(11 days)
The ATR5000, LC5000 and Implant System are intended for the preparation of intra-oral bone for implantology procedures. These ATR Surgical Micromotors are intended to prepare Intraoral bone for implantology procedures.
The ATR Surgical Micromotors ATR5000, LC5000 and Implant System, are microprocessor driven surgical micromotors used in implantology procedures. The ATR Surgical Micromotors ATR5000, LC5000 and Implant System consists of a microprocessor controlled unit, foot pedal, electric micromotor, support rods and sterile, disposable irrigation tubes. The units also house the peristaltic pump. The ATR Surgical Micromotors ATR5000, LC5000 and Implant System provide electronic control of velocity and torque. The ATR Surgical Micromotors ATR5000, LC5000 and Implant System can be programmed and retain programs in memory. The ATR Surgical Micromotors ATR5000, LC5000 and Implant System are fully operational from foot pedal.
This document describes acceptance criteria and testing for the ATR Surgical Micromotors (ATR5000, LC5000, and Implant System), which are microprocessor driven surgical micromotors used in implantology procedures.
Here's the breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The device's performance is demonstrated by its compliance with a series of electrical, safety, and electromagnetic compatibility (EMC) standards. The acceptance criteria essentially are compliance with these listed standards.
| Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|
| UL 2601-1 Standard for Medical Electrical Equipment, General requirements for Safety (IEC 60601-1 with U.S. deviations) | Complies with the standard |
| UL 601-1 Medical Electrical Equipment, General Requirement for Safety | Complies with the standard |
| IEC 60601-1 Medical Electrical Equipment Part 2, "General Safety Norms" | Complies with the standard |
| IEC 60601-1-2 Medical Electrical Equipment Part 1 EMC | Complies with the standard |
| EMC Conducted RF emissions IEC 55011 | Complies with the standard |
| Degrees of protection provided by enclosures (IP Code) IEC 60529 | Complies with the standard |
| IEC 61000-4-2 EMC Electrostatic Discharge Immunity | Complies with the standard |
| IEC 61000-4-3 Radiated RF immunity | Complies with the standard |
| IEC 61000-4-4 Fast Transient Immunity | Complies with the standard |
| IEC 61000-4-5 Pulse immunity | Complies with the standard |
| IEC 61000-4-6 Conducted RF immunity | Complies with the standard |
| IEC 61000-4-11 Supply Voltage Hole Immunity | Complies with the standard |
| Transformer safety IEC 61558-1 | Complies with the standard |
| IEC 61558-2-6 Transformer safety-Particular prescriptions | Complies with the standard |
| ISO 11498:1997 Dental handpieces Dental low-voltage electrical motors | Complies with the standard |
| Withstanding voltage test | Performed in conformance with UL 2601-1 requirements |
| Leak current test | Performed in conformance with UL 2601-1 requirements |
2. Sample size used for the test set and the data provenance
- Sample Size: The document does not specify a numerical "sample size" in terms of number of devices tested. It refers to "All evaluation of the ATR 5000, LC5000 and Implant System" being performed. This implies at least one unit of each model (ATR5000, LC5000, and Implant System) was tested.
- Data Provenance: The tests were "performed by certified body Laboratory," but the specific country of origin for the data is not explicitly stated. The submitting company (ATR srl) is based in Italy, suggesting the testing might have been done in Italy or a European certified lab. The predicate device was also evaluated "with the same regulation standard," indicating historical consistency in testing approach.
- Retrospective or Prospective: Not applicable in the context of device performance testing against engineering standards. This is a prospective testing of the manufactured device models.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable to this type of device submission. The "ground truth" here is compliance with established engineering and safety standards, not a medicaldiagnosis or interpretation by experts. The "certified body Laboratory" acts as the independent verifier of compliance.
4. Adjudication method for the test set
This information is not applicable. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies where conflicting expert interpretations need resolution. For compliance testing against engineering standards, the results are objectively measured against the standard's specifications.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. MRMC studies and the concept of "human readers" improving with "AI assistance" are relevant to diagnostic imaging devices or other AI-enabled tools for interpretation. This submission is for surgical micromotors, which are hand-held mechanical and electronic devices, not AI-driven diagnostic tools.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable. This device is a surgical micromotor, not a software algorithm. Its performance is evaluated through physical and electrical tests, not algorithmic performance.
7. The type of ground truth used
The "ground truth" in this context is the specific technical requirements and limits defined by the listed international and national standards (e.g., UL 2601-1, IEC 60601-1, IEC 60601-1-2, IEC 61000 series, ISO 11498). The device is deemed safe and effective if its measured performance falls within the acceptable parameters set by these standards.
8. The sample size for the training set
This information is not applicable. Surgical micromotors are hardware devices that do not use machine learning algorithms, and therefore, there is no "training set."
9. How the ground truth for the training set was established
This information is not applicable for the same reason as above; there is no training set for this type of medical device.
{0}------------------------------------------------
NOV 2 5 2003
ATR Surgical Micromotors 510(k): Device Modification
SECTION 14: SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMD 1990 and CFR 807.92.
SUBMITTER INFORMATION 14.1
| a. | Company Name | Advanced Technology Research(A.T.R.)srl |
|---|---|---|
| b. | Company Address | Via del Pescino, 651100 PISTOIAItaly |
| c. | Company phoneCompany fax | +39 0573 364 254+39 0573 364 002 |
| d. | Contact Person | Daniele PoliPresident |
| e. | Date Summary Prepared | June 11, 2003 |
DEVICE IDENTIFICATION 14.2
| a. | Trade/Proprietary Names: | ATR5000, LC5000, ImplantSystem |
|---|---|---|
| b. | Classification Name: | Dental Handpieces andAccessories 21 CFR 872.4200 |
| c. | Common Names: | Surgical Micromotor forimplantology |
14.3 IDENTIFICATION OF PREDICATE DEVICE
| Company | Device | <10(k) No. | CDate C.leared |
|---|---|---|---|
| ATR | TR3000 | K991401 | 1 . T.1999ﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ |
14.4 DEVICE DESCRIPTION
The ATR Surgical Micromotors ATR5000, LC5000 and Implant System, are microprocessor driven surgical micromotors used in implantology procedures. The ATR Surgical Micromotors ATR5000, LC5000 and Implant System consists of a microprocessor controlled unit, foot pedal, electric micromotor, support rods and sterile, disposable irrigation tubes.
{1}------------------------------------------------
KWized "13103
ATR Surgical Micromotors 510(k): Device Modification
The units also house the peristaltic pump.
The ATR Surgical Micromotors ATR5000, LC5000 and Implant System provide electronic control of velocity and torque.
The ATR Surgical Micromotors ATR5000, LC5000 and Implant System can be programmed and retain programs in memory.
The ATR Surgical Micromotors ATR5000, LC5000 and Implant System are fully operational from foot pedal.
14.5 SUBSTANTIAL EQUIVALENCE
The ATR Surgical Micromotors ATR5000, LC5000 and Implant System are substantially equivalent to the ATR3000 Surgical Micromotor in commercial distribution by ATR. The fundamental technical characteristics of the ATR5000, LC5000 and Implant System are similar to those of the predicate device and are listed on the comparison chart provided in this 510(k) submission.
The ATR5000, LC5000 and Implant System and the predicate device have adjustable speed, torque and reduction rates, and are programmable. The micromotor handpiece of ATR5000, LC5000 and Implant System and the one of the predicate device are autoclavable.
14.6 INTENDED USE
The ATR5000, LC5000 and Implant System are intended for the preparation of intra-oral bone for implantology procedures.
14.7 TECHNICAL CHARACTERISTICS
The ATR5000, LC5000 and Implant System were designed and developed to provide a microprocessor controlled surgical system with similar performances compared to predicate device.
ATR5000, LC5000 and Implant System are equivalent in design and materials to the predicate device, have adjustable speeds and torque values (related to the reduction rate of the handpiece selected) that are fully customisable by the end user, have automatic motor shutdown system which provides to switch off the motor whenever torque set is reached, have a micromotor sterilizable according the recommended protocols, and can be operated bv foot control.
Any of these feature is found in the predicate device also.
ATR5000, LC5000 and Implant System use for the central unit housing an ABS material with thickness sufficient to be classified V0 in flammability class as the predicate device.
ATR5000, LC5000 use the same micromotor and housing of the predicate device; Implant System uses the same material for micromotor housing with different dimension to fit the new brushless motor.
{2}------------------------------------------------
ATR Surgical Micromotors 510(k): Device Modification
ATR5000, LC5000 and Implant System were designed to use the same disposable tubing of the predicate device.
The difference between ATR5000, LC5000, Implant System and the predicate device are that :
Implant System uses a Brushless motor driving instead of a DC motor driving that is designed for the same intended use of the predicate device.
ATR5000/LC5000 and ATR ImplantSystem have new features.
All the three devices are equipped with a wider display with larger characters, and with a new larger keyboard.
ATR5000/LC5000 are equipped with the beep approaching set torque function, that warns the user when reaching 75% of set torque value. This can aid the user because increasing applied torque may depends by reaching the lower cortical part of the bone, or by reaching the mechanical breaking limit of the burr and otherwise reminds that when reaching the torque value set, the micromotor automatically shuts off.
ATR 5000/LC5000 and Atr ImplantSystem footpedal control have the same control functions as predicate device.
ATR5000 and LC 5000 are equipped with I/O devices such parallel port and serial port for printer and PC connection, and smart card reader. Those devices were not present on the predicate devices. In order to maintain safety a designed hardware prevent the use of I/O devices when motor runs.
Those new features are useful to create patients archive on software or paper support, so user can quickly remind patient status and operations advancement.
Safety recommends about I/O use, and safety system that prevent the use of I/O devices when motor runs, are reported in ATR5000/LC5000 and ATR ImplantSystem user manuals. In order to verify new devices safety issues they were designed and tested following standard requirements:
- UL 2601-1 Standard for Medical Electrical Equipment, General ● requirements for Safety (IEC 60601-1 with U.S. deviations)
- IEC 60601-1 Medical Electrical Equipment Part 2,"Genaral Safety Norms"
- IEC 60601-1-2 Medical Electrical Equipment Part 1 EMC
No I/O devices in Implant System model as in the predicate device.
{3}------------------------------------------------
ATR Surgical Micromotors 510(k): Device Modification
PERFORMANCE DATA 14.8
The ATR 5000, LC 5000 and Implant System were tested in accordance with the technical requirements of UL 2601-1 (IEC 60601-1 with U.S. deviations), IEC 601-1 and IEC 60601-1-2 and successive variants .
We remember here that LC5000 unit is different from ATR5000 only for frontal plate graphics and model label.
Whitstanding voltage test and leak current test were performed on devices in conformance with requirements from UL 2601-1.
All evaluation of the ATR 5000, LC5000 and Implant System were performed by certified body Laboratory, and all the results comply to standard listed below.
Also the predicate device has been evaluated with the same regulation standard. So ATR5000/LC5000 and ATR ImplantSystem and predicate device can be considered substantially equivalent in performance.
The conclusions drawn from the performance test are that ATR 5000, LC5000 and Implant System complies with:
- UL 2601-1 Standard for Medical Electrical Equipment, General requirements . for Safety (IEC 60601-1 with U.S. deviations)
- UL 601-1 Medical Electrical Equipment, General Requirement for Safety ●
- EMC Conducted RF emissions IEC 55011 ●
- Degrees of protection provided by enclosures (IP Code) IEC 60529 .
- IEC 60601-1 Medical Electrical Equipment Part 2,"Genaral Safety Norms" .
- IEC 60601-1-2 Medical Electrical Equipment Part 1 EMC ●
- IEC 61000-4-2 EMC Electrostatic Discharge Immunity .
- IEC 61000-4-3 Radiated RF immunity ●
- IEC 61000-4-4 Fast Transient Immunity ●
- IEC 61000-4-5 Pulse immunity
- IEC 61000-4-6 Conducted RF immunity ●
- IEC 61000-4-11 Supply Voltage Hole Immunity .
- Transformer safety IEC 61558-1 ●
- IEC 61558-2-6 Transformer safety-Particular prescriptions .
- ISO 11498:1997 Dental handpieces Dental low-voltage electrical motors .
and are effective and safe to use.
Declaration of conformity to a standard in Chapter 12.1 of this submission.
14.9 510(k) CHECKLIST
This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer's Checklist is provided in this submission.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle's head, with three parallel lines forming the head and beak.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 5 2003
Advanced Technology Research (A.T.R) S.R.L C/O Ms. Elizabeth Drew Responsible Third Party Official Underwriters Laboratories, Incorporated 1655 Scott Boulevard Santa Clara, California 95050
Re: K033597
Trade/Device Name: Surgical Micromotors Regulation Number: 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EKX Dated: November 12, 2003 Received: November 14, 2003
Dear Ms. Drew:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Ms. Drew
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours. ·
Susan Runos
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
ATR Surgical Micromotors Special Premarket 510(k) Notification
INDICATIONS FOR USE
510(k) Number:
K02359 To be Assigned by FDA
Device Name:
Surgical Micromotors
Model:
ATR5000 IC 5000 Implant System
Indications for Use:
These ATR Surgical Micromotors are intended to prepare Intraoral bone for implantology procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Betz DDS for Dr. Runner
1-Off)
Resthesiology, General Hospital, ction Control, Dental L
510(k) Number
V Prescription Use
OR Over-The -Counter Use _________
(Per 21 CFR 801.109)
6.3
§ 872.4200 Dental handpiece and accessories.
(a)
Identification. A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.(b)
Classification. Class I.