K Number
K042418

Validate with FDA (Live)

Date Cleared
2005-03-21

(199 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KD-575 Memory Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive, technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 8.6614 inches to 13.78 inches

Device Description

KD-575 Fuily Automatic Electronic Blood Pressure Monitor is a Non-invasive blood pressure measurement system for only one person each time. Based on oscillometric and silicon integrate pressure sensor technology, this devise is used to monitor systolic, diastolic blood pressure and pulse rate which will be shown on a LCD with an electronic interface module. Swathing the air cuff around the left upper arm 1-2cm above elbow joint automatically inflated and released by an internal pump, the device can analyze the signals promptly and display the results. It can storage and show 120 times measuring result with the month, day, time of measuring.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the KD-575 Fully Automatic Electronic Blood Pressure Monitor:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Standard)Reported Device Performance (Compliance)
ANSI/AAMI SP10-1992 AccuracyMeets or exceeds accuracy requirements

2. Sample Size Used for the Test Set and Data Provenance:

The document states, "Clinical tests were performed and complied the accuracy requirements of ANSI/AAMISP10-1992." However, it does not specify the sample size for the test set or the data provenance (e.g., country of origin, retrospective/prospective nature of the data).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

The document does not specify the number of experts used or their qualifications. For blood pressure monitoring devices, the "ground truth" for accuracy studies typically involves a reference method (e.g., auscultatory measurement by trained observers) rather than expert consensus on images.

4. Adjudication Method for the Test Set:

The document does not mention any adjudication method for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

A multi-reader multi-case (MRMC) comparative effectiveness study was not performed, nor is it applicable for this type of device. This study design is typically used for AI-powered diagnostic imaging tools where human readers interpret medical images. Blood pressure monitors are standalone devices that provide direct measurements.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance:

Yes, a standalone performance study was done. The device itself is an automated blood pressure monitor, and the clinical tests evaluated its direct measurement accuracy against a standard (ANSI/AAMI SP10-1992). This means the device's algorithm and hardware were assessed in isolation to determine its performance.

7. Type of Ground Truth Used:

The ground truth used for assessing the device's accuracy was based on the requirements of the ANSI/AAMI SP10-1992 standard. For blood pressure monitors, this standard typically involves comparing the device's readings against simultaneous auscultatory measurements performed by trained observers using a mercury sphygmomanometer as the reference standard. While the document doesn't explicitly state "auscultatory measurements," adherence to this standard implies such a method.

8. Sample Size for the Training Set:

The document does not mention a training set size. This is expected as this is not an AI/machine learning device that requires a separate training phase with a distinct dataset. Its "algorithm" is embedded in the device's design and calibration.

9. How the Ground Truth for the Training Set Was Established:

As there is no mention of a training set in the context of this traditional medical device, the concept of establishing ground truth for a training set is not applicable. The device's calibration and design are likely based on engineering principles and validated against known standards, rather than by training on a labeled dataset in the machine learning sense.

{0}------------------------------------------------

K042418

Page 1 of 2

510(k) Summary

Identification of the submitter:

Submitter:Kodon (Tianjin) Electronic & Electrical ApparatusCo., LTDNo 31, Changjiang Road, Nankai District, Tianjin,P.R. China, 300193
Telephone number:86-22-60526082
Fax number:86-22-60526162
Contact:Liu Yi
Date of Application:08/25/04

Identification of the product:

Device proprietary Name:KD-575 Memory Automatic Electronic BloodPressure Monitor
Common name:Noninvasive blood pressure measurement systems
Classification name:Noninvasive blood pressure measurement systemClass II per 21 CFR 870.1130

Marketed Devices to which equivalence is claimed:

Devicemanufacture510(k) number
KD-622Kodon (Tianjin) Electronic and ElectricalApparatus Co., Ltd.K030358

Device description:

KD-575 Fuily Automatic Electronic Blood Pressure Monitor is a Non-invasive blood pressure measurement system for only one person each time. Based on oscillometric and silicon integrate pressure sensor technology, this devise is used to monitor systolic, diastolic blood pressure and pulse rate which will be shown on a LCD with an electronic interface module. Swathing the air cuff around the left upper arm 1-2cm above elbow joint automatically inflated and released by an internal pump, the device can analyze the signals promptly and display the results. It can storage and show 120 times measuring result with the month, day, time of measuring.

Intended use:

{1}------------------------------------------------

use by medical professionals or at home to monitor and display d systolic blood pressure and pulse rate on anyone each time, except infants and unconscious people, with the air cuff around the left upper arm according to the instruction in the user's guide manual.

Comparison of technological characteristics of new device to predicate devices:

KD-575 Memory Automatic Electronic Blood Pressure Monitor can storage and show 120 times measuring results with the month, day, time of measuring. It can help patients to memorize the measuring data of their measuring results. Therefore the patients can analyze the measuring data and inspect their physical conditions. The time and the date are the advantages of KD-575 compare with the predicate device.

Clinical Tests:

Clinical tests were performed and complied the accuracy requirements of ANSI/AAMISP10-1992. The results meet or exceed the accuracy requirements of ANSI/AAMISP10-1992.

Non-clinical Tests:

tests the following standards, including non-clinical coincide All Electromagnetic Compatibility test.

IEC601-1 (1988)

Medical electrical equipment --- Part 1:General requirements for safety IEC601-1 (1988)

Amendment 2

IEC 60602-2-30: 1995

Medical electrical equipment-part2: Particular requirements for the safety of automatic cycling indirect blood pressure monitoring equipment. GB9706.1-1995 ISO 1833-1977

{2}------------------------------------------------

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized eagle or bird-like figure with three curved lines representing its wings or feathers.

MAR 2 1 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Kondon (Tianjin) Electronic & Electrical Apparatus Co., Ltd. c/o Mr. Liu Yi President, Board of Directors No. 31. Changjiang Road Tianjin CHINA 300193

Re: K042418

Trade/Device Name: KD-575 Fully Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: January 18, 2005 Received: January 21, 2005

Dear Mr. Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Mr. Liu Yi

or any Federal statutes and regulations administered by other Federal agencies. You must or ally Federal statutes and regulations and admited to: registration and listing (21 comply with an the Act 3 requirements, morealing, were and 800 manufacturing as set 80 miles of as set CFK Part 807), labeling (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radianon control provisions (Declons of I device as described in your Section 5 10(k) I his letter will anow you to obgin maneting your antial equivalence of your device to a legally premits notification. THC PDA miding of castomination for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrive ior your as (240) 276-0120. Also, please note the regulation entitled, Contact the Online of Comphane at (210) 21 Top Part 807.97). You may obtain Misoranding by reference to premanteering in the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bhimmarfor

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K042418

Device Name: KD-575 Fully Automatic Electronic Blood Pressure Monitor

Indications For Use:

The KD-575 Memory Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive, technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 8.6614 inches to 13.78 inches

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bhummer

(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K042418

Page 1 of 1

Yes

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).