K Number
K051108
Manufacturer
Date Cleared
2005-07-08

(67 days)

Product Code
Regulation Number
862.1440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Piccolo Lactate Dehydrogenase Test System (presently contained on the Piccolo Basic Metabolic Panel Plus Reagent Disc) used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of lactate dehydrogenase activity in heparinized plasma or serum in a clinical laboratory setting or point-of-care location.

Lactate dehydrogenase measurements are used in the diagnosis and treatment of liver diseases such as acute viral hepatitis and cirrhosis; cardiac diseases such as myocardial infarction; malignant diseases; and tissue alterations of the heart, kidney, liver, and muscle.

Device Description

The Piccolo Basic Metabolic Panel Plus Reagent Disc (which contains the Piccolo Lactate Dehydrogenase Test System) is designed for heparinized plasma and serum, only. The disc meters the required quantity of sample and diluent, mixes the sample with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted samples mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer. Alternately, the disc may also be used with serum.

AI/ML Overview

Here's an analysis of the provided text to extract information about the device's acceptance criteria and the supporting study, formatted as requested:

Device: Piccolo® Lactate Dehydrogenase Test System


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined "acceptance criteria" in terms of specific performance targets (e.g., "linearity must be >0.99"). Instead, it presents the results of linearity and precision studies. The implicit acceptance criteria would be for these results to demonstrate performance comparable to the predicate device and suitable for its intended use.

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Piccolo System)Predicate Device Information (Synchron LX20)
LinearityComparable to predicate/good clinical practiceNot explicitly stated for predicate
- SlopeN/A1.012N/A
- InterceptN/A+0.253N/A
- Correlation CoefficientN/A (implied >0.99 for good linearity)0.998N/A
Precision (Within-Run)Comparable to predicate/good clinical practiceNot explicitly stated for predicate
- Control Level 1 MeanN/A87 U/LN/A
- Control Level 1 SDN/A3.0N/A
- Control Level 1 %CVN/A (implied low %CV for good precision)3.4N/A
- Control Level 2 MeanN/A350 U/LN/A
- Control Level 2 SDN/A3.8N/A
- Control Level 2 %CVN/A (implied low %CV for good precision)1.1N/A
Precision (Total)Comparable to predicate/good clinical practiceNot explicitly stated for predicate
- Control Level 1 MeanN/A87 U/LN/A
- Control Level 1 SDN/A4.4N/A
- Control Level 1 %CVN/A (implied low %CV for good precision)5.0N/A
- Control Level 2 MeanN/A350 U/LN/A
- Control Level 2 SDN/A7.0N/A
- Control Level 2 %CVN/A (implied low %CV for good precision)2.0N/A
Sample Type ComparabilityHeparinized plasma and serum results comparableEstablished for Lactate DehydrogenaseN/A

2. Sample Size Used for the Test Set and Data Provenance

  • Linearity: Sample size not explicitly stated beyond the reported slope, intercept, and correlation coefficient.
  • Precision:
    • Sample Size: n = 80 for both within-run and total precision studies.
  • Sample Type Comparison: Sample size not explicitly stated.
  • Data Provenance: Not explicitly stated as retrospective or prospective, or country of origin. Given it's a 510(k) submission for a US market, studies would typically be conducted to meet US regulatory requirements. It is a non-clinical study since it involves comparing the device to a predicate device in a laboratory setting.

3. Number of Experts Used and Their Qualifications for Ground Truth

Not applicable. This is a non-clinical study evaluating the performance of a laboratory diagnostic device, not a diagnostic imaging or AI algorithm requiring expert interpretation for ground truth. The "ground truth" for linearity and precision would be derived from certified reference materials or established high-accuracy laboratory methods.


4. Adjudication Method for the Test Set

Not applicable, as this is a non-clinical laboratory device performance study, not a clinical trial or AI study involving human interpretation.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. This study focuses on the analytical performance (linearity, precision) of a laboratory assay, not on the effectiveness of human readers with or without AI assistance.


6. Standalone Performance (Algorithm Only without Human-in-the-Loop)

Yes, the studies reported (Linearity, Precision, Sample Type Comparison) represent the standalone performance of the Piccolo® Lactate Dehydrogenase Test System. These are analytical studies of the device itself, without human interpretation as part of the performance metric.


7. Type of Ground Truth Used

  • Linearity, Precision, and Sample Type Comparison: The ground truth would be established using reference methods or certified calibrators/control materials with known concentrations/activities of Lactate Dehydrogenase. For precision, the "true" value is the mean value obtained over multiple runs. For sample type comparison, the "true" value is derived from a comparative method or the predicate device if it's considered the reference.

8. Sample Size for the Training Set

Not applicable. This is a non-AI/ML medical device submission. Therefore, there is no "training set" in the context of machine learning. The device's calibration is based on "Bar code with factory calibrated lot specific data."


9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for an AI/ML algorithm. The device uses "factory calibrated lot specific data" for its calibration, which would be established during manufacturing and quality control processes using reference materials and established laboratory standards.

{0}------------------------------------------------

JUL - 8 2005

3240 Whipple Road, Union City, CA 94587 510 · 675-6500 Fax 510 · 441-6150 Phone

Image /page/0/Picture/4 description: The image shows the logo for ABAXIS. The logo is in black and white and features the company name in bold, sans-serif font. A stylized oval shape encircles the letters 'XIS' of the company name, adding a visual element to the logo.

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________

1. Applicant Information:

Date Prepared:April 29, 2005
Name:Abaxis, Inc.
Address:3240 Whipple RoadUnion City, CA 94587
Contact Person:Dennis M. Bleile, PhD
Phone Number:(510) 675-6515
Fax Number:(510) 441-6150

Device Information: 2.

ClassificationClass II
Trade Name:Piccolo® Lactate Dehydrogenase Test System

Classification Name: Lactate Dehydrogenase Test system 862.1440

Identification of legally marketed device to which the submitter claims 3. equivalence:

The following table identifies the legally marketed device to which Abaxis claims equivalence:

Predicate Device
Predicate DeviceManufacturer510(k)NumberDate of SEDetermination
Lactate DehydrogenaseReagents on theSynchron LX20Chemistry SystemBeckman Coulter,Inc. (Brea, CA)K0112135/16/01

Description of the Device: 4.

The Piccolo Basic Metabolic Panel Plus Reagent Disc (which contains the Piccolo The Thoolo Dacto MetaTest System) is designed for heparinized plasma and serum, only. The disc meters the required quantity of sample and diluent, mixes

{1}------------------------------------------------

Summary of Safety and Effectiveness (continued)

the sample with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted samples mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer. Alternately, the disc may also be used with serum.

5. Statement of Intended Use:

The Piccolo Lactate Dehydrogenase Test System (presently contained on the Basic Metabolic Panel Plus Reagent Disc) used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of lactate dehydrogenase activity in heparinized plasma or serum in a clinical laboratory setting or point-of-care location.

Summary of the technological characteristics of the new device in comparison 6. to those of the predicate device:

outlines the technological characteristics of the Piccolo Lactate ← Table Dehydrogenase Test System in comparison to the legally marketed predicate device.

Piccolo Point-of-CareChemistry AnalyzerSynchron LX20 ChemistrySystem
Intended UseQuantitative analysis ofLactate DehydrogenaseQuantitative analysis ofLactate Dehydrogenase
MethodologyEnzymatic rate reactionEnzymatic rate reaction
Sample TypeHeparinized plasma and serumHeparinized plasma and serum
Sensitivity50 U/L5 U/L
ReagentsDry test-specific reagent beadsand liquid diluent; reconstitutionperformed by analyzerActive ingredients:LactateNicotinamide adeninedicucleotide (NAD+)Diaphorasep-Iodonitrotetrazolium Violet(INT)Liquid reagentsActive ingredients:LactateNicotinamide adeninedicucleotide (NAD+)
Temperature ofReaction37°C37°C
CalibrationBar code with factorycalibrated lot specific dataCalibration not required.
Assay Range50 - 1,000 U/L5 - 750 U/L(600 - 2,700 U/L ORDAC*)
Testing EnvironmentProfessional useProfessional use

Table 1: Specification Comparison: Piccolo Lactate Dehydrogenase Test System

{2}------------------------------------------------

Summary of Safety and Effectiveness (continued)

Table 1: Specification Comparison: Piccolo Lactate Dehydrogenase Test System (continued)

Piccolo Point-of-CareChemistry AnalyzerSynchron LX20 ChemistrySystem
Sample SizeApproximately 100 µL13 µL(3 µL ORDAC*)

Brief discussion of the clinical and nonclinical tests relied on for a 7. determination of substantial equivalence.

Tables 2 summarize the results of clinical and non-clinical tests performed using the Piccolo Lactate Dehydrogenase Test System.

Linearity:

Table 2: Summary of Linearity

LactateDehydrogenase
Slope1.012
Intercept+0.253
Corr. Coefficient0.998

{3}------------------------------------------------

Summary of Safety and Effectiveness (continued)

Precision:

Precision studies were designed to evaluate within-run and total precision of the Lactate Dehydrogenase Test System when run on the Piccolo Point-of-Care Chemistry Analyzer.

Table 3: Within-Run and Total Precision for Lactate Dehydrogenase, Assayed on the Piccolo Point-of-Care Chemistry Analyzer

AnalyteWithin-Run(n = 80)Total(n = 80)
Lactate Dehydrogenase (U/L)
Control Level 1
Mean8787
SD3.04.4
%CV3.45.0
Control Level 2
Mean350350
SD3.87.0
%CV1.12.0

Sample Type Comparison:

A study was conducted to examine and compare results for heparinized plasma and serum on the Piccolo® Point-of-Care Chemistry Analyzer.

Heparinized plasma and serum comparability were established for Lactate Dehydrogenase.

Conclusions 8.

The clinical and non-clinical tests performed for Lactate Dehydrogenase, when run on the Piccolo Point-of-Care Chemistry Analyzer, demonstrate that the test system is as safe, effective and performs as well as the legally marketed device identified above.

{4}------------------------------------------------

.

:

Abaxis Confidential

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the top half of the circle.

Public Health Service

JUL - 8 2005

Food and Drug Administrat 2098 Gaither Road Rockville MD 20850

Dennis M. Bleile Ph.D. Director of Assay Performance &Compliance Abaxis, Inc 3240 Whipple Road Union City, CA 94587

Re: K051108

K031100
Trade/Device Name: Piccolo® Lactate Dehydrogenase Test Regulation Number: 21 CFR 862.1440 Regulation Name: Lactate dehydrogenase test system Regulatory Class: Class II Product Code: CFJ Dated: April 29, 2005 Received: May 2, 2005

Dear Dr. Bleile:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for the stated in the encrosure) to regard nent date of the Medical Device Amendments, or to commerce prof to May 20, 1977, the enacements with the provisions of the Federal Food, Drug, devices mail have been recuire approval of a premarket approval application (PMA). alle Cosmetic Act (71cc) market the device, subject to the general controls provisions of the Act. The 1 ou may, mercrore, market the act include requirements for annual registration, listing of general controls provisions of aactice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device it may be subject to Sach additions (CFR), Parts 800 to 895. In addition, FDA can oc nound in This 21, Courtents concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be advised that i Dr i loseants of the complies with other requirements of the Act that IDA has made a determinations administered by other Federal agencies. You must of any reactal statutes and regaranents ancluding, but not limited to: registration and listing (21 Comply with an the Free Stocker Parts 801 and 809); and good manufacturing practice CI K Part 807), tabeling systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your section 5 10(k) I his letter will anow you to begin manketing your antial equivalence of your device to a legally premarked notification. The PDT Intuing of basicance of the super and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, If you desire specific intornation advartising of your device, please contact the Office of In
or questions on the promotion and advertising of your device, please contact th of questions on the promotion and Safety at (240) 276-0484. Also, please note the VIIro Diagnostic Device Development and Sales ... (21CFR Part 807.97).
regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). regulation entitled, "Misolanding of responsibilities under the Act from the You may other other general informations on your Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Carol C. Benem

Carol C. Benson, M.A. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known):K051108
------------------------------------

Piccolo® Lactate Dehydrogenase Test System Device Name: _

Indications For Use:

The Piccolo Lactate Dehydrogenase Test System (presently contained on the The Piccolo Lactate Denyarogenato Treat Systems with the Piccolo Point-of-Care Precolo Basic Metabolic Panel Place Lused for the in vitro quantitative Chemilstly Analyzer 15 Interneta determination or oratory setting or point-of-care location.

Lactate dehydrogenase measurements are used in the diagnosis and treatment Lactate denydrogenase measurements and circhosis; cardiac diseases
of liver diseases such as acute viral hepatitis and cirches hart kidney, liver of liver diseases such as acute virus nopality and tissue alterations of the heart, kidney, liver, and muscle.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Page 1 of

510(k) K051108

§ 862.1440 Lactate dehydrogenase test system.

(a)
Identification. A lactate dehydrogenase test system is a device intended to measure the activity of the enzyme lactate dehydrogenase in serum. Lactate dehydrogenase measurements are used in the diagnosis and treatment of liver diseases such as acute viral hepatitis, cirrhosis, and metastatic carcinoma of the liver, cardiac diseases such as myocardial infarction, and tumors of the lung or kidneys.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.