(165 days)
The Mercury Medical Negative Inspiratory Force (NIF) Disposable Manometer is used to indicate inspiratory force.
The Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer is designed to connect to an endotracheal tube. To record the maximum inspiratory pressure, the occlusion knob is held down during patient inhalation. The knob is released immediately after measurement. A memory pointer identifies and remains at the maximum inspiratory pressure to -60cm H2O. Rotating the memory pointer knob counterclockwise resets the memory pointer.
The provided text describes a Premarket Notification (510(k)) for the Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer. This submission focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a detailed study with specific acceptance criteria and performance metrics of a novel device. As such, the information required to fully answer your request regarding acceptance criteria and a study proving the device meets them is not explicitly present in the provided text.
Specifically, the document states:
- "Non-Clinical Data: Performance and specifications of the device are consistent with all requirements for this device type specified by ISO 5356-1: 1987 - Anesthetic and Respiratory Equipment-Conical connectors-Part 1: Cones and Sockets. ASTM F1054 - Standard Specification for Conical Fittings of 15mm and 22mm sizes."
- "The comparison to the predicate devices demonstrates that the proposed device is safe and effective and is substantially equivalent to the predicate device."
This indicates that the "acceptance criteria" are adherence to established standards (ISO 5356-1 and ASTM F1054) and demonstrating substantial equivalence to the DHD Medical Products Negative Inspiratory Force Monitoring Kit. The "study" proving this seems to be a comparison to the predicate device and adherence to these standards, but the specifics of how this comparison was performed, what data was collected, or what quantitative acceptance thresholds were used (beyond general consistency with the standards) are not detailed.
Therefore, many of the requested points cannot be answered from the provided text.
Here's a breakdown of what can be inferred or explicitly stated, and what is missing:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Consistency with ISO 5356-1: 1987 | Device performance and specifications are consistent. |
| Consistency with ASTM F1054 | Device performance and specifications are consistent. |
| Substantial Equivalence to Predicate Device | Device is demonstrated to be safe, effective, and substantially equivalent to the DHD Medical Products NIF Monitoring Kit. |
| Maximum Inspiratory Pressure Range | Reads up to -60cm H2O. |
| Measurement of Inspiratory Force | Indicates inspiratory force. |
Note: The specific quantitative thresholds or detailed performance metrics required by ISO 5356-1 and ASTM F1054 are not elaborated upon in this summary.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not provided. The text does not detail any specific clinical or non-clinical test set, sample sizes, or data provenance for a study. It refers to adherence to standards and comparison to a predicate device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable/Not provided. This type of information is typically relevant for studies involving subjective human assessment (e.g., image interpretation). For a manometer, ground truth would likely be established by calibrated instruments, not expert consensus in the human sense. The text does not mention any expert involvement in establishing "ground truth" for performance.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable/Not provided. Given the nature of the device (a pressure manometer) and the focus on standards and substantial equivalence, an adjudication method for a "test set" in the context of human interpretation is not relevant here and is not mentioned.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This device is a simple mechanical manometer, not an AI-powered diagnostic tool. MRMC studies are not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is a physical, disposable manometer, not an algorithm. Its "standalone" performance would be its inherent accuracy and compliance with standards, which is generally what the non-clinical data section alludes to. No specific "standalone study" akin to an algorithm's performance is described.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Inferred: Calibrated measurement devices and established industry standards. For a manometer, the "ground truth" for negative pressure measurements would be provided by highly accurate and calibrated reference instruments. The text mentions consistency with ISO and ASTM standards, which would define the acceptable performance against such a ground truth.
8. The sample size for the training set
- Not applicable/Not provided. This device does not involve machine learning or a "training set."
9. How the ground truth for the training set was established
- Not applicable/Not provided. This device does not involve machine learning or a "training set."
In summary, the provided document focuses on demonstrating that the Mercury Medical NIF Disposable Manometer is "substantially equivalent" to an existing legally marketed device (the DHD Medical Products NIF Monitoring Kit) and adheres to relevant international and industry standards. It does not provide details of a quantitative study with specific acceptance criteria and performance metrics in the way one might expect for a novel or more complex device. The "proof" lies in its conformance to these established benchmarks and standards.
{0}------------------------------------------------
ENCLOSURE L FEB - 3 2004
Premarket Notification (510(k) - Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer
Summary of Safety and Effectiveness
Non-Confidential Summary of Safety and Effectiveness
Page 1 of 2 January 29, 2004
Mercury Enterprises, Inc./Mercury Medical 11300 49th St. N. Clearwater, Fl. 33762 Tel: (800) 237-6418 Fax: (727) 572-4501
Official Contact: Karen Seltzer QA/RA Manager
Proprietary or Trade Name: Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer
Common/Usual Name: NIF Manometer
Classification: Class II, BXR, 21 CFR 868.1780
Classification Name: Airway Pressure Meter (Inspiratory Force)
Device: Negative Inspiratory Force (NIF) Disposable Manometer
Predicate Devices: DHD Medical Products Negative Inspiratory Force Monitoring Kit
Device Description:
The Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer is designed to connect to an endotracheal tube. To record the maximum inspiratory pressure, the occlusion knob is held down during patient inhalation. The knob is released immediately after measurement. A memory pointer identifies and remains at the maximum inspiratory pressure to -60cm H2O. Rotating the memory pointer knob counterclockwise resets the memory pointer.
{1}------------------------------------------------
ENCLOSURE L
Premarket Notification (510(k) - Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer
Non-Confidential Summary of Safety and Effectiveness
Page 2 of 2 January 29, 2004
Indicated Use:
The Mercury Medical® Negative Inspiratory Force (NIF) Disposable Manometer is used to indicate inspiratory force. Federal law (USA) restricts this device to sale by or on the order of a physician.
Technical Characteristics: The device has the same technical characteristics for measuring negative pressure as the DHD Medical Products NIF Monitoring kit predicate device.
Non-Clinical Data: Performance and specifications of the device are consistent with all requirements for this device type specified by ISO 5356-1: 1987 - Anesthetic and Respiratory Equipment-Conical connectors-Part 1: Cones and Sockets. ASTM F1054 - Standard Specification for Conical Fittings of 15mm and 22mm sizes.
Environment of Use: Hospital/Transport
Conclusions: The comparison to the predicate devices demonstrates that the proposed device is safe and effective and is substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of three abstract shapes resembling waves or stylized human figures, stacked on top of each other.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 3 2004
Mr. Wayne Glover QA/RA Engineer Mercury Medical Inc. 11300 49th St. N. Clearwater, FL 33762
Re: K032593
Trade/Device Name: Mercury Medical Negative Inspiratory Force (NIF) Disposable Manometer Regulation Number: 21 CFR 868.1780 Regulation Name: Meter, Airway Pressure (Inspiratory Force) Regulatory Class: Class II Product Code: 73 BXR Dated: December 18, 2003 Received: December 19, 2003
Dear Mr. Glover:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration
{3}------------------------------------------------
Page 2 - Mr. Wayne Glover
and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Chris Lins, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indication for Use
| 510(k) Number (if known): | K032593 |
|---|---|
| Device Name: | Negative Inspiratory Force (NIF) Disposable Nanometer |
| Intended Use: | The Mercury Medical Negative Inspiratory Force (NIF) Disposable Manometer is used to indicate inspiratory force. |
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L.A. Winterhausen
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
2 510(k) Number:
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
§ 868.1780 Inspiratory airway pressure meter.
(a)
Identification. An inspiratory airway pressure meter is a device used to measure the amount of pressure produced in a patient's airway during maximal inspiration.(b)
Classification. Class II (performance standards).