(24 days)
The Medtronic Trillium™ BioPump Plus is indicated for use only with the Medtronic Bio-Console to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary byrass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g. valvuloplasty, circulatory support during mitral valve reoperation, surgery of vena cava or aorta, liver transplants, etc.)
The Trillium® BioPump Plus is a centrifugal blood pump which is a single use, disposable, non-pvrogenic device designed to move blood through the extracorporeal circuit by centrifugal forces created by smooth rotating cones. It is the disposable portion of the pumping system and is electromechanically coupled to the BioConsole which is the instrument that monitors and displays the flow and pressure of the blood flow. Venous blood enters the inlet port of the Bio-Pump. The smooth cones rotate in the polycarbonate housing. As the cones rotate, energy in the form of pressure and velocity is transferred from the cones to the blood. The blood is gently accelerated towards the outlet of the pump. The Bio-Pump moves the blood through the circuit at a desired pressure and flow rate by increasing or decreasing the speed of the rotating cones. This is accomplished by adjusting the BioConsole's RPM (revolutions per minute). The arterial blood is returned to the patient from the extracorporeal circuit.
The provided text describes a "SPECIAL 510(k)" submission for a medical device called the Trillium® BioPump Plus. This submission aims to demonstrate that a modified version of an already marketed centrifugal pump is substantially equivalent to existing devices, primarily due to the addition of a Trillium coating to blood-contacting surfaces.
Here's an analysis based on the provided text, addressing your specific points:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" in a quantified, numerical table format, nor does it provide a direct "reported device performance" in terms of specific metrics. Instead, it relies on a comparative approach, stating that the Trillium® BioPump Plus should perform comparably to predicate devices.
The core "acceptance criteria" can be inferred as:
- Safety and Effectiveness Equivalence: The Trillium® BioPump Plus does not significantly affect safety and effectiveness when compared to the predicate devices.
- No New Safety/Effectiveness Questions: The technological characteristic (Trillium coating) does not raise new types of safety or effectiveness questions.
- Acceptable Scientific Methods: Acceptable scientific methods exist for assessing the effects of this new technological characteristic.
The reported device performance is qualitative: "The biocompatibility and in vitro bench testing demonstrated that when compared to the predicate devices, the Trillium®BioPump Plus does not significantly affect safety and effectiveness and is substantially equivalent to other commercially distributed extracorporeal cardiopulmonary devices."
Table of Inferred Acceptance Criteria and Reported Performance
| Acceptance Criteria (Inferred) | Reported Device Performance (Qualitative) |
|---|---|
| No significant adverse effect on safety and effectiveness | "The biocompatibility and in vitro bench testing demonstrated that... does not significantly affect safety and effectiveness..." |
| No new types of safety or effectiveness questions | "...these 'technological characteristics do not raise new types of safety or effectiveness questions'." |
| Existence of acceptable scientific methods for assessment | "...there are acceptable scientific methods which exist for assessing effects of these new technological characteristics." |
| Substantial Equivalence to predicate devices (BPX80 BioPump Plus, AFFINITY® Hollow Fiber Oxygenator with Trillium® Biopassive Surface) | "...is substantially equivalent to other commercially distributed extracorporeal cardiopulmonary devices." and "These 'performance data demonstrate' that the Trillium® BioPump Plus is substantially equivalent to other marketed extracorporeal cardiopulmonary bypass devices." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "biocompatibility and in vitro bench testing," which implies a test set was used. However, it does not specify the sample size for these tests (e.g., number of devices tested, number of in vitro samples).
The data provenance is not explicitly stated regarding country of origin, retrospective/prospective nature, or whether it involved human or animal subjects (though "biocompatibility" often involves in vitro and sometimes in vivo studies). Given it's a "SPECIAL 510(k)" for a modification to an existing device, the testing would likely be focused on the impact of the coating.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is not applicable to the provided document. The Trillium® BioPump Plus is a mechanical medical device (a centrifugal blood pump), not an AI or diagnostic imaging device that would typically involve human expert interpretation to establish a ground truth for a test set. The validation relies on engineering performance parameters, biocompatibility, and material equivalence rather than expert consensus on diagnostic interpretations.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is not applicable for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are used in studies where a "ground truth" needs to be established from subjective expert opinions, typically in diagnostic tasks like image interpretation. This device's testing would involve objective measurements against engineering specifications and biocompatibility standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable. An MRMC study is relevant for diagnostic devices (especially those involving AI for interpretation) where the performance of human readers, with and without AI assistance, is compared across multiple cases. The Trillium® BioPump Plus is a therapeutic/support device, not a diagnostic one involving human "readers."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This section is not applicable. The device is a physical, mechanical blood pump, not an algorithm. The concept of "standalone" performance without human-in-the-loop is relevant for AI algorithms. The device functions as part of a human-controlled extracorporeal circuit.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Given the nature of the device, the "ground truth" would be established through objective scientific and engineering principles:
- Biocompatibility Standards: Compliance with recognized biocompatibility standards (e.g., ISO 10993 series) for materials in blood contact.
- In Vitro Bench Testing Parameters: Specific performance metrics like flow rates, pressure generation, hemolysis indices, thrombogenicity, and durability under simulated physiological conditions.
- Material Equivalence: Verification that the Trillium coating behaves as expected and is comparable to existing applications (e.g., AFFINITY® Hollow Fiber Oxygenator with Trillium® BioSurface) in terms of blood compatibility.
The document implicitly states that "acceptable scientific methods exist for assessing effects of these new technological characteristics," which refers to these types of objective tests and standards.
8. The sample size for the training set
This section is not applicable. The Trillium® BioPump Plus is a physical device, not an AI model that requires a "training set."
9. How the ground truth for the training set was established
This section is not applicable for the same reasons as point 8.
{0}------------------------------------------------
KOSD109
SUMMARY OF SAFETY AND EFFECTIVENESS
COMPANY AND CONTACT PERSON
Medtronic Perfusion Systems 7611 Northland Drive Minneapolis, MN 55428 (763) 391-9533 Tel: FAX: (763) 391-9603
Preeti Jain, Director, Regulatory/Clinical Affairs
DEVICE NAME
Trillium® BioPump Plus
NAME OF PREDICATED OR LEGALLY MARKETED DEVICE
BPX80 BioPump Plus (K973011) AFFINITY® Hollow Fiber Oxygenator with Trillium® Biopassive Surface (K973760)
DESCRIPTION OF DEVICE
The Trillium® BioPump Plus is a centrifugal blood pump which is a single use, disposable, non-pvrogenic device designed to move blood through the extracorporeal circuit by centrifugal forces created by smooth rotating cones. It is the disposable portion of the pumping system and is electromechanically coupled to the BioConsole which is the instrument that monitors and displays the flow and pressure of the blood flow. Venous blood enters the inlet port of the Bio-Pump. The smooth cones rotate in the polycarbonate housing. As the cones rotate, energy in the form of pressure and velocity is transferred from the cones to the blood. The blood is gently accelerated towards the outlet of the pump. The Bio-Pump moves the blood through the circuit at a desired pressure and flow rate by increasing or decreasing the speed of the rotating cones. This is accomplished by adjusting the BioConsole's RPM (revolutions per minute). The arterial blood is returned to the patient from the extracorporeal circuit.
STATEMENT OF INTENDED USE
The Medtronic Trillium® BioPump Plus is indicated for use only with the Medtronic Bio-Console to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g. valvuloplasty, circulatory support during mitral valve reoperation, surgery of vena cava or aorta, liver transplants, etc.)
Trillium® BioPump Plus
{1}------------------------------------------------
STATEMENT OF INTENDED USE OF PREDICATED/MARKETED DEVICE
The Medtronic BioPump Plus is indicated for use only with the Medtronic Bio-Console to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory pupport for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six lhours) not requiring complete cardiopulmonary bypass (e.g. valvuloplasty, circulatory nually having mitral valve reoperation, surgery of vena cava or aorta, liver transplants, etc.)
STATEMENT OF TECHNOLOGICAL CHARACTERISTICS COMPARISON
Information regarding technological characteristics comparison is provided in the following section, "Determination of Substantial Equivalence".
DETERMINATION OF SUBSTANTIAL EQUIVALENCE
This "SPECIAL 510(k)" is being submitted for a modification to the BioPump Plus. The modification to the currently marketed centrifugal pump is to coat the blood contact surfaces with Trillium.
The Trillium™ BioPump Plus is being compared to the following marketed devices:
- BPX80 BioPump Plus Centrifugal Blood Pump (K973011) .
- AFFINITY® Hollow Fiber Oxygenator with Trillium® Biopassive Surface (K973760) .
The Trillium®BioPump Plus has the same indications statement and intended uses as the presently marketed BioPump Plus.
The Trillium® BioPump Plus has "no new technological characteristics (e.g., materials and manufacturing processes)" from the currently marketed centrifugal pumps. The technological characteristic is solely the coating material of the blood pathway:
-
Trillium® .
The technological characteristic of the Trillium® BioSurface is common to other medical devices (hollow fiber oxygenators) currently in commercial distribution as follows: -
AFFINITY® Hollow Fiber Oxygenator with Trillium® BioSurface (K973760) .
This technological characteristic "could affect the safety and effectiveness of the device". However, these "technological characteristics do not raise new types of safety or effectiveness questions". In addition, "there are acceptable scientific methods which exist for assessing effects of these new technological characteristics".
"Performance data to assess the effects of these new technological characteristics" has been performed. These "performance data demonstrate" that the Trillium® BioPump Plus is substantially equivalent to other marketed extracorporeal cardiopulmonary bypass devices.
{2}------------------------------------------------
The biocompatibility and in vitro bench testing demonstrated that when compared to the predicate devices, the Trillium®BioPump Plus does not significantly affect safety and effectiveness and is substantially equivalent to other commercially distributed extracorporeal cardiopulmonary devices.
.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image is a logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a bird with outstretched wings.
FEB 1 1 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medtronic Perfusion Systems c/o Mr. Preeti Jain Director, Regulatory/Clinical Affairs 7611 Northland Drive N Minneapolis, MN 55428-1088
K050109 Re:
Trillium™ BioPump Plus Regulation Number: 21 CFR 870.4370 Regulation Name: Cardiopulmonary Bypass Non-Roller type blood pump Regulatory Class: Class III (three) Product Code: KFM Dated: January 14, 2005 Received: January 18, 2005
Dear Mr. Jain:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We nave reviewed your Section >10(x) premainer is substantially equivalent (for the indications ferenced above and nave determined the are are are are and and marketed in interstate for use stated in the encrosule for regars manatt date of the Medical Device Amendments, or to commerce prior to May 26, 1776, the enating with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Costnetic Act (Act) that do not require appear al controls provisions of the Act. The You may, therefore, market me actives, policements for annual registration, listing of general controls provisions of the rist labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) the existing major regulations affecting your device can may be subject to such additional controlior Linesting of Parts 800 to 898. In addition, FDA may be found in the Ood or reasis concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Preeti Jain
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease oc actived that i Drimination that your device complies with other requirements of the Act that I DA has made a actor regulations administered by other Federal agencies. You must or any I carated and the Act's requirements, including, but not limited to: registration and listing (21 Comply with an the 11et 0704 and manufacturing practice requirements as set CI K rat 6077, moonling (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provention as described in your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premaint in the motions in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific of Compliance at (301) 594-4646. Additionally, for questions on the contact the Office of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to (2017) 27 1055. 1100, pressed and 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Duna R. Vochner
Image /page/4/Picture/5 description: The image shows what appears to be a signature. The signature is composed of curved lines and strokes, forming a unique and stylized design. The overall impression is that of a handwritten mark, possibly representing a person's name or initials.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): Ko 5 010 9
Device Name:__________________________________________________________________________________________________________________________________________________________________
Indications For Use:
The Medtronic Trillium™ BioPump Plus is indicated for use only with the Medtronic Bio-Console to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary byrass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g. valvuloplasty, circulatory support during mitral valve reoperation, surgery of vena cava or aorta, liver transplants, etc.)
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
oning R. Jackine
Ovision Sign-Off) n ision of Cardiovascula
510(K) Number_KoS 0 Lo 9
§ 870.4360 Nonroller-type blood pump.
(a)
Nonroller-type cardiopulmonary and circulatory bypass blood pump —(1)Identification. A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (
i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
(2)
Classification —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
Nonroller-type temporary ventricular support blood pump —(1)Identification. A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
Classification. Class III (premarket approval).(c)
Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.