(157 days)
The DETECT™ Surgical Pacing and Mapping Tool is a handheld, single use device designed to provide temporary cardiac pacing or monitoring.
The Medtronic® Detect™ Temporary Pacing and Mapping Device Description: Electrode Probe consists of a handle, a malleable stainless steel shaft with a flouropolymer sheath ending in a textured ball tip electrode, and a cable for connection to diagnostic device. Sterile, Nonpyrogenic, Disposable, Single use only.
The Grounding Electrode consists of a needle and a cable for connection to a diagnostic device. Sterile, Nonpyrogenic, Disposable, Single use only.
The Detect™ Electrode Probe is compatible with the Medtronic External Temporary Pacemaker (Model 5388), and the Medtronic Programmer/ Analyzer (Model 2090/2290).
The provided 510(k) summary does not contain detailed acceptance criteria and a study proving the device meets them in the way a modern AI/ML device submission would. This document pertains to a medical device (a temporary pacing and mapping tool) from 2004, which predates the widespread regulatory frameworks for AI/ML-driven devices.
The "study" referenced is a regulatory review for substantial equivalence to a predicate device, rather than a performance study with detailed metrics and statistical significance for a novel AI algorithm.
However, based on the provided text, I can infer and extract information to address your points as best as possible within the context of a 2004 medical device submission.
Here's an analysis of the provided text in response to your questions:
1. A table of acceptance criteria and the reported device performance
Based on the document, the "acceptance criteria" are not explicitly defined as pass/fail thresholds for specific quantitative performance metrics of a novel algorithm. Instead, the acceptance is based on the device being deemed "safe and effective" and "substantially equivalent" to a predicate device, primarily through compliance with established medical device standards and preliminary guidance.
| Acceptance Criteria (Inferred from regulatory context) | Reported Device Performance (Inferred) |
|---|---|
| Safety and Effectiveness: Meets general requirements for medical devices. | The device has been "tested and are considered safe and effective." |
| Substantial Equivalence: Similar in principle and intended use to a predicate device. | The device is deemed "substantially equivalent" to legally marketed predicate devices (specifically, the Medtronic® Model 6494 Unipolar Temporary Myocardial Pacing Wire). |
| Compliance with Standards: Adheres to relevant international and FDA guidance documents.** | The device meets standards set by: - "Electrode Recording Catheter Preliminary guidance" (FDA, 1995) - IEC 60601-1 - IEC 60601-27 |
| Intended Use Fulfilled: Capable of temporary cardiac pacing or monitoring. | The device is designed and demonstrated to provide "temporary cardiac pacing or monitoring." |
| Compatibility: Works with specified Medtronic devices. | Compatible with Medtronic External Temporary Pacemaker (Model 5388) and Medtronic Programmer/ Analyzer (Model 2090/2290). |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not describe a "test set" in the context of an AI/ML algorithm's performance evaluation using patient data. It refers to testing of the physical device for safety and effectiveness, which would typically involve bench testing, biocompatibility testing, electrical safety testing, and potentially some limited animal or human clinical data for a novel device, but not a "test set" for an algorithm.
Therefore, this information is not applicable or available in the provided text.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
As there is no mention of a "test set" for an algorithm's performance, there is no discussion of experts establishing ground truth for such a set. The "ground truth" here is the regulatory assessment of device safety and effectiveness and substantial equivalence by regulatory bodies (FDA) based on submitted data and adherence to standards.
Therefore, this information is not applicable or available in the provided text.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Similarly, since there's no "test set" for algorithm performance described, there's no adjudication method mentioned.
Therefore, this information is not applicable or available in the provided text.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not describe an MRMC comparative effectiveness study or any study involving human readers interacting with an AI system. This is a physical device (an electrode probe), not an AI diagnostic or assistive tool.
Therefore, this information is not applicable or available in the provided text.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This refers to an AI algorithm's performance. The device described is a physical medical tool, not an AI algorithm.
Therefore, this information is not applicable or available in the provided text.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For this type of traditional medical device, the "ground truth" for regulatory approval would be established through a combination of:
- Bench Testing Data: Verification of electrical performance, mechanical integrity, material compatibility, and sterile barrier integrity against predefined specifications.
- Biocompatibility Testing: Data demonstrating the device's materials do not cause adverse biological reactions.
- Performance Data: Evidence (often from preclinical studies or, if applicable, limited clinical data) that the device performs its intended function (pacing or monitoring) effectively and safely.
- Comparison to Predicate: The predicate device itself serves as a "ground truth" reference for safety and effectiveness, meaning the new device must demonstrate comparable performance characteristics.
The studies mentioned in the document are compliance with IEC standards, which cover aspects like electrical safety and electromagnetic compatibility.
No human expert consensus, pathology, or specific outcomes data for defining an "algorithm ground truth" is mentioned.
8. The sample size for the training set
The concept of a "training set" is specific to machine learning algorithms. This document describes a physical medical device, not an AI/ML algorithm.
Therefore, this information is not applicable or available in the provided text.
9. How the ground truth for the training set was established
As there is no "training set" for an AI/ML algorithm, the method for establishing its ground truth is not applicable.
Therefore, this information is not applicable or available in the provided text.
In summary: The provided 510(k) summary is for a traditional physical medical device from 2004 and does not address the criteria typically associated with AI/ML device submissions (e.g., test sets, training sets, AI performance metrics, expert adjudication for algorithms). The "proof" for this device lies in its substantial equivalence to a legally marketed predicate and its compliance with established safety and performance standards.
{0}------------------------------------------------
510 (K) SUMMARY of Safety and Effectiveness
(In accordance with SMDA of 1990 and pursuant with 21 CFR 807.92)
l. Applicant Information:
SEP - 2 2004
| Date Prepared: | July 29, 2004 |
|---|---|
| Submitter: | Medtronic, Inc. |
| Address: | 710 Medtronic Parkway, NEMinneapolis, MN 55432-5604 |
| EstablishmentRegistration No. | 2135394 |
| Contact Person: | David D. CoxPrinciple Regulatory Affairs Specialist |
| Telephone Number: | (763) 391-9251 |
| Fax Number: | (763) 391-9279 |
= Device Information:
| Trade Name: | Detect™ Mapping and Pacing Tool |
|---|---|
| Common Name: | Detect™ Electrode Probe |
| Classification Name: | Electrode, Pacemaker, Temporary |
|---|---|
| Classification: | Class II, 21 CFR 870.3680 |
| Product Code: | LDF |
Streamline™ 6494 Unipolar Temporary Myocardial Pacing Wire Predicate Device: 510(k) No. K012459, Reg. No. 870.3680; Product Code: LDF
The Model 6494 Unipolar Temporary Device Intended Use: Myocardial Pacing Wire is intended for temporary atrial and ventricular pacing and sensing for a contemplated implant duration of 7 days or less. The device is supplied sterile and is intended for SINGLE USE ONLY .
{1}------------------------------------------------
Detect™ Mapping and Pacing Tool, model 10650 Pre-market notification-510(K)
The Medtronic® Detect™ Temporary Pacing and Mapping Device Description: Electrode Probe consists of a handle, a malleable stainless steel shaft with a flouropolymer sheath ending in a textured ball tip electrode, and a cable for connection to diagnostic device. Sterile, Nonpyrogenic, Disposable, Single use only.
The Grounding Electrode consists of a needle and a cable for connection to a diagnostic device. Sterile, Nonpyrogenic, Disposable, Single use only.
The Detect™ Electrode Probe is compatible with the Medtronic External Temporary Pacemaker (Model 5388), and the Medtronic Programmer/ Analyzer (Model 2090/2290).
- The DETECTIM Surgical Pacing and Mapping Tool is a Intended Use: handheld, single use device designed to provide temporary cardiac pacing or monitoring.
III. SUBSTANTIAL EQUIVALENCE TESTING SUMMARY
The DETECT™ Mapping and Pacing Tool and the Grounding Needle Electrode have been tested and are considered safe and effective per the "Electrode Recording Catheter Preliminary guidance", Mark Massi, Pacing and Electrophysiology Device Branch, Division of Cardiovascular, Respiratory and Neurological Devices, Office of Device Evaluation FDA, CDRH, 1995.
The DETECT™ Mapping and Pacing Tool and the Grounding Needle Electrode have been tested and are considered safe and effective per "Standard for Medical Equipment; Part 1: General requirements, IEC 60601-1; IEC 60601-27
{2}------------------------------------------------
Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design featuring three abstract shapes resembling birds in flight, arranged in a stacked formation.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 2 2004
Medtronic, Inc c/o David D. Cox, Ph.D. Principle Regulatory Affairs Specialist Cardiac Surgery Technologies 710 Medtronic Parkway, NE Minneapolis, MN 55432-5604
Re: K040812
Trade Name: Detect™ Mapping and Pacing Tool Regulation Number: 21 CFR 870.3680 Regulation Name: Cardiovascular Permanent or Temporary Pacemaker Electrode Regulatory Class: II (two) Product Code: LDF Dated: July 30, 2004 Received: August 02, 2004
Dear Dr. Cox:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - David D. Cox, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of R Fart 877) adoming (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you dosine of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Neil R., Oph.
Bram D. Zuckerman, M.D.
Bram D. Zuckerman, M Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K040812
Detect™ Surgical Pacing and Mapping Tool, Model 10650 Device Name:
Indications for Use:
The DETECT™ Surgical Pacing and Mapping Tool is a handheld, single use device designed to provide temporary cardiac pacing or monitoring.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Ogle
(Division Sign-Off)
Division of Cardiovascular Devices
for bDz
510(k) Number_
Page 1 of 1
§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.
(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.