K Number
K033897
Date Cleared
2004-07-19

(216 days)

Product Code
Regulation Number
868.2700
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MEDICYL-E-Lite is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MRI safe, MRIcompatible, and intended for use during MR imaging for MRI systems up to 3.0T.

For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.

Device Description

The MEDICYL-E-Lite is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MRI safe, MRI-compatible, and intended for use during MR imaging for MRI systems up to 3.0T. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.

The intearated system includes an aluminum cylinder, valve, requlator and flow meter. The device offers low flow settings that may be clinically appropriate for certain classes of patients.

AI/ML Overview

This document describes a 510(k) submission for the AGA Linde Healthcare MEDICYL-E-Lite Portable Oxygen Delivery System. This is a medical device, and the information provided is for regulatory clearance based on substantial equivalence to predicate devices, rather than a study proving the device meets specific performance criteria in the way an AI/ML device would be assessed.

Therefore, many of the requested categories for AI/ML device performance studies are not directly applicable to this document. I will extract the information that is present and explain why other sections are not covered.


1. Table of acceptance criteria and the reported device performance

For a medical device like the MEDICYL-E-Lite, "acceptance criteria" are typically related to meeting regulatory standards, safety requirements, and functional specifications, often demonstrated through testing and comparison to predicate devices. The document focuses on showing substantial equivalence and compliance with relevant regulations.

ParameterAcceptance Criteria (Predicate)Reported Device Performance (MEDICYL-E-Lite MRI)
Valve/Regulator
Low Flow SettingsYes (MEDICYL-E-Lite) / No (Praxair Grab n' Go)Yes
Flow Between SettingsNo (MEDICYL-E-Lite & Praxair Grab n' Go)No
Cylinder On/OffYes (MEDICYL-E-Lite) / No (Praxair Grab n' Go)Yes
Filling PortActive (MEDICYL-E-Lite & Praxair Grab n' Go)Active
Contents GaugeNon-active (MEDICYL-E-Lite) / Active (Praxair Grab n' Go)Non-active
Filters3 (MEDICYL-E-Lite) / 1 (Praxair Grab n' Go)3
Pressure Design4350psi (MEDICYL-E-Lite) / 3000psi (Praxair Grab n' Go)4350psi
Excess Flow DeviceYes (MEDICYL-E-Lite) / No (Praxair Grab n' Go)Yes
Single stage piston styleYes (MEDICYL-E-Lite & Praxair Grab n' Go)Yes
Guard
Hand grip2 grip (MEDICYL-E-Lite) / 1 grip (Praxair Grab n' Go)2 grip
Access PortsYes (MEDICYL-E-Lite) / No (Praxair Grab n' Go)Yes
Flow selector/hose barb/gauge alignedYes (MEDICYL-E-Lite) / No (Praxair Grab n' Go)Yes
ColorGreen (MEDICYL-E-Lite & Praxair Grab n' Go)Green
Height6.75" (MEDICYL-E-Lite) / 8" (Praxair Grab n' Go)6.75"
Cylinder
SizesD, E (MEDICYL-E-Lite) / E (Praxair Grab n' Go)D, E
Weight (E)900gr (MEDICYL-E-Lite) / 1060gr (Praxair Grab n' Go)900gr
Materials/constructionAluminum (MEDICYL-E-Lite) / Steel (Praxair Grab n' Go)Aluminum
MRI Compatibility
MRI SafeNo (Predicate MEDICYL-E-Lite & Praxair Grab n' Go)Yes; tested up to 3.0T
MRI CompatibleNo (Predicate MEDICYL-E-Lite & Praxair Grab n' Go)Yes; tested up to 3.0T

The primary "acceptance criteria" here is substantial similarity to the predicate devices, with the added functionality of MRI compatibility for the "MEDICYL-E-Lite (MRI)" version. The document explicitly states: "The manufacturer believes that the technological characteristics of the MEDICYL-E-Lite portable oxygen system is substantially similar to those of the predicate devices."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided. For a device like an oxygen delivery system, "test set" would typically refer to the units tested for physical performance and safety. The document mentions "safety and performance testing" and compliance with "voluntary standards" and "21 CFR49 § 178.46, Specification SAL seamless aluminum cylinders." It also states the MRI compatibility was "tested up to 3.0T" and evaluated "in accordance with the draft CDRH Magnetic Resonance Working Group document, A Primer on Medical Device Interactions with Magnetic Resonance Imaging Systems, dated February 7, 1997." However, the sample size of devices tested or the specific provenance of that test data is not detailed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as this is a mechanical device, not an AI/ML diagnostic tool requiring expert ground truth for image interpretation or similar. The "truth" here relates to physical specifications and safety standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable for this type of device submission. Adjudication methods are typically used in clinical trials or studies for diagnostic devices where subjective interpretations might differ.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. This is a physical medical device, not an AI/ML system, and therefore no MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as this is a physical medical device, not an algorithm or AI system.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for this device is based on engineering specifications, regulatory compliance, performance testing results against predefined standards, and demonstration of substantial equivalence to legally marketed predicate devices. For example:

  • Physical dimensions and materials: Measured against design specifications.
  • Pressure design: Tested to meet the stated psi.
  • Flow settings: Verified to operate as specified.
  • MRI compatibility: Tested against established protocols outlined in the CDRH Magnetic Resonance Working Group document.
  • Cylinder specifications: Conformity to "21 CFR49 § 178.46, Specification SAL seamless aluminum cylinders."

8. The sample size for the training set

This is not applicable. There is no AI/ML model for which a "training set" would be used for this device.

9. How the ground truth for the training set was established

This is not applicable as there is no AI/ML model and therefore no "training set."

{0}------------------------------------------------

AGA Linde Healthcare

Image /page/0/Picture/2 description: The image shows the word "Linde" in a stylized, cursive font. The word is written in black ink on a white background. The "L" has a large, decorative swirl at the top, and the other letters are connected in a flowing script.

510(k) Summary MEDICYL-E-Lite Portable Oxygen System 510(k) Number: K033897

Submitted in accordance with the requirements of SMDA 1990 and 21CFR807.92.

1. APPLICANT'S INFORMATION:
Edward W. Siurek III
Dir., Operational Quality, IBD Global Operations
PH:(216) 328-9801
FX:(216) 328-9818
Internet:http://www.agalindehealthcare.com
Medical Establishment Registration No.:pending
2. SUBMITTER'S INFORMATION
James Jochen Rogers
General Manager
Coastal Consulting Group, Ltd.
P.O. Box 391117
Solon, OH 44139
PH/FX:440.546.4936
Mobile:724.713.2298
E-mail:coastalcg@earthlink.net
Internet:http://www.immersive-reality.biz
3. Date:December 15, 2003
4. DEVICE INFORMATION
Trade/Proprietary Name:MEDICYL-E-Lite Portable Oxygen delivery system
Common Name:MEDICYL-E-Lite Portable Oxygen delivery system
DEVICE NAME:Cylinder, Compressed Gas, and Valve
Classification Panel:Cardiovascular and Respiratory Devices
Classification Number:868.2700
Product Nomenclature:Regulator, Pressure, Gas Cylinder
Product Code(s):CAN
Classification Number:868.2610
Product Nomenclature:Gauge, Gas Pressure, Cylinder/Pipeline
Product Code(s):BXH
Classification Number:unclassified
Product Nomenclature:Cylinder, Compressed Gas, and Valve
Product Code(s):ECX
Classification Number:unclassified
Product Nomenclature:Cylinder, Gas (Empty)
Product Code(s):KGA

Helping You Care More Effectively
6055 Rockside Woods Blvd.
Corporate Office (216) 642-6600

{1}------------------------------------------------

AGA Linde Healthcare

Image /page/1/Picture/1 description: The image shows the word "Linde" in a stylized, cursive font. The letters are connected, and the "L" has a large, decorative swirl at the top. The word appears to be a logo or brand name.

5. DEVICE CLASSIFICATION:

Empty compressed gas cylinders and compressed gas cylinder with valve assemblies are unclassified devices, and reviewed by the Anesthesiology and Respiratory Devices Branch, Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices.

Gas cylinder pressure regulators and gas pressure gauges are Class I devices and exempted from pre-market notification.

6. PREDICATE DEVICE(s):

  • MEDICYL-E-Lite Portable Oxygen System .
  • . Praxair Grab n' Go Portable Medical Oxygen System

7. DEVICE DESCRIPTION:

The MEDICYL-E-Lite is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MRI safe, MRI-compatible, and intended for use during MR imaging for MRI systems up to 3.0T. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.

The intearated system includes an aluminum cylinder, valve, requlator and flow meter. The device offers low flow settings that may be clinically appropriate for certain classes of patients.

8. INDICATIONS FOR USE:

The MEDICYL-E-Lite is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MRI safe, MRI-compatible, and intended for use during MR imaging for MRI systems up to 3.0T. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.

9. TECHNOLOGICAL CHARACTERISTICS:

A summary comparison of technological characteristics, including design and materials is provided in the table below:

ParameterMEDICYL-E-Lite(MRI)MEDICYL -E-LitePraxair Grab n'Go
Valve/Regulator
Low Flow Settingsyesyesno
Flow Between Settingsnonono
Cylinder On/Offyesyesno
Filling Portactiveactiveactive
Contents Gaugenon-activenon-activeactive
Filters331
Pressure Design4350psi4350psi3000psi
Excess Flow Deviceyesyesno

{2}------------------------------------------------

AGA Linde Healthcare

Image /page/2/Picture/1 description: The image shows the word "Linde" in a stylized, cursive font. The word is written in black ink and appears to be a logo or brand name. The letters are connected, and the "L" has a decorative swirl at the top.

ParameterMEDICYL-E-Lite(MRI)MEDICYL -E-LitePraxair Grab n'Go
Single stage pistonstyleyesyesyes
Guard
Hand grip2 grip2 grip1 grip
Access Portsyesyesno
Flow selector/hosebarb/gauge alignedyesyesno
Colorgreengreengreen
Height6.75"6.75"8"
Cylinder
SizesD, ED, EE
Weight (E)900gr900gr1060gr
Materials/constructionAluminumAluminumSteel
MRI Compatibility
MRI Safeyes; tested up to3.0Tnono
MRI Compatibleyes; tested up to3.0Tnono

The manufacturer believes that the technological characteristics of the MEDICYL-E-Lite portable oxygen system is substantially similar to those of the predicate devices.

    1. PERFORMANCE DATA:
      The aluminum cylinders conform to the requirements of 21 CFR49 § 178.46, Specification SAL seamless aluminum cylinders.

The MEDICYL-E-Lite portable oxygen delivery system has been evaluated in accordance with the draft CDRH Magnetic Resonance Working Group document, A Primer on Medical Device Interactions with Magnetic Resonance Imaging Systems, dated February 7, 1997.

11. STATEMENT OF SUBSTANTIAL EQUIVALENCE:

Based upon the safety and performance testing and compliance with voluntary standards, the manufacturer believes that the MEDICYL-E-Lite portable oxygen delivery system is substantially equivalent to the predicate devices, and does not raise any new questions of safety or effectiveness.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the department's name around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 9 2004

AGA-Linde Healthcare, Incorporated C/O Mr. James Jochen Rogers General Manager Coastal Consulting Group, Limited P.O. Box 391117 Solon, Ohio 44139

Re: K033897

Trade/Device Name: Medicyl-E-Lite Portable Oxygen Delivery System Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: ECX Dated: May 15, 2004 Received: May 19, 2004

Dear Mr. Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 -Mr. Rogers

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Cluls

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K033897

Device Name: Medicyl-E-Lite Portable Oxygen Delivery System

Indications for Use:

The MEDICYL-E-Lite is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults. The device is MRI safe, MRIcompatible, and intended for use during MR imaging for MRI systems up to 3.0T.

For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.

Sta y/t
7/16/2001

Division of Anesthesiology. General Hospital. Infection Control. Dental Devices

510(k) Number:

X Prescription Use (Part 21 CFR 801 Subpart D)

AND

X Over-the-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 868.2700 Pressure regulator.

(a)
Identification. A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.