(20 days)
The HORIZON 9000 WS (Cathlab) is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature. Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
The Cathlab is capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature. Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats. The Cathlab runs on a UNIX/Sun Solaris* operating system on a SUN® computer that utilizes spowerful, real-time software to control the system operation and to process the vital patient sign data measurements acquired from the Physiological Front End or entered manually at the keyboard. The Thermal Array Chart Recorder provides a continuous recording of all monitored vital signs, patient ID, time and date during the procedures. A Laser Printer is provided in addition to the Chart Recorder in the central console. This provides printouts of textual and graphical summaries of all patient data and catheterization procedures.
This document describes a Special 510(k) for a device modification of the Horizon 9000WS Cathlab, specifically replacing the Patient Front End (PFE) with a Cathlab Front End (CFE). The submission aims to demonstrate that the modified device is substantially equivalent to the predicate device.
Here's an analysis of the provided text, focusing on acceptance criteria and supporting studies:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as "acceptance criteria" but are embedded within the product specifications for the Cathlab Front End (CFE). The device performance is reported as meeting these specifications.
| Parameter | Acceptance Criteria (CFE Specification) | Reported Device Performance |
|---|---|---|
| Input Circuit Parameters | ||
| Chassis Leakage Current | < 50 µA, Meets or exceeds ANSI standard: "Safe Current Limits for Electromedical Apparatus," (SCLE) Dec, 1978 item 2.1.1. | < 50 µA, Meets or exceeds ANSI standard: "Safe Current Limits for Electromedical Apparatus," (SCLE) Dec, 1978 item 2.1.1. |
| ECG | ||
| Frequency Response (Monitor) | 0.5 to 40 Hz | 0.5 to 40 Hz |
| Frequency Response (Diagnostic) | 0.05 to 150 Hz, -3 dB | 0.05 to 150 Hz, -3 dB |
| Frequency Response (Exercise) | 1 to 25 Hz, -3 dB | 1 to 25 Hz, -3 dB |
| Input Impedance | Typical 20 MΩ; Min > 5 MΩ differential (DC to 10 Hz); 2.5 MΩ differential (10 to 100 Hz); 3 MΩ differential (at 10 Hz) | Typical 20 MΩ; Min > 5 MΩ differential (DC to 10 Hz); 2.5 MΩ differential (10 to 100 Hz); 3 MΩ differential (at 10 Hz) |
| Common Mode Rejection | At least 100 dB at 50/60 Hz (Without lead misbalance); 86 dB (with lead misbalance); In accordance with ANSI/AAMI EC11 (9) Para. 3.2.14. | At least 100 dB at 50/60 Hz (Without lead misbalance); 86 dB (with lead misbalance); In accordance with ANSI/AAMI EC11 (9) Para. 3.2.14. |
| Input Dynamic Range | ±5mV p-p at a rate up to 320mV/sec, as per ANSI/AAMI EC13(8) Para. 3.2.9.1. | ±5mV p-p at a rate up to 320mV/sec, as per ANSI/AAMI EC13(8) Para. 3.2.9.1. |
| Input Offset | ± 300mV, as per ANSI/AAMI EC13 Para. 3.2.9.1. | ± 300mV, as per ANSI/AAMI EC13 Para. 3.2.9.1. |
| Noise | Less than 30 μ V p-p referenced to input | Less than 30 μ V p-p referenced to input |
| Pacemaker Pulse Rejection | Reject pulses from 2.0 mV to 700 mV with 0.2 to 2.0 mSec pulse widths and ≥3.0mV for 0.1mSec pulse width | Reject pulses from 2.0 mV to 700 mV with 0.2 to 2.0 mSec pulse widths and ≥3.0mV for 0.1mSec pulse width |
| Defibrillator Protection | Up to 5 KV. Amplifier Recovery time: < 3 seconds | Up to 5 KV. Amplifier Recovery time: < 3 seconds |
| QRS Detection | 0.25 to 5.0 mV, 70-120 msec width | 0.25 to 5.0 mV, 70-120 msec width |
| Synchronous Defibrillation Signal | Pulse Width: 100 ms; Amplitude: 5 Vdc into 500Ω, short-circuit proof | Pulse Width: 100 ms; Amplitude: 5 Vdc into 500Ω, short-circuit proof |
| ECG Analog Output | 1 Volt / mVolt | 1 Volt / mVolt |
| Heart Rate | ||
| Range | 20 to 350 bpm | 20 to 350 bpm |
| Accuracy | Within 2 bpm | Within 2 bpm |
| Response Time | Less than 7 sec for step change of 60 bpm from a base of 60 bpm | Less than 7 sec for step change of 60 bpm from a base of 60 bpm |
| Blood Pressure | ||
| Input Sensitivity | 5 μvolts/volt/mmHg | 5 μvolts/volt/mmHg |
| Transducer Excitation | 5 Volt | 5 Volt |
| Ranges | -50 to +300 mmHg | -50 to +300 mmHg |
| Max Variation during Zero | ± 2 mmHg | ± 2 mmHg |
| Zero Accuracy | ± 0.2 mmHg | ± 0.2 mmHg |
| Zero Drift | Less than +/- 0.2 mmHg in 24 hours | Less than +/- 0.2 mmHg in 24 hours |
| Transducer Load Impedance | 300 – 600Ω | 300 – 600Ω |
| Linearity | Better than 1% of full scale | Better than 1% of full scale |
| Common Mode Rejection | 80 dB minimum (reference to chassis 50/60Hz) | 80 dB minimum (reference to chassis 50/60Hz) |
| Frequency Response | DC to 12 Hz (DC to 40 Hz optional) | DC to 12 Hz (DC to 40 Hz optional) |
| Cardiac Output | ||
| Range | 0.5 to 20 liter/minute | 0.5 to 20 liter/minute |
| Frequency Response | DC to 15 Hz | DC to 15 Hz |
| Blood Temperature Range | 27 °C to 45°C | 27 °C to 45°C |
| Injectate Temperature Range | 0°C to 25°C. (32°F to 77°F) | 0°C to 25°C. (32°F to 77°F) |
| Accuracy | Blood Temp ± 0.05°C; Inj. Temp ± 0.2 °C | Blood Temp ± 0.05°C; Inj. Temp ± 0.2 °C |
| Linearity | Better than 1% of full scale | Better than 1% of full scale |
| Temperature | ||
| Range | 27 °C to 45°C | 27 °C to 45°C |
| Accuracy | ± 0.2°C | ± 0.2°C |
| Respiration | ||
| Frequency Response | 0.13 to 2.5 Hz., 3 dB bandwidth | 0.13 to 2.5 Hz., 3 dB bandwidth |
| Range | 8 to 150 bpm | 8 to 150 bpm |
| Excitation | 65 kHz | 65 kHz |
| Pulse Oximetry (SpO2) | ||
| Probe Type | Masimo™ reusable or disposable | Masimo™ reusable or disposable |
| Range | 0% to 100% | 0% to 100% |
| Pulse Rate Range | 20-250 bpm, below 20 displays zero | 20-250 bpm, below 20 displays zero |
| Rate Accuracy | ± 3 bpm | ± 3 bpm |
| SpO2 Accuracy (Adult) | ±2 digits between 70% and 100%; ±3 digits between 50% and 70% | ±2 digits between 70% and 100%; ±3 digits between 50% and 70% |
| SpO2 Accuracy (Neonatal) | ±3 digits between 70% and 95% | ±3 digits between 70% and 95% |
| Auxiliary Inputs | ||
| Input Voltage | +/-5 Volt | +/-5 Volt |
| Frequency Response | DC to 120 Hz | DC to 120 Hz |
| Non-Invasive Blood Pressure (NIBP) | ||
| Method | Oscillometric | Oscillometric |
| Initial Inflation | 150 mmHg (adult); 120 mmHg (pediatric) | 150 mmHg (adult); 120 mmHg (pediatric) |
| Pressure Accuracy | Overall ± 3 mmHg, full scale | Overall ± 3 mmHg, full scale |
| Defib. Pulse Protection | 5KV as per ANSI/AAMI EC13 (9), clause 3.2.2.2 and per IEC 60601-2-27 (12), clauses 17, 101 and 102 | 5KV as per ANSI/AAMI EC13 (9), clause 3.2.2.2 and per IEC 60601-2-27 (12), clauses 17, 101 and 102 |
| Degree of protection against electrical shock | Type CF and BF (ECG, IBP and CO = CF; NIBP and SpO2 = BF) | Type CF and BF (ECG, IBP and CO = CF; NIBP and SpO2 = BF) |
| Electrosurgical Interference Suppression | Yes | Yes |
2. Sample size used for the test set and the data provenance
The document does not specify a separate "test set" in the context of image-based AI or diagnostic algorithm evaluation. Instead, it states that "The Cathlab with the CFE has been subject to extensive safety and performance testing to ensure that the signals/waveforms sent by the CFE to the Cathlab have the same characteristics as those sent by the PFE to the Cathlab." This implies a comparative testing approach rather than a traditional test set for an algorithm.
The data provenance is not explicitly detailed. It can be inferred that the testing was performed by "Mennen Medical Ltd." (Israel) and an "independent testing laboratory" for electrical safety and EMC. There's no mention of country of origin of data in terms of patient population or images, as this is a physiological monitoring device rather than an imaging or diagnostic device. The testing describes the performance characteristics of the hardware and software, not historical patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a physiological monitor, not an AI or imaging diagnostic device that would require expert-established ground truth for a test set. The "ground truth" here is adherence to technical specifications and industry standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable, as there is no specific "test set" for diagnostic accuracy with human-expert adjudication in this context. The adjudication method for meeting technical standards would typically involve verification by qualified engineers/testers against established metrics.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document pertains to a hardware modification for a physiological monitoring device and does not involve AI for interpretation or improvement of human reader performance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physiological monitor; its performance is described in terms of its ability to acquire and display patient data, not as a standalone diagnostic algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" used for this device modification is based on technical specifications, industry standards (e.g., ANSI/AAMI, IEC), and the performance characteristics of the predicate device (Horizon 9000WS Cathlab with PFE). The goal was to ensure the modified CFE matched or exceeded the performance of the existing PFE.
8. The sample size for the training set
Not applicable. This document describes a hardware device modification, not a machine learning model that would require a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for a machine learning model described.
{0}------------------------------------------------
OCT 1 5 2003
Food and Drug Administration Device Modification - Horizon 9000WS Cathlab: Special 510(k) for new PFE (Patient Front End) for Cathlab
Image /page/0/Picture/2 description: The image shows the logo for Mennen Medical. The logo consists of a lowercase letter 'm' inside of a black circle. The text 'mennenmedical' is written below the logo, with 'mennen' in black and 'medical' in a lighter color.
Mennen Medical Ltd., 4 Hayarden Street, Yavne PO Box 102, Rehovot 76100 Israel
Date: 16 September 2003
Topic: 510(k) Safety and Effectiveness Summary as per 21 CFR Section 807.92(c) Special 510(k): Device Modification - Horizon 9000WS Cathlab
Establishment Name, Registration Number and Address:
| Name: | Mennen Medical Ltd. |
|---|---|
| Registration Number | 9611022 |
| Operator Number: | 9011766 |
| Address: | 4 Hayarden Street, Yavne, 81228, Israel |
| Postal Address: | PO Box 102,Rehovot, 76100, Israel |
| Tel: | +972-8-9323333 |
| Fax: | +972-8-9328510 |
| Contact person: | Asher Kassel, Director of Regulatory Affairs |
| To: | Food and Drug AdministrationCenter for Devices and Radiological HealthDocument Mail Center (HFZ-401)9200 Corporate BoulevardRockville MD, 20850 |
| Attn.: | Document Control Clerk |
| From: | Asher Kassel, Director of Regulatory Affairs |
| Product Name: | |
| Proprietary: | Horizon 9000WS |
| Common: | Cathlab |
| Mennen Medical Part Number: | 960-100-020 (full system) |
| FDA Classification of Cathlab: | |
| Classification Name: | Programmable diagnostic computer |
| Classification Number: | 21 CFR 870.1425 |
| Classification: | Class II |
| Product Code: | DXG |
{1}------------------------------------------------
Performance Standards: None promulgated
Voluntary Standards: IEC 60601-1: General Requirement for Safety for Medical Electrical Systems - part 1, (1988); Amendment 1 - 1991-11 Amendment 2 - 1995-03
IEC 60601-1-2 (2001):
Medical electrical equipment. General requirements for Safety. Electromagnetic Compatibility Requirements and Tests.
IEC 60602-2-27 (1994):
Medical electrical equipment, Part 2, Requirements for safety of electrocardiograph monitoring equipment.
IEC 60601-2-30 (1995):
Medical electrical equipment, Part 2 - requirements for safety of automatic cycling indirect blood pressure monitoring equipment
IEC 60601-2-34 (1994):
Medical electrical equipment, Part 2 - Particular requirements for the safety of direct blood pressure monitoring equipment
IEC 60601-2-49 (2001):
Particular Requirements for the safety of multifunction patient monitoring equipment
Terminology:
PFE = Patient Front End, and refers to the data acquisition module of the Horizon 9000WS, the predicate device before modification
CFE = Cathlab Front End, and refers to the data acquisition module of the Horizon 9000WS after modification, the subject of this Special 510(k).
Predicate Device:
{2}------------------------------------------------
General Description of the Horizon 9000WS Cathlab
The Cathlab is capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature. Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
The Cathlab runs on a UNIX/Sun Solaris* operating system on a SUN® computer that utilizes spowerful, real-time software to control the system operation and to process the vital patient sign data measurements acquired from the Physiological Front End or entered manually at the keyboard. The Thermal Array Chart Recorder provides a continuous recording of all monitored vital signs, patient ID, time and date during the procedures. A Laser Printer is provided in addition to the Chart Recorder in the central console. This provides printouts of textual and graphical summaries of all patient data and catheterization procedures.
Base Configuration: Cathlab parameters
- 4 Invasive Blood Pressure channels
- . Diagnostic 7 or 12 Lead ECG
- Non-invasive Blood Pressure .
- Thermodilution cardiac output
- Pulse Oximetry (SpO2)
- 24 channel thermal array chart recorder .
Horizon 9000 WS Options:
- Full Disclosure ●
- Off-line workstations ●
- Remote Interactive terminal ●
- Angiography Analysis Package ●
- Cardiology Data Base and Inventory ●
- Optomagnetic or CDR drive .
- Choice of Console Table regular, enhanced or compact .
Intended Use of the Horizon 9000WS Cathlab:
The Horizon 9000 WS Cathlab is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature.
Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
{3}------------------------------------------------
K032997
page 4/10
Summary of the technological characteristics of the CFE of the Horizon 9000WS Cathlab
The following tables summarize data on the CFE (modified device) of the Horizon 9000WS Cathlab:
| CFE of Horizon 9000WS Cathlab | |
|---|---|
| Part/Option Number | 960-551-010 (CFE) |
| Dimensions (H x W x D) | 24 x 22 x 10 cm (9 x 11 x 4") |
| Input Circuit Parameters | CFE |
| Chassis Leakage Current | All patient signal inputs fully isolated (<50 $\mu$ A).Meets or exceeds ANSI standard: "Safe Current Limits forElectromedical Apparatus," (SCLE) Dec, 1978 item 2.1.1. |
| CFE | |
|---|---|
| ECG | 7 or 12 leads |
| Frequency Response | Monitor Mode: 0.5 to 40 Hz |
| Diagnostic: 0.05 to 150 Hz, | |
| Exercise: 1 to 25 Hz, -3 dB | |
| Input Impedance: | Typical 20 MΩ |
| Minimum greater than: | |
| 5 MΩ differential, DC to 10 Hz; 2.5 MΩ differential 10 to 100Hz. 3 MΩ differential at 10 Hz | |
| Common Mode Rejection: | At least 100 dB at 50/60 Hz Without lead misbalance86 db with lead misbalance |
| The common mode rejection ratio is in accordance withANSI/AAMI EC11 (9) Para. 3.2.14. | |
| Input Dynamic Range: | ±5mV p-p at a rate up to 320mV/sec, as per ANSI/AAMI EC13(8)Para. 3.2.9.1. |
| Input offset | ± 300mV, as per ANSI/AAMI |
| EC13 Para. 3.2.9.1. | |
| Gain: | Manual selection of 250, 500, 1000, 2000, 4000 and 8000 x ECG.Signal impressed across selected lead |
| Noise: | Less than 30 μ V p-p referenced to input |
{4}------------------------------------------------
K 032997
page 5/10
| ECG | CFE |
|---|---|
| Pacemaker PulseRejection: | Reject pulses from:2.0 mV to 700 mV pulses of 0.2 to 2.0 mSec pulse widths and≥3.0mV for 0.1mSec pulse width |
| Defibrillator Protection: | Up to 5 KV. AmplifierRecovery time: < 3 seconds |
| Lead Fault Sense: | On any ECG electrode |
| QRS Detection: | 0.25 to 5.0 mV, 70-120 msec width |
| Synchronous DefibrillationSignal: | Pulse Width: 100 ms.Amplitude: 5 Vdc amplitude into 500Ω, short-circuit proof |
| ECG Analog Output: | 1 Volt / mVolt |
{5}------------------------------------------------
| Heart Rate | CFE |
|---|---|
| Range: | 20 to 350 bpm |
| Accuracy: | Within 2 bpm. |
| Response Time: | Less than 7 sec for step change of 60 bpm from a base of 60 bpm |
| Blood Pressure | |
| Input Sensitivity: | 5 μvolts/volt/mmHg |
| Transducer Excitation: | 5 Volt |
| Ranges: | -50 to +300 mmHg |
| Maximum variation duringzero: | ± 2 mmHg |
| Zero Accuracy: | ± 0.2 mmHg |
| Zero Drift: | Less than +/- 0.2 mmHg in 24 hours |
| Transducer Load Impedance: | 300 – 600Ω |
| Linearity: | Better than 1% of full scale |
| Common Mode Rejection: | 80 dB minimum (reference to chassis 50/60Hz) |
| Frequency Response: | DC to 12 Hz(DC to 40 Hz optional) |
| Cardiac Output | CFE |
| Range: | 0.5 to 20 liter/minute |
| Frequency Response: | DC to 15 Hz. |
| Blood Temperature Range: | 27 °C to 45°C |
| Injectate Temperature Range: | 0°C to 25°C. (32°F to 77°F) |
| Accuracy: | Blood Temp ± 0.05°C; Inj. Temp ± 0.2 °C |
| Linearity: | Better than 1% of full scale |
| Temperature | CFE |
| Range: | 27 °C to 45°C. |
| Accuracy: | ± 0.2°C. |
| Respiration | CFE |
| Frequency Response: | 0.13 to 2.5 Hz., 3 dB bandwidth. |
| Range: | 8 to 150 bpm. |
| Excitation: | 65 kHz |
| Pulse Oximetry (SpO2) | CFE |
| Probe Type: | Masimo™ reusable or disposable |
| Range: | 0% to 100% |
| Pulse Rate Range: | 20-250 bpm, below 20 displays zero |
| Rate Accuracy: | ± 3 bpm |
| SpO2 Accuracy: | Determined by specific sensor:Adult: ±2 digits between 70% and 100%±3 digits between 50% and 70%.Neonatal: ±3 digits between 70% and 95% |
| Auxiliary Inputs | CFE |
| Input Voltage: | +/-5 Volt |
| Frequency Response: | DC to 120 Hz |
{6}------------------------------------------------
{7}------------------------------------------------
| Non-Invasive BloodPressure (NIBP) | CFE |
|---|---|
| Method: | Oscillometric |
| Initial Inflation: | 150 mmHg (adult)120 mmHg (pediatric). |
| Pressure Accuracy: | Overall ± 3 mmHg, full scale. |
| Defib. Pulse Protection | 5KV as per ANSI/AAMI EC13 (9), clause 3.2.2.2 and perIEC 60601-2-27 (12), clauses 17,101 and 102 |
| Degree of protectionagainst electrical shock | Type CF and BF.ECG, IBP and CO = CFNIBP and SpO2 = BF |
| ElectrosurgicalInterference Suppression | Yes |
| Displayed Waveforms | CFE |
| ECG | Up to 12 lead |
| BP | Up to 4, separate or superimposed |
| Respiration | 1 |
| SpO2 | 1 |
| Displayed NumericParameters | 1 |
| Heart Rate | Yes |
| Respiration Rate | Yes |
| SpO2 | Yes |
| BP - Systolic, Diastolic,Mean | Yes |
| Temperature | 2 |
{8}------------------------------------------------
| Alarm Indications | No |
|---|---|
| Display Functions | CFE |
| Change ECG LeadSelection | YES |
| Display of ArrhythmiaInformation | YES |
| Data Review: Trends | YES |
| Data Review: Tabular | YES |
| User defined ConfigurationSetup | YES |
| User defined DefaultSettings | YES |
| Accessories | Compatible with Mennen Medical Envoy patient monitor |
Conclusion of technological characteristics:
We consider the Horizon 9000WS Cathlab with the CFE to be substantially equivalent to the Horizon 9000WS Cathlab with the PFE. We submit that any differences between the two modules:
- fall within the scope of a Special 510(k) Device Modification and .
- do not raise any new issues of safety and effectiveness ●
Testing
The Cathlab with the CFE has been subject to extensive safety and performance testing in order to ensure that the signals/waveforms sent by the CFE to the Cathlab have the same characteristics as those sent by the PFE to the Cathlab.
Final testing for the whole Cathlab system included various performance tests designed to ensure that the device meets all functional requirements and performance specifications. Electrical Safety testing and EMC testing were performed by an independent testing laboratory to ensure that the device complies to applicable industry and safety standards.
Indications for Use
There is no change to the Indications for Use for the Cathlab with CFE. Mennen intends to use the CFE of the Cathlab for "acquiring and displaying essential patient data" according to the approved "Indications for Use" in K991775. The full "Indications for Use" appear on page 9 below.
{9}------------------------------------------------
K032997
page 10/10
INDICATIONS FOR USE
The HORIZON 9000 WS (Cathlab) is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature.
Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and ,numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three stripes extending from its head, symbolizing health, hope, and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 5 2003
Mennen Medical Ltd. c/o Mr. Asher Kassel Department of Regulatory Affairs P.O. Box 102 Rehovot 76100 ISRAEL
Re: K032997 Trade Name: Horizon 9000WS Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DXG Dated: September 24, 2003 Received: September 25, 2003
Dear Mr. Kassel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
Page 2 - Mr. Asher Kassel
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
MHepernar
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
INDICATIONS FOR USE
The HORIZON 9000 WS (Cathlab) is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature.
Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
UMefelmay
(Division Sign-Off)
Division of Cardiovascular Devi
510(k) Number
§ 870.1435 Single-function, preprogrammed diagnostic computer.
(a)
Identification. A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.(b)
Classification. Class II (performance standards).