K Number
K041105
Device Name
MACROPORE HYDROSORB SPINE SYSTEM
Date Cleared
2004-07-28

(89 days)

Product Code
Regulation Number
888.3060
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The MacroPore Hydrosorb Spine System, in conjunction with traditional rigid fixation, is intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. The device is not intended for load bearing indications.
Device Description
The MacroPore Hydrosorb Spine System is a resorbable graft containment system composed of various sized porous plates / sheet, non-porous plates / sheets, and associated fixation screws manufactured from polylactic acid (PLA). The MacroPore Hydrosorb Spine System is composed of MacroPore Sheets, Plates, and Screws provided with and without USP barium sulfate beads (18mg - 20mg/bead) imbedded into the PLA polymer for radiopacity. MacroPore Sheets and Plates can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPore Plates / Sheets to the desired shape or size. The MacroPore Sheet and Plates are fully malleable when heated to approximately 55°C (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. Screws range in size from 2.4mm to 6.5mm in outer diameter with lengths ranging from 4mm to 30mm. The MacroPore Sheets and Plates are provided in various sizes ranging from 1.0mm to 5.0mm in thickness according to the region to be treated. The MacroPore Sheets and Plates range in size from as small as 24mm x 18mm to as large as 100mm x 100mm. The MacroPore Sheets and Plates are provided with and without macroporous holes. The macroporous holes range in size from 500 microns to 3,000 microns in diameter. The radiopaque barium sulfate beads have an approximate nominal mass of 18mg - 20mg and range in size from 0.5mm - 2.0mm in diameter. All configurations are to be within a mass of 100 grams of polymer. Various manual instruments (screw drivers, taps, drill bit, etc.) are used in conjunction with the MacroPore Hydrosorb Spine System to assist in the installation process.
More Information

Not Found

No
The document describes a resorbable graft containment system made of PLA with no mention of software, algorithms, or data processing that would indicate AI/ML.

Yes
The device is intended for use in spinal fusion procedures to maintain the relative position of weak bony tissue, which is a therapeutic purpose.

No

The device is a resorbable graft containment system used in spinal fusion procedures to maintain the position of bony tissue. It is a physical implant, not a tool for diagnosis.

No

The device description clearly outlines physical components made of polylactic acid (PLA), including plates, sheets, and screws, as well as associated manual instruments. This indicates a hardware-based medical device, not a software-only one.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states the device is for use in spinal fusion procedures to maintain the relative position of weak bony tissue. This is a surgical implant used directly on the patient's body.
  • Device Description: The description details a resorbable graft containment system made of plates, sheets, and screws. These are physical components implanted during surgery.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens derived from the human body (like blood, urine, tissue samples, etc.) to provide information for diagnosis, monitoring, or screening.

IVD devices are used outside the body to analyze biological samples. This device is an implantable medical device used within the body during surgery.

N/A

Intended Use / Indications for Use

The MacroPore Hydrosorb Spine System, in conjunction with traditional rigid fixation, is intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. The device is not intended for load bearing indications.

Product codes

KWQ

Device Description

The MacroPore Hydrosorb Spine System is a resorbable graft containment system composed of various sized porous plates / sheet, non-porous plates / sheets, and associated fixation screws manufactured from polylactic acid (PLA). The MacroPore Hydrosorb Spine System is composed of MacroPore Sheets, Plates, and Screws provided with and without USP barium sulfate beads (18mg - 20mg/bead) imbedded into the PLA polymer for radiopacity. MacroPore Sheets and Plates can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPore Plates / Sheets to the desired shape or size. The MacroPore Sheet and Plates are fully malleable when heated to approximately 55 degrees C (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation.

Screws range in size from 2.4mm to 6.5mm in outer diameter with lengths ranging from 4mm to 30mm. The MacroPore Sheets and Plates are provided in various sizes ranging from 1.0mm to 5.0mm in thickness according to the region to be treated. The MacroPore Sheets and Plates range in size from as small as 24mm x 18mm to as large as 100mm x 100mm. The MacroPore Sheets and Plates are provided with and without macroporous holes. The macroporous holes range in size from 500 microns to 3,000 microns in diameter. The radiopaque barium sulfate beads have an approximate nominal mass of 18mg - 20mg and range in size from 0.5mm - 2.0mm in diameter. All configurations are to be within a mass of 100 grams of polymer.

Various manual instruments (screw drivers, taps, drill bit, etc.) are used in conjunction with the MacroPore Hydrosorb Spine System to assist in the installation process.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

spinal

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

In Vitro Testing
Because the MacroPore Hydrosorb Spine Plates / Sheets are intended to be heated in the surgical suite to temperatures above the material's glass transition temperature to facilitate shaping to anatomic structures, testing was performed to determine the effect of prolonged heating in saline at 60 degrees C on inherent viscosity. The testing demonstrates that viscosity stayed within an appropriate range over 120 minutes. Therefore, the relatively brief exposure anticipated during the surgical preparation of Hydrosorb Spine Plates / Sheets is not expected to have a significant effect on its mechanical properties.

Aging studies were performed on MacroPore Hydrosorb Spine System components. Testing demonstrated that the MacroPore Hydrosorb Spine Sheet is as rigid and as strong as the predicate after 6 month of exposure. Mechanical testing was performed on the MacroPore Hydrosorb Spine Sheets and MacroPore Hydrosorb Spine Screws. Testing determined the MacroPore Hydrosorb Spine System to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions.

Crystallinity was tested for by DSC (differential scanning calorimetery). This test measures the amount of heat energy that is absorbed by a material. A crystalline material will require more energy once it reaches it's melting point. This release of heat energy can be seen on a graph as a sharp spike and is referred to as a "melting endotherm". The tests ran on the sterile and nonsterile samples revealed no endothermic spikes, indicating that the implants are amorphous and non-crystalline.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K010911, K021039, K000943, K002063

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.

0

JUL 28 2004

MacroPore Hydrosorb Spine System

Page 1 of 3

ADMINISTRATIVE INFORMATION

| Manufacturer Name: | MacroPore Biosurgery, Inc.
6740 Top Gun Street
San Diego, CA 92121 |
|--------------------|----------------------------------------------------------------------------------------------------------|
| Official Contact: | Kenneth K. Kleinhenz
Director of Regulatory Affairs
Telephone (858) 458-0900
Fax (858) 458-0994 |

DEVICE NAME

510K Summary

Classification Name:Appliance, Fixation, Spinal Intervertebral Body
Trade/Proprietary Name:MacroPore Hydrosorb Spine System

ESTABLISHMENT REGISTRATION NUMBER

2031733

DEVICE CLASSIFICATION AND PRODUCT CODE

As shown in 21CFR 888.3060 Appliance, Fixation, Spinal Intervertebral Body devices intended for use in spinal plating procedures are classified as Class II. They have been assigned Product Code KWQ.

INTENDED USE

The MacroPore Hydrosorb Spine System, in conjunction with traditional rigid fixation, is intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. The device is not intended for load bearing indications.

1

1041105

Design Characteristics

The MacroPore Hydrosorb Spine System is a resorbable graft containment system composed of various sized porous plates / sheet, non-porous plates / sheets, and associated fixation screws manufactured from polylactic acid (PLA). The MacroPore Hydrosorb Spine System is composed of MacroPore Sheets, Plates, and Screws provided with and without USP barium sulfate beads (18mg - 20mg/bead) imbedded into the PLA polymer for radiopacity. MacroPore Sheets and Plates can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPore Plates / Sheets to the desired shape or size. The MacroPore Sheet and Plates are fully malleable when heated to approximately 55°C (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation.

Screws range in size from 2.4mm to 6.5mm in outer diameter with lengths ranging from 4mm to 30mm. The MacroPore Sheets and Plates are provided in various sizes ranging from 1.0mm to 5.0mm in thickness according to the region to be treated. The MacroPore Sheets and Plates range in size from as small as 24mm x 18mm to as large as 100mm x 100mm. The MacroPore Sheets and Plates are provided with and without macroporous holes. The macroporous holes range in size from 500 microns to 3,000 microns in diameter. The radiopaque barium sulfate beads have an approximate nominal mass of 18mg - 20mg and range in size from 0.5mm - 2.0mm in diameter. All configurations are to be within a mass of 100 grams of polymer.

Various manual instruments (screw drivers, taps, drill bit, etc.) are used in conjunction with the MacroPore Hydrosorb Spine System to assist in the installation process.

Material Composition

The Hydrosorb Spine System is fabricated from polylactic acid (PLA).

In Vitro Testing

Because the MacroPore Hydrosorb Spine Plates / Sheets are intended to be heated in the surgical suite to temperatures above the material's glass transition temperature to facilitate shaping to anatomic structures, testing was performed to determine the effect of prolonged heating in saline at 60℃ on inherent viscosity. The testing demonstrates that viscosity stayed within an appropriate range over 120 minutes. Therefore, the relatively brief exposure anticipated during the surgical preparation of Hydrosorb Spine Plates / Sheets is not expected to have a significant effect on its mechanical properties.

Aging studies were performed on MacroPore Hydrosorb Spine System components. Testing demonstrated that the MacroPore Hydrosorb Spine Sheet is as rigid and as strong as the predicate after 6 month of exposure. Mechanical testing was performed on the MacroPore Hydrosorb Spine Sheets and MacroPore Hydrosorb Spine Screws. Testing determined the MacroPore Hydrosorb Spine System to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions.

Crystallinity was tested for by DSC (differential scanning calorimetery). This test measures the amount of heat energy that is absorbed by a material. A crystalline material will require more energy once it reaches it's melting point. This release of heat energy can be seen on a graph as a sharp spike and is referred to as a "melting endotherm". The tests ran on the sterile and nonsterile samples revealed no endothermic spikes, indicating that the implants are amorphous and non-crystalline.

2

510K Summary

EQUIVALENCE TO MARKETED PRODUCT

The MacroPore Hydrosorb System shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to pre-amendment devices: MacroPore OS Spinal System (K010911), the DePuy AcroMed BowTi Anterior Buttress Staple System (K021039), Exactech Tecres Cemex (K000943), and the Biotrace Model 1730 (K002063); Class II medical devices that were cleared for Bryan marketing in the United States.

Indications For Use

The MacroPore Hydrosorb Spine System shares identical indications for use with the predicated devices; MacroPore OS Spinal System (K010911) and the DePuy AcroMed BowTi Anterior Buttress Staple System (K021039) while sharing indications for use principles with the Exactech Tecres Cemex (K000943) and the Bryan Biotrace Model 1730 (K002063) as they are both indicated for use as a radiopaque additive in bone repair procedures.

Design and Materials

The physical designs of the MacroPore Hydrosorb Spine System and the predicate devices [MacroPore OS Spinal System (K010911), and the DePuy AcroMed BowTi Anterior Buttress Staple System (K021039)], are substantially equivalent as they all utilize a solid plate or sheet that is secured in place with screws. The MacroPore Hydrosorb Spine System and the MacroPore OS Spinal System (K010911) predicate are constructed from the same resorbable raw material. The physical design of the MacroPore Hydrosorb Spine System is substantially equivalent to the DePuy AcroMed BowTi Anterior Buttress Staple System (K021039) with respect to a screw sizes up to 6.5mm in diameter. The design features of the MacroPore Hydrosorb Spine System are substantially equivalent to the MacroPore OS Spinal System (K010911) predicate device as both devices utilize flat sheet designs of similar shapes and sizes containing multiple holes that accept a screw. The MacroPore Hydrosorb Spinc System and the predicate MacroPore OS Spinal System (K010911) are substantially equivalent as the holes in both of the respective products' plate / sheet devices contain a counter sink feature that assures precision engagement between the plate and the screw. The counter sink design of both the MacroPore Hydrosorb Spine System sheets and the predicate device's sheets share of allowing the screw to be thish with the surface of the plate when properly installed. The counter sink design of both the MacroPore Hydrosorb Spine System sheeting and the predicate device's sheets also share the feature of assuring maximum contact between the screw and sheet as the tapered regions of the sheet and screw neck are designed to engage for maximum contact. The mechanical characteristics of the MacroPore Hydrosorb Spine System are substantially equivalent to the predicate device as the MacroPore Hydrosorb Spine System has the same or greater mechanical strength than the MacroPore OS Spinal System predicate (K010911). The materials used in both the MacroPore Hydrosorb System and the predicate device are identical as both devices are fabricated from the same polylactic acid raw material.

The physical designs of the MacroPore Hydrosorb Spine System and the Exactech Tecres Cemex (K000943) and the Bryan Biotrace Model 1730 (K002063) predicate device are substantially equivalent as they both utilize irradiation sterilized USP barium sulfate in combination with a bone repair device. The barium sulfate beads utilized in the MacroPore Hydrosorb Spine System are fabricated from the identical USP grade barium sulfate as the Exactech Tecres Cemex (K000943) and the Bryan Biotrace predicate device (K002063). The MacroPore Hydrosorb Spine System utilized a maximum amount of 0.3grams of barium sulfate (16 beads) while the predicate devices utilize as much as 5-10 grams of barium sulfate in a single procedure.

3

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes, oriented diagonally.

Public Health Service

JUL 2 8 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Kenneth K. Kleinhenz Director of Regulatory Affairs MacroPore Biosurgery, Inc. 6740 Top Gun Street San Diego, California 92121

Re: K041105

Trade/Device Name: MacroPore Hydrosorb Spine System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: KWQ Dated: April 26, 2004 Received: April 30, 2004

Dear Mr. Kleinhenz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

4

Page 2 - Mr. Kenneth K. Kleinhenz

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely, yours,

Mark N. Millikan

Celia M. Witten, Ph.D., M.D. Director

Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

04/105

Device Name: MacroPore Hydrosorb Spine System

Indications for Use:

The MacroPore Hydrosorb Spine System, in conjunction with traditional rigid fixation, is intended for use in spinal fusion procedures as a means to maintain the relative position of weak bony tissue such as allografts or autografts. The device is not intended for load bearing indications.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NECESSARY) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

R Over-The-Counter Use
l Mark N Mellkum

(Division Sign-用 Division of General, Restorative, and Neurological Devices 041105

510(k) Number