K Number
K050613
Manufacturer
Date Cleared
2005-08-16

(159 days)

Product Code
Regulation Number
866.5630
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Quantia Beta-2 Microglobulin is intended as a latex particle enhanced immunoturbidimetric assay for The Quantitative determination of beta-2-microglubulin concentration in human serum or plasma the ?? Viro quarklative accension in the diagnosis of active rheumatoid arthritis and kidney disease.

Quantia PROTEINS Control is intended for use in monitoring the quality control of results obtained with the Quantia Beta-2 Microglobulin and Quantia A1-AT reagents by turbidimetry. (NOTE: This control has been also 510(k) FDA submitted for use with A1-AT) For in vitro diagnostic use

Quantia Beta-2 Microglobulin Standard is intended for use in establishing the calibration curve for the Quantia Beta-2 Microglobulin reagents by turbidimetry. For in vitro diagnostic use.

Device Description

The Quantia Beta-2 Microglobulin is intended as a latex particle enhanced immunoturbidimetric assay for the in vitro quantitative determination of beta-2-microglubulin concentration in human serum or plasma (EDTA) on the AEROSET® instrument as an aid in the diagnosis of active rheumatoid arthritis and kidney disease.

Quantia PROTEINS Control is intended for use in monitoring the quality control of results obtained with the Quantia Beta-2 Microglobulin and Quantia A1-AT reagents by turbidimetry. (NOTE: This control has been also 510 (k) FDA submitted for use with Quantia A1-AT) For in vitro diagnostic use.

Quantia Beta-2 Microglobulin Standard is intended for use in establishing the calibration curve for the Quantia Beta-2 Microglobulin reagents by turbidimetry. For in vitro diagnostic use.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Quantia Beta-2 Microglobulin device based on the provided text:

Quantia Beta-2 Microglobulin Acceptance Criteria and Performance Study

This document describes the performance characteristics of the Quantia Beta-2 Microglobulin, a latex particle enhanced immunoturbidimetric assay for the quantitative determination of beta-2-microglobulin in human serum or plasma.

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implied/Standard for Assay Types)Reported Device Performance
Method ComparisonSubstantial equivalence to predicate device (implied by 510(k) process and statement)Slope of 0.876 and correlation coefficient (r) of 0.9986 against predicate device
Within-run Precision (CV)Typically < 5-10% (implied standard for assays)Control I: 1.1% (at 1.00 mg/L) Control II: 0.9% (at 5.46 mg/L) Third Control: 0.9% (at 13.61 mg/L)
Total Precision (CV)Typically < 10-15% (implied standard for assays)Control I: 1.9% Control II: 1.3% Third Control: 1.1%
Linearity Range (without rerun)A specified range where results are proportional to concentration (implied)0.25 to 16 mg/L
Linearity Range (with rerun)A specified range where results are proportional to concentration (implied)0.02 to 100 mg/L
Limit of Quantification (without rerun)Minimum quantity measurable with defined precision and error (CV < 20%, error < ± 20%)0.25 mg/L
Limit of Quantification (with rerun)Minimum quantity measurable with defined precision and error (CV < 20%, error < ± 20%)0.02 mg/L
Interference: LipemiaNo significant interference up to a certain level (implied)No significant interference up to sample absorbance of 2.1 AU/cm at 660 nm
Interference: TriglyceridesNo significant interference up to a certain concentration (implied)No significant interference up to 1300 mg/dL
Interference: BilirubinNo significant interference up to a certain concentration (implied)No significant interference up to 20 mg/dL
Interference: HemoglobinNo significant interference up to a certain concentration (implied)No significant interference up to 480 mg/dL
Interference: RFInterference below a certain percentage up to a certain concentration (implied)Interference below 10% up to 288 IU/mL

Note: The document directly states the results of the performance studies rather than explicitly listing "acceptance criteria." The acceptance criteria listed above are implied based on common regulatory expectations for in vitro diagnostic devices and the fact that the device received 510(k) clearance, indicating its performance was deemed acceptable for its intended use.

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Method Comparison: 105 serum samples.
    • Precision: Not explicitly stated as a "test set" for precision, but control materials were used. Control I, Control II, and a third control (spiked with Beta-2 Microglobulin) were run in duplicate twice a day over twenty days.
    • Linearity, Limit of Quantification, Interference: Sample sizes for these specific studies are not explicitly provided in the text beyond the control materials or general statements about the analytes tested.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The study was conducted by Biokit S.A. in Spain. Therefore, it is likely the data was generated in Spain or a similar geographical region. The text does not specify if the studies were retrospective or prospective, but assay performance studies are typically prospective evaluations.

3. Number of Experts and Qualifications for Ground Truth

Not applicable. This device is an in vitro diagnostic assay that measures a biomarker concentration. Ground truth is established by the analytical method itself (e.g., accuracy against a reference method or known concentrations) rather than expert interpretation of images or clinical data.

4. Adjudication Method for the Test Set

Not applicable. As described above, this is an analytical performance study for an in vitro diagnostic assay, not a study requiring human adjudication of results or interpretations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is an analytical performance study for an in vitro diagnostic assay. It does not involve human readers interpreting cases or AI assistance.

6. Standalone (Algorithm Only) Performance

Yes, the studies described are for the standalone performance of the Quantia Beta-2 Microglobulin assay on the Abbott AEROSET® instrument. There is no human-in-the-loop component mentioned in these performance evaluations.

7. Type of Ground Truth Used

  • Method Comparison: The predicate device, IL Test Beta-2-Microglobulin, served as the comparative "ground truth" or reference for evaluating performance equivalence.
  • Precision: Ground truth was established by the known concentrations of Beta-2 Microglobulin in the control materials (e.g., purified Beta-2 Microglobulin for spiking, or characterized control levels).
  • Linearity, Limit of Quantification, Interference: Ground truth was based on known concentrations of the analyte and interferents, or reference measurements.

8. Sample Size for the Training Set

Not applicable. The Quantia Beta-2 Microglobulin is an immunoturbidimetric assay, not an AI or machine learning algorithm that requires a "training set" in the conventional sense. The assay's performance is determined by its chemical and optical reagents and the instrument's calibration and measurement principles.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As noted above, this device does not utilize a training set in the context of machine learning. The "ground truth" for calibrating and developing such an assay typically involves using highly purified standards of known concentration and validating methods against established reference procedures.

{0}------------------------------------------------

AUG 1 6 2005

Ko50 6/3

Section 3 Quantia Beta-2 Microglobulin 510(k) Summary (Summary of Safety and Effectiveness)

Submitted by:

Biokit S.A. Can Male, Llissa d'Amunt Barcelona 08186 Spain

Contact Person:

Contact: Joan Guixer, Quality Assurance and Regulatory Affairs Director Phone: 34 - 93 860 90 00

Summary Prepared:

August 10th, 2005

Name of the device:

Quantia Beta-2 Microglobulin

Classification name(s):

866.5630Beta-2-microglobulin immunological test systemClass I
JZGSystem, Test, Beta-2-Microglobulin Immunological

ldentification of predicate device(s):

IL Test Beta-2-Microglobulin (Instrumentation Laboratory Co.) K943686

Description of the device/intended use(s):

The Quantia Beta-2 Microglobulin is intended as a latex particle enhanced immunoturbidimetric assay for the in vitro quantitative determination of beta-2-microglubulin concentration in human serum or plasma (EDTA) on the AEROSET® instrument as an aid in the diagnosis of active rheumatoid arthritis and kidney disease.

Quantia PROTEINS Control is intended for use in monitoring the quality control of results obtained with the Quantia Beta-2 Microglobulin and Quantia A1-AT reagents by turbidimetry. (NOTE: This control has been also 510 (k) FDA submitted for use with Quantia A1-AT) For in vitro diagnostic use.

Quantia Beta-2 Microglobulin Standard is intended for use in establishing the calibration curve for the Quantia Beta-2 Microglobulin reagents by turbidimetry. For in vitro diagnostic use.

Statement of Technological Characteristics of the Device Compared to Predicate Device:

Quantia Beta-2 Microglobulin is substantially equivalent to the commercially available predicate device, IL Test Beta-2-Microglobulin, in performance and intended use.

{1}------------------------------------------------

Summary of Performance Data:

In a method comparison study 105 serum samples with Beta-2 Microglobulin levels ranging from 1.01 to 101.1 mg/L were evaluated on the Abbott AEROSET® instrument. The slope was 0.876 and the correlation coefficient (r) was 0.9986 for the Quantia Beta-2 Microglobulin versus the predicate device.

Within run precision was assessed using the Quantia PROTEINS Controls I and II (low and high) and a third control spiked with pure Beta-2 Microglobulin at 10-15 mg/L, run in duplicate twice a day over twenty days (following the NCCLS EP5-A guideline), on the Abbott AEROSET® instrument. Control I gave a CV of 1.1 % (at a mean of 1.00 mg/L), Control II gave a CV of 0.9 % (at a mean of 5.46 mg/L) and the third control gave a CV of 0.9 % (at a mean of 13.61 mg/L). Total run precision gave a CV of 1.9% for Control I, 1.3% for Control II and 1.1% for the third control.

Linearity was assessed according to NCCLS EP6-A guideline. The assay was found to be linear within the following ranges:

  • 0.25 to 16 mg/L without the automatic rerun capability.
  • 0.02 to 100 mg/l with the automatic rerun capability.

The Limit of Quantification was defined as the minimum quantity of analyte that can be measured with a within-run CV below 20% and an error below ± 20 %. Without the automatic rerun capability, the Limit of Quantification for the Beta-2 Microglobulin assay is 0.25 mg/L and, with the automatic rerun capability, 0.02 mg/L.

The Interference study was assessed using the NCCLS guideline EP7-A. Interference testing is summarized as follows:

No significant interference from lipemia up to sample absorbance of 2.1 AU/cm at 660 nm. No significant interference from triglycerides up to concentrations of 1300 mg/dL. No significant interference from bilirubin up to concentrations of 20 mg/dL. No significant interference from haemoglobin up to concentrations of 480 mg/dL. RF interference is below 10% up to 288 IU/mL.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image shows the text "Biokit S.A. c/o Ms Joan Guixer Quality Assurance & Regulatory Affairs Director Can Malé". The text appears to be an address or contact information for a company or individual. The text is black and is on a white background.

Barcelona 08186, Spain

Lliçà d'Amunt

AUG 1 6 2005

Food and Drive Administration 2098 Gaither Road Rockville MD 20850

Re: K050613 Trade/Device Name: Ouantia Beta-2 Microglobulin Quantia Proteins Control Quantia Beta-2 Microglobulin Standard Regulation Number: 21 CFR 866.5630 Regulation Name: Beta-2-microglobulin Immunological Test System Regulatory Class: Class II Product Code: JZG, JJS, JJX Dated: March 3, 2005 Received: March 10, 2005

Dear Ms. Guixer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally

{3}------------------------------------------------

Page 2

marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you attitle Office of Compliance at (240) 276-0131. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other Whoormation on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Robert L. Becker h

Robert L. Becker, Jr., M.D., PhD Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K050613

Device Name: Quantia Beta-2 Microglobulin

Indications for Use:

The Quantia Beta-2 Microglobulin is intended as a latex particle enhanced immunoturbidimetric assay for The Quantitative determination of beta-2-microglubulin concentration in human serum or plasma the ?? Viro quarklative accension in the diagnosis of active rheumatoid arthritis and kidney disease.

Quantia PROTEINS Control is intended for use in monitoring the quality control of results obtained with the Quantia Beta-2 Microglobulin and Quantia A1-AT reagents by turbidimetry. (NOTE: This control has been also 510(k) FDA submitted for use with A1-AT) For in vitro diagnostic use

Quantia Beta-2 Microglobulin Standard is intended for use in establishing the calibration curve for the Quantia Beta-2 Microglobulin reagents by turbidimetry. For in vitro diagnostic use.

Prescription Use × (Part 21 CFR 801 Subpart D) OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mana Chan
Division Sign Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K050613

§ 866.5630

Beta -2-microglobulin immunological test system.(a)
Identification. Abeta -2-microglobulin immunological test system is a device that consists of the reagents used to measure by immunochemical techniquesbeta -2-microglobulin (a protein molecule) in serum, urine, and other body fluids. Measurement ofbeta -2-microglobulin aids in the diagnosis of active rheumatoid arthritis and kidney disease.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.