section of a sensor on a forehead. The image shows the different layers of the forehead, including the periosteal, dura mater, and superior sagittal sinus. The image also shows the light source and the distances of 30mm and 40mm.
For cerebral monitoring, a disposable sensor of medical grade materials is applied to the patient's forehead (Figure 1). The sensor incorporates a light source and two return signal detectors at different predetermined distances from the light source. The signal detector nearest the light source (3 cm) is considered
the "shallow detector" and the further detector from the light source (4 cm) the "deep detector."
While the light reaching the deep detector has sampled about the same amount of skin, scalp, and skull as the light reaching the shallow detector, it has sampled more brain tissue. This difference is used to help separate out the brain signal and suppress anatomical differences in patients. The additional information unique to the deep signal return is predominately from brain tissue blood which is composed mostly of venous blood. The information contained in the shallow and deep signal returns is processed by an algorithm to measure changes in hemoglobin oxygen saturation in a small region of tissue beneath the sensor, predominately in the brain.
For non-cerebral monitoring, a somatic sensor which is similar to the cerebral sensor is placed on skeletal muscle in an area free of large fat deposits, bony protuberances and hair. Since the tissue being interrogated is relatively homogeneous, the resultant rSQ2 index is an average of the hemoglobin oxygen saturation of blood in the region of tissue below the sensor.
The SomaSensor is connected to a preamplifier (1.4 x 7.65 x 3.75 in.) which is placed close to the patient and amplifies the rSO2 signal. The signal is then carried to a display unit (8.4 x 9.6 x 8.5 in.) where the values and trends are displayed on the screen. The display unit controls all functions of the system with selections made by keys with onscreen labels. The system will operate for up to 120 minutes on battery, enabling patient transport without loss of data.
{2}------------------------------------------------
Substantial Equivalence:
The INVOS 5100B is substantially equivalent to the INVOS 5100 (K001842) in that the method of measurement, electronics, packaging and physical configuration are identical but the indications for use are expanded. The newer INVOS 5100B will operate with the SSFB Somatic SomaSensor to monitor skeletal muscle tissue. The indications for use are substantially equivalent to the indications for the Spectros Corporation T-Stat™ 303 Microvascular Tissue Oximeter (K040684) and the Hutchinson InSpectra (K963903).
Nonclinical Testing:
The INVOS 5100B is identical to the INVOS 5100 Adult/Pediatric Cerebral Oximeter System (K001842) and the SSFB Somatic SomaSensor is identical to the SAFB (K001842) except that it is supplied with a larger adhesive pad and revised labeling. The result of previous validations, including software validations for a moderate level of concern, confirms that the model 5100B performs according to published specifications.
Clinical Testing:
- An adult volunteer hypoxia study (K971628) demonstrated that the INVOS System was . able to accurately track changes in cerebral blood oxygen saturation as compared with co-oximetry analysis of blood samples over a wide range of arterial saturation repeated at two levels of arterial carbon dioxide.
- A study in the pediatric cath lab (K001842) demonstrated that the INVOS System also . accurately tracked changes in cerebral oxygen saturation as compared with co-oximetry analysis of blood samples in children with congenital heart defects during sedation and during supplemental oxygen administration.
- . An animal study using isolated perfused porcine limbs demonstrated that the INVOS System was able to track changes in blood oxygen saturation in skeletal muscle tissue over a range of 5% to 95% with a bias of 1 and standard deviation of 7.
- . Multiple peer-reviewed studies have shown the INVOS System provides skeletal muscle somatic oxygen saturation readings in low and no flow ischemic states in human and animal trials.
- . Two published studies demonstrate the ability of the INVOS System somatic saturation to reflect the intensity of exercise.
- Somatic saturation has been shown to correlate with mixed venous oxygen saturation . and has the potential to predict the onset of anaerobic metabolism.
Conclusion:
The INVOS 5100B and SSFB Somatic SomaSensor were found to be substantially equivalent to the previously marketed INVOS 5100 and SAFB SomaSensor (K001842) and the indications for use are substantially equivalent to the Spectros Corporation T-Stat 303 (K040684). No new questions of safety or effectiveness were raised.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human figures.
Public Health Service
NOV - 4 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Ronald A. Widman Vice President, Medical Affairs Somanetics Corporation 1653 East Maple Road Troy, Michigan 48083-4208
Re: K051274
Trade/Device Name: INVOS 5100B Oximeter System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: MUD Dated: October 6, 2005 Received: October 7, 2005
Dear Mr. Widman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Widman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Qur
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of _1
510(k) Number (if known) K051274
Device Name: Somanetics INVOS® 5100B System and SSFB Somatic SomaSensor®
Indications For Use:
The noninvasive INVOS 5100B is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain of an individual. It is also intended for use as an adjunct trend monitor of hemoglobin oxygen saturation of blood in a region of skeletal muscle tissue beneath the sensor in infants, children, or adults at risk for reduced-flow or no-flow ischemic states.
The prospective clinical value of data from the INVOS System has not been demonstrated in disease states. The INVOS System should not be used as the sole basis for diagnosis or therapy.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---|---|
| Division of Anesthesiology, General Hospital, | |
| Infection Control, Dental Devices | |
| 510(k) Number: | K051274 |
| Prescription Use | X | OR | Over-The-Counter Use |
|---|---|---|---|
| (Per 21 CFR 801.109) |
(Optional Format 1-2-96)
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).