(79 days)
For In Vitro Diagnostic Use. For Professional Use Only.
Cystatin C Immunoparticles are intended for the quantitative determination of cystatin C in human serum, heparinized plasma and EDTA plasma by turbidimetry and nephelometry. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases.
Cystatin C Control Set is an assayed bi-level control intended to monitor and evaluate the precision and accuracy of the quantitative immunological determination of human cystatin C by turbidimetry or nephelometry.
The Cystatin C Calibrator is intended for establishing calibration curves for the quantitative immunological determination of human cystatin C by turbidimetry or nephelometry.
Cystatin C Immunoparticles are purified immunoglobulin fraction of rabbit antiserum directed against cystatin C covalently coupled to uniform polystyrene particles. The Cystatin C Immunoparticles are provided as a suspension preserved with 15 mmol/L sodium azide and are supplied ready for use. Recombinant human cystatin C produced in E. coli was used as an immunogen for raising the antibody coupled to the polystyrene particles.
In the DakoCytomation Cystatin C Assay, human serum or plasma is mixed with the Cystatin C Immunoparticles. The resulting immune complexes are measured by turbidimetry or nephelometry. The generated signal is correlated with the concentration of cystatin C in the sample. By interpolation on a standard curve, the concentration of cystatin C in the sample is calculated.
Cystatin C Control Set is a bi-level control. The controls are liquid pools of delipidated human serum enriched with recombinant human cystatin C produced in E. coli and preserved with 15 mmol/L sodium azide. Each donor has been tested and found negative for hepatitis B virus surface antigen, antibodies to hepatitis C virus and human immunodeficiency virus 1 and 2. The Cystatin C Control Set is supplied ready for use.
The cystatin C value assignment has been carried out by turbidimetry using the DakoCytomation Cystatin C Calibrator, code No. X 0974, as reference.
Cystatin C Calibrator is a liquid pool of delipidated human serum enriched with recombinant human cystatin C produced in E. coli and preserved with 15 mmol/L sodium azide. Each donor has been tested and found negative for hepatitis B virus surface antigen, antibodies to hepatitis C virus and human immunodeficiency virus 1 and 2. The Cystatin C Calibrator is supplied ready for use.
The cystatin C value assignment has been carried out by turbidimetry using a precise transfer protocol ensuring traceability to a pure recombinant human cystatin C reference preparation, where the cystatin C concentration was established by dry mass determination.
The provided text describes a 510(k) premarket notification for three in vitro diagnostic devices: Cystatin C Immunoparticles, Cystatin C Control Set, and Cystatin C Calibrator. The focus of the submission is to demonstrate substantial equivalence to previously marketed predicate devices rather than proving a new performance characteristic against a fixed set of acceptance criteria. Therefore, the information regarding acceptance criteria and performance studies is presented differently than would be expected for a device with novel performance claims.
Here's an analysis of the provided information:
1. A table of acceptance criteria and the reported device performance
The document does not present a formal table of distinct acceptance criteria with corresponding performance results similar to what might be found for a device with a specific diagnostic accuracy claim (e.g., sensitivity, specificity, AUC). Instead, the performance evaluations were conducted to demonstrate "substantial equivalence" to predicate devices. The performance characteristics studied include:
| Performance Characteristic | Description / Reported Performance |
|---|---|
| Precision | Studied and found to contribute to substantial equivalence. Specific quantitative results (e.g., CV%) are not provided in this summary. |
| Accuracy | Studied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary. |
| Linearity | Studied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary. |
| Sensitivity | Studied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary. |
| Specificity | Studied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary. |
| Security Ranges | Studied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary. |
| Interferences | Studied and found to contribute to substantial equivalence. Details on specific interferents and their effects are not provided. |
| Stability Testing | Studied and found to contribute to substantial equivalence. Details on study methods and results are not provided. |
| Comparability (Turbidimetry/Nephelometry) | Assay results obtained with various commercially available turbidimeters and the IMMAGE® Immunochemistry System (nephelometer) were analyzed and found to be comparable. |
| Comparability to Predicate (Dade Behring) | A comparison of performance demonstrated substantial equivalence to the Dade Behring, Inc. N Latex Cystatin C Test Kit (K003503) and N Protein Standard UY (K003501). |
| Comparability to Predicate (Roche Diagnostics) | A comparison of performance demonstrated substantial equivalence to the Roche Diagnostics Corp. Creatinine Plus assay (K953239). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the sample sizes used for the test sets for precision, accuracy, linearity, etc. It mentions human serum or plasma samples being used. There is no information regarding the country of origin of the data or whether the studies were retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This type of information (expert review, ground truth establishment) is typically relevant for diagnostic imaging AI devices, rather than for an in vitro diagnostic (IVD) immunoassay like Cystatin C measurement. For IVD devices, the "ground truth" often refers to a reference method or validated standard, not expert human interpretation of an image. The document mentions value assignment for controls and calibrators by turbidimetry and traceability to a pure recombinant human cystatin C reference preparation where the concentration was established by dry mass determination. This constitutes the "ground truth" for the calibrators and controls.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Adjudication methods are typically associated with human reader studies for subjective interpretations (e.g., radiology). This concept does not apply to the performance testing of an automated IVD assay where results are quantitative measurements. The "adjudication" in this context would be the validation of the reference methods and calibrator value assignments.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. These are typically for AI-assisted diagnostic imaging where human readers interpret cases. This device is an in vitro diagnostic reagent and not an AI-assisted imaging device. The device itself is intended to quantitatively determine Cystatin C, and its performance is evaluated against predicate devices and analytical criteria, not human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to the inherent performance of the assay itself. The performance characteristics listed (precision, accuracy, linearity, sensitivity, specificity, interferences, stability) are all "standalone" performance evaluations of the assay components (Immunoparticles, Control Set, Calibrator) as they would be run on an automated analyzer. The device's primary function is quantitative measurement, not providing AI assistance to a human interpreter.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the DakoCytomation Cystatin C Assay, the "ground truth" for the calibrators and controls was established by:
- Turbidimetry using the DakoCytomation Cystatin C Calibrator (X 0974) as reference for the control set.
- Traceability to a pure recombinant human cystatin C reference preparation, where the cystatin C concentration was established by dry mass determination for the calibrator.
- For the overall performance, comparison to legally marketed predicate devices (Dade Behring N Latex Cystatin C Test Kit, Roche Diagnostics Corp. Creatinine Plus assay) served as a comparative "ground truth" in demonstrating substantial equivalence.
8. The sample size for the training set
The document does not describe a "training set" in the context of machine learning or AI. This device is an immunoassay, not an AI/ML product. The term "training set" is not applicable here. The development of the immunoparticles would involve reagent optimization and validation, but this is not typically referred to as a "training set" in this context.
9. How the ground truth for the training set was established
As there is no "training set" in the AI/ML sense, this question is not applicable. The development of the assay components would involve internal validation and standardization processes. For the purpose of regulatory substantial equivalence, the "ground truth" for comparison for the overall device performance was established by demonstrating comparable performance to the established predicate devices, which are legally marketed and are considered "safe and effective."
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with a stylized letter 'K' inside of an oval shape. To the right of the logo is the text 'Design & Development'. The text is in a simple sans-serif font and appears to be slightly blurred, possibly due to the image quality or resolution.
SEP - 3 2004
510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92(c).
Assigned 510(k) number
K04 627 ----------------------------------------------------------------------------------------------------------------------------------------------------------------------
Submitter's Information
Name: Address:
Phone: Fax Contact person: Date of preparation:
Device Name
Trade Name: Common/Usual name: Classification name:
Trade Name: Common/Usual name: Classification name:
DakoCytomation Denmark A/S Produktionsvej 42 DK-2600 Glostrup Denmark + 45 44 85 95 00 + 45 88 30 59 98 Dana Olsen, Requiatory Affairs June 10, 2004
Cystatin C Immunoparticles (code No. LX 002) Cystatin C Reagent Test, Cystatin C (21 CFR - 862.1225)
Cystatin C Control Set (code No. X 0973) Controls Single (Specified) Analyte Controls (Assayed and Unassayed) (21 CFR 862.1660)
Cystatin C Calibrator (code No. X 0974) Trade Name: Common/Usual name: Calibrator Classification name: Calibrator, Secondary (21 CFR 862.1150)
Predicate Devices
For Cystatin C Immunoparticles:
N Latex Cystatin C Test Kit, Dade Behring, Inc. (K003503) Creatinine Plus, Roche Diagnostics Corp. (K953239)
For Cystatin C Control Set: N Cystatin C Control, Dade Behring, Inc. (K003503)
For Cystatin C Calibrator: N Protein Standard UY, Dade Behring, Inc. (K003501)
Cystatin C Immunoparticles - Cystatin C Control Set - Cystatin C Calibrator DakoCytomation Denmark A/S
{1}------------------------------------------------
Device Description
Cystatin C Immunoparticles are purified immunoglobulin fraction of rabbit antiserum directed against cystatin C covalently coupled to uniform polystyrene particles. The Cystatin C Immunoparticles are provided as a suspension preserved with 15 mmol/L sodium azide and are supplied ready for use. Recombinant human cystatin C produced in E. coli was used as an immunogen for raising the antibody coupled to the polystyrene particles.
In the DakoCytomation Cystatin C Assay, human serum or plasma is mixed with the Cystatin C Immunoparticles. The resulting immune complexes are measured by turbidimetry or nephelometry. The generated signal is correlated with the concentration of cystatin C in the sample. By interpolation on a standard curve, the concentration of cystatin C in the sample is calculated.
Cystatin C Control Set is a bi-level control. The controls are liquid pools of delipidated human serum enriched with recombinant human cystatin C produced in E. coli and preserved with 15 mmol/L sodium azide. Each donor has been tested and found negative for hepatitis B virus surface antigen, antibodies to hepatitis C virus and human immunodeficiency virus 1 and 2. The Cystatin C Control Set is supplied ready for use.
The cystatin C value assignment has been carried out by turbidimetry using the DakoCytomation Cystatin C Calibrator, code No. X 0974, as reference.
Cystatin C Calibrator is a liquid pool of delipidated human serum enriched with recombinant human cystatin C produced in E. coli and preserved with 15 mmol/L sodium azide. Each donor has been tested and found negative for hepatitis B virus surface antigen, antibodies to hepatitis C virus and human immunodeficiency virus 1 and 2. The Cystatin C Calibrator is supplied ready for use.
The cystatin C value assignment has been carried out by turbidimetry using a precise transfer protocol ensuring traceability to a pure recombinant human cystatin C reference preparation, where the cystatin C concentration was established by dry mass determination.
Intended Use
For In Vitro Diagnostic Use. For Professional Use Only.
Cystatin C Immunoparticles are intended for the quantitative determination of cystatin C in human serum, heparinized plasma and EDTA plasma by turbidimetry and nephelometry. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases
Cystatin C Control Set is an assayed bi-level control intended to monitor and evaluate the precision and accuracy of the quantitative immunological determination of human cystatin C by turbidimetry or nephelometry.
The Cystatin C Calibrator is intended for establishing calibration curves for the quantitative immunological determination of human cystatin C by turbidimetry or nephelometry.
Cystatin C Immunoparticles - Cystatin C Control Set - Cystatin C Calibrator DakoCytomation Denmark AJS
{2}------------------------------------------------
Substantial Equivalence Comparison
The DakoCytomation Cystatin C Immunoparticles are substantially equivalent to the Dade Behring, Inc., N Latex Cystatin C Test Kit (K003503) with respect to indications for use, device design, materials and operational principles. The basic differences between the new device and the Dade Behring predicate device is the instruments used for testing. The DakoCytomation device can be used on commercially available turbidimetry and nephelometry analyzers, while the Dade Behring, Inc. product is applicable only on the Dade Behring, Inc. Nephelometer Systems. The DakoCytomation Cystatin C Immunoparticles are also claimed substantially equivalent to the Roche Diagnostics Corp. Creatinine Plus (K953239) in the intended use and results obtained. The difference in technology does not introduce new issues of safety and effectiveness.
The DakoCytomation Cystatin C Control Set is substantially equivalent to the Dade Behring N Cystatin C Control (K003503), with regard to the technological characteristics and in terms of the intended use.
The DakoCytomation Cystatin C Calibrator is substantially equivalent to the Dade Behring N Protein Standard UY (K003501), with regard to the technological characteristics and in terms of the intended use.
Performance Characteristics and Data
Performance characteristics and data presented in support of the Cystatin C Immunoparticles together with Cystatin C Control Set and Cystatin C Calibrator include precision, accuracy, linearity, sensitivity, specificity, security ranges, interferences and stability testing. Assay results obtained with various commercially available turbidimeters and IMMAGE® Immunochemistry System (nephelometer) were analyzed and found to be comparable.
A comparison of performance between the DakoCytomation Cystatin C Immunoparticles, Cystatin C Control Set and Cystatin C Calibrator with the Dade Behring, Inc. N Latex Cystatin C Test Kit (K003503) and N Protein Standard UY (K003501) demonstrated substantial equivalence. Additionally, a comparison of performance between the DakoCytomation Cystatin C Assay, and the Roche Diagnostics Corp. Creatinine Plus assay (K953239) demonstrated substantial equivalence.
Substantial Equivalence
Based on the information provided in this premarket notification, DakoCytomation concludes that the new devices, Cystatin C Immunoparticles, Cystatin C Control Set and Cystatin C Calibrator, are safe, effective and substantially equivalent to the Dade Behring predicate devices N Latex Cystatin C Test Kit (K003503), N Cystatin C Control (K003503) and N Protein Standard UY (K003501) respectively, with regard to the indications for use, device design, materials and operational principles. Additionally, the Cystatin C Immunoparticles are substantially equivalent to Roche Diagnostics Corp. Creatinine Plus (K953239) with regard to the intended use.
Cystatin C Immunoparticles - Cystatin C Control Set - Cystatin C Calibrator DakoCytomation Denmark A/S
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake entwined around it, overlaid on three human figures. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
SEP - 3 2004
Ms. Dana Olsen Regulatory Affairs DakoCytomation Denmark A/S Produktionsvej 42 DK-2600 Glostrup Denmark
Re:
Trade/Device Name: Cystatin C Immunoparticles Cystatin C Control Set Cystatin C Calibrator Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine Test system
Regulatory Class: Class II Product Code: NDY, JJX, JIT Dated: June 10, 2004 Received: June 16, 2004
Dear Ms. Olsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Jean M. Cooper MS, DVM.
Jean M. Cooper MS DVM
Jean M. Cooper, MS. D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number K041627
Indications for Use
Device Name: Cystatin C Immunoparticles
Indications for Use:
For In Vitro Diagnostic Use. For Professional Use Only.
Cystatin C Immunoparticles are intended for the quantitative defermiof cystation of cystatin C in human serum, heparinized plasma and EDTA nation of of overbidimetry and nephelometry. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Carol C Benem
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K041627
Page 1 of 3
Cystatin C Immunoparticles - Cystatin C Control Set - Cystatin C Calibrator DakoCytomation Denmark A/S
{6}------------------------------------------------
510(k) Number K04 1627
Indications for Use
Device Name: Cystatin C Control Set
Indications for Use:
For In Vitro Diagnostic Use. For Professional Use Only.
Cystatin C Control Set is an assayed bi-level control intended to monitor and evaluate the precision and accuracy of the quantitative immunological determination of human cystatin C by turbidimetry or nephelometry.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Carol C. Benson
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safe
510(k) K041627
Page 2 of 3
Cystatin C Immunoparticles - Cystatin C Control Set - Cystatin C Calibrator DakoCytomation Denmark A/S
{7}------------------------------------------------
K641627 510(k) Number
Indications for Use
Device Name: Cystatin C Calibrator
Indications for Use:
For In Vitro Diagnostic Use. For Professional Use Only.
Cystatin C Calibrator is intended for establishing calibration curves for the quantitative immunological determination of human cystatin C by turbidimetry or nephelometry.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
Please Do Not WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Carol C. Benson
inn Sien-Oft
Office of In Vitro Diagnostic Device Evaluation and Safe
510(k) K041627
Page 3 of 3
Cystatin C Immunoparticles - Cystatin C Control Set - Cystatin C Calibrator DakoCytomation Denmark A/S
§ 862.1225 Creatinine test system.
(a)
Identification. A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.(b)
Classification. Class II.