(101 days)
Not Found
No
The document describes the device's function as acquiring, recording, storing, and providing interpretation of ECG and heart sound data. While it mentions "analysis" and "interpretation," it does not explicitly state or imply the use of AI or ML algorithms for these functions. The description focuses on the hardware components and the acquisition of electrical and acoustic signals. There is no mention of training data, test data, or performance metrics typically associated with AI/ML models.
No
The device is used for acquiring, recording, and storing cardiac data for diagnosis, not for treatment.
Yes
The device acquires, records, and stores cardiac data, and the data is reviewed and used by trained medical personnel in the diagnosis of patients with various cardiac conditions (LVH, MI, S3/S4 heart sounds). It also provides interpretations of the data for consideration by physicians.
No
The device description explicitly mentions hardware components like the Mortara ELI 200+ electrocardiograph, an Audicor Adaptor, and an Audicor Sensor, which are integral to the device's function of acquiring ECG and heart sound data.
Based on the provided information, this device is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs analyze samples taken from the body. The ELI 200+ Audicor Option Electrocardiograph acquires electrical and acoustic signals from the surface of the skin of a patient. It does not analyze biological samples like blood, urine, or tissue.
- The device is used for direct patient measurement. It is applied to the patient's skin to acquire physiological signals. This is characteristic of in vivo diagnostic devices, not in vitro ones.
The device is clearly described as an electrocardiograph that acquires and analyzes physiological signals directly from the patient's body. This falls under the category of in vivo diagnostic devices.
N/A
Intended Use / Indications for Use
The ELI 200+ Audicor Option Electrocardiograph is intended to acquire, record and then store acquired cardiac data of symptomatic patients. The ELI 200+ Audicor Option can also be used to acquire both 12-lead ECG and heart sound data at the same time. The cardiac data is reviewed, confirmed, and used by trained medical personnel in the diagnosis of LVH, acute and age undetermined MI and detection of S3/S4 heart sounds. The S3/S4 heart sound is useful in the identification of cardiac conditions associated with Left Ventricular dysfunction.
- The device is indicated for use to acquire, analyze, display and orint ECG and heart sound (COR) data.
- The device is indicated for use to provide interpretation of the d sta for consideration by physicians.
- The device is indicated for use in a clinical setting, by a physici: n or by trained personnel . who are acting on the orders of a licensed physician. It is not inlended as a sole means of diagnosis.
- The interpretations of ECG and heart sound data (COR) offerer by the device are only significant when used in conjunction with physician overread as well as consideration of all other relevant patient data.
- The device is indicated for use on adult populations, typically sumptomatic.
- The device is not intended to be used as a vital signs physiolog cal monitor. .
- Evaluation of cardiac conditions such as LVH, Acute and Age Undetermined MI and detection of S3 and S4 heart sounds.
Product codes (comma separated list FDA assigned to the subject device)
DPS
Device Description
The ELI 200+ Audicor is an interpretive electrocardiograph that provides simult: neous 12-lead acquisition and is designed to be portable. It also employs Audicor Correlated Audicelectrir. Cardiography (COR) technology from Inovise Medical, Inc. that will enable acquisition of both 12-lear. ECG and heart sound data at the same time. This will be accomplished with a Mortara ELI 200+ electrocar iograph and the use of an Audicor Adaptor and Audicor Sensor. In use, the Audicor Sensor will replace tt = standard V4 electrode and connect to the Audicor Adaptor, enabling simultaneous acquisition of the ECG . gnal and an acoustical sional.
The Audicor sensor is a single-use, non-invasive, disposable medical device th: t has the ability to acquire a single-lead ECG (electrical) signal and a single-lead acoustic signal simultaneo: sly from the skin surface of a patient. It is designed to transfer these signals to an attached Audicor Adaptor, which itself connects and transfers the signals through patient leads to compatible cardiographs such as ine Mortara ELL 200+ electrocardiograph for acquisition and analysis.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Adult populations
Intended User / Care Setting
Clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.2340 Electrocardiograph.
(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).
0
JUL 2 5 2003
Image /page/0/Picture/1 description: The image shows a logo with a heart shape at the top and the word "Mortara" below it. The heart shape is white and is surrounded by a black border. The word "Mortara" is in a bold, sans-serif font and is also black. The logo is simple and clean, with a focus on the heart shape and the company name.
510(k) Notification: Mortara ELI 200+ Audicor
510(k): ELI 200+ Audicor Device Summary
Submitter:
Date: February 5, 2003
Harlan Van Matre, Manager of Quality Assurance / Regulatory Affairs Mortara Instrument, Inc. 7865 N. 86" Street Milwaukee, WI 53224 Fax: (414) 354-4760 Phone: (414) 354-1600 Contact: Harlan Van Matre (see above)
Trade Name: | ELI 200+ Audicor |
---|---|
Common Name: | Electrocardiograph |
Classification Name: | Electrocardiograph |
(Per 21 CFR 870.2340) |
Legally marketed devices to which S. E. is claimed
The Mortara Instrument's ELI 200+ Audicor Electrocardiograph is a current tecli nology evolution of the Mortara ELI 200 and is substantially equivalent to these legally marketed predic ate devices:
- ELI 200 by Mortara Instrument (K920626) ●
- . Inovise's Cardiovise Interpretive Software (K001349)
- HP's 1514A ECG/Phono System ●
The proposed ELL 200+ Audicor is a direct evolution of this Mortara predicate d'uice. It will combine ELI 200 technologies with current technologies including Audicor Correlated Audioelect c Cardiography (COR) technology, resulting in the Mortara ELI 200+ Audicor.
Description:
The ELI 200+ Audicor is an interpretive electrocardiograph that provides simult: neous 12-lead acquisition and is designed to be portable. It also employs Audicor Correlated Audicelectrir. Cardiography (COR) technology from Inovise Medical, Inc. that will enable acquisition of both 12-lear. ECG and heart sound data at the same time. This will be accomplished with a Mortara ELI 200+ electrocar iograph and the use of an Audicor Adaptor and Audicor Sensor. In use, the Audicor Sensor will replace tt = standard V4 electrode and connect to the Audicor Adaptor, enabling simultaneous acquisition of the ECG . gnal and an acoustical sional.
The Audicor sensor is a single-use, non-invasive, disposable medical device th: t has the ability to acquire a single-lead ECG (electrical) signal and a single-lead acoustic signal simultaneo: sly from the skin surface of a patient. It is designed to transfer these signals to an attached Audicor Adaptor, which itself connects and transfers the signals through patient leads to compatible cardiographs such as ine Mortara ELL 200+ electrocardiograph for acquisition and analysis.
Intended use:
The ELI 200+ Audicor Option Electrocardiograph is intended to acquire, record ind then store acquired cardiac data of symptomatic patients. The ELI 200+ Audicor Option can also be used to acquire both 12-lead ECG and heart sound data at the same time. The cardiac data is reviewed, cor irmed, and used by trained medical personnel in the diagnosis of patients with LVH, acute and age undeter nined MI and detection of S3/S4 heart sounds. The S3/S4 heart sound is useful in the identification of carr.iac conditions associated with Left Ventricular dysfunction.
1
510(k) Notification: Mortara ELI 200+ Audicor
Indications for use:
- . The device is indicated for use to acquire, analyze, display and orint ECG and heart sound (COR) data.
- The device is indicated for use to provide interpretation of the d sta for consideration by physicians.
- The device is indicated for use in a clinical setting, by a physici: n or by trained personnel . who are acting on the orders of a licensed physician. It is not inlended as a sole means of diagnosis.
- The interpretations of ECG and heart sound data (COR) offerer by the device are only significant when used in conjunction with physician overread as well as consideration of all other relevant patient data.
- . The device is indicated for use on adult populations, typically sumptomatic.
- The device is not intended to be used as a vital signs physiolog cal monitor. .
- . Evaluation of cardiac conditions such as LVH, Acute and Age Undetermined MI and detection of S3 and S4 heart sounds.
2
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines that suggest the head and wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 5 2003
Mortara Instrument, Inc. c/o Ms. Chantel Carson Group Ldr Engineer Underwriters Laboratories, Inc. 333 Pfingsten Road Northbrook, IL 60062-2096
Re: K031182
Trade Name: ELI 200+ Audicor Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: July 8, 2003 Received: Jul 9, 2003
Dear Ms. Carson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 – Ms. Chantel Carson
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
l. D. Zuber, MD
ram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Page 1 of 1
510(k) Number (if known):
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
ELI 200+ AUDICOR Device Name:
The ELI 200+ with Audicor is a resting electrocardiograph that provides simultaneous 12-lead ECG acquisition and heart sound data (COR - Correlated Audioelectric Cardiography).
The ELI 200+ Audicor Option Electrocardiograph is intended to acquire, record and then store acquired cardiac data of symptomatic patients. The ELI 200+ Audicor Option can also be used to acquire both 12-lead ECG and heart sound data at the same time. The cardiac data is reviewed, confirmed, and used by trained medical personnel in the diagnosis of patients with LVH, Acute and Age Undetermined MI and detection of S3/S4 Heart Sounds. The S3 and S4 Heart Sounds are useful in the identification of cardiac conditions associated with Left Ventricular dysfunction.
This device is appropriate for the indications listed below:
- The device is indicated for use to acquire, analyze, display and print ECG and heart sound . (COR) data.
- . The device is indicated for use to provide interpretation of the data for consideration by physicians.
- . The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.
- The interpretations of ECG and heart sound data (COR) offered by the device are only • significant when used in conjunction with physician overread as well as consideration of all other relevant patient data.
- . The device is indicated for use on adult populations, typically symptomatic.
- The device is not intended to be used as a vital signs physiological monitor. .
- Evaluation of cardiac conditions such as LVH, Acute and Age Undetermined MI and detection . of S3 and S4 heart sounds.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) | |
---|---|
(Division Sign-Off) | |
Division of Cardiovascular Devices | |
510(k) Number | K031182 |
Prescription Use (Per 21CFR801.109) | OR | Over-The-Counter Use (Optional Format 1-2-96) |
---|---|---|
------------------------------------- | ---- | ----------------------------------------------- |
(Optional Format 1-2-96)