K Number
K032177
Device Name
GASTROINTESTINAL VIDEOSCOPE XGIF-Q14OM, XGIF-2T140M COLONOVIDEOSCOPE XCF-Q14ML/I, XPCF-160AML/I
Date Cleared
2003-10-02

(78 days)

Product Code
Regulation Number
876.1500
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
GASTROINTESTINAL VIDEOSCOPE XGIF-Q140M, XGIE-2T140M This instrument has been designed to be used with the OLYMPUS Video System Center, Light Source, Documentation Equipment, Video Monitor, Electrosurgical unit, Endo-Therapy Accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach and duodenum). COLONOVIDEOSCOPE XCF-Q140ML/L XPCF-160AML/I This instrument has been designed to be used with the OLYMPUS Video System Center, Light Source, Documentation Eguipment, Video Monitor, Electrosurgical unit, Endo-Therapy Accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, and sigmoid colon, colon, and ileocecal valve).
Device Description
The subject devices. XGIF-Q140M. XGIF-2T140M. XCF-Q140ML/I. XPCF-160AML/I are basically identical to the predicate devices, GIF-Q140, GIF-2T140, CF-Q140L/I, PCF-140L/I except that the mechanical structure of variable stiffness* and Multi-bending function has been added. *This feature has been added to XPCF-160AML/I only. While XPCF-160AML/I has the mechanical structure of variable stiffness (Refer to Attachment 2. "Flexibility Adiustment Function") this feature is identical to those of another predicate device, the EVIS EXERA Colonovideoendoscopes (#K001241). The addition of the variable stiffness function does not require any new insertion method or technique. The Multi-bending function has two independent bending sections in the insertion portion of the scope. This new function provides an easier approach in reaching the lesion and also a better view of the lesion from the front. Some lesions in the upper digestive tract were difficult to perform a perfect biopsy and endoscopy with the conventional forward-viewing gastroscope. (Refer to Appendix-3, "Clinical Literature"). The subject devices improve the approach to the lesion in the lower digestive tract compared with the predicate devices. We have evaluated the approach performance of the XPCF-160AMI compared to the predicate device using a colon model. (Refer to Attachment 1, "Comparison between approach performance of Predicate and Subject Device using a colon model").
More Information

Not Found

No
The description focuses on mechanical improvements (variable stiffness, multi-bending function) and does not mention any software or image processing features that would suggest AI/ML.

No
The device is described as an instrument for endoscopy and endoscopic surgery, used with accessories like biopsy forceps. Its primary function is diagnostic (visualization) and to facilitate therapeutic procedures performed by other tools, rather than being therapeutic itself.

Yes

Explanation: The device is a videoscope designed for endoscopy within the digestive tract, which is a diagnostic procedure used to visualize internal organs and detect abnormalities. It also mentions accessories like biopsy forceps, indicating its use in obtaining tissue samples for further diagnostic analysis.

No

The device description explicitly details mechanical structures and physical components (variable stiffness, multi-bending function, insertion portion of the scope), indicating it is a hardware device, not software-only.

Based on the provided text, these devices are not IVDs (In Vitro Diagnostics).

Here's why:

  • Intended Use: The intended use clearly describes the devices as instruments for endoscopy and endoscopic surgery within the digestive tract. This involves direct visualization and intervention within the body.
  • Device Description: The description focuses on the mechanical aspects of the videoscopes, such as variable stiffness and multi-bending functions, designed to improve navigation and visualization inside the body.
  • Lack of IVD Characteristics: There is no mention of the devices being used to examine specimens outside the body (in vitro), such as blood, tissue, or other bodily fluids, which is the defining characteristic of an IVD.

These devices are clearly intended for in-vivo (within the living body) procedures.

N/A

Intended Use / Indications for Use

GASTROINTESTINAL VIDEOSCOPE XGIF-Q140M, XGIE-2T140M

This instrument has been designed to be used with the OLYMPUS Video System Center, Light Source, Documentation Equipment, Video Monitor, Electrosurgical unit, Endo-Therapy Accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach and duodenum).

COLONOVIDEOSCOPE XCF-Q140ML/L XPCF-160AML/I

This instrument has been designed to be used with the OLYMPUS Video System Center, Light Source, Documentation Eguipment, Video Monitor, Electrosurgical unit, Endo-Therapy Accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, and sigmoid colon, colon, and ileocecal valve).

Product codes (comma separated list FDA assigned to the subject device)

78-FDS, 78FDF

Device Description

The subject devices. XGIF-Q140M. XGIF-2T140M. XCF-Q140ML/I. XPCF-160AML/I are basically identical to the predicate devices, GIF-Q140, GIF-2T140, CF-Q140L/I, PCF-140L/I except that the mechanical structure of variable stiffness* and Multi-bending function has been added.

*This feature has been added to XPCF-160AML/I only.

While XPCF-160AML/I has the mechanical structure of variable stiffness (Refer to Attachment 2. "Flexibility Adiustment Function") this feature is identical to those of another predicate device, the EVIS EXERA Colonovideoendoscopes (#K001241). The addition of the variable stiffness function does not require any new insertion method or technique.

The Multi-bending function has two independent bending sections in the insertion portion of the scope. This new function provides an easier approach in reaching the lesion and also a better view of the lesion from the front. Some lesions in the upper digestive tract were difficult to perform a perfect biopsy and endoscopy with the conventional forward-viewing gastroscope. (Refer to Appendix-3, "Clinical Literature"). The subject devices improve the approach to the lesion in the lower digestive tract compared with the predicate devices. We have evaluated the approach performance of the XPCF-160AMI compared to the predicate device using a colon model. (Refer to Attachment 1, "Comparison between approach performance of Predicate and Subject Device using a colon model").

In conclusion, the subject device is substantially equivalent to the predicate devices. A comparison table of the subject device and predicate devices is found in Attachment 1.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Upper digestive tract (including the esophagus, stomach and duodenum).
Lower digestive tract (including the anus, rectum, and sigmoid colon, colon, and ileocecal valve).

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

#K954451, #K001241

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

0

K 032177

Pg 1 of 3

SMDA 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR, Part 807, Subpart E, Section 807.92.

A. GENERAL INFORMATION

1. Applicant:OLYMPUS OPT-ELECTRONICS CO., LTD. (New Company Name: AIZU OLYMPUS CO., LTD)
Address:500 Aza Muranishi Ooaza Niidera, Monden-machi Aizuwakamatsu-shi Fukushima, Japan, 965-8520
Establishment Registration No.:9610595
2. Submission Correspondent:Tina Steffanie-Oak
Title:Senior Regulatory Affairs Analyst OLYMPUS AMERICA INC.
Address:Two Corporate Center Drive, Melville, NY 11747-3157
Telephone:631-844-5477
Facsimile:631-844-5416
E-mail address:Tina.Steffanie-Oak@olympus.com
Establishment Registration No.:2429304

3. Initial Impoter: Olympus America Inc. Two Corporate Center Drive, Melville, NY 11747-3157 Address: Establishment Registration No .: 2429304

B. DEVICE IDENTIFICATION

1. Common/Usual Name

GASTROINTESTINALVIDEOSCOPE / COLONOVIDEOSCOPE

2. Device Name

GASTROINTESTINALVIDEOSCOPE XGIF-Q140M, XGIF-2T140M COLONOVIDEOSCOPE XCF-Q140ML/I, XPCF-160AML/I

3. Classification Name

| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Company of the program of the comments of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the contribution of

A Clair | | . |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| CARLING COMMENT CAR
A 2 4 4 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
programs and only of the consisted on the program and any and any and
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | · GAMUSCANDO
AND STORESCOLIBE

Complete of the can and management comments on any a more and considered to the control of the first and the may be | ---------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | i
1
ALL BRANDER AT MALLER A LE BREAK E & REPORTER FREE S
|

1

Image /page/1/Picture/0 description: The image shows handwritten text on a white background. The text at the top reads "K032177", which appears to be a reference or identification number. Below this, the text "Pg 2 of 3" indicates that this is page 2 of a 3-page document. The handwriting is cursive and slightly stylized.

C. IDENTIFICATION OF LEGALLY MARKETED DEVICES WHICH WE CLAIM SUBSTANTIAL EQUIVALENCE

The following listed devices are considered as predicate devices in consideration of their characteristics, and the following table shows their regulatory histories.

| Model | 510(k)# | Manufacturer | Class | Product
Code |
|--------------------------------------|----------|---------------------------|-------|-----------------|
| GIF-Q140
EVIS-140 Series Scopes | #K954451 | Olympus Optical Co., Ltd. | II | 78-FDS |
| GIF-2T140
EVIS-140 Series Scopes | #K954451 | Olympus Optical Co., Ltd. | II | 78-FDS |
| CF-Q140L/I
EVIS-140 Series Scopes | #K954451 | Olympus Optical Co., Ltd. | II | 78-FDS |
| PCF-140L/I
EVIS-140 Series Scopes | #K954451 | Olympus Optical Co., Ltd. | II | 78-FDS |
| EVIS EXERA
Colonovideoendoscopes | #K001241 | Olympus Optical Co., Ltd. | II | 78FDF |

D. DEVICE DESCRIPTION

1. Summary

The subject devices. XGIF-Q140M. XGIF-2T140M. XCF-Q140ML/I. XPCF-160AML/I are basically identical to the predicate devices, GIF-Q140, GIF-2T140, CF-Q140L/I, PCF-140L/I except that the mechanical structure of variable stiffness* and Multi-bending function has been added.

*This feature has been added to XPCF-160AML/I only.

While XPCF-160AML/I has the mechanical structure of variable stiffness (Refer to Attachment 2. "Flexibility Adiustment Function") this feature is identical to those of another predicate device, the EVIS EXERA Colonovideoendoscopes (#K001241). The addition of the variable stiffness function does not require any new insertion method or technique.

The Multi-bending function has two independent bending sections in the insertion portion of the scope. This new function provides an easier approach in reaching the lesion and also a better view of the lesion from the front. Some lesions in the upper digestive tract were difficult to perform a perfect biopsy and endoscopy with the conventional forward-viewing gastroscope. (Refer to Appendix-3, "Clinical Literature"). The subject devices improve the approach to the lesion in the lower digestive tract compared with the predicate devices. We have evaluated the approach performance of the XPCF-160AMI compared to the predicate device using a colon model. (Refer to Attachment 1, "Comparison between approach performance of Predicate and Subject Device using a colon model").

In conclusion, the subject device is substantially equivalent to the predicate devices. A comparison table of the subject device and predicate devices is found in Attachment 1.

2

K032177

Pg 3 of =

2. Design

XGIF-Q140M, XGIF-2T140M, XCF-Q140MLII, XPCF-160AMLII has been designed, manufactured and tested in compliance with voluntary safety standards. It meets the requirement of IEC60601-1, IEC60601-1-1and IEC60601-2-18.

3. Materials

All the patient contacting materials used in this endoscope and ancillary equipment are identical materials that have been cleared in the past 510(k) submissions. And all materials have been confirmed with ISO 10993-1.

4. Intended Use of the device

GASTROINTESTINALVIDEOSCOPE XGIF-Q140M, XGIF-2T140M

This instrument has been designed to be used with the OLYMPUS Video System Center, Light Source, Documentation Equipment, Video Monitor, Electrosurgical unit, Endo-Therapy Accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach and duodenum).

COLONOVIDEOSCOPE XCF-Q140ML/I, XPCF-160AML/I

This instrument has been designed to be used with the OLYMPUS Video System Center, Light Source, Documentation Equipment, Video Monitor, Electrosurgical unit, Endo-Therapy Accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, and sigmoid colon, colon, and ileocecal valve).

5. Summary including conclusion drawn form Non-clinical Tests

When compared to the preamendment/predicate devices, XGIF-Q140M. XGIF-2T140M. XCF-Q140ML/I, XPCF-160AML/I does not incorporate any significant changes in the intended use, method of operation, material, or design that could affect the safety and effectiveness. Therefore clinical data is not necessary for its evaluation of safety and efficacy.

3

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a stylized eagle or bird-like figure with three curved lines representing its wings or body.

OCT - 2 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OL YMPUS OPT-ELECTRONICS CO., LTD./AIZU OLYMPUS CO., LTD. c/o Ms. Tina Steffanie-Oak Senior Regulatory Affairs Analyst Olympus America Inc. Two Corporate Center Drive MELVILLE NY 11747-3157

Re: K032177

Trade/Device Name: Gastrointestinal Videoscope, XGIF-Q140M, XGIF-2T140M and Colonovideoscope, XCF-Q140ML/I, XPCF-160AML/I Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FDS and FDF Dated: July 11, 2003 Received: July 16, 2003

Dear Ms. Steffanie-Oak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

4

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

K 032/77

Pg1

510(k) Number(if known): GASTROINTESTINALVIDEOSCOPE XGIF-Q140M, XGIE-2T140M Device Name: COLONOVIDEOSCORE XCF-Q140ML/L XPCF-160AML/I

Indications for Use:

GASTROINTESTINAL VIDEOSCOPE XGIF-Q140M, XGIE-2T140M

This instrument has been designed to be used with the OLYMPUS Video System Center, Light Source, Documentation Equipment, Video Monitor, Electrosurgical unit, Endo-Therapy Accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach and duodenum).

COLONOVIDEOSCOPE XCF-Q140ML/L XPCF-160AML/I

This instrument has been designed to be used with the OLYMPUS Video System Center, Light Source, Documentation Eguipment, Video Monitor, Electrosurgical unit, Endo-Therapy Accessories such as a Biopsy Forceps and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, and sigmoid colon, colon, and ileocecal valve).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter Use

(Per 21 CFR 801.109)

(Optional Format 1-2-96)

Gabriel A. Segarra

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number.