K Number
K031481
Date Cleared
2003-05-29

(17 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments.

EASI 12-lead ECG is only for use on adult and pediatric patients.

ST Segment monitoring is restricted to adult patients only.

The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.

Device Description

The Philips MP60, MP70, and MP90 IntelliVue Patient Monitor. The modification is updated software of the Philips Medical System MP60, MP70 and MP90 IntelliVue patient monitor devices.

AI/ML Overview

The provided text describes the Philips MP60, MP70, and MP90 IntelliVue Patient Monitors and their clearance as substantially equivalent to previously marketed devices. However, the document does not contain a detailed study report with specific acceptance criteria and device performance metrics in the format requested.

Here's an analysis based on the available information:

1. A table of acceptance criteria and the reported device performance

The document explicitly states: "Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence." It does not provide specific quantitative acceptance criteria (e.g., accuracy, precision) or numerical reported device performance values. It broadly states that the new device "meets all reliability requirements and performance claims."

Acceptance CriteriaReported Device Performance
Not explicitly stated in quantitative terms. Based on specifications cleared for the predicate device."Meets all reliability requirements and performance claims."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample size used for the test set. It does not mention the country of origin of the data or whether the data was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document. As this is a patient monitoring device, the ground truth would likely be established through direct physiological measurements or clinical observation, rather than expert interpretation of images. However, the exact methodology for ground truth establishment for testing is not detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe any adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or described. This type of study is typically relevant for interpretative diagnostic devices (like imaging AI), not for physiological monitors where the device itself measures and presents data. There is no mention of "human readers" or "AI assistance" in the context of improving human performance.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The entire study described is essentially a standalone performance assessment of the device. The "updated software" is an integral part of the patient monitor, and its performance is evaluated as a standalone system against the predicate device's specifications. The evaluation is focused on the device's ability to accurately monitor physiological parameters.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The document does not explicitly state the type of ground truth used. For physiological monitors, ground truth is typically established by using highly accurate reference measurement systems (e.g., a calibrated pressure transducer for blood pressure, a reference ECG machine for arrhythmia detection) or direct observation of physiological events.

8. The sample size for the training set

The document does not mention a training set or its sample size. This is a common characteristic of device clearances that involve software updates for existing devices, where the focus is on verification and validation against established specifications, rather than development of new algorithms requiring extensive training data.

9. How the ground truth for the training set was established

Since no training set is mentioned, there is no information on how ground truth for a training set was established.

Summary of the Study:

The study described is a verification and validation effort to demonstrate that the updated software for the Philips MP60, MP70, and MP90 IntelliVue Patient Monitors maintains the same performance, functionality, and reliability as the legally marketed predicate devices. The methodology involved "system level tests, performance tests, and safety testing from hazard analysis." The "Pass/Fail criteria were based on the specifications cleared for the predicate device." The conclusion was that "test results showed substantial equivalence" and the device "meets all reliability requirements and performance claims." This type of submission falls under the 510(k) pathway, which primarily focuses on demonstrating substantial equivalence to a predicate device, rather than proving novel clinical effectiveness through extensive clinical trials with detailed statistical reporting of performance metrics.

{0}------------------------------------------------

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.

    1. The submitter of this premarket notification is: Hauke Schik Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-2031 Fax: ++49 7031 463-2442 e-mail: hauke.schik@philips.com
      This summary was prepared on May 05, 2003.
    1. The names of the devices are the Philips MP60, MP70, and MP90 IntelliVue Patient Monitor. Classification names are as follows:
Device PanelClassificationProCodeDescription
CirculatorySystem Devices(12625)$870.1025, IIIDSIDetector and alarm, arrhythmia
$870.1025, IIIMLDMonitor, ST Segment with Alarm
$870.1025, IIIMHXMonotor, Physiological, Patient(with arrhythmia detection oralarms)
$870.1100, IIDSJAlarm, Blood Pressure
$870.1110, IIDSKComputer, Blood Pressure
$870.1130, IIDXNSystem, Measurement, Blood-Pressure, Non-Invasive
$870.1435, IIDXGComputer, Diagnostic, Pre-Programmed, Single-Function
$870.1915, IIKRBProbe, Thermodilution
$870.2060, IIDRQAmplifier and SignalConditioner, Transducer Signal
$870.2300, IIDRTMonitor, Cardiac (incl.Cardiotachometer & Rate Alarm)
$870.2340, IIDPSElectrocardiograph
$870.2340, IIMLCMonitor, ST Segment
$870.2350, IIDRWElectrocardiograph, LeadSwitching Adapter
$870.2370, IIKRCTester, Electrode, Surface,Electrocardiograph
$870.2450, IIDXJDisplay, Cathode-Ray Tube,Medical
$870.2600, IDRJSystem, Signal Isolation
$870.2700, IIDQAOximeter
$870.2770, IIDSBPlethysmograph, Impedance
$870.2800, IIDSHRecorder, Magnetic tape,Medical
$870.2810, IDSFRecorder, Paper Chart
$870.2850, IIDRSExtravascular Blood PressureTransducer
$870.2900, IDSACable, Transducer andElectrode, incl. PatientConnector

{1}------------------------------------------------

MSXSystem, Network andCommunication, PhysiologicalMonitors
Anesthesiologyand RespiratoryTherapy (12624)$868.1400, IICCKAnalyzer, Gas, Carbon Dioxide,Gaseous-Phase
$868.1500, IICBQAnalyzer, Gas, Enflurane,Gaseous-Phase (AnestheticConcentration)
$868.1500, IINHOAnalyzer, Gas, Desflurane,Gaseous-Phase (AnestheticConcentration)-
$868.1500, IINHPAnalyzer, Gas, Sevoflurane,Gaseous-Phase (AnestheticConcentration)
$868.1500, IINHQAnalyzer, Gas, Isoflurane,Gaseous-Phase (AnestheticConcentration)
$868.1620, IICBSAnalyzer, Gas, Halothane,Gaseous-Phase (AnestheticConcentration)
$868.1700, IICBRAnalyzer, Gas, Nitrous Oxide,Gaseous-Phase (AnestheticConcentration)
$868.1720, IICCLAnalyzer, Gas, Oxygen, Gaseous-Phase
$868.2375, IIBZQMonitor, Breathing Frequency
$868.2480, IILKDMonitor, Carbon Dioxide,Cutaneous
$868.2500, IIKLKMonitor, Oxygen, Cutaneous, forInfant not under Gas Anesthesia
General Hospitaland Personal Use(12520)$880.2910, IIFLLThermometer, Electronic,Clinical
Neurological(12513)$882.1400, IIGWRElectroencephalograph
$882.1420, IGWSAnalyzer, Spectrum,Electroencephalogram Signal
    1. The new devices are substantially equivalent to previously cleared Philips devices marketed pursuant to K001664, K021778 and 030038.
    1. The modification is updated software of the Philips Medical System MP60, MP70 and MP90 IntelliVue patient monitor devices.
    1. The new devices have the same intended use as the legally marketed predicate devices. When used in the hospital environment, they are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates.
    1. The new devices have the same technological characteristics as the legally marketed predicate devices.
    1. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new device with respect to the

{2}------------------------------------------------

K031481
page 3 of 3

Testing involved system level tests, performance predicate. tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that Component Compact Monitor meets all reliability requirements and performance claims.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 9 2003

Philips Medizin Systeme Boeblingen GmbH c/o Mr. Hauke Schik Sr. Regulatory Affairs Engineer Cardiac and Monitoring Systems Hewlett-Packard Str. 2 Boeblingen Germany D-71034

Re: K031481

Trade Name: The Philips MP60, MP70 and MP90 IntelliVue Patient Monitors Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm Regulatory Class: Class III (three) Product Code: MHX Dated: May 5, 2003 Received: May 12, 2003

Dear Mr. Schik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Hauke Schik

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

031481

Page 1 of 2

510(k) Number (if known):

Device Name: The Philips IntelliVue MP60, MP70, and MP90 Patient Monitors, Release A.20.

K D3 14

Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in hospital environments.

EASI 12-lead ECG is only for use on adult and pediatric patients.

ST Segment monitoring is restricted to adult patients only.

The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseXOROver-The-Counter
(Per 21 CFR 801.109)
(Optional Format 1-2-96)

(Division Sign-Off)

Division of Cardiovascular Devices

510(k) NumberK03148/
------------------------

10

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.