K Number
K050303
Date Cleared
2005-02-25

(17 days)

Product Code
Regulation Number
880.2910
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device measures the body temperature of a patient by means of a sensor (transducer) coupled with an electronic signal amplification, conditioning and digital 1.CD (display) unit. The device is reusable and intended for oral, axillary or rectal temperature measurements.

Device Description

The 5 Sec Digital Thermometer, models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X, and ST8B30X/ST8B40X, are the electronic thermometers by using a thermistor as the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) -- controlled circuit, which is considered the hard-wire control instead of programmable control.

From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts.

This system uses a 1.5V DC battery for operation of complete system whenever the battery is low, the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Regarding the performance of 5 Sec Digital Thermometer, models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X and ST8B30X/ST8B40X it was designed and verified according to the US standard ASTM E 1112-00.

AI/ML Overview

The provided text describes the 510(k) summary for the Mesure 5 Sec Digital Thermometer. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study proving the device meets specific acceptance criteria with a novel AI algorithm. Therefore, many of the requested details related to AI-specific studies (e.g., multi-reader multi-case studies, standalone algorithm performance, AI training/test set details) are not applicable or findable in this document.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance data in the typical sense of a clinical trial for an AI-powered device. Instead, it states that the device was designed and verified according to the US standard ASTM E 1112-00. This standard defines performance requirements for electronic thermometers.

Acceptance Criteria (from ASTM E 1112-00, as implied by compliance):

Acceptance Criteria CategorySpecific Criteria (Implied by ASTM E 1112-00 compliance)Reported Device Performance
Accuracy (In Vivo)Expected to meet stated accuracy requirements for clinical thermometers (e.g., ±0.1°C or ±0.2°F for certain temperature ranges, ±0.2°C or ±0.3°F for others). Precision, repeatability, and response time are also parts of this standard.The document states, "Regarding the performance of 5 Sec Digital Thermometer... it was designed and verified according to the US standard ASTM E 1112-00." This implies that the device meets the accuracy and other performance specifications defined by ASTM E 1112-00. Specific numerical performance values beyond compliance with the standard are not provided in this summary.
Safety and EMC (Non-Clinical)Compliance with EN 60601-1 (general safety), EN 60601-1-1 (safety requirements for medical electrical systems), and EN 60601-1-2 (electromagnetic compatibility)."Compliance to applicable voluntary standards includes ASTM E1112: 2000, as well as EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement." This indicates the device meets these safety and EMC standards.
Technological CharacteristicsThe new device should maintain safety and effectiveness and not alter the fundamental scientific technology compared to the predicate device."The Mesure 5 Sec Digital Thermometer... have the same intended use and technological characteristics as the cleared device... Moreover, verification and validation tests contained in this submission demonstrate that the difference... could maintain the same safety and effectiveness as that of cleared device."
Intended UseThe device should measure body temperature for oral, axillary, or rectal measurements for home use on people of all ages.The stated intended use is identical to the predicate device, and the changes (sensor wire material and wiring construction) are described as small and not affecting intended use.

Study Proving Acceptance Criteria:

The study proving the device meets the acceptance criteria is referred to as "verification and validation tests" conducted according to the US standard ASTM E 1112-00 and other relevant safety standards (EN 60601-1 series). These are typically non-clinical laboratory tests to ensure accuracy, precision, and safety parameters.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified. For a digital thermometer conforming to ASTM E 1112-00, the "test set" would typically refer to the number of individual thermometer units tested, the number of temperature measurements taken under controlled conditions (e.g., water bath), and potentially a small number of human subjects for limited clinical validation (though in-depth clinical trials are not usually required for substantial equivalence of such devices). The document does not provide these numbers.
  • Data Provenance: Not specified. However, the company is based in Taiwan (Mesure Technology Co., Ltd., Taipei Hsien, Taiwan, R.O.C.), suggesting testing likely occurred in Taiwan or through certified labs. The submission is to the FDA for the US market. The term "retrospective or prospective" is not applicable as this concerns a medical device's physical performance, not an AI model's data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This query is typically relevant for AI/imaging devices where ground truth is established by expert consensus (e.g., radiologists). For a digital thermometer, the "ground truth" for temperature measurement is established by reference standards (e.g., highly accurate temperature baths traceable to national standards or calibration standards). Experts involved would be metrology engineers, technicians, and quality control personnel within the manufacturing and testing facilities, ensuring compliance with standards like ASTM E 1112-00. The document does not specify the number or qualifications of such experts.

4. Adjudication Method for the Test Set

Not applicable in the context of a digital thermometer's performance testing. Adjudication methods (e.g., 2+1, 3+1) are for resolving discrepancies among human readers interpreting medical images or data. For a thermometer, performance is objectively measured against a reference standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. An MRMC study is relevant for AI systems assisting human readers in interpreting medical data (e.g., radiology). This is a simple digital thermometer, not an AI-powered diagnostic device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, in essence. The "device" (thermometer) itself is a standalone measurement tool. Its performance is evaluated intrinsically through laboratory testing against metrological standards, not in conjunction with human interpretation in the way an AI algorithm might be. However, this is not an "algorithm" in the sense of a complex AI model. The "algorithm" here is simply the ASIC (Application Specific IC) controlling the circuit to calculate and display the temperature. Its standalone performance is what's verified against ASTM E 1112-00.

7. The Type of Ground Truth Used

The ground truth for a digital thermometer's accuracy and precision testing is established using calibrated reference standards, such as highly stable and accurate temperature-controlled water baths, whose temperatures are known with high certainty and are traceable to national/international measurement standards.

8. The Sample Size for the Training Set

Not applicable. This device is not an AI algorithm that undergoes a "training phase." Its function is determined by its hardware design (thermistor, ASIC circuit) which translates a physical measurement into a digital reading based on fundamental physics and electronics, not machine learning from a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable, as explained in point 8.

{0}------------------------------------------------

Mesure Technalpgy J ి గ్రామాల

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92, and the relevant 510(k) submission ouidance.

The assigned 510(k) number is: ಗಡಿಸಿ 0 3ರ 3 Submitter's Identifications: Company: Mesure Technology Co., Ltd. Address: 7F, No. 88, Sec. 1, KwangFu Road, Sanchung City, Taipei Hsien, Taiwan, R.O.C. Contact person: John Chen / Ph. D.

    1. Name of the Device: 5 Sec Digital Thermometer, models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X and ST8B30X/ST8B40X.
    1. Information of the 510(k) Cleared Device (Predicate Device): , Thermometer, model ST8631 (K021048) and 10-second Digital Thermometer, model ST87, ST89, ST80, and ST8B (K040156).

3. Device Description:

The 5 Sec Digital Thermometer, models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X, and ST8B30X/ST8B40X, are the electronic thermometers by using a thermistor as the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) -- controlled circuit, which is considered the hard-wire control instead of programmable control.

From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts.

This system uses a 1.5V DC battery for operation of complete system whenever the battery is low, the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Regarding the performance of 5 Sec Digital Thermometer, models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X and ST8B30X/ST8B40X it was designed and verified according to the US standard ASTM E 1112-00.

4. Intended Use:

The device measures the body temperature of a patient by means of a sensor coupled with an electronic signal amplification, conditioning and digital LCD (display) unit. The device is reusable and intended for oral, axially or rectal temperature measurements for home use on people of all ages.

5. Comparison to the 510(k) Cleared Device (Predicate Device):

Since the new models 5 Sec Digital Thermometer, models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X, and ST8B30X/ST8B40X were developed from the cleared device ST8631 through the design control procedures of Measure Technology Co., Ltd. with only the small change in sensor wire material and wiring construction so as to shorten the measuring time, the new device is substantial equivalence to that of device being modified, ST8631(K021048) and 10-second Digital Thermometer, model ST87, ST89, ST89, and ST8B.(K040156).

{1}------------------------------------------------

Mesure Technology Co., Ltd.

    1. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
      Compliance to applicable voluntary standards includes ASTM E1112: 2000, as well as EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement.

7 Conclusions

The Mesure 5 Sec Digital Thermometer, models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X, and ST8B30X/ST8B40X have the same intended use and technological characteristics as the cleared device of Measure's model ST8631. Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate could maintain the same safety and effectiveness as that of cleared device.

In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device ST8631 (K021048) and 10-second Digital Thermometer, model ST87, ST89, ST80, and ST8B (K040156).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a circular logo for the Department of Health & Human Services - USA. The logo features the department's name in a circular arrangement around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird in flight, composed of three curved lines representing wings and two curved lines representing the body.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 5 2005

Mr. John Chen Mesure Technology Company Limited 7F, No. 86/88 Sec 1, Kwang Fu Road Sanchung City, Taipei Hsien, TAIWAN

Re: K050303 Trade/Device Name: 5 Sec Digital Clinical Thermometer / Models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST840X, and ST8B30X/ST8B40X. Regulation Number: 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: February 5, 2005 Received: February 8, 2005

Dear Mr. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Chen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications For Use

14050303 510(k) Number (if known):

Device Name: 5 Sec Digital Clinical Thermometer / Models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X, and ST8B30X/ST8B40X.

Indications For Use:

The device measures the body temperature of a patient by means of a sensor (transducer) coupled with an electronic signal amplification, conditioning and digital 1.CD (display) unit. The device is reusable and intended for oral, axillary or rectal temperature measurements.

Prescription Use (Part 21 CFR 801 Subpart D) OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Cim

Seneral Hospital.

Page 1 of 1

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.