(93 days)
Not Found
Not Found
No
The summary describes a mechanical spinal fixation system and does not mention any AI or ML components.
No
This device is a pedicle screw fixation system used for immobilization and stabilization of spinal segments, primarily acting as an adjunct to fusion rather than actively treating a disease or condition therapeutically. It is intended to be removed after solid fusion.
No
The device description clearly states it is a "pedicle screw system" designed for "fixation" and "stabilization of spinal segments as an adjunct to fusion." Its function is mechanical support during the healing process following spinal surgery, not to diagnose medical conditions.
No
The device description explicitly lists hardware components such as bone screws, lock screws, connectors, swivels, and rods, which are physical implants used in spinal surgery.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health.
- Device Description: The description clearly states the device consists of "bone screws, lock screws, connectors, swivels, and rods" used for "pedicle screw fixation" and "immobilization and stabilization of spinal segments." These are physical implants used within the body for structural support.
- Intended Use: The intended use describes the conditions the device is used to treat (degenerative disc disease, spondylolisthesis, fracture, etc.) by providing mechanical support to the spine. This is a therapeutic use, not a diagnostic one.
The device is a surgical implant, not a diagnostic test performed on samples outside the body.
N/A
Intended Use / Indications for Use
When intended for pedicle screw fixation from L1-S1, the indications include immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, and failed previous fusion.
As a pedicle screw system placed between L3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autologous bone graft, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is established.
After solid fusion occurs, these devices serve no functional purpose and should be removed. In most cases, removal is indicated because the implants are not intended to transfer or support forces developed during normal activities. Any decision to remove the device must be made by the physician and the patient, taking into consideration the patient's general medical condition and the potential risk to the patient of a second surgical procedure.
Product codes
KWQ, MNI, MNH, KWP, NKB
Device Description
InSight constructs consist of bone screws, lock screws, connectors, swivels, and rods, which when assembled create a polyaxial range of motion. InSight only allows the placement of 5.5 mm titanium rods. InSight screws may be implanted in an open manner; or via cannulation, which allows a subset of procedures to be completed through a mini open procedure.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
lumbar and sacral spine, L1-S1, L3 and S1, posterior lumbosacral spine
Indicated Patient Age Range
skeletally mature patients
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Mechanical testing data, including data collected in accordance with ASTM 1717 and ASTM 1798, was collected to verify that the design changes met established design requirements.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Not Found
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 888.3060 Spinal intervertebral body fixation orthosis.
(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.
0
K 631855
Spinal Concepts, Inc. InSight™ Pedicle Screw System
142
510(k) Summary of Safety and Effectiveness
| SUBMITTED BY | Spinal Concepts, Inc.
5301 Riata Park Court, Bldg. F
Austin, TX 78727 | | | | |
|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-----------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| ESTABLISHMENT
REGISTRATION NUMBER | 1649384 | | | | |
| CONTACT PERSON | PrimaryAlternateLisa Peterson
Regulatory Affairs Specialist
Phone: 512-533-1080
Fax: 512-249-6734David Hooper, Ph.D.
Director, Clinical and
Regulatory Affairs
Phone: 512-533-1038
Fax: 512-249-6734 | Primary | Alternate | Lisa Peterson
Regulatory Affairs Specialist
Phone: 512-533-1080
Fax: 512-249-6734 | David Hooper, Ph.D.
Director, Clinical and
Regulatory Affairs
Phone: 512-533-1038
Fax: 512-249-6734 |
| Primary | Alternate | | | | |
| Lisa Peterson
Regulatory Affairs Specialist
Phone: 512-533-1080
Fax: 512-249-6734 | David Hooper, Ph.D.
Director, Clinical and
Regulatory Affairs
Phone: 512-533-1038
Fax: 512-249-6734 | | | | |
| DATE PREPARED | June 11, 2003 | | | | |
| CLASSIFICATION NAME | KWQ 888.3060- Spinal Intervertebral Body Fixation
Orthosis
KWP 888.3050- Spinal Interlaminal Fixation Orthosis
MNI 888.3070- Pedicle Screw Spinal System
MNH 888.3070 - Spondylolisthesis Spinal Fixation System
NKB 888.3070 - Pedicle Screw Fixation System,
Degenerative Disc Disease | | | | |
| COMMON NAME | Spinal Fixation System | | | | |
| PROPRIETARY NAME | Spinal Concepts Inc. InSight ™ Pedicle Screw System | | | | |
| SUBSTANTIAL
EQUIVALENCE | The Spinal Concepts Inc. InSight Pedicle Screw System
was determined to be substantially equivalent to several
commercially available systems. | | | | |
DEVICE DESCRIPTION
InSight constructs consist of bone screws, lock screws, connectors, swivels, and rods, which when assembled create a polyaxial range of motion. InSight only allows the placement of 5.5 mm titanium rods. InSight screws may be implanted in an open manner; or via cannulation, which allows a subset of procedures to be completed through a mini open procedure.
.
1
INDICATIONS:
When intended for pedicle screw fixation from L1-S1, the indications include immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, and failed previous fusion.
As a pedicle screw system placed between L3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autologous bone graft, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is established.
After solid fusion occurs, these devices serve no functional purpose and should be removed. In most cases, removal is indicated because the implants are not intended to transfer or support forces developed during normal activities. Any decision to remove the device must be made by the physician and the patient, taking into consideration the patient's general medical condition and the potential risk to the patient of a second surgical procedure.
MECHANICAL TEST DATA
Mechanical testing data, including data collected in accordance with ASTM 1717 and ASTM 1798, was collected to verify that the design changes met established design requirements.
2
Image /page/2/Picture/1 description: The image shows a logo with a stylized bird in flight. The bird is composed of three curved lines, suggesting movement and freedom. The logo is encircled by text, though the specific words are illegible due to the image quality. The overall design is simple and iconic, likely representing an organization or company.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 7 2003
Ms. Lisa Peterson Regulatory Affairs Specialist Spinal Concepts, Inc. 5301 Riata Park Court, Bldg. F Austin, TX 78727
Re: K031855 Trade/Device Name: InSight™ Pedicle Screw System Regulation Numbers: 21 CFR 888.3060, 888.3070, 888.3050 Regulation Names: Spinal intervertebral body fixation orthosis. Pedicle screw spinal system, Spinal interlaminal fixation orthosis Regulatory Class: III Product Codes: KWQ, MNI, MNH, KWP, NKB Dated: June 11, 2003
Dear Ms. Peterson:
Received: June 24, 2003
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
3
Page 2 - Ms. Lisa Peterson
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mach N Melleman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Ku 31852
INDIC ATIONS FOR USE STATEMENT
Page I of I
510(k) Number (if known):
Device Name:
Spinal Concepts, Inc. InSightTM Pedicle Screw System
Indications for Use:
When intended for pedicle screw fixation from L1-S1, the indications include immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the lumbar and sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, and failed previous fusion.
As a pedicle screw system placed between L3 and S1, the indications include Grade 3 or Grade 4 spondylolisthesis, when utilizing autologous bone graft, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is established.
After solid fusion occurs, these devices serve no functional purpose and should be removed. In most cases, removal is indicated because the implants are not intended to transfer or support forces developed during normal activities. Any decision to remove the device must be made by the physician and the patient, taking into consideration the patient's general medical condition and the potential risk to the patient of a second surgical procedure.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: (Per 21 CFR 801.109)
OR Over-The-Counter: __
(Optional Format 1-2-96)
for Mark N. Millem
(Division Sign-Off)
Division of General, Restorative and Neurological Devices
510(k) Number K031853