(87 days)
The CryoCor Model 3100 Series Sheath Kits are intended to facilitate the percutaneous introduction of cardiovascular catheters into the heart, while preventing backflow of blood.
The Model 3100 Series Sheath Introducer Kit consists of various sheath introducers, each packaged in a kit together with a dilator, introducer catheter and guide wire.
The sheath introducer is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A side port with three-way stopcock is provided for air aspiration, fluid infusion, blood sampling and pressure monitoring. A radioopaque marker at the tip of the sheath will aid during placement while utilizing fluoroscopic guidance. There are side holes at the distal end to allow for injection of contrast media.
The dilator is tapered at the distal tip to allow for use of standard needles for trans-septal procedures, and will accommodate a guide wire or needle no larger than 0.038".
The introducer catheter is gradually tapered distally to allow for gradual dilation of entry into vasculature without producing damage.
The guide wire is 0.038" in diameter, 150 cm in length with a "I"-tip configuration. It is supplied in a hoop dispenser with a "J" straightener attached to the distal end. The proximal end of the hoop dispenser has a luer hub to facilitate flushing of the wire.
The provided 510(k) summary for K040552 describes a medical device, the CryoCor Model 3100 Series Sheath Kits, which are Catheter Introducers. The summary focuses on showing substantial equivalence to predicate devices, rather than presenting a detailed study with specific acceptance criteria and performance metrics traditionally associated with validating new clinical claims or complex algorithms.
Therefore, many of the requested categories (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, ground truth types, training set details) are not applicable or not provided in this type of submission.
Here's the information that can be extracted from the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria (Explicitly Stated/Implied) | Reported Device Performance |
|---|---|
| Design and Performance Requirements | Met all of its design and performance requirements. |
| Performance vs. Predicate Device | Performed at least as well as the predicate device. |
| Biocompatibility (blood-contacting materials) | All materials were found to be biocompatible (according to ISO 10993-1). |
| Sterilization Efficacy | Sterilized using a validated E-beam radiation sterilization cycle. |
Study Information
-
Sample size used for the test set and the data provenance:
- Sample Size: Not specified.
- Data Provenance: Not specified. The studies appear to be bench and pre-clinical comparisons, rather than human clinical trials.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not provided. The "test set" likely refers to engineering or bench test samples, not clinical patient data requiring expert ground truth.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable/Not provided.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is a physical catheter introducer, not an AI-powered diagnostic tool. MRMC studies are not relevant.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm.
-
The type of ground truth used (expert concensus, pathology, outcomes data, etc):
- For performance testing, the "ground truth" would be the engineering specifications and established performance metrics for catheter introducers. For biocompatibility, it's compliance with ISO standards. For sterilization, it's a validated cycle.
-
The sample size for the training set:
- Not applicable/Not provided. As above, this is a physical device, not an AI model.
-
How the ground truth for the training set was established:
- Not applicable/Not provided.
{0}------------------------------------------------
MAY 2 8 2004
510(k) Summary of Substantial Equivalence
| Date Prepared | March 01, 2004 |
|---|---|
| SubmitterAddress | CryoCor, Inc.9717 Pacific Heights Blvd.San Diego, Ca 92121 |
| Contact | Jami MillerRegulatory Affairs Specialist |
| Phone | 858-909-2231 |
| FaxE-Mail Address | 858-909-2350jmiller@cryocor.com |
| Device Trade Name/ Model Numbers | Model 3100 Series Sheath KitsModel 3110Model 3130 |
| Device Common Name | Catheter Introducer |
| Device Classification Name | Catheter Introducer |
| Classification Regulation | Class II, C.F.R. Section 870.1340 |
| Product Code | DYB |
Device Description
The Model 3100 Series Sheath Introducer Kit consists of various sheath introducers, each packaged in a kit together with a dilator, introducer catheter and guide wire.
The sheath introducer is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A side port with three-way stopcock is provided for air aspiration, fluid infusion, blood sampling and pressure monitoring. A radioopaque marker at the tip of the sheath will aid during placement while utilizing fluoroscopic guidance. There are side holes at the distal end to allow for injection of contrast media.
The dilator is tapered at the distal tip to allow for use of standard needles for trans-septal procedures, and will accommodate a guide wire or needle no larger than 0.038".
The introducer catheter is gradually tapered distally to allow for gradual dilation of entry into vasculature without producing damage.
{1}------------------------------------------------
The guide wire is 0.038" in diameter, 150 cm in length with a "I"-tip configuration. It is supplied in a hoop dispenser with a "J" straightener attached to the distal end. The proximal end of the hoop dispenser has a luer hub to facilitate flushing of the wire.
Intended Use
The CryoCor Model 3100 Series Sheath Kits are intended to facilitate the perculaneous introduction of cardiovascular catheters into the heart, while preventing backflow of blood.
Substantially Equivalent Devices
The Model 3100 Series Sheath Kits are substantially equivalent to the Daig Fast-Cath Intra-Cardiac Introducer cleared under K973840 and the Daig Fast-Cath Transseptal Introducer cleared under K964518.
The physical characteristics, the intended use, and the mode of use of the Model 3100 Series Sheath Kits are similar to the predicate device.
Test Summarv
Performance Testing
Bench testing confirmed that the Model 3100 Series Sheath Kits met all of its design and performance requirements.
Pre-Clinical Studies
Comparison testing found that the Model 3100 Series Sheath Kits performed at least as well as the predicate device.
Biocompatibility Information
Biocompatibility testing was performed on the materials which are blood contacting according to ISO 10993-1. All materials were found to be biocompatible.
Sterilization Validation
The Model 3100 Series Sheath Kits are sterilized using a validated E-beam radiation sterilization cycle.
Conclusion
CryoCor, Inc. considers the Model 3100 Series Sheath Kits to be substantially equivalent to their legally marketed predicate device based on the data and information presented within this application.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure with three wing-like shapes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 8 2004
CryoCor, Inc. c/o Ms. Jami Miller Regulatory Affairs Specialist 9717 Pacific Heights Blvd. San Diego, CA 92121
K040552 Re: Model 3100 Series Sheath Kits Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: March 1, 2004 Received: March 2, 2004
Dear Ms. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, arorely manns of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or dayses to de of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Ms. Jami Miller
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA 3 issualite of a verse complies with other requirements of the Act that FDA has made a decertifications administered by other Federal agencies. You must or any Federal statutes and regulations damands. but not limited to: registration and listing (21 Comply with an the Fiel 31 cquinements, 110, good manufacturing practice requirements as set CFR Fart 807), adoning (21 OF R Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation oom or provision arketing your device as described in your Section 510(k) This letter will anow you to begin maxicong your antial equivalence of your device to a legally premarket notheation. The PDF intelling sizes cation for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific acries 101 ) 594-4648. Also, please note the regulation entitled, Connact the Office or Comparket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mionnational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
R. Vachner
Br
Di
Bram D. Zuckerman, M.D. Director Office of Device Evaluation
Division of Cardiovascular Devices Center for Devices and Radiological Health
{4}------------------------------------------------
Indications for Use
510(k) Number: K (140557
Device Name: Model 3100 Series Sheath Kits
Indications for Use:
The CryoCor Model 3100 Series Sheath Kits are intended to facilitate the percutaneous introduction of cardiovascular catheters into the heart, while preventing backflow of blood.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duna R. Vachner
(Division Sign
(Division Sign-Off) Statution Olgh-Oll)
Division of Cardiovascular Devices
510(k) Number_K040552
§ 870.1340 Catheter introducer.
(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).