(17 days)
The AutoPulse™ Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
AutoPulse Resuscitation System Model 100 ("Device") is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. The Device consists of a single use chest compression assembly (CCA) that includes a patient liner, and a reusable platform that contains a user control panel, a drive mechanism, a control system, and a power system (rechargeable battery).
This device description is for the AutoPulse Resuscitation System Model 100, an automated chest compression device.
Based on the provided text, there is no specific acceptance criteria or detailed study results that prove the device meets acceptance criteria mentioned.
The document states:
- "Appropriate product testing was conducted to evaluate conformance to product specification and substantial equivalence to predicate devices." (Section 1)
However, it does not provide any data, metrics, or specific acceptance criteria that were used in this testing. It only concludes that the device is "substantially equivalent" to predicate products based on indications for use, basic overall function, and materials.
Therefore, I cannot populate the requested table or provide details for most of the other questions.
Here's an assessment based on the available information:
- Table of acceptance criteria and the reported device performance:
- Acceptance Criteria: Not specified in the provided text.
- Reported Device Performance: Not specified in the provided text.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not provided | Not provided |
- Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not provided.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as no test set details or ground truth establishment are described.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is an automated mechanical device, not an AI-based diagnostic or assistive system for human readers.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document describes the device itself as an "automated, portable, battery powered device that compresses the chest." While its performance would inherently be standalone in its function, no specific "standalone performance study" details (like accuracy or efficacy metrics) are provided. It operates as an adjunct to manual CPR, implying it's often used with human oversight, but its core function is automated.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not provided.
- The sample size for the training set: Not applicable, as this is a mechanical device, not a machine learning model with a training set.
- How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
MAR 1 1 2004
510(k) Summary
General Information
| Classification | Class III |
|---|---|
| Trade Name | AutoPulse™ Resuscitation System Model 100 |
| Submitter | Revivant Corporation775 Palomar AvenueSunnyvale, CA 94085 |
| 408-524-3500 | |
| Contact | Bob H. KatzVice President, Operations |
Intended Use
The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
Predicate Devices
| K851139 | Thumper™ Model 1005 |
|---|---|
| K011046 | AutoPulse™ Resuscitation System Model 100 |
| K022345 | AutoPulse™ Resuscitation System Model 100 |
| K032852 | AutoPulse™ Resuscitation System Model 100 |
Device Description
AutoPulse Resuscitation System Model 100 ("Device") is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. The Device consists of a single use chest compression assembly (CCA) that includes a patient liner, and a reusable platform that contains a user control panel, a drive mechanism, a control system, and a power system (rechargeable battery).
Materials
All materials used in the manufacture of the AutoPulse Resuscitation System Model 100 are suitable for this use and have been used in numerous previously cleared products.
{1}------------------------------------------------
Testing
Appropriate product testing was conducted to evaluate conformance to product specification and substantial equivalence to predicate devices.
Summary of Substantial Equivalence
The AutoPulse Resuscitation System Model 100 is equivalent to the predicate products. The indications for use, basic overall function, and materials used are substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Public Health Service
MAR 1 1 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Revivant Corporation c/o Mr. Bob H. Katz Vice President and Chief Operating Officer 775 Palomar Avenue Sunnyvale, CA 94085
Re: K040453
AutoPulse™ Resuscitation System Model 100 Regulation Number: 21 CFR 870.5200 Regulation Name: External Cardiac Compressor Regulatory Class: Class III (three) Product Code: DRM Dated: February 20, 2004 Received: February 27, 2004
Dear Mr. Katz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Bob H. Katz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire snecific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Duna R. Vachner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): Ko40453
AutoPulse™ Resuscitation System Model 100 Device Name:
Indications For Use:
The AutoPulse Resuscitation System Model 100 is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
Prescription Use X (Part 21 CFR 801 Subpart D)
.
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Diana R. Vachner
ivision Sir -Off) Division of Cardiovascular Devices
510/k) Nimber K040453
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
§ 870.5200 External cardiac compressor.
(a)
Identification. An external cardiac compressor is an externally applied prescription device that is electrically, pneumatically, or manually powered and is used to compress the chest periodically in the region of the heart to provide blood flow during cardiac arrest. External cardiac compressor devices are used as an adjunct to manual cardiopulmonary resuscitation (CPR) when effective manual CPR is not possible (e.g., during patient transport or extended CPR when fatigue may prohibit the delivery of effective/consistent compressions to the victim, or when insufficient EMS personnel are available to provide effective CPR).(b)
Classification. Class II (special controls). The special controls for this device are:(1) Nonclinical performance testing under simulated physiological conditions must demonstrate the reliability of the delivery of specific compression depth and rate over the intended duration of use.
(2) Labeling must include the following:
(i) The clinical training necessary for the safe use of this device;
(ii) Adjunctive use only indication prominently displayed on labels physically placed on the device and in any device manuals or other labeling;
(iii) Information on the patient population for which the device has been demonstrated to be effective (including patient size and/or age limitations,
e.g., adult, pediatric and/or infant); and(iv) Information on the time necessary to deploy the device as demonstrated in the performance testing.
(3) For devices that incorporate electrical components, appropriate analysis and testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(4) Human factors testing and analysis must validate that the device design and labeling are sufficient for effective use by the intended user, including an evaluation for the time necessary to deploy the device.
(5) For devices containing software, software verification, validation, and hazard analysis must be performed.
(6) Components of the device that come into human contact must be demonstrated to be biocompatible.