K Number
K030936
Date Cleared
2003-04-01

(7 days)

Product Code
Regulation Number
880.5725
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RITA® Clips are used with Harvard 2 Syringe Pump to increase the syringe holding capacity of the Pump from two to five syringes to allow one Harvard 2 Syringe Pump to supply local delivery of saline through multiple irrigation ports of one RITA® Electrosurgical Device of the RITA® System. The RITA System is intended for use in percutaneous. laparoscopic, or intraoperative coagulation and ablation of soft tissue, including the partial or complete ablation of non-resectable liver lesions.

Device Description

The RITA® Clips are a syringe holder accessory to the Harvard 2 Syringe Pump to accommodate five (5) 20-cc syringes.

AI/ML Overview

The provided text describes the 510(k) summary for the RITA® Clips, an accessory for the Harvard 2 Syringe Pump. The primary purpose of this submission is to demonstrate substantial equivalence to a predicate device and establish the safety and effectiveness of the RITA® Clips.

Here's an analysis based on the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document doesn't explicitly state quantitative acceptance criteria in a typical "pass/fail" numerical fashion for a and clinical study. Instead, it mentions functional tests.

Acceptance Criteria CategoryStated VerificationReported Device Performance
Flow Rate VerificationVerify the flow rate of the RITA® Clips combined with the Harvard 2 Syringe Pump."The RITA® Clips and Harvard 2 Syringe Pump underwent testing to verify flow rate... The devices passed the test criteria."
Alarm Function VerificationVerify the alarm function of the RITA® Clips combined with the Harvard 2 Syringe Pump."The RITA® Clips and Harvard 2 Syringe Pump underwent testing to verify... alarm function. The devices passed the test criteria."
Biocompatibility (Patient Contact)Not applicable as the device does not directly contact the patient."The RITA® Clips do not directly contact the patient therefore no biocompatibility testing was required."

Interpretation: The acceptance criteria were broadly focused on ensuring the RITA® Clips did not negatively impact the fundamental functions (flow rate and alarm) of the associated Harvard 2 Syringe Pump. The document states these tests were "passed," implying the device met these functional requirements.

2. Sample Size Used for the Test Set and Data Provenance

The document describes performance data related to functional testing of the device rather than a clinical study with human subjects.

  • Sample Size for Test Set: The document does not specify a numerical sample size for the "testing to verify flow rate and alarm function." It simply states "The RITA® Clips and Harvard 2 Syringe Pump underwent testing." This likely implies testing of a sufficient number of units to ensure consistent performance, but the exact quantity isn't provided.
  • Data Provenance: The data provenance is prospective testing conducted by RITA Medical Systems to verify the functionality of the RITA® Clips when used with the Harvard 2 Syringe Pump. There is no mention of country of origin for, or retrospective use of, data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This question is not applicable to the provided information. The "testing to verify flow rate and alarm function" is a mechanical/electrical performance test, not a clinical study requiring expert assessment of ground truth. Ground truth in this context would be the actual flow rate or alarm activation, which is measured by instruments, not human experts.

4. Adjudication Method for the Test Set

This question is not applicable. Since the testing involved mechanical performance verification (flow rate and alarm function) rather than human interpretation or clinical outcomes, an adjudication method for a test set (e.g., 2+1, 3+1) is not relevant.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The RITA® Clips are a mechanical accessory (syringe holder) for an infusion pump. The study described is a verification of its mechanical function (flow rate and alarm) rather than a clinical study evaluating diagnostic accuracy or reader improvement. Therefore, the concept of human readers, AI assistance, or effect size for such improvement is not relevant to this submission.

6. If a Standalone Performance (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable in the context of the RITA® Clips.

  • The RITA® Clips are a mechanical device, not an algorithm or software.
  • The performance described is the physical functionality of the clips when attached to the pump, not an algorithm's output.
  • There is no "human-in-the-loop" interaction in the described performance testing; it's about the device's inherent function.

7. The Type of Ground Truth Used

The ground truth used for the performance testing was objective mechanical measurements.

  • For flow rate: The ground truth would be the accurately measured volume/time output from the pump system with the RITA® Clips installed, compared against the expected or desired flow rate.
  • For alarm function: The ground truth would be the correct activation of the pump's alarm under specific (e.g., occlusion, end-of-infusion) conditions.

These are verifiable physical phenomena, not expert consensus, pathology, or outcomes data typically associated with diagnostic or therapeutic clinical trials.

8. The Sample Size for the Training Set

This question is not applicable. The RITA® Clips are a physical medical device, not an AI/ML algorithm or system that requires a "training set" in the computational sense. The described testing is performance verification of hardware.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable, as there is no training set for this type of mechanical device.

{0}------------------------------------------------


510(K) SUMMARY OF SAFETY AND EFFECTIVENESS Y 030 936 9.0

ContactInformationRITA Medical Systems
APR 01 2003
967 N. Shoreline Blvd.
Mountain View, CA 94043
Phone: 650 314-3400 Fax: 650 890-3905
Contact Person: Vicki Hacker, VP of Clinical Affairs
GeneralProvisionsTrade Name: RITA® Clips
Common/Classification Name: Infusion Pump Accessory
Name ofPredicate DeviceHarvard Clinical Technology - Harvard 2 Syringe Pump
ClassificationClass II
PerformanceStandardsPerformance standards have not been established for infusion pump accessory(syringe holder) under section 514 of the Food, Drug and Cosmetic Act.
Intended UseThe RITA® Clips are used with Harvard 2 Syringe Pump to increase the syringeholding capacity of the Pump from two to five syringes to allow one Harvard 2Syringe Pump to supply local delivery of saline through multiple irrigation ports ofone RITA® Electrosurgical Device of the RITA® System. The RITA® System isintended for use in percutaneous, laparoscopic, or intraoperative coagulation andablation of soft tissue, including the partial or complete ablation of non-resectableliver lesions.
DeviceDescriptionThe RITA® Clips are a syringe holder accessory to the Harvard 2 Syringe Pump toaccommodate five (5) 20-cc syringes.
PerformanceDataThe RITA® Clips and Harvard 2 Syringe Pump underwent testing to verify flowrate and alarm function. The devices passed the test criteria. The RITA® Clips donot directly contact the patient therefore no biocompatibility testing was required.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image contains the words "Public Health Service". The text is in a simple, sans-serif font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line.

Image /page/1/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right. Above the profiles are three curved lines that resemble feathers or wings.

APR 0 1 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

RITA Medical Systems Incorporated C/O Mr. Charles Mack Responsible Third Party Official Underwriters Laboratories Incorporated 2600 N.W. Lake Road Camas, Washington 98601-8542

Re: K030936

Trade/Device Name: RITA ® Clips for Infusion Pumps Regulation Number: 880.5725 Regulation Name: Infusion Pump Accessory Regulatory Class: II Product Code: MRZ Dated: March 19, 2003 Received: March 25, 2003

Dear Mr. Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Mack

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Raanes

Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

INTENDED USE 3.0

Indications for Use Statement

510(K) Number (if known)

RITA® Clips for Infusion Pumps Device Name

The RITA® Clips are used with Harvard 2 Syringe Pump to increase the syringe holding capacity of the Pump from two to five syringes to allow one Harvard 2 Syringe Pump to supply local delivery of saline through multiple irrigation ports of one RITA® Electrosurgical Device of the RITA® System. The RITA System is intended for use in percutaneous. laparoscopic, or intraoperative coagulation and ablation of soft tissue, including the partial or complete ablation of non-resectable liver lesions.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use v (per 21 CFR 801).

OR

Over the Counter Use _________________________________________________________________________________________________________________________________________________________

Talcus Cicerite

Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

510(k) Number:

March 6, 2003

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).