K Number
K033245
Manufacturer
Date Cleared
2003-11-06

(30 days)

Product Code
Regulation Number
876.4400
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bard® UltraView™ Multiple Band Ligator is used for endoscopic ligation of esophageal varices.

Device Description

The Bard® UltraView™ Multiple Band Ligator is comprised of a ligating unit that fits over the distal end of an endoscope with seven premounted rubber latex ligating bands. The ligating unit is attached to the handle by the activating sheath. The bands are 1.9 mm thick with a 5.1 mm outer diameter and a 2.0 mm diameter. Four endoscope adapters and a centering sleeve are provided to allow the ligating unit to fit securely on endoscopes ranging from 8.5 mm to 11.5 mm. The small adapter fits 8.5 to 9.3 mm scopes. The medium adapter fits 9.3 to 10.2 mm scopes. The large adapter fits 10.2 to 10.8 mm scopes. The X-large adapter fits 10.8 to 11.5 mm scopes. A scope gauge is included to determine the required adapter for the endoscope. The modified design differs from the predicate device in that the ligating tip has been modified to improve the sealing integrity of the lip seal during aspiration and to improve the smooth actuation of the Band Driver over the Band Carrier by the addition of 12 ribs to the Band Driver. The sheath attaches to the handle assembly, which can be held in the physician's hand or attached to the endoscope sheath outside of the patient. The bands are deployed by depressing the thumb paddle on the handle assembly causing the bands to be pushed/deployed off of the band carrier. The modified device mounts flush with the tip of the endoscope, as does the predicate device, to allow maximum visualization. The modified and current device tips extend beyond the distal tip of the endoscope allowing variceal tissue to be suctioned into the band carrier cylinder for ligation.

AI/ML Overview

The provided text describes modifications to an existing medical device, the Bard® UltraView™ Multiple Band Ligator, and demonstrates its substantial equivalence to the predicate device. This is a 510(k) premarket notification, which typically focuses on demonstrating equivalence rather than establishing new performance criteria for a novel device.

Here's an analysis of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state formal "acceptance criteria" in a quantitative sense as might be seen for a new device's efficacy study. Instead, the performance data focuses on demonstrating that the modified device is substantially equivalent to the predicate device and remains safe and effective for its intended use.

Acceptance Criterion (Implied)Reported Device Performance
Biocompatibility: No adverse tissue reactions from patient-contacting components.Cytotoxicity testing of the colored scope adaptors (the only newly patient-contacting components) showed no evidence of toxicity. All other material changes were to non-patient-contacting components.
Functionality: The device operates as intended for ligation.Functionality testing demonstrated that the modified Bard® UltraView™ Multiple-Band Ligator is substantially equivalent to the current Bard® UltraView™ Multiple-Band Ligator. The modified design specifically aimed to "improve the sealing integrity of the lip seal during aspiration and to improve the smooth actuation of the Band Driver over the Band Carrier." The device mounts flush and allows variceal tissue to be suctioned for ligation, similar to the predicate.
Tensile Strength: The bands and components maintain integrity under expected forces.Tensile testing demonstrated that the modified Bard® UltraView™ Multiple-Band Ligator is substantially equivalent to the current Bard® UltraView™ Multiple-Band Ligator.
Aspiration Performance: The device can effectively aspirate tissue for ligation.Aspiration testing demonstrated that the modified Bard® UltraView™ Multiple-Band Ligator is substantially equivalent to the current Bard® UltraView™ Multiple-Band Ligator.
Safety and Effectiveness: The device is safe and effective for its intended use and patient population.Based on the biocompatibility and engineering performance tests (functionality, tensile, aspiration), the manufacturer concluded that the device is safe for its intended use and patient population and is substantially equivalent to the predicate.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not provide details on sample sizes for the functionality, tensile, aspiration, or cytotoxicity tests. These are typically engineering and laboratory tests, not clinical studies with patient populations. The data provenance (country of origin, retrospective/prospective) is not specified, but these would be internal validation tests conducted by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable to the type of testing described. The "tests" mentioned are engineering performance tests (e.g., tensile strength, aspiration efficacy, functionality) and biocompatibility tests (cytotoxicity), not tests requiring expert interpretation of diagnostic outcomes or clinical ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods are typically used in clinical studies where multiple human readers or experts assess and come to a consensus on "ground truth" for medical images or clinical outcomes. The tests described are laboratory and engineering performance tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study and no mention of AI assistance. This document describes a traditional medical device (a ligator) and its modification, not an AI-powered diagnostic or therapeutic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable. The device is a physical medical instrument, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The concept of "ground truth" as typically applied in diagnostic or AI studies is not directly relevant here. The "truth" for the performance claims is based on:

  • Biocompatibility Standards: For cytotoxicity, adherence to established standards for determining toxicity.
  • Engineering Specifications: For functionality, tensile strength, and aspiration, performance is measured against internal engineering specifications and, by extension, the characteristics of the predicate device to demonstrate "substantial equivalence." The "ground truth" is whether the device performs as specified and as comparably to the predicate as required for safety and effectiveness.

8. The sample size for the training set

This information is not applicable. There is no mention of a "training set" as this is not an AI/machine learning device. The development process for this type of device involves engineering design, prototyping, and testing to meet functional requirements.

9. How the ground truth for the training set was established

This information is not applicable as there is no "training set."

In summary: This 510(k) submission details modifications to an existing physical medical device and relies on engineering and laboratory performance testing (biocompatibility, functionality, tensile, aspiration) to demonstrate substantial equivalence to its predicate device. It does not involve AI, clinical efficacy studies with human subjects, or the establishment of "ground truth" in the diagnostic sense. The "acceptance criteria" are implied by the need to show that the modified device remains safe and effective and performs at least as well as the predicate, which was confirmed through these specific tests.

{0}------------------------------------------------

K033245 Pgz 1 of 2

NOV - 6 2003

Bard Endoscopic Technologies C.R. Bard, Inc. 129 Concord Road P.O. Box 7031 Billerica, MA 01821-7031

978-663-8989

Image /page/0/Picture/3 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct and modern appearance. The font style is consistent throughout the word, creating a unified and recognizable logo or brand name.

VI 510(k) SUMMARY SAFETY AND EFFECTIVENESS INFORMATION

As required by the Safe Medical Devices Act of 1990, codified under Section 513, Part (i)(3)(A) of the Food Drug and Cosmetic Act, a summary of the safety and effectiveness information upon which substantial equivalence determination is based follows.

A. Submitter Information
Submitter's Name:Bard Endoscopic TechnologiesC.R. Bard, Inc.
Address:129 Concord Road, Bldg. #3Billerica, MA 01821
Phone:(978) 262 - 4868
Fax:(978) 262 - 4878
Contact Person:Michael A. Patz
Date of Preparation:September 29, 2003
B. Device Name
Trade Name:Bard® UltraView™ Multiple Band Ligator
Common/Usual Name:Classification Name:Ligator, Esophageal
C. Predicate Device Name(s)
Trade Name:Bard® UltraView™ Multiple Band Ligator
  • D. Device Description:
    The Bard® UltraView™ Multiple Band Ligator is comprised of a ligating unit that fits over the distal end of an endoscope with seven premounted rubber latex ligating bands.

{1}------------------------------------------------

K033245 Page 2 of 2

The ligating unit is attached to the handle by the activating sheath. The bands are 1.9 mm thick with a 5.1 mm outer diameter and a 2.0 mm diameter.

Four endoscope adapters and a centering sleeve are provided to allow the ligating unit to fit securely on endoscopes ranging from 8.5 mm to 11.5 mm. The small adapter fits 8.5 to 9.3 mm scopes. The medium adapter fits 9.3 to 10.2 mm scopes. The large adapter fits 10.2 to 10.8 mm scopes. The X-large adapter fits 10.8 to 11.5 mm scopes. A scope gauge is included to determine the required adapter for the endoscope.

  • E. Intended Use:
    The Bard® UltraView™ Multiple Band Ligator is used for endoscopic ligation of esophageal varices.

  • F. Technological Characteristics Summary:
    The modified Bard UltraView ligator is comprised of similar medical grade plastics, stainless steels as the predicate and uses the identical band material.

The modified design differs from the predicate device in that the ligating tip has been modified to improve the sealing integrity of the lip seal during aspiration and to improve the smooth actuation of the Band Driver over the Band Carrier by the addition of 12 ribs to the Band Driver. The sheath attaches to the handle assembly, which can be held in the physician's hand or attached to the endoscope sheath outside of the patient. The bands are deployed by depressing the thumb paddle on the handle assembly causing the bands to be pushed/deployed off of the band carrier. The modified device mounts flush with the tip of the endoscope, as does the predicate device, to allow maximum visualization. The modified and current device tips extend beyond the distal tip of the endoscope allowing variceal tissue to be suctioned into the band carrier cylinder for ligation.

  • G. Performance Data
    Biocompatibility tests were not completed as all material changes made were to nonpatient contacting components except for the color coded scope adaptors. Results of cytotoxicity testing of the colored scope adaptors showed no evidence of toxicity. Functionality testing, tensile testing and aspiration testing have demonstrated that the modified Bard® UltraView™ Multiple-Band Ligator is substantially equivalent to the current Bard® UltraView™ Multiple-Band Ligator and that the device is safe for its intended use and patient population.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

NOV - 6 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Michael A. Patz Regulatory Affairs Specialist Bard Endoscopic Technologies C. R. Bard, Inc. 129 Concord Road P.O. Box 7031 BILLERICA MA 01821-7031

Re: K033245

Trade/Device Name: Bard® UltraView™ Multiple Band Ligator Regulation Number: 21 CFR §876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: II Product Code: 78 MND Dated: October 6, 2003 Received: October 7, 2003

Dear Mr. Patz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx. 1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2

{4}------------------------------------------------

K03 3245

Indications for Use Statement

Ko33245 510(k) Number (if known): なおり Device Name: Bard® UltraView™ Multiple Band Ligator Indications For Use: The Bard® UltraView™ Multiple Band Ligator is used for endoscopic ligation of esophageal varices.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use V (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

Nancy C. hodon

(Division Sign-C Division of Reproductive, Abdon and Radiological Devices 510(k) Number

§ 876.4400 Hemorrhoidal ligator.

(a)
Identification. A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.(b)
Classification. Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.