K Number
K031705
Device Name
SKYLIGHT IMAGING SYSTEM
Manufacturer
Date Cleared
2003-06-17

(15 days)

Product Code
Regulation Number
892.1200
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The SKYLight Imaging System is intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.
Device Description
The modified SKYLight Imaging System offers all the features of the existing SKYLight Imaging System (K013496) while expanding the Concurrent Imaging feature for Multi-Energy/Multi-Isotope Imaging and the addition of DICOM to the JETStream™ acquisition system. The basic underlying algorithms for concurrent imaging remain the same. Software modifications to the Multi-Energy/Multi-Isotope application extend the current functionality to include acquisition of more image events simultaneously. Acquisition data can be exported to the Pegasys or any other processing station via DICOM by translating image data from the native XML format to a DICOM 3.0 complaint format. The addition of DICOM to the acquisition software will enhance workflow by providing DICOM Worklist and the capability to import patient information for scheduling purposes. The SKYLight is designed to provide extended imaging functionality relative to a ring style gantry. The SKYLight was designed for single or dual detector nuclear imaging accommodating a broad range of emission computed tomography (ECT) studies. The device includes a gantry frame, two detector arms (with detectors), a collimator storage structure with an acquisition computer unit, a patient imaging table, and a remote hand controller. The modified SKYLight is capable of accommodating two separate patient tables. The gantry is "open" so that a high degree of imaging flexibility is available to image patients sitting, standing or lying down, with or without the included patient imaging table. The patient imaging tables are mechanized to allow for patient loading access and then raised to an imaging height. The tables do not move during imaging since the gantry is flexible enough to perform all required motions for non-circular orbits. The imaging pallet includes removable arm, leg, breast, and headrest supports for patient positioning during studies that require support. SKYLight® is an 'open frame' imaging system, the 'open frame' consisting of an overhead rectangular structure that is supported by 4 (or more) columns that are bolted to the imaging room floor. A detector arm support structure is mounted onto the open frame, which allows for the detector arm support structure (and both detector arms) to move in the longitudinal motion (X-axis). Arms that can telescope up and down (Z-axis) support the detectors. These detector arms are supported by a structure that allows for each detector arm to move toward or away from each other (Y-axis). Each detector has the ability to rotate and acquire images independently The combination of these four motions (Z, Y, X, and detector rotate), permit the detectors to perform all standard non-circular and circular ECT studies in both the 90-degree and 180-degree relative position. The detectors use two standard collimators, or MCD shields, which may be exchanged using an automatic exchange unit. The modified JETStream SKYLight acquisition computer uses the same acquisition software (subsystem) as existing SKYLight to acquire ECT imaging studies and interfaces with a Pegasys computer system (workstation). The acquisition CPU is used to install the camera software and to archive and restore acquisition parameters and correction tables. The acquisition software is used to complete pre-programmed gantry motions for setup of imaging studies and exchange of collimators; to perform quality assurance tests on the SK YLight system; and to program image parameters and gantry for clinical studies. When using either a single detector or dual detectors placed in a relative 90-degree or relative 180-degree positions (as study appropriate), SKYLight can be used to perform static, dynamic, gated, total body, circular-orbit and non-circular orbit SPECT studies, coincidence studies, gated SPECT (circular and non-circular) studies, computerprogrammed protocol strings, and reference scans (dual detectors only). SPECT and total body acquisitions are routinely acquired with two detectors. There are some planar procedures such as bone statics and lung scan also use two detectors. There are many additional nuclear medicine procedures that only use one detector at a time. These single detector procedures are typically renal, gastric emptying, hepatobiliary, flow studies, GI bleed, thyroid, and delayed static views. This acquisition sub-system interfaces with a Pegasys computer system for entry into the database. The Pegasys workstation also includes software used for image processing, database utilities, and archiving utilities.
More Information

No
The description focuses on hardware modifications, expanded concurrent imaging capabilities based on existing algorithms, and DICOM integration for workflow enhancement and data export. There is no mention of AI or ML being used for image processing, analysis, or any other function.

No
The device is described as an "Imaging System" intended "to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel." It is designed for diagnostic imaging purposes and does not describe any therapeutic function.

Yes

The device "is intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel," which indicates its purpose is to aid in diagnosis by providing visual information about internal bodily structures and functions.

No

The device description explicitly lists multiple hardware components including a gantry frame, detector arms, collimator storage structure, patient imaging table, and a remote hand controller. While software is mentioned for acquisition and processing, it is part of a larger hardware system.

Based on the provided information, the SKYLight Imaging System is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • Intended Use: The intended use is to "produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel." This describes an in vivo imaging system, where the radioisotopes are administered to the patient and the device detects their distribution within the living body.
  • Device Description: The description details a physical imaging system with a gantry, detectors, patient table, and associated software for acquiring and processing images of the human body. This is consistent with an in vivo imaging device used in nuclear medicine.
  • Lack of In Vitro Activities: There is no mention of the device being used to examine specimens (like blood, urine, tissue) outside of the body. IVD devices are specifically designed for testing these types of samples.

Therefore, the SKYLight Imaging System is an in vivo imaging device, not an IVD.

N/A

Intended Use / Indications for Use

The SKYLight Imaging System is intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.

Product codes

90 KPS

Device Description

The modified SKYLight Imaging System offers all the features of the existing SKYLight Imaging System (K013496) while expanding the Concurrent Imaging feature for Multi-Energy/Multi-Isotope Imaging and the addition of DICOM to the JETStream™ acquisition system. The basic underlying algorithms for concurrent imaging remain the same. Software modifications to the Multi-Energy/Multi-Isotope application extend the current functionality to include acquisition of more image events simultaneously. Acquisition data can be exported to the Pegasys or any other processing station via DICOM by translating image data from the native XML format to a DICOM 3.0 complaint format. The addition of DICOM to the acquisition software will enhance workflow by providing DICOM Worklist and the capability to import patient information for scheduling purposes.

The SKYLight is designed to provide extended imaging functionality relative to a ring style gantry. The SKYLight was designed for single or dual detector nuclear imaging accommodating a broad range of emission computed tomography (ECT) studies. The device includes a gantry frame, two detector arms (with detectors), a collimator storage structure with an acquisition computer unit, a patient imaging table, and a remote hand controller. The modified SKYLight is capable of accommodating two separate patient tables. The gantry is "open" so that a high degree of imaging flexibility is available to image patients sitting, standing or lying down, with or without the included patient imaging table. The patient imaging tables are mechanized to allow for patient loading access and then raised to an imaging height. The tables do not move during imaging since the gantry is flexible enough to perform all required motions for non-circular orbits. The imaging pallet includes removable arm, leg, breast, and headrest supports for patient positioning during studies that require support.

SKYLight® is an 'open frame' imaging system, the 'open frame' consisting of an overhead rectangular structure that is supported by 4 (or more) columns that are bolted to the imaging room floor. A detector arm support structure is mounted onto the open frame, which allows for the detector arm support structure (and both detector arms) to move in the longitudinal motion (X-axis). Arms that can telescope up and down (Z-axis) support the detectors. These detector arms are supported by a structure that allows for each detector arm to move toward or away from each other (Y-axis). Each detector has the ability to rotate and acquire images independently The combination of these four motions (Z, Y, X, and detector rotate), permit the detectors to perform all standard non-circular and circular ECT studies in both the 90-degree and 180-degree relative position. The detectors use two standard collimators, or MCD shields, which may be exchanged using an automatic exchange unit.

The modified JETStream SKYLight acquisition computer uses the same acquisition software (subsystem) as existing SKYLight to acquire ECT imaging studies and interfaces with a Pegasys computer system (workstation). The acquisition CPU is used to install the camera software and to archive and restore acquisition parameters and correction tables. The acquisition software is used to complete pre-programmed gantry motions for setup of imaging studies and exchange of collimators; to perform quality assurance tests on the SK YLight system; and to program image parameters and gantry for clinical studies.

When using either a single detector or dual detectors placed in a relative 90-degree or relative 180-degree positions (as study appropriate), SKYLight can be used to perform static, dynamic, gated, total body, circular-orbit and non-circular orbit SPECT studies, coincidence studies, gated SPECT (circular and non-circular) studies, computerprogrammed protocol strings, and reference scans (dual detectors only). SPECT and total body acquisitions are routinely acquired with two detectors. There are some planar procedures such as bone statics and lung scan also use two detectors. There are many additional nuclear medicine procedures that only use one detector at a time. These single detector procedures are typically renal, gastric emptying, hepatobiliary, flow studies, GI bleed, thyroid, and delayed static views. This acquisition sub-system interfaces with a Pegasys computer system for entry into the database. The Pegasys workstation also includes software used for image processing, database utilities, and archiving utilities.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

single photon and positron emitting radioisotopes

Anatomical Site

human body

Indicated Patient Age Range

Not Found

Intended User / Care Setting

medical personnel

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K013496

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.

0

KD31705

510(k) SUMMARY SAFETY AND EFFECTIVENESS

| A. | Submitted By:
ADAC Laboratories
540 Alder Dr.
Milpitas, CA 95035 | Contact: Coleen Coleman
Tel: (408) 468-3051
Fax: (408) 468-3050 |
|----|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| B. | Device Trade Name:
Common Name:
Classification Name:
Device Class:
Product Code: | SKYLight Imaging System
Gamma Camera Systems
Emission Computed Tomography System
21 CFR 892.1200, Class II
90 KPS |
| C. | Date prepared: | May 8, 2003 |
| D. | Predicate Device: | SKYLight® Imaging System (K013496) |
| E. | Intended Use: | |

The SKYLight Imaging System is intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.

F. Device Description:

The modified SKYLight Imaging System offers all the features of the existing SKYLight Imaging System (K013496) while expanding the Concurrent Imaging feature for Multi-Energy/Multi-Isotope Imaging and the addition of DICOM to the JETStream™ acquisition system. The basic underlying algorithms for concurrent imaging remain the same. Software modifications to the Multi-Energy/Multi-Isotope application extend the current functionality to include acquisition of more image events simultaneously. Acquisition data can be exported to the Pegasys or any other processing station via DICOM by translating image data from the native XML format to a DICOM 3.0 complaint format. The addition of DICOM to the acquisition software will enhance workflow by providing DICOM Worklist and the capability to import patient information for scheduling purposes.

The SKYLight is designed to provide extended imaging functionality relative to a ring style gantry. The SKYLight was designed for single or dual detector nuclear imaging accommodating a broad range of emission computed tomography (ECT) studies. The device includes a gantry frame, two detector arms (with detectors), a collimator storage structure with an acquisition computer unit, a patient imaging table, and a remote hand controller. The modified SKYLight is capable of accommodating two separate patient tables. The gantry is "open" so that a high degree of imaging flexibility is available to image patients sitting, standing or lying down, with or without the included patient imaging table. The patient imaging tables are mechanized to allow for patient loading access and then raised to an imaging height. The tables do not move during imaging since the gantry is flexible enough to perform all required motions for non-circular orbits. The imaging pallet includes removable arm, leg, breast, and headrest supports for patient positioning during studies that require support.

1

SKYLight® is an 'open frame' imaging system, the 'open frame' consisting of an overhead rectangular structure that is supported by 4 (or more) columns that are bolted to the imaging room floor. A detector arm support structure is mounted onto the open frame, which allows for the detector arm support structure (and both detector arms) to move in the longitudinal motion (X-axis). Arms that can telescope up and down (Z-axis) support the detectors. These detector arms are supported by a structure that allows for each detector arm to move toward or away from each other (Y-axis). Each detector has the ability to rotate and acquire images independently

The combination of these four motions (Z, Y, X, and detector rotate), permit the detectors to perform all standard non-circular and circular ECT studies in both the 90-degree and 180-degree relative position. The detectors use two standard collimators, or MCD shields, which may be exchanged using an automatic exchange unit.

The modified JETStream SKYLight acquisition computer uses the same acquisition software (subsystem) as existing SKYLight to acquire ECT imaging studies and interfaces with a Pegasys computer system (workstation). The acquisition CPU is used to install the camera software and to archive and restore acquisition parameters and correction tables. The acquisition software is used to complete pre-programmed gantry motions for setup of imaging studies and exchange of collimators; to perform quality assurance tests on the SK YLight system; and to program image parameters and gantry for clinical studies.

When using either a single detector or dual detectors placed in a relative 90-degree or relative 180-degree positions (as study appropriate), SKYLight can be used to perform static, dynamic, gated, total body, circular-orbit and non-circular orbit SPECT studies, coincidence studies, gated SPECT (circular and non-circular) studies, computerprogrammed protocol strings, and reference scans (dual detectors only). SPECT and total body acquisitions are routinely acquired with two detectors. There are some planar procedures such as bone statics and lung scan also use two detectors. There are many additional nuclear medicine procedures that only use one detector at a time. These single detector procedures are typically renal, gastric emptying, hepatobiliary, flow studies, GI bleed, thyroid, and delayed static views. This acquisition sub-system interfaces with a Pegasys computer system for entry into the database. The Pegasys workstation also includes software used for image processing, database utilities, and archiving utilities.

G. Technological Comparison:

The modified SKYLight Imaging System and existing SKYLight Imaging System have identical intended use and indications for use. The modified SKYLight and the existing SKYLight are technologically identical. They share the identical mechanical and electrical components. No changes or modifications have been made to the mechanical or electrical components. All the features provided on the existing SKYLight are provided on the modified SKYLight. The only change has been to the JETStream acquisition software, which provides the modified SKYLight Imaging System with the expanded feature of Concurrent Multi-Energy/Multi-Isotope Imaging and DICOM Export/Worklist.

2

H. Conclusion

The modified SK YLight® Imaging System is substantially equivalent to the predicate device SKYLight based upon identical indications for use, technological comparison and overall system performance.

3

Image /page/3/Picture/1 description: The image shows a black and white logo. The logo features a stylized bird with three curved lines representing its body and wings. The bird is facing right. The logo is surrounded by a circular border with text, though the text is not legible due to the image quality.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 1 7 2003

ADAC Laboratories % Mr. Morten Christensen Principal Reviewer/Office Coordinator 510(k) Review Program Underwriters Laboratories, Inc. 1655 Scott Boulevard SANTA CLARA CA 95050-4169

Re: K031705

Trade/Device Name: SKYLight™ Imaging System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: 90 KPS Dated: May 29, 2003 Received: June 2, 2003

Dear Mr. Christensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

4

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

d.

INDICATIONS FOR USE STATEMENT

510 (k) NUMBER (IF KNOWN): KD31705

DEVICE NAME:

SKYlight™ Imaging System

SPONSOR NAME:

ADAC Laboratories

INDICATIONS FOR USE:

The Skylight Imaging System is intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109) OR

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number

Over-The-Counter-Use (Optional Format 1-2-96)

0001