(63 days)
Diazyme Lithium Enzymatic Assay Kit is for quantitative in vitro determination of lithium in human serum and plasma. Measurements of lithium are carried out essentially to ensure that proper drug dosage is administered in the treatment of patient suffering from bipolar disorder and to avoid toxicity.
Diazyme's lithium enzymatic assay is a spectrophotometric method, which can be adapted to most automated clinical chemistry analyzers. In Diazyme's lithium enzymatic assay, Lithium is determined spectrophotometrically through a kinetic coupling assay system involving a Diazyme's proprietary phosphatase whose activity is sensitive to lithium concentration (IC50=0.1mM). Through enzymatic coupling, the phosphatase substrate, adenosine biphosphate (PAP) is converted to hypoxanthine by a series of enzymatic reactions to generate uric acid and hydrogen peroxide (H2O2). H2O2 generated reacts with N-Ethyl-N-(2-hydroxy-3-sulfopropyl)-3-m-toluidine (EHSPT) and 4-aminoantipyrine (4-AA) in the presence of peroxidase (POD) to form a quinone dye which has maximal absorbance at 556nm. The rate of the quinine dye formation is inversely proportion to the concentration of lithium in serum samples.
The provided document (K033360) describes the Diazyme Lithium Enzymatic Assay Kit, a device for quantitative in vitro determination of lithium in human serum and plasma. The document is a 510(k) premarket notification, which seeks to demonstrate substantial equivalence to a legally marketed predicate device.
Here's an analysis of the acceptance criteria and study data based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated in a quantitative manner as "pass/fail" thresholds. Instead, the document presents performance characteristics and asserts that they demonstrate excellent accuracy and safety/effectiveness. The comparison is primarily against a predicate device and existing methods.
| Performance Metric | Implicit Acceptance Criteria (based on predicate comparison/general clinical needs) | Reported Device Performance |
|---|---|---|
| Precision | Low coefficient of variation (CV%) indicating good reproducibility. | 1.0mM Li+: Within Precision CV%=4.7%, Total Precision CV%=6.9% 2.4mM Li+: Within Precision CV%=3.3%, Total Precision CV%=5.5% |
| Measuring Range | Clinically relevant range for lithium therapy monitoring (typically 0.6-1.2 mM for maintenance, up to 1.5mM for acute treatment, toxicity > 1.5mM). | 0 to 3 mmol/L |
| Correlation with Predicate Device (Trace colorimetric method) | Strong positive correlation (e.g., correlation coefficient > 0.95) | Correlation coefficient of 0.97 |
| Correlation with Reference Method (ISE method) | Strong positive correlation (e.g., correlation coefficient > 0.95) | Correlation coefficient of 0.97 |
| Interference | Minimal or no significant interference from common biological substances and medications. | Little interference found at indicated concentrations for various substances (NH4Cl, KPi, CaCl2, NaCl, KCl, CuCl2, FeCl3, ZnCl2, Triglyceride, Ascorbic Acid, Bilirubin). |
| Accuracy / Substantial Equivalence | No significant deviation compared to the predicate device and good agreement with reference methods. | "excellent accuracy and is safe and effective." "No significant deviation between the results obtained by Diazyme's Lithium Enzymatic assay and legally marketed predicate Trace Lithium Reagent when testing clinical patient serum samples." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document mentions "clinical patient serum samples" for the comparison analysis with the predicate device but does not specify the exact number of samples used in the test set.
- Data Provenance: The document does not explicitly state the country of origin. The studies appear to be retrospective as they involve testing existing clinical patient serum samples and comparing them to established methods.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of information is generally not applicable to an in vitro diagnostic (IVD) assay for a quantitative analyte like lithium. The "ground truth" for lithium concentration would be established by established analytical methods (e.g., ISE, atomic absorption spectrophotometry, predicate device) or reference materials, not by expert consensus on interpretation. Therefore, there were no experts in the sense of clinicians or radiologists used to establish ground truth.
4. Adjudication Method for the Test Set
Not applicable. As noted above, the "ground truth" is analytical, not interpretive.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an in vitro diagnostic assay for quantifying a chemical analyte (lithium), not an imaging device or AI-driven diagnostic that involves human reader interpretation. Therefore, MRMC studies and "human reader improvement with AI" are not relevant to this device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This statement applies to the Diazyme Lithium Enzymatic Assay as it is a standalone assay. It provides a quantitative result for lithium concentration based solely on the enzymatic reaction and spectrophotometric measurement, without human intervention for interpretation beyond standard laboratory procedures (e.g., sample handling, instrument operation). The "algorithm" here is the chemical reaction and measurement protocol that generates the result.
7. The Type of Ground Truth Used
The ground truth was established by:
- Comparison to Legally Marketed Predicate Device: Trace Lithium Reagent (K003583).
- Comparison to Established Analytical Methods: Ion-selective electrode (ISE) method.
- Reference Materials/Spiking Studies: For linearity (implied by "wide measuring range") and interference studies, where samples were spiked with known concentrations of lithium or interfering substances.
- Known Concentrations: For precision studies (1.0mM Li+ and 2.4mM Li+).
8. The Sample Size for the Training Set
The document does not specify a training set size. For an enzymatic assay like this, "training set" is not typically a concept used in the same way as for machine learning algorithms. The assay development would involve optimizing reagents and conditions, which is analogous to "training" but doesn't involve a distinct, quantified "training set" of patient samples in the way an AI model would have.
9. How the Ground Truth for the Training Set Was Established
As explained above, a traditional "training set" with established ground truth in the context of an AI/machine learning model is not directly applicable. The "ground truth" for optimizing the assay would come from:
- Chemical principles and known reactions: The design of the enzymatic coupling system.
- Known concentrations of lithium: To establish the dose-response curve and sensitivity.
- Reference materials: For calibration and verification of the assay's performance during development.
- Comparison to existing methods: During the development and optimization phase to ensure the new assay aligns with established ways of measuring lithium.
{0}------------------------------------------------
K03360
Image /page/0/Picture/2 description: The image shows the word "DIAZYME" in bold, block letters. Above the word is a four-pointed star-like shape, with each point resembling a diamond. The text and the star are both in black, contrasting with the white background.
Submitter's name:
Summary
Diazyme Laboratories Division, General Atomics
3550 General Atomics Court Submitter's address: San Diego, CA 92121 858-455-4754 Phone: 858-455-4760 Fax: Chong Yuan, Ph.D. Name of Contact Person: Diazyme Laboratories Division General Atomics 3550 General Atomics Court San Diego, CA 92121 Phone: 858-455-4754
October 16, 2003 Date the summary was prepared:
| Name of the device: | Lithium Enzymatic Assay |
|---|---|
| Trade Name: | Diazyme Lithium Enzymatic Assay |
| Common/Usual Name: | Enzymatic Assay, Lithium |
| Classification Name: | Single (Specified) Analyte Controls (Per 21CFR section 862.1660) |
| Device Class: | II |
Fax: 858-455-4760
Predicate Device:
The legally marketed device to which we are claiming equivalence [807.92(a)(3)]: Trace Lithjum Reagent (K003583) manufactured by Trace America, Inc., Arlington, Texas.
Description of the devices
Currently, the most commonly used methods to detect serum lithium are ion-selective electrode (ISE) atomic absorption spectrophotometry, and flame emission photometry. However, routine maintenance of these analyzers requires much effort and sometimes can be cumbersome. Diazyme's lithium enzymatic assay is a spectrophotometric method, which can be adapted to most automated clinical chemistry analyzers.
In Diazyme's lithium enzymatic assay, Lithium is determined spectrophotometrically through a kinetic coupling assay system involving a Diazyme's proprietary phosphatase whose activity is sensitive to lithium concentration (IC50=0.1mM). Through enzymatic coupling, the phosphatase substrate, adenosine biphosphate (PAP) is converted to hypoxanthine by a series of enzymatic reactions to generate uric acid and hydrogen peroxide (H2O2). H2O2 generated reacts with N-Ethyl-N-(2-hydroxy-3-sulfopropyl)-3-m-toluidine (EHSPT) and 4-aminoantipyrine (4-AA) in the
{1}------------------------------------------------
presence of peroxidase (POD) to form a quinone dye which has maximal absorbance at 556nm. The rate of the quinine dye formation is inversely proportion to the concentration of lithium in serum samples.
Intended Use of the Device:
Diazyme Lithium Enzymatic Assay Kit is for quantitative in vitro determination of lithium in human serum and plasma. Measurements of lithium are carried out essentially to ensure that proper drug dosage is administered in the treatment of patient suffering from bipolar disorder and to avoid toxicity.
Performance Characteristics
Diazyme's lithium enzymatic assay is a two reagent (R1 and R2) based kinetic assay system. The results are obtained in 10 min by measuring absorbance at 550 nm. No off line pretreatment is needed. The assay has a wide measuring range from 0 to 3 mmol/L. The assay offers excellent precision as shown in the table below:
| 1.0mM Li+ | 2.4mM Li+ | |
|---|---|---|
| Within Precision | CV%=4.7% | CV%=3.3% |
| Total Precision | CV%=6.9% | CV%=5.5% |
Diazyme's Lithium Enzymatic assay has good correlation with both Trace colorimetric method and ISE method (correlation coefficient of 0.97 for both methods). We have conducted interference study by spiking the substances to be tested to the pooled human sera and found little interference at the indicated concentrations
| Interference | Concentration |
|---|---|
| NH4Cl | 0.5 mM |
| KPi | 1.5 mM |
| CaCl2 | 5 mM |
| NaCl | 200 mM |
| KCl | 10mM |
| CuCl2 | 0.25 mM |
| FeCl3 | 0.25mM |
| ZnCl2 | 0.25 mM |
| Triglyceride | 250mg/dl |
| Ascorbic Acid | 5 mM |
| Bilirubin | 45 mg/dl |
Conclusion: Comparison analysis presented in the 510K submission for this device in the comparison section, together with linearity, precision and interference study presented demonstrated that the Diazyme's Lithium Enzymatic assay has excellent accuracy and is safe and effective. There is no significant deviation between the results obtained by Diazyme's Lithium Enzymatic assay and legally marketed predicate Trace Lithium Reagent when testing clinical patient serum samples. Therefore, Diazyme's Lithium Enzymatic assay is substantially
{2}------------------------------------------------
similar to the commercially available products to measure lithium levels in human serum samples.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. Above the caduceus, there are three parallel lines that curve upwards, resembling the wings of a bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 2 3 2003
Chong Yuan, Ph.D. Managing Director General Atomics Diazyme Laboratories 3550 General Atomics Court San Diego, CA 92121-1194
K033360 Re:
Trade/Device Name: Diazyme Lithium Enzymatic Assay Kit Regulation Number: 21 CFR 862.3560 Regulation Name: Lithium test system Regulatory Class: Class II Product Code: JII Dated: October 16, 2003 Received: October 29, 2003
Dear Dr. Yuan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
FDA INDICATIONS FOR USE FORM
510(k) Number (if known): K033360____________________________________________________________________________________________________________________________________________
Device Name: Diazyme Lithium Enzymatic Assay Kit
Indications for Use:
Diazyme Lithium Enzymatic Assay Kit is for quantitative in vitro determination of Diazyme Dithium Entzyments of lithium are carried out essentially to ensure that proper drug dosage is administered in the treatment of patient suffering from bipolar disorder and to avoid toxicity.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---|---|
| -------------------------------------------------------- | -- |
| Prescription Use | |
|---|---|
| (Per 21 CFR 801.109) |
OR
| Over-The-Counter Use | |
|---|---|
| ---------------------- | -- |
| Division Sign-Off | Carol Benson |
|---|---|
| (Optional Format 1-2-96) |
Office of In Vitro Diagnostic Device
Evaluation and Safety
| 510(k) | K033360 |
|---|---|
| -------- | --------- |
§ 862.3560 Lithium test system.
(a)
Identification. A lithium test system is a device intended to measure lithium (from the drug lithium carbonate) in serum or plasma. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).(b)
Classification. Class II.