(29 days)
ABP 90121 RMS, ABP 90219 RMS
No
The description focuses on standard software functionalities like data acquisition, report generation, and simple statistical calculations (mean, standard deviation). There is no mention of AI/ML terms or capabilities.
No
The device is a software application that manages, processes, and reports data from a blood pressure monitor, but it does not directly provide therapy. It is for diagnostic support, not therapeutic intervention.
Yes
The device is described as providing "data to qualified medical personnel for the purpose of assessing the patient's cardiac health via blood pressure readings." While it states that "Patient diagnosis is not to be performed solely based on the results of this device," the primary function of providing data for health assessment aligns with the definition of a diagnostic device, which aids in determining the nature of a disease or condition.
Yes
The device is described as a "PC-based software application" that accepts data from a separate hardware monitor. Its function is data management, analysis, and reporting, not direct hardware control or data acquisition from the patient.
Based on the provided information, the ABP 92506 RMS is not an IVD (In Vitro Diagnostic) device.
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening.
- ABP 92506 RMS Function: The ABP 92506 RMS is a software application that processes data collected from an ambulatory blood pressure monitor. This monitor takes blood pressure readings directly from the patient's body, not from a specimen taken from the body.
- Intended Use: The intended use clearly states it's for assessing cardiac health via blood pressure readings taken during daily activity. This is a physiological measurement, not an analysis of a biological specimen.
- Device Description: The description focuses on the software's ability to acquire, process, and report blood pressure data from the monitor. It doesn't mention any interaction with biological samples.
Therefore, the ABP 92506 RMS falls under the category of a physiological monitoring device rather than an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The ABP 92506 RMS is used to provide data to qualified medical personnel for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to a 48-hour period.
The data is intended to provide preliminary blood pressure data from ambulatory patients for non-time critical applications only.
It is for use in hospitals, clinics or physicians offices by a qualified physician or trained staff member under the supervision of that physician.
Patient diagnosis is not to be performed solely based on the results of this device.
Product codes
DXN
Device Description
The Spacelabs Medical ABP 92506 RMS is a PC-based software application that accepts data collected from one of Spacelabs Medical's ABP monitors through an electrically-isolated interface cable.
The ABP 92506 RMS has a graphical user interface (GUI) similar to most Microsoft applications such as Word, Excel, etc. that provides common windowing controls such as menus, icons, mouse support, wizards, etc. These controls allow the operator to communicate with the ABP monitor.
Additionally, the ABP 92506 RMS provides necessary initialization and configuration of operating parameters, as well as acquires the monitor's stored blood pressure reading data. These readings are combined with patient demographic entry data to create an ABP report that can be reviewed, edited, stored, confirmed and printed by a physician.
Furthermore, the ABP 92506 RMS application also performs simple summary trending calculations from the individual blood pressure readings to present global statistical results, such as mean and standard deviations for systolic, diastolic, MAP, heart rate and pulse pressure. These are calculated for intervals such as the entire recording period, day, night or period of time as defined by operator.
Finally, the ABP 92506 RMS application can import reports generated by the ABP 90121 RMS and ABP 90219 RMS, predecessors of the ABP 92506 RMS.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
hospitals, clinics or physicians offices by a qualified physician or trained staff member under the supervision of that physician.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
JUN - 6 2003
K03/479 1 OF 2
510(k) Summary of Safety and Effectiveness Ambulatory Blood Pressure 92506 Report Management System
Proprietary: | Ambulatory Blood Pressure (ABP) 92506 Report Management System (RMS) |
---|---|
Common: | Noninvasive blood pressure measurement system, accessory |
Classification: | Class II – 21 CFR §870.1130 |
Panel: | Cardiovascular |
Contact Person: | Al Van Houdt, Director of Regulatory Affairs and Quality |
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 1992.
The Spacelabs Medical ABP 92506 Report Management System (RMS) is substantially equivalent to the following currently marketed device(s):
| Spacelabs Medical Ambulatory Blood Pressure Monitor
(includes 90219 RMS) | K855127 |
---|---|
Spacelabs Medical Ambulatory Blood Pressure Monitor | |
(includes 90121 RMS) | K855127 (Letter to File) |
The Spacelabs Medical ABP 92506 RMS is a PC-based software application that accepts data collected from one of Spacelabs Medical's ABP monitors through an electrically-isolated interface cable.
The ABP 92506 RMS has a graphical user interface (GUI) similar to most Microsoft applications such as Word, Excel, etc. that provides common windowing controls such as menus, icons, mouse support, wizards, etc. These controls allow the operator to communicate with the ABP monitor.
Additionally, the ABP 92506 RMS provides necessary initialization and configuration of operating parameters, as well as acquires the monitor's stored blood pressure reading data. These readings are combined with patient demographic entry data to create an ABP report that can be reviewed, edited, stored, confirmed and printed by a physician.
Furthermore, the ABP 92506 RMS application also performs simple summary trending calculations from the individual blood pressure readings to present global statistical results, such as mean and standard deviations for systolic, diastolic, MAP, heart rate and pulse pressure. These are calculated for intervals such as the entire recording period, day, night or period of time as defined by operator.
Finally, the ABP 92506 RMS application can import reports generated by the ABP 90121 RMS and ABP 90219 RMS, predecessors of the ABP 92506 RMS.
The Spacelabs Medical ABP 92506 RMS is designed to comply with the following and Federal Codes:
0000006
1
| 21 CFR §11 | Applicable Sections of Department of Health and Human Services
— Electronic records; electronic signatures |
|-------------|---------------------------------------------------------------------------------------------------------------|
| 45 CFR §164 | Applicable Sections of Department of Health and Human Services
— Administrative requirements |
The ABP 90121 RMS and the ABP 92506 RMS are substantially equivalent in design concepts, technologies, materials and intended uses. The ABP 92506 RMS was developed following a robust software development process, and was fully specified and validated through rigorous testing that, in part, support the compliance of the ABP 92506 RMS to the above mentioned Federal Codes.
The ABP 92506 RMS is the next generation in the Spacelabs Medical ABP RMS family of products.
000007
KO 314 79
EOFE
2
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized eagle symbol. The eagle is depicted with three overlapping profiles, creating a sense of depth and unity. The overall design is simple and professional, reflecting the department's role in public health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN - 6 2003
Datex-Ohmeda, Inc. Spacelabs Medical c/o Mr. Al Van Houdt Director, Regulatory Affairs & Quality 5150 220th Ave SE Issaquah, WA 98029
Re: K031479
Trade Name: Ambulatory Blood Pressure (ABP) 92506 Report Management System (RMS) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system. Regulatory Class: Class II (two) Product Code: DXN Dated: May 8, 2003 Received: May 12, 2003
Dear Mr. Van Houdt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 - Mr. Van Houdt
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
for
Br
Di
Mefacmeyn
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Office of Device Evaluation 5 Center for Devices and Radiological Health
Enclosure
4
Indications for Use
510(k) Number (if known): | K031479 |
---|---|
Device Name: | Ambulatory Blood Pressure (ABP) 92506 Report Management |
System (RMS) | |
Indications for Use: | The ABP 92506 RMS is used to provide data to qualified |
medical personnel for the purpose of assessing the patient's | |
cardiac health via blood pressure readings taken during daily | |
activity for up to a 48-hour period. |
The data is intended to provide preliminary blood pressure data
from ambulatory patients for non-time critical applications only.
It is for use in hospitals, clinics or physicians offices by a
qualified physician or trained staff member under the supervision
of that physician.
Patient diagnosis is not to be performed solely based on the
results of this device. |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Mekimyer for B. Zuckerman
(Medi-Cal Sign Off)
in Sign-()It) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number:
Prescription Use (Per 21CFR801.109)
OR Over-The-Counter Use
Image /page/4/Picture/8 description: The image shows a sequence of numbers. The numbers are all zeros, except for the last digit, which is an eight. The sequence of numbers is "000008".