K Number
K050185
Manufacturer
Date Cleared
2005-05-26

(119 days)

Product Code
Regulation Number
880.5970
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PowerHohn and PowerLine Catheters are indicated for short or long term access to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal and power injection of contrast media. The maximum recommended infusion rate is 5cc/sec. The maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi.

Device Description
  • The PowerHohn and PowerLine Catheters are open-ended radiopaque polyurethane . catheters.
  • Catheter size is 5 Fr SL with 50 cm usable length. .
  • The catheter has a reverse taper design. ●
  • Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a . reference for the catheter insertion point.
  • PowerHohn and PowerLine Catheters are provided in sterile tray configurations. .
  • Purple colorants were added to the catheter materials to provide the catheter with an ● I urpic colorants were added to arer to differentiate the PowerHohn and PowerLine as power injectable from other central venous catheters.
  • The molded hub is labeled to identify the catheter as PowerHohn and PowerLine. .
  • The catheter extension leg and clamp are labeled with information to facilitate proper use of . the device.
AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the 5 Fr SL PowerHohn™ and PowerLine™ Catheters:

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary is a high-level summary and does not explicitly list specific numerical acceptance criteria or detailed performance data in a tabular format. Instead, it makes general statements about meeting predetermined criteria and substantial equivalence.

Acceptance Criterion (Inferred from text)Reported Device Performance
Safety and Effectiveness"met all the predetermined performance criteria of design verification evaluations" (implies safety and effectiveness were considered)
Equivalence to Predicate Devices"substantially equivalent to the predicate devices"
Intended Use"The PowerHohn and PowerLine Catheters met all the acceptance criteria of the testing performed" (specific to their stated intended use)
Power Injectability (Max Infusion Rate)"maximum recommended infusion rate is 5cc/sec"
Power Injectability (Max Pressure)"maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi"
Material Biocompatibility"There are no new materials therefore, no new biocompatibility is required." (Implies predicate devices' biocompatibility is acceptable)
Compliance with Standards"The FDA's Guidance on Premarket Notification [510(k)] Submission... and relevant ISO 10555 Standards were used"

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify the sample size used for the test set. It mentions "verification testing was performed" but provides no numerical details.
  • Data Provenance: The document does not provide information on the country of origin of the data or whether it was retrospective or prospective. It is a submission for regulatory clearance in the US, so the testing would typically be conducted by or for the manufacturer.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document describes a medical device (catheter) rather than a diagnostic algorithm or imaging system. Therefore, the concept of "ground truth" established by experts in the context of diagnostic interpretation (e.g., radiologists) is not directly applicable here. The "ground truth" for a device like this would be its physical properties and functional performance, measured through engineering tests against established specifications.

4. Adjudication Method for the Test Set

As this is a device performance study rather than a diagnostic interpretation study, adjudication methods like 2+1 or 3+1 are not relevant. Device performance is typically determined by measurements and physical tests against predefined criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. This is a 510(k) submission for a medical device (catheter), not an AI algorithm. Therefore, an MRMC comparative effectiveness study, which is common for AI tools in diagnostics, was not performed. The study aimed to demonstrate substantial equivalence to predicate devices, not improvement over human performance with AI assistance.

6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) was Done

N/A. This is a physical medical device, not an algorithm. Therefore, the concept of a "standalone" algorithm performance study does not apply.

7. The Type of Ground Truth Used

For this type of device, the "ground truth" would be established through:

  • Engineering specifications and design requirements: These define the physical and functional properties the catheter must possess.
  • Performance metrics derived from relevant standards: Such as ISO 10555 for intravascular catheters, which would specify test methods and acceptance limits (e.g., tensile strength, flow rates, pressure limits, material properties).
  • Predicate device characteristics: The performance of the predicate devices (PowerPICC™ Catheter and Broviac® Catheter) would serve as a benchmark for substantial equivalence.

The summary states: "Verification testing was performed according to protocols based on the above-referenced guidance document recommendations and additional standards." This indicates that the "ground truth" was based on a combination of regulatory guidance, industry standards, and the characteristics of the predicate devices.

8. The Sample Size for the Training Set

N/A. This is a physical medical device, not a machine learning algorithm. Therefore, there is no "training set" in the sense of data used to train an AI model. The "training" for a physical device involves its design and manufacturing processes, and the verification testing confirms its performance.

9. How the Ground Truth for the Training Set Was Established

N/A. As there is no "training set" for an AI model, this question is not applicable. The "ground truth" for the device's design and manufacturing is established through engineering principles, material science, and adherence to quality systems and regulatory standards.

{0}------------------------------------------------

MAY 2 6 2005

510(k) Summary of Safety and Effectiveness Information

5 Fr SL PowerHohn™ and PowerLine™ Catheters

Submitter Information 6.1

Submitter Name:Bard Access Systems, Inc. (BAS)[Subsidiary of C. R. Bard, Inc.]
Address:5425 W. Amelia Earhart DriveSalt Lake City, UT 84116
Telephone Number:(801) 595-0700, Ext. 5541
Fax Number:(801) 595-4903
Contact Person:Charles Morreale
Date of Preparation:January 17, 2005

Device Name 6.2

Device Name:PowerHohn™ and PowerLine™ Catheter
Trade Name:PowerHohn™ and PowerLine™ Catheter
Common/Usual Name:Central Venous Catheter
Classification Name:Class II, 80 LJS - Long Term Intravascular Catheter
Classification Panel:General Hospital

Predicate Device(s): 6.3

Device Name: Trade Name: Common/Usual Name: Classification Name: Classification Panel: Premarket Notification:

Device Name: Trade Name: Common/Usual Name: Classification Name: Classification Panel: Premarket Notification: PowerPICC™ Catheter PowerPICCTM Catheter Peripherally Inserted Central Catheter (PICC) Class II, 80 LJS - Long Term Intravascular Catheter General Hospital K033389, Substantial Equivalence: April 14, 2004

Broviac® Catheter Broviac® Catheter Central Venous Catheter Class II, 80 LJS - Long Term Intravascular Catheter General Hospital K830256, Substantial Equivalence: March 1, 1983

{1}------------------------------------------------

K050185

2 of 3

6.4 Device Description

  • The PowerHohn and PowerLine Catheters are open-ended radiopaque polyurethane . catheters.
  • Catheter size is 5 Fr SL with 50 cm usable length. .
  • The catheter has a reverse taper design. ●
  • Catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a . reference for the catheter insertion point.
  • PowerHohn and PowerLine Catheters are provided in sterile tray configurations. .
  • Purple colorants were added to the catheter materials to provide the catheter with an ● I urpic colorants were added to arer to differentiate the PowerHohn and PowerLine as power injectable from other central venous catheters.
  • The molded hub is labeled to identify the catheter as PowerHohn and PowerLine. .
  • The catheter extension leg and clamp are labeled with information to facilitate proper use of . the device.

Intended Use 6.5

PowerHohn™ and PowerLine™ Catheters are indicated for short or long term access to the central I owcirrom - and rower hasigned for administering I.V. fluids, blood products, drugs and parenteral venous system. "They are designed withdrawal and power injection of contrast media. The maximum recommended infusion rate is 5cc/sec. The maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi.

Technological Characteristics Summary 6.6

New device is compared to Marketed Device

Yes.

Does the new device have the same indication statement?

Yes. However, the Indications For Use were expanded to provide instructions for chest placement, percutaneous introduction and tunneling techniques for central venous catheters.

Do the differences alter the intended therapeutic/diagnostic/etc. effect (i.e. deciding may consider impact on safety and effectiveness)?

No, the differences do not alter the intended use of the device.

Does the new device have the same technological characteristics, e.g. design, material, etc.?

Yes. The PowerHohn and PowerLine catheters are identical to the PowerPICC catheter in terms of catheter design, materials, color (purple), and manufacture, with the exception of the addition of a SureCuff® Tissue Ingrowth Cuff to the PowerLine Catheter and an optional VitaCuff® Antimicrobial Cuff to the PowerHohn Catheter. The basic fundamental scientific technology of the catheter has not changed.

Could the new characteristics affect safety or effectiveness?

Yes. The new characteristics could affect safety or effectiveness of the device.

Do the new characteristics raise new types of safety and effectiveness questions?

{2}------------------------------------------------

No. There are no new types of safety and effectiveness questions.

Do accepted scientific methods exist for assessing effects of the new characteristics?

Yes. The FDA's Guidance on Premarket Notification [510(k)] Submission for Short-Term and I cs. The I Drive Catheters, dated 3/16/95, and relevant ISO 10555 Standards were used toong Term Intrasterning the methods for evaluating the modified device's performance.

There are no new materials therefore, no new biocompatibility is required.

Are performance data available to assess effects of new characteristics?

Yes. Verification testing was performed according to protocols based on the above-referenced guidance document recommendations and additional standards.

Do performance data demonstrate equivalence?

Yes. Performance data gathered in design verification testing demonstrated that the PowerHohn and PowerLine Catheter is substantially equivalent to the predicate devices.

Conclusion

The PowerHohn and PowerLine Catheters met all the predetermined performance criteria of design verification evaluations and is substantially equivalent to the predicate devices.

6.7 Conclusion

The PowerHohn and PowerLine Catheters meet all the acceptance criteria of the testing performed and, based on FDA 's decision tree, is substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the symbol.

MAY 2 6 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Charles Morreale Senior Manager Quality Systems/ Regulatory Affairs Bard Access Systems, Incorporated 5425 West Ameila Earhart Drive Salt Lake City, Utah 84116

Re: K050185

Trade/Device Name: 5 Fr SL PowerHohn and PowerLine Catheters Regulation Number: 880.5970 Regulation Name: Percutaneous Implanted Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: April 1, 2005 Received: April 8, 2005

Dear Mr. Morreale:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Morreale

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Sydette J. Michailoms.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K050185

1 of 1

PowerHohn™ and PowerLine™ Catheters 510(k) Premarket Notification

STATEMENT OF INDICATION FOR USE

510(k) Number (if known): K050185

Device Name: 5 Fr SL PowerHohn and PowerLine Catheters

Indications For Use:

PowerHohn and PowerLine Catheters are indicated for short or long term access to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal and power injection of contrast media. The maximum recommended infusion rate is 5cc/sec. The maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi.

Prescription Use X (Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Rumm for MW

wision Sign-Off) Mislon Sign-Off)
Wision of Anesthesiology, General Hospital, Intection Control, Dental Devices

2(k) Number: KCSC183

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”