(154 days)
The device is intended for the intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home.
The TD-1107/TD-1110 ear/skin/surface IR thermometer is characterized by measuring human body temperature and object's temperature in the ear canal and at the surface, respectively. It utilizes infrared technology to measure either infrared energy emitted from the eardrum and surrounding tissues or the surface radiation of the object when making a temperature measurement. It is able to detect skin temperature (only as a reference) when aimed at the target surface of human body. In addition, it is together with an ambient temperature sensor to monitor room temperature. A clock is integrated into the thermometer to provide what time it is.
The provided text describes a 510(k) summary for the TaiDoc TD-11 series ear/skin/surface IR thermometer. It outlines the device's intended use, comparison to a predicate device, and compliance with various consensus standards. However, it does not provide detailed information about specific acceptance criteria or the specifics of a study proving the device meets those criteria, beyond stating general compliance with standards.
Here's a breakdown of what can and cannot be extracted from the provided document based on your request:
1. Table of acceptance criteria and the reported device performance
The document states that the device conforms to consensus standards, specifically mentioning ASTM E1965-98 for infrared thermometers and IEC 60601-1 / IEC 60601-1-2 for safety. It also states: "For clinical results, differences were within clinical acceptability and repeatability was statistically and clinically acceptable."
- Acceptance Criteria (Implicit from standards cited):
- ASTM E1965-98: This standard sets specifications for infrared thermometers used for determining patient temperature. Specific numerical tolerances for accuracy and precision would be defined within this standard.
- IEC 60601-1: General requirements for safety of medical electrical equipment.
- IEC 60601-1-2: Electromagnetic compatibility requirements.
- Reported Device Performance:
- "differences were within clinical acceptability"
- "repeatability was statistically and clinically acceptable"
A table cannot be constructed with specific numerical acceptance criteria and reported device performance from the provided text, as those details are not explicitly given. The document only references compliance with general standards.
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not specified. The document only states "The clinical and non-clinical studies were conducted to validate the effectiveness of use."
- Data Provenance (e.g., country of origin, retrospective or prospective): Not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This information is not provided in the document. The document refers to "clinical acceptability" and "statistically and clinically acceptable repeatability" but does not detail how these were determined or what ground truth method was used.
4. Adjudication method for the test set
- This information is not provided in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This is not applicable as the device is an infrared thermometer, not an AI-assisted diagnostic tool requiring human reader interpretation of images or data.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This is not applicable as the device is a physical thermometer with direct output, not an algorithm, and is intended for use by individuals (human-in-the-loop, but not as an AI interpretation). The "standalone" performance here would refer to the device's accuracy and repeatability as a measurement tool. The document implies standalone performance was evaluated through compliance with standards.
7. The type of ground truth used
- While not explicitly stated as "ground truth," for a thermometer, the "ground truth" for accuracy and repeatability studies typically involves comparison against a highly accurate reference thermometer (e.g., a calibrated laboratory standard thermometer or a rectal thermometer in clinical settings). The document mentions "clinical acceptability" and "repeatability," suggesting such comparisons were made, but it does not specify the exact methodology or reference standard used.
8. The sample size for the training set
- Not applicable/Not provided. This device is a traditional infrared thermometer where "training sets" in the AI sense are not relevant. Its performance is based on its physical design and calibration, not machine learning.
9. How the ground truth for the training set was established
- Not applicable/Not provided for the same reasons as #8.
{0}------------------------------------------------
JUL 2 7 2005
1 of 3
510 (k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA 1990 and CFR 807.92.
- Company Information
| Company name: | TaiDoc Technology Corporation |
|---|---|
| Address: | 4F, No.88, Sec. 1, Kwang-Fu RD.,San-Chung, Taipei County, Taiwan, R.O.C., 241 |
| Phone: | 886-2-6635-8080 |
| FAX: | 886-2-6635-5959 |
-
- Device Identification
TaiDoc TD-11 series ear/skin/surface IR thermometer Model no.:
- Device Identification
CLEVER TD-1107 ear/skin/surface IR thermometer CLEVER TD-1110 ear/skin/surface IR thermometer
-
- Predicate Device
AViTA TS-802 3 in 1 Ear/Forehead/Room thermometer (K031503) by AVITA Corporation.
- Predicate Device
-
- Device Description
The TD-1107/TD-1110 ear/skin/surface IR thermometer is characterized by measuring human body temperature and object's temperature in the ear canal and at the surface, respectively. It utilizes infrared technology to measure either infrared energy emitted from the eardrum and surrounding tissues or the surface radiation of the object when making a temperature measurement. It is able to detect skin temperature (only as a reference) when aimed at the target surface of human body. In addition, it is together with an ambient temperature sensor to monitor room temperature. A clock is integrated into the thermometer to provide what time it is.
- Device Description
-
- Intended Use
The TD-1107/TD-1110 ear/skin/surface IR thermometer is an electronic thermometer using an infrared sensor to detect human body temperature from the ear canal in the home. It is also available to detect object's surface
- Intended Use
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains handwritten text. The text at the top reads "2 of 3". Below that, there is another line of text that appears to be "K050463". The handwriting is somewhat rough, but the characters are generally legible.
temperature including human skin temperature. It is for use on people of all ages.
-
- Comparison to Predicate Device
The TD-1107/TD-1110 ear/skin/surface IR thermometer is substantially equivalent in intended use, technology characteristics, and performance characteristics to the AViTA TS-802 3 in 1 Ear/Forehead/Room thermometer.
- Comparison to Predicate Device
7. Compliance with Consensus Standards
The TD-1107/TD-1110 ear/skin/surface IR thermometer conformed to consensus standards and other standards that include:
-
ASTM E1965-98 �
intermittent -Standard specification for infrared thermometers for determination of patient temperature -
IEC 60601-1:1988 �
-Medical electrical equipment -part 1 general requirements for safety, 1998; Amendment 1, 1991-11, Amendment 2, 1995-03 -
IEC 60601-1-2:2001 �
-
Medical electrical equipment -part 1 general requirements for safety; electromagnetic compatibility –requirements and tests
-
� EN 1441:1997
-Medical device Risk analysis
ISO 13485:2003 •
-Medical devices-Quality management systems-Requirements for requlatory purposes
� ISO 15223:2000
-Medical devices-Symbols to be used with medical device labels, labeling and information to be supplied
ISO 14971:2000 +
-Medical devices-Application of risk management to medical devices
8. Performance Studies
The clinical and non-clinical studies were conducted to validate the effectiveness of use. The results were compliance to applicable voluntary standards includes ASTM E1965-98, as well as IEC 60601-1 and IEC
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a handwritten note with the text "3 of 3" at the top. Below that, there is a sequence of characters that appear to be "KO504063". The handwriting is somewhat rough, but the characters are still legible.
60601-1-2 requirements. For clinical results, differences were within clinical acceptability and repeatability was statistically and clinically acceptable.
-
- Conclusion
The TD-1107/TD-1110 ear/skin/surface IR thermometer:
- Conclusion
-
conforms to consensus standards that are applicable to infrared ear thermometers
-
has the similar intended use and technological characteristics as the 510(k) predicate device of AViTA model TS-802
Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of the predicate device. Those engineering changes do not (1) affect the intended use or (2) alter the fundamental scientific of the device.
ү
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three heads, representing health, human services, and well-being. The eagle is surrounded by a circular border with the words "U.S. Department of Health and Human Services, USA" printed around it.
JUL 27 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Shu-Mei Wu Project Manager TaiDoc Technology Corporation 4F, No. 88, Sec. 1, Kwang-Fu Road, San-Chung, Taipei CHINA (Taiwan) 241
Re: K050463
Trade/Device Name: TaiDoc TD-11 Series Ear/Skin/Surface IR Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: May 24, 2005 Received: May 26, 2005
Dear Mr. Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Wu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Signature
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "1 of 1" on the top line and "K050463" on the second line. The text is written in a handwritten style. The numbers are clearly visible and the letters are legible.
Indications for Use
510(k) Number:
Device Name: TaiDoc TD-11 series ear/skin/surface IR thermometer
Indications for Use:
The device is intended for the intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Shila H. Murphy, MD for AD Watson
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number. K050463
Page 1 of ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
§ 880.2910 Clinical electronic thermometer.
(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.