K Number
K032554
Manufacturer
Date Cleared
2003-12-31

(134 days)

Product Code
Regulation Number
866.3390
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APTIMA Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal ribonucleic acid (rRNA) from Chlamydia trachomatis and/or Neisseria gonorrhoeae in clinician-collected endocervical, vaginal, and male urethral swab specimens, patient-collected vaginal swab specimens* and male and female urine specimens. The assay may be used to test specimens from symptomatic and asymptomatic individuals to aid in the diagnosis of gonococcal and/or chlamydial urogenital disease.

*Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use.

The APTIMA Vaginal Swab Specimen Collection Kit is for use with the APTIMA Assays for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae. The APTIMA Vaginal Swab Specimen Collection Kit is intended to be used for clinician and patient collection of vaginal swab specimens according to the instructions provided. Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The APTIMA Vaginal Swab Specimen Collection Kit is not for home use.

Device Description

Clearance of this premarket notification extends the clinical performance claims of the commercially available GEN-PROBE APTIMA Combo 2 Assay to include clinician-collected and patient-collected vaginal swabs (in a medical setting) as acceptable testing specimens. The ancillary kit formulated for this specific application is the GEN-PROBE APTIMA Vaginal Swab Specimen Collection Kit. The components of the APTIMA Vaginal Swab Specimen Collection Kit include: (1) a sterile swab for the collection of vaginal specimens and (2) a transport tube containing transport media with a penetrable cap.

The APTIMA Vaginal Swab Specimen Collection Kit is for use with the APTIMA Assays for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae. The APTIMA Vaginal Swab Specimen Collection Kit is intended to be used for clinician and patient collection of vaginal swab specimens according to the instructions provided. The APTIMA Vaginal Swab Specimen Collection Kit not for home use.

The APTIMA Combo 2 Assay incorporates the technologies of target capture, in vitro nucleic acid amplification, and hybridization of target amplicons with acridinium ester-labeled DNA probes to specifically detect and differentiate both C. trachomatis and N. gonorrhoeae nucleic acids in clinical specimens. GEN-PROBE's proprietary technologies are combined in this product to allow qualitative detection of C. trachomatis rRNA and N. gonorrhoeae rRNA.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the GEN-PROBE® APTIMA® Combo 2 Assay, based on the provided text:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied through the sensitivity and specificity values reported, which are compared against a "patient infected status algorithm" as the ground truth. While explicit numerical acceptance thresholds are not stated in the provided text (common in older 510(k) summaries which often relied on substantial equivalence to predicate devices), the study aims to demonstrate high performance.

Chlamydia trachomatis (CT)

MetricAcceptance Criteria (Implied by High Performance)Reported Performance (All Vaginal Swab Specimens)Reported Performance (Patient-Collected Vaginal Swab)Reported Performance (Clinician-Collected Vaginal Swab)
SensitivityHigh96.6% (95% CI: 92.8 - 98.8%)96.6% (95% CI: 92.8 - 98.8%)96.7% (95% CI: 92.9 - 98.8%)
SpecificityHigh97.8% (95% CI: 96.8 - 98.5%)97.8% (95% CI: 96.8 - 98.5%)97.1% (95% CI: 96.0 - 97.9%)

Neisseria gonorrhoeae (NG)

MetricAcceptance Criteria (Implied by High Performance)Reported Performance (All Vaginal Swab Specimens)Reported Performance (Patient-Collected Vaginal Swab)Reported Performance (Clinician-Collected Vaginal Swab)
SensitivityHigh98.7% (95% CI: 93.0 - 100%)98.7% (95% CI: 93.0 - 100%)96.2% (95% CI: 89.2 - 99.2%)
SpecificityHigh99.6% (95% CI: 99.0 - 99.8%)99.6% (95% CI: 99.0 - 99.8%)99.4% (95% CI: 98.8 - 99.7%)

2. Sample Size Used for the Test Set and Data Provenance

The study was a multi-center clinical study involving 1,464 symptomatic and asymptomatic female subjects.

  • Sample Size:
    • For CT analysis: 2,868 vaginal swab test results.
    • For NG analysis: 2,867 vaginal swab test results.
    • These numbers represent the combined results from patient-collected and clinician-collected vaginal swabs.
  • Data Provenance: The data was collected from subjects attending STD, OB/GYN, teen, and family planning clinics. The text does not explicitly state the country of origin, but given it's a US FDA submission, it's highly likely to be a prospective multi-center clinical study conducted in the United States.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The ground truth was established using an algorithm, not directly by human experts interpreting images or test results. Therefore, the concept of "number of experts" and "qualifications of those experts" as typically applied to expert review in imaging studies, does not directly apply here. The accuracy of the reference standard relies on the performance of the commercially available NAATs used.

4. Adjudication Method for the Test Set

The adjudication method for establishing the patient infected status was algorithmic (rule-based):

  • Subjects were considered infected with C. trachomatis or N. gonorrhoeae if two of the four reference NAAT results were positive (one specimen testing positive in each NAAT: APTIMA Combo 2 Assay and another commercially available NAAT, using endocervical swab and urine specimens).
  • Subjects were considered non-infected if less than two reference NAAT results were positive.

This method uses concordant results from multiple tests/specimens to define the "true" infection status.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This study is for an in vitro diagnostic (IVD) assay detecting biomarkers, not an AI-assisted diagnostic imaging device that involves human reader interpretation. Therefore, the concept of human readers improving with AI vs. without AI assistance does not apply.

6. Standalone Performance Study

Yes, a standalone performance study was done for the algorithm (the APTIMA Combo 2 Assay). The entire clinical study described evaluates the performance of the device on its own, comparing its results from vaginal swab specimens to the defined patient infected status algorithm. The reported sensitivity and specificity values are for the device operating as a standalone diagnostic tool for vaginal swab specimens.

7. Type of Ground Truth Used

The ground truth used was an algorithmic consensus based on multiple reference nucleic acid amplification tests (NAATs). Specifically:

  • Endocervical swab and urine specimen results from the commercially-available APTIMA Combo 2 Assay and another commercially-available NAAT were used.
  • Infection status was determined by requiring a positive result from both reference NAATs.
  • Culture was not used as a reference test for this specific study, as the APTIMA Combo 2 Assay had previously been evaluated against culture for other specimen types (as per K003395).

8. Sample Size for the Training Set

The document does not provide information regarding a distinct "training set" sample size. For an IVD assay like this, development typically involves analytical validation (sensitivity, specificity, interference etc.) and then clinical validation with a distinct set of clinical samples. The analytical studies (like analytical sensitivity, specificity, recovery, interference) might be considered analogous to early-stage development/testing, but a formal "training set" in the machine learning sense is not explicitly mentioned or typically applicable for this type of assay in this context. The clinical study described served as the primary performance evaluation.

9. How the Ground Truth for the Training Set Was Established

As noted above, the concept of a separate "training set" with established ground truth in the machine learning context is not detailed. The primary clinical study used the algorithmic consensus described in point 4 as its ground truth for evaluating the device's performance.

{0}------------------------------------------------

Summary of Safety and Effectiveness K032554

DEC 3 1 2003

2. SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________

Gen-Probe Incorporated

Confidential

{1}------------------------------------------------

GEN-PROBE® APTIMA® Combo 2 Assay GEN-PROBE® APTIMA® Vaginal Swab Specimen Collection Kit

Sponsor/Contact Information

Submitted By: Gen-Probe Incorporated 10210 Genetic Center Drive San Diego, California 92121 (858) 410-8000 (858) 410-8622 (FAX)

Company Contact: Alan Maderazo, Ph.D., RAC Regulatory Affairs Specialist (858) 410-8332 (858) 410-8622(FAX) alanma@gen-probe.com

{2}------------------------------------------------

Summary of Safety and Effectiveness K032554

General Information

Trade Name:GEN-PROBE® APTIMA® Combo 2 Assay
Common or Usual Name:Ribosomal RNA (rRNA) target-amplified nucleic acid probetest for the in vitro diagnostic detection of Chlamydiatrachomatis and/or Neisseria gonorrhoeae
Classification Names:DNA Probe, Nucleic Acid Amplification, ChlamydiaDNA Reagents, Neisseria

APTIMA Combo 2 Assay

Device DescriptionDNA Probe, Nucleic Acid Amplification, Chlamydi
Medical SpecialtyMicrobiology
Product CodeMKZ
Device Class1
Regulation number866.3120
Device DescriptionDNA Reagents, Neisseria
Medical SpecialtyMicrobiology
Product CodeLSL
Device Class2
Regulation number866.3390

Substantially Equivalent Devices:

APTIMA Combo 2 Assay; K003395

{3}------------------------------------------------

Device Description

Clearance of this premarket notification extends the clinical performance claims of the commercially available GEN-PROBE APTIMA Combo 2 Assay to include clinician-collected and patient-collected vaginal swabs (in a medical setting) as acceptable testing specimens. The ancillary kit formulated for this specific application is the GEN-PROBE APTIMA Vaginal Swab Specimen Collection Kit. The components of the APTIMA Vaginal Swab Specimen Collection Kit include: (1) a sterile swab for the collection of vaginal specimens and (2) a transport tube containing transport media with a penetrable cap.

The APTIMA Vaginal Swab Specimen Collection Kit is for use with the APTIMA Assays for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae. The APTIMA Vaginal Swab Specimen Collection Kit is intended to be used for clinician and patient collection of vaginal swab specimens according to the instructions provided. The APTIMA Vaginal Swab Specimen Collection Kit not for home use.

The background information describing the relevant diseases, Chilamydia trachomatis and Neisseria gonorrhoeae, and the principles of the APTIMA Combo 2 Assay are provided in the Summary of Safety and Effectiveness for the APTIMA Combo 2 Assay (K003395)

{4}------------------------------------------------

Intended Use

AC2 Assay package insert:

The APTIMA Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal ribonucleic acid (rRNA) from Chlamydia trachomatis and/or Neisseria gonorrhoeae in clinician-collected endocervical, vaginal, and male urethral swab specimens, patient-collected vaginal swab specimens* and male and female urine specimens. The assay may be used to test specimens from symptomatic and asymptomatic individuals to aid in the diagnosis of gonococcal and/or chlamydial urogenital disease.

*Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use.

Ancillary Kit package insert:

The APTIMA Vaginal Swab Specimen Collection Kit is for use with the APTIMA Assays for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae. The APTIMA Vaginal Swab Specimen Collection Kit is intended to be used for clinician and patient collection of vaginal swab specimens according to the instructions provided. Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The APTIMA Vaginal Swab Specimen Collection Kit is not for home use.

{5}------------------------------------------------

Summary of Technological Characteristics of the APTIMA Combo 2 Assay

The APTIMA Combo 2 Assay incorporates the technologies of target capture, in vitro nucleic acid amplification, and hybridization of target amplicons with acridinium ester-labeled DNA probes to specifically detect and differentiate both C. trachomatis and N. gonorrhoeae nucleic acids in clinical specimens. GEN-PROBE's proprietary technologies are combined in this product to allow qualitative detection of C. trachomatis rRNA and N. gonorrhoeae rRNA.

Analytical Sensitivity

Chlamydia trachomatis analytical sensitivity (limits of detection) was determined by directly comparing dilutions of C. trachomatis organisms in cell culture and in the assay. The analytical sensitivity claim for the assay is one Inclusion-Forming Unit (IFU) per assay (7.25 IFU/swab, 5 IFU/mL urine) for all 15 C. trachomatis serovars. However, dilutions of less than one IFU/assay of all serovars tested positive in the APTIMA Combo 2 Assay.

Neisseria gonorrhoeae analytical sensitivity was determined by directly comparing dilutions of 57 different clinical isolates in culture and in the APTIMA Combo 2 Assay. The analytical sensitivity claim for the assay is 50 cells/assay (362 cells/swab, 250 cells/mL urine). However, all strains tested were positive at less than 50 cells/assay.

{6}------------------------------------------------

Analytical Specificity

A total of 154 culture isolates were evaluated using the APTIMA Combo 2 Assay. These isolates included 86 organisms that may be isolated from the urogenital tract and 68 additional organisms that represent a phylogenetic cross-section of organisms. The tested organisms included bacteria, fungi, yeast, parasites, and viruses. All organisms except C. psittaci, C. pneumoniae, and the viruses were tested at 1.0x106 cells/assay in both swab and urine transport medium. C. psittaci and C. pneumoniae were tested at 1.0x10 TFU/assay. The viruses were tested as follows: (a) herpes simplex viruses I and II: 6.0x104 TCID50/assay, (b) human papilloma virus 16: 2.9x106 DNA copies/assay and (c) cytomegalovirus: 1x106 infected cell culture cells/assay. Only C. trachomatis and N. gonorrhoeae samples produced positive results in the APTIMA Combo 2 Assay. The list of organisms tested is shown in the following table.

{7}------------------------------------------------

Analytical Specificity

ORGANISMORGANISMORGANISM
Achromobacter xerosisEscherichia coliNeisseria mucosa (3)
Acinetobacter calcoaceticusFlavobacterium meningosepticumNeisseria sicca (3)
Acinetobacter IwoffiFusobacterium nucleatumNeisseria subflava (14)
Actinomyces israeliiGardnerella vaginalisNeisseria perflava
Actinomyces pyogenesGemella haemolysansNeisseria polysaccharea
Aerococcus viridansHaemophilus ducreyiParacoccus denitrificans
Aeromonas hydrophilaHaemophilus influenzaePeptostreptococcus anaerobius
Agrobacterium radiobacterHerpes simplex virus IPeptostreptococcus productus
Alcaligenes faecalisHerpes simplex virus IIPlesiomonas shigelloides
Bacillus subtilisHuman papilloma virus 16Propionibacterium acnes
Bacteriodes fragilisKingella dentrificansProteus mirabilis
Bacteriodes ureolyticusKingella kingaeProteus vulgaris
Bifidobacterium adolescentisKlebsiella oxytocaProvidencia stuartii
Bifidobacterium breviKlebsiella pneumoniaePseudomonas aeruginosa
Branhamella catarrhalisLactobacillus acidophilusPseudomonas fluorescens
Brevibacterium linensLactobacillus brevisPseudomonas putida
Campylobacter jejuniLactobacillus jensoniiRahnella aquatilis
Candida albicansLactobacillus lactisRhodospirillum rubrum
Candida glabrataLegionella pneumophila (2)Saccharomyces cerevisiae
Candida parapsilosisLeuconostoc paramensenteroidesSalmonella minnesota
Candida tropicalisListeria monocytogenesSalmonella typhimurium
Chlamydia pneumoniaeMicrococcus luteusSerratia marcescens
Chlamydia psittaci (2)Moraxella lacunataStaphylococcus saprophyticus
Chromobacterium violaceumMoraxella osloensisStaphylococcus aureus
Citrobacter freundiiMorganella morganiiStaphylococcus epidermidis
Clostridium perfringensMycobacterium smegmatisStreptococcus agalactiae
Corynebacterium genitaliumMycoplasma genitaliumStreptococcus bovis
Corynebacterium xerosisMycoplasma hominisStreptococcus mitis
Cryptococcus neoformansN. meningitidis Serogroup AStreptococcus mutans
CytomegalovirusN. meningitidis Serogroup BStreptococcus pneumoniae
Deinococcus radioduransN. meningitidis Serogroup C (4)Streptococcus pyogenes
Derxia gummosaN. meningitidis Serogroup DStreptococcus salivarius
Eikenella corrodensN. meningitidis Serogroup YStreptococcus sanguis
Enterobacter aerogenesN. meningitidis Serogroup W135Streptococcus griseinus
Enterobacter cloacaeNeisseria cinerea (4)Trichomonas vaginalis
Enterococcus aviumNeisseria dentrificansUreaplasma urealyticum
Enterococcus faecalisNeisseria elongata (3)Vibrio parahaemolyticus
Enterococcus faeciumNeisseria flavaYersinia enterocolitica
Erwinia herbicolaNeisseria flavescens (2)
Erysipelothrix rhusiopathiaeNeisseria lactamica (9)

(n) = number of strains tested

(1) - hamber or earline tested in the APTIMA Combo 2 Assay based on kinetic profile type and RLU.

{8}------------------------------------------------

Interference Studies

The following table lists the commonly encountered substances found in vaginal swab specimens that were tested in the assay. All were tested for potential assay interference in the absence and presence of C. trachomatis and N. gonorrhoeae at the estimated rRNA equivalent of one C. trachomatis IFU/assay (5 fg/assay) and 50 N. gonorrhoeae cells/assay (250 fg/assay). The rRNA equivalents were calculated based on the genome size and estimated DNA:RNA ratio/cell of each organism.

SwabUrine
10% Blood30% Blood
Contraceptive jellyUrine analytes:
SpermicideProtein
MoisturizerGlucose
Hemorrhoidal anestheticKetones
Body oilBilirubin
PowderNitrate
Anti-fungal creamUrobilinogen
Vaginal lubricantspH 4 (acidic)
Feminine spraypH 9 (alkaline)
Leukocytes (1 x 106 cells/mL)Leukocytes (1 x 106 cells/mL)
Cellular debris
Vitamins
Minerals
Acetaminophen
Aspirin
Ibuprofen

Interfering Substances Testing

No interference was observed with any of the tested substances.

Gen-Probe Incorporated

{9}------------------------------------------------

Recovery

Lactobacillus acidophilus, Gardnerella vaginalis, Bacteroides ureolyticus and Staphylococcus epidermidis (1 x 108 cells/assay) were added to pooled negative vaginal swab samples containing the rRNA equivalent of approximately one C. trachomatis IFU (5 fg) and 50 N. gonorrhoeae cells (250 fg). These additions did not interfere with the amplification and detection of C. trachomatis or N. gonorrhoeae rRNA using the APTIMA Combo 2 Assay.

Vaginal Swab Specimen Clinical Study Results

In the vaginal swab specimen multi-center clinical study, 1,464 symptomatic and asymptomatic female subjects attending STD, OB/GYN, teen, and family planning clinics were enrolled into the clinical study. Subjects were classified as symptomatic if symptoms such as discharge, dysuria, and pelvic pain were reported by the subjects were classified as asymptomatic if the subject did not report symptoms. Of the 646 asymptomatic subjects enrolled in the study, two were less than 16 years of age, 158 were between the ages of 16 and 20, 231 were between the ages of 21 and 25, and 255 were greater than 25 years of age. Of the 818 symptomatic subjects enrolled in the study, 160 were between the ages of 16 and 20, 324 were between the ages of 21 and 25, and 334 were greater than 25 years of age.

Five specimens were collected from each eligible subject; one urine specimen, one patientcollected vaginal swab, one clinician-collected vaginal swab, and two randomized endocervical swabs. APTIMA Combo 2 Assay results were generated from the two vaginal swabs, one of the endocervical swabs, and an aliguot of the urine specimen. The second endocervical swab and a second aliquot of the urine specimen were tested using another commercially-available nucleic acid amplification test (NAAT) for C. trachomatis and another commercially-available NAAT for N. gonorrhoeae. Endocervical swab and urine specimens tested in the APTIMA Combo 2 Assay and the other commercially available NAATs were used as reference NAATs to determine infected status for each subject in the vaginal swab specimen clinical study.

Gen-Probe Incorporated

Confidential

{10}------------------------------------------------

Specimen testing was conducted either at the site of subject enrollment or at an external testing site.

All performance calculations were based on the total number of APTIMA Combo 2 Assay patient- and clinician-collected vaginal swab results compared to a patient infected status algorithm. A total of 2,868 C. trachomatis and 2,867 N. gonorrhoeae vaginal swab test results were used in the data analysis. In the algorithm, the designation of a subject as being infected or not infected with C. trachomatis or N. gonorrhoeae was based on endocervical swab and urine specimen results from the commercially-available APTIMA Combo 2 Assay and the other commercially-available NAAT. Subjects were considered infected with C. trachomatis or N. gonorrhoeae if two of the four endocervical swab and urine specimens tested positive in the APTIMA Combo 2 Assay and the other reference NAAT (one specimen testing positive in each NAAT). Subjects were considered non-infected if less than two reference NAAT results were positive. Tables 1 and 2 summarize the number of results from symptornatic and asymptomatic subjects designated as infected or non-infected with C. trachomatis or N. gonorrhoeae, respectively, according to the patient infected status algorithm. For this clinical study, two commercially available NAATs were used to determine GC-infected status. Culture was not used as a reference test since the APTIMA Combo 2 Assay has already been evaluated against culture for other specimen types (refer to K003395).

Sensitivity and specificity for C. trachomatis by gender, specimen type and symptom status are presented in Table 3. Table 4 shows the APTIMA Combo 2 Assay sensitivity, specificity, and predictive values for C. trachomatis compared to patient infected status for each clinical site and overall. Sensitivity and specificity for detection of N. gonorrhoeae by gender, specimen type and symptom status are presented in Table 5. Table 6 shows the N. gonorrhoeae sensitivity, specificity, and predictive values for the APTIMA Combo 2 Assay compared to patient infected status for each clinical site and overall. Samples that were APTIMA Combo 2 Assay positive and infected patient status negative (i.e., apparent false positives) were tested in alternate TMA assays for C. trachomatis and N. gonorrhoeae that targeted sequences unique

Gen-Probe Incorporated

Confidential

{11}------------------------------------------------

from those targeted in APTIMA Combo 2. The results of the alternate TMA assays were not used to change the original patient categorizations (Tables 3 and 5).

Of the 1,464 subjects enrolled, there were 13 subjects with unknown CT patient infected status and 14 subjects with unknown GC patient infected status. Subjects were designated with an unknown patient infected status if results were missing that prevented conclusive determination of infected status. These subjects' results were not included in any performance calculations. Of the 5,782 APTIMA Combo 2 Assay vaginal swab results from the multi-center clinical study, there was a small percentage (28, 0.5%) of vaginal swab specimens that initially tested invalid or equivocal for CT or GC. Upon repeat testing only three C. trachomatis results and two N. gonorrhoeae results were equivocal and were excluded from the analysis. No specimens tested invalid on repeat testing.

{12}------------------------------------------------

EffectivenessandSafetyof
.maryK032554
r
NAAT 1NAAT 2 (APTIMA Combo 2 Assay)APTIMA Combo 2 AssaySymptom StatusTotal
Patient Infected StatusEndocervical SwabUrineEndocervical SwabUrinePatient-Collected Vaginal SwabClinician-Collected Vaginal SwabSymptomaticAsymptomatic
Infected++++++7943122
Infected++++++011
Infected++++N/A+8513
Infected+++-N/A+101
Infected+++=++101
Infected++++++8311
Infected++++++123
Infected+++N/A++101
Infected++++++538
Infected++++++505
Infected++++++202
Infected++++++145
Non-infected------011
Non-infected------044
Non-infected------213
Non-infected----N/A-6410
Non-infected-+-+-+101
Non-infected-+-+-+426
Non-infected-+-+-+101
Non-infected-+-+-+202
Non-infected-+-+-+123
Non-infected-+-+-+235
NAATT 2 (APTIMA Combo Assay)NAPTIMA Combo 2 Assa/mptom Stat
tient Infecte StatusndocervicSwabUrineEndocervicaSwabUrinePatient- Collected Iginal SwaClinician- Collected Taginal SwamptomaticAsymptomatГоса
-+-
l+ਟਾ ਕੇ ਉੱದ ೧೯೭೪ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧೯೧೭ ೧ារ ប្រជាជនជាតិ ។ ស
---)
--N/A0
-1
1--+0
l-1ll861
----る。 そのため、その他 その他の あなたのです。 その他の人気がない ないかなかなかなかなかったのです。 その他の人気がないということです。 その他の人気がない。 その他の人気がない。 その他の人気がない。 その他の人気がない。 その他の人気がない。 その他の人気がない。 その他の人気がない。 その他の人気がない。 その他の人気がない。 その他N/AI0
-l1lll
ll1==+
l-1-নেP
-VIN VIN1-N/A-00 NO Do
lll-ll6N
----N/A
=-l-1
tlN/A-l0 ທ
*-N/Al--p6
-ll+
ll
Total811હ્વા્વર્સ

able 1: C. trachomatis Patient- and Clinician-Collected Vaginal Swab Specimen Analysis for Female Patient Infected Statu

Confidentia

Gen-Probe Incorporated

13

{13}------------------------------------------------

. . .mary of Safety and Effectivene
K032554

N/A = Specimen not obtained or available for testing
= represents equivocal on repeat testing

  • Represents equivalent on repeat testing

Confidential

Gen-Probe Incorporate

{14}------------------------------------------------

Summary of Safety and Effectiveness
K032554
Patient InfectedStatusNAAT 1NAAT 2 (APTIMA Combo 2 Assay)APTIMA Combo 2 AssaySymptom StatusTotal
EndocervicalSwabUrineEndocervicalSwabUrinePatient-CollectedVaginal SwabClinician-CollectedVaginal SwabSymptomaticAsymptomatic
Infected++++++441559
Infected++++++101
Infected++++N/A+011
Infected++++++101
Infected++++++224
Infected++++++112
Infected++++++112
Infected++++++011
Infected++++++101
Infected++++++516
Non-infected+-+-++101
Non-infected++++++527
Non-infected++++++011
Non-infected++++++213
Non-infected++++++202
Non-infected++++++224
Non-infected------6985771275
Non-infected+N/A+N/AN/A=022
Non-infected+N/A+N/AN/AN/A15924
Non-infected+N/A+N/AN/AN/A101
Non-infected+N/A+N/AN/AN/A224
Non-infected+N/A+N/AN/AN/A111021

Table 2: N. gonorrhoeae Patient- and Clinician-Collected Vaginal Swab Specimen Analysis for Female Patient Infected Statu

Confidentia

Gen-Probe Incorporated

15

{15}------------------------------------------------

mary of Safety and Effectivenes ﺮ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍ

2554
03
Patient InfectedStatusNAAT 1NAAT 2 (APTIMA Combo 2 Assay)APTIMA Combo 2 AssaySymptom StatusTotal
EndocervicalSwabUrineEndocervicalSwabUrinePatient-CollectedVaginal SwabClinician-CollectedVaginal SwabSymptomaticAsymptomatic
Non-infected-=---N/A011
Non-infected=-----112
Non-infected--N/A---011
Non-infected-----N/A549
Non-infected--=---112
Total8106401450

N/A = Specimen not obtained or available for testing
= represents equivocal on repeat testing

PERSPECTIVES ON
ISSUES IN INTEGRATED
STUDIES

Confidential

Gen-Probe Incorporated

{16}------------------------------------------------

Summary of Safety and Effectiveness
K032554
SpecimerSymptom StatusTDFPTNFNSensiti vity (95% C.I.Specificity(95% C.I
Patient-Collected Vaginal SwatAsymptomatic288રહ। ୧.રવેવ08:491.5 - 100 (92.8 - 98.897.2(95.4 - 98.4)
All42317228b217996.697.896.8 - 98.5)
inician-Collected Vaginal SwabSymptomatic80011323°રવેતPતે છે. 696.7(95.1 - 97.9) (96.0 - 98.7) (96.0 - 97.9)
Asymptomaticર 3661l Agર રતેCતેરૂ જિલ્લામાં આવેલું એક ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે91.5 - 99.1 189.0 - 99.6 192.9 - 98.897.6
144537°22896.797 l
CT TMA Alternate Amplification results: # positive results/# specimens test
a: 13/16 b: 24/28 c: 15/23 d: 13/14 e: 28/37

ble 3: C. trachomatis Sensitivity and Specificity: APTIMA Combo 2 Assay Vaginal Swab Specimens vs. Patient Infected Statu

Gen-Probe Incorporate

:

Confidential

{17}------------------------------------------------

and Effectiveness
Safetyof
.larySL.K032554
Table 4: C. trachomatis Performance by Clinical Site: APTIMA Combo 2 Assay Vaginal Swab Specimens vs. Patient Infected Status
SpecimenSiteNTPFPTNFNPrev (%)Sensitivity (95% C.I.)Specificity (95% C.I.)PPV (%)NPV (%)
Patient-CollectedVaginal Swab1218346177116.197.1 (85.1 - 99.9)96.7 (93.0 - 98.8)85.099.4
2195506137226.796.2 (86.8 - 99.5)95.8 (91.1 - 98.4)89.398.6
311110110009.0100 (69.2 - 100)99.0 (94.6 - 100)90.9100
426120423618.095.2 (76.2 - 99.9)98.3 (95.8 - 99.5)83.399.6
519913318306.5100 (75.3 - 100)98.4 (95.4 - 99.7)81.3100
6290325251211.794.1 (80.3 - 99.3)98.0 (95.5 - 99.4)86.599.2
71021029009.8100 (69.2 - 100)97.8 (92.4 - 99.7)83.3100
847314306.4100 (29.2 - 100)97.7 (88.0 - 99.9)75.0100
ALL1423172281217612.596.6 (92.8 - 98.8)97.8 (96.8 - 98.5)86.099.5
Clinician-CollectedVaginal Swab1227358182216.394.6 (81.8 - 99.3)95.8 (91.9 - 98.2)81.498.9
2196505139226.596.2 (86.8 - 99.5)96.5 (92.1 - 98.9)90.998.6
311310210108.8100 (69.2 - 100)98.1 (93.2 - 99.8)83.3100
4262201023118.095.2 (76.2 - 99.9)95.9 (92.5 - 98.0)66.799.6
519913218406.5100 (75.3 - 100)98.9 (96.2 - 99.9)86.7100
6296348254011.5100 (89.7 - 100)96.9 (94.1 - 98.7)81.0100
7102919119.890.0 (55.5 - 99.7)98.9 (94.1 - 100)90.098.9
850314606.0100 (29.2 - 100)97.9 (88.7 - 99.9)75.0100
ALL1445174371228612.596.7 (92.9 - 98.8)97.1 (96.0 - 97.9)82.599.5

Confidential

Gen-Probe Incorporated

18

{18}------------------------------------------------

Status
-
Property
1
1
Vaginal Swab Specimens vs. Patient Infected Status------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
a at 10 alle the la
1 1993 1
ombe
-
DITTS
necificity:
1
MITS UNIVERSITY CAS CALLERY CAR CALTACA WA
Ancitivity and
ver-
Onom044.
able 5: N. g
C
and and and and and the commend of the commend of the commend of the comments of the first and the mail of the first and the first and the former the former the forme the for
SpecimerSymptom StatusN1 LTTFPTNFNSensitivity95% C.I.specificit95% C.I.
Patient-Collected Vaginal SwabAsymptomatic659------------------------------------------------------------------------------------------------------------------------------------------------------------------------------22くなখ୧୦ર10010084.6 -99.7100198.8 -
All42376ep34098.710093.0 -99.699.0 - 99.8
'linician-Collected Vaginal Swab Symptomatic807ਟ ਤੇર્દ ૮747196.4તેત્વે છ87.5 - .99.398.5 - 99.8)
Asymptomatic63722p &QIIવેરે 778.1 - 99.9,વેત્તે રે98.6 - 99.9
Al444758.35896.289.2 - 99.299.498.8 - 99.7
(1) Marte Alexanda Amerificantian maultar # manimano rach

'GC TMA Alternate Amplification results: # positive results/# specimens tested
a: 2/2 b: 6/6 c: 5/5 d: 2/3 e: 7/8

28.66.55.22.33.7

Confidential

Gen-Probe Incorporated

{19}------------------------------------------------

able 6: N. gonorrhoeae Performance by Clinical Site: APTIMA Combo 2 Assay Vaginal Swab Specimens vs. Patient Infected Stati

SpecimeSiteNTPFPTNFNPrev (%)Sensitivity(95% C.I.)Specificity(95% C.I.)PPVNPV
(%)(%)
217I 320392.9ರಿಗೆ ಕೆಳಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರ्२० ।100
atient-Collect aginal SwabCો તેર164-00000000100ಕ್ಷಣ ಸಿಕ್ಷಣ ಸ್ಥಳದ ಪ್ರತಿ ಸ್ಥೆಗಳು ಸೇರಿ ಸೇರಿ ಸೇರಿ ಸೇರಿ ಸೇರಿ ಸ일 중 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공 공್ದ ಕ್ಷತ್ರ ಕ್ರಿ ಕೃತ ಕೃತ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ರಿ ಕ್ಷಿತ್ರ ಪ್ರಸಿ
100
261100
SI dd100
9-----------ទ្ធសន្ធមាន ១៩ នាក់ ១៩ នាក់ ១៩ នាក់ ១៩ នាក់ ១៩ នាក់ ១៩ នាក់ ១៩ នាក់ ១៩ នោះ ១៩ នោះ ១៩ នោះ ១៩ នោះ ១៩ នោះ ១៩ នោះ ១៩ នោះ ១៩ នោះ ១៩ នោះ ១៩ នោះ ១៩ នោះ ១៩ នាក់ ១៩ នាក់ ១៩ នាក់ ១៩ នាള്ള ജ്വീട്ടില്ല് ജീവിച്ചിരിച്ചു. അവലംബം വേട്ടി ജീവിച്ച വിവ
201 0671/11 00 T
471001
ALL142314 4 6 0 0 0 0 0 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1 2 6 1340ហូតមួយ ១០០០ ដក001 10011 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1ക്രിക്കുന്നത്. അവലംബം അവലംബം പ്രവർത്തിച്ചു. അവലംബം കുറച്ചു വിവരുന്നു വാഹ്യത്തിൽ നിന്നും വിശ്വാസ്ത്രിക്കുന്നു. അവലംബം അവലംബം അവലംബം
nician-Collec227100100100
aginal Swa100
ర్లాల75.06 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6- $61. - 88.8.8 - 78.82
P100100)
S100100(47.8 - 15.8 -റ്റും ഉപ്പുഴി സംഭവ ഗ്ഗങ്ങളുടെ ഭാഗം ഗ്ഗങ്ങളുടെ ഭ001 1666 (001 6666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666666ತ್ತಿದ್ದ ಸ್ನೇಶ್ವಿ ಪ್ರಶಸ್ತಿ ಶ್ವಿತ್ವತ ಕ್ರಿ ಪ್ಲೆ ಶ್ವಿತ ಶ್ವಿ ಶ್ವಿ ಶ್ವಿ ಶ್ವಿ ಶ್ವಿ ಶ್ರೀ ಶೇಖ
9001 901 918699.6)
1001
ន គួរ និង និង និង ន7 – 68 – 68 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 – 2 –000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000កម្ពុជា និងប្រជាជនជាតិ00-0000000ស់ ហុស្សនា ១០០ នាក់ ប្រជាជន ប្រជាជន ប្រជាជន និង ប្រជាជន និង ប្រជាជន និង ប្រជាជន និង ប្រជាជន និង ប្រជាជន និង ប្រជាជន និង ប្រជាជន និង ប្រជាជន និង ប្រជាជន និង ប្រជាជន និង ប្រជាារ សារ លោក68.3 - 98.8)តែ និង និង និង និង ន
ALL44496.299.498.8 - 99.790.4ర్తి 8

Confidentia

Gen-Probe Incorporate

.

{20}------------------------------------------------

Prevalence

The prevalence of C. trachomatis and/or N. gonorrhoeae disease in patient populations depends on risk factors such as age, gender, the presence of symptoms, the type of clinic, and the test method. A summary of the prevalence of three C. trachomatis and N. gonorrhoeae disease outcomes as determined by the APTIMA Combo 2 Assay is shown in Table 7 for the vaginal swab specimen multi-center clinical study by clinical site and overall.

Confidential

{21}------------------------------------------------

Summary of Safety and Effectiveness
K032554
SitePatient-Collected Vaginal Swab, % Prevalence (# positive/# tested)Clinician-Collected Vaginal Swab, % Prevalence (# positive/# tested)
CT+/GC+CT+/GC-CT-/GC+CT+/GC+CT+/GC-CT-/GC+
11.8(4/220)16.4(36/220)4.1(9/220)3(7/230)15.7(36/230)3.5(8/230)
29.6(19/198)18.7(37/198)6.6(13/198)9.5(19/199)18.1(36/199)7(14/199)
30.9(1/111)9(10/111)2.7(3/111)0.9(1/113)9.7(11/113)1.8(2/113)
40.4(1/266)9(24/266)1.9(5/266)0.4(1/267)11.2(30/267)2.2(6/267)
50.5(1/199)7.5(15/199)0.5(1/199)0.5(1/199)7(14/199)0.5(1/199)
62.8(8/290)10(29/290)5.5(16/290)2(6/296)12.2(36/296)5.4(16/296)
70(0/102)11.8(12/102)0(0/102)0(0/102)9.8(10/102)0(0/102)
80(0/48)8.3(4/48)2.1(1/48)0(0/51)7.8(4/51)2(1/51)
All2.4(34/1434)11.6(167/1434)3.3(48/1434)2.4(35/1457)12.1(177/1457)3.3(48/1457)

0

Table 7: Prevalence of C. trachomatis and/or N. gonorrhoeae Disease as Deternined by the APTIMA Combo 2 Assay Results by Clinical Site

Gen-Probe Incorporate

{22}------------------------------------------------

Positive and Negative Predictive Values for Hypothetical Prevalence Rates

The estimated positive and negative predictive values (PPV and NPV) for different hypothetical prevalence rates using the APTIMA Combo 2 Assay are shown in Tables 8 and 9 for C. trachomatis and N. gonorrhoeae, respectively. These calculations are based on a hypothetical prevalence and the overall sensitivity and specificity calculated from the patient infected status for two multi-center clinical studies; the endocervical swab, male urethral swab, and urine specimen clinical study and the vaginal swab specimen clinical study. The overall sensitivity and specificity for C. trachomatis was 96.1% and 97.9%, respectively (Table 8). The overall sensitivity and specificity for N. gonorrhoeve was 98.0% and 99.1%, respectively (Table 9). The actual PPV and NPV calculated using the clinical study data are shown in Tables 4 and 6 for C. trachomatis and N. gonorrhoeae, respectively

Table 8: Positive and Negative Predictive Values for Hypothetical Prevalence Rates- C. trachomatis

Prevalence Rate (%)Sensitivity (%)Specificity (%)Positive PredictiveValue (%)Negative PredictiveValue (%)
295.998.252.199.9
595.998.273.799.8
1095.998.285.599.5
1595.998.290.499.3
2095.998.293.099.0
2595.998.294.798.6
3095.998.295.898.2
Table 9: Positive and Negative Values for Hypothetical Prevalence Rates- N. gonorrhoeae
Prevalence Rate (%)Sensitivity (%)Specificity (%)Positive PredictiveValue (%)Negative PredictiveValue (%)
297.898.964.5100
597.898.982.499.9
1097.898.990.899.8
1597.898.994.099.6
2097.898.995.799.4
2597.898.996.799.3
3097.898.997.499.1

{23}------------------------------------------------

Assay Control Performance

Control RLU data were generated as part of the vaginal swab specimen clinical study. A summary of the APTIMA Positive Control, CT/Negative Control, GC and APTIMA Positive Control, GC/Negative Control, CT performance during the clinical study is presented in Table 10.

ControlStatisticsTotal RLU (x1000)
Positive, CT/Negative, GCMaximum1996
75th Percentile1279
Median1135
25th Percentile933
Minimum174
Positive, GC/Negative, CTMaximum1420
75th Percentile1255
Median1169
25th Percentile1084
Minimum249

Table 10: Distribution of Total RLU of the APTIMA Assay Controls from the Vaginal Swab Specimen Clinical Study

{24}------------------------------------------------

Conclusions from the Clinical Data

The non-clinical and clinical study results support the use of clinician-collected and patient-collected vaginal swab specimens in the GEN-PROBE APTIMA Combo 2 Assay for the detection of C. trachomatis and N. gonorrhoeae. The GEN-PROBE APTIMA Vaginal Swab Specimen Collection Kit provides the necessary materials to allow for the testing of clinician-collected or patient-collected vaginal specimens in the APTIMA Combo 2 Assay. Use of this ancillary kit broadens the application of the APTMA Combo 2 Assay as a diagnostic tool to provide information that measurably contributes to a diagnosis of C. trachomatis and N. gonorrhoeae infection.

The results of the clinical study demonstrate reasonable evidence that when the APTIMA Combo 2 Assay and the APTIMA Vaginal Swab Specimen Collection Kit are labeled as proposed, the GEN-PROBE APTIMA Combo 2 Assay is safe and effective for its stated intended use.

Application of the APTIMA Vaginal Swab Specimen Collection kit in the APTIMA Combo 2 Assay provides performance that is substantially equivalent to that of the previously cleared claims (K003395) for the detection of C. trachomatis and N. gonorrhoeae which is indicative of it's safety and effectiveness.

Contraindications and Cautions

There are no contraindications or cautions.

{25}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/25/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a stylized symbol resembling three overlapping lines, with the bottom line having a wavy appearance. The symbol is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the symbol.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 3 1 2003

Alan Maderazo, Ph.D. RAC Regulatory Affairs Specialist Gen-Probe Incorporated 10210 Genetic Center Drive San Diego, CA 92121-1589

Re: K032554

Trade/Device Name: GEN-PROBE® APTIMA® Combo 2 Assay Regulation Number: 21 CFR 866.3390 Regulation Name: Neisseria spp. Direct Serological Test Reagents Regulatory Class: Class II Product Code: LSL, MKZ Dated: December 24, 2003 Received: December 29, 2003

Dear Dr. Maderazo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{26}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{27}------------------------------------------------

Indications for Use

K032554 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name:

GEN-PROBE® APTIMA® Combo 2 Assay

Indications for Use (AC2 Assay package insert):

The APTIMA Combo 2 Assay is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection and differentiation of ribosomal ribonucleic acid (rRNA) from Chlamydia trachomatis and/or Neisseria gonorrhoeae in clinician-collected endocervical, vaginal, and male urethral swab specimens, patient-collected vaginal swab specimens* and male and female urine specimens. The assay may be used to test specimens from symptomatic and asymptomatic individuals to aid in the diagnosis of gonococcal and/or chlamydial urogenital disease.

*Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use.

Indications for Use (Ancillary Kit package insert):

The APTIMA Vaginal Swab Specimen Collection Kit is for use with the APTIMA Assays for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae. The APTIMA Vaginal Swab Specimen Collection Kit is intended to be used for clinician and patient collection of vaginal swab specimens according to the instructions provided. Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The APTIMA Vaginal Swab Specimen Collection Kit is not for home use.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Saqartzonl 12/31/03

Division Sien-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)_K0325554

§ 866.3390

Neisseria spp. direct serological test reagents.(a)
Identification. Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identifyNeisseria spp. from cultured isolates. Additionally, some of these reagents consist ofNeisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence ofNeisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusNeisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.(b)
Classification. Class II (performance standards).