(155 days)
The diaDexus PLAC™ test is an enzyme immunoassay for the quantitative determination of Lp-PLA2 (lipoprotein-associated phospholipase A2) in human plasma, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predicting risk for coronary heart disease.
The diaDexus PLAC™ test is based on the principle of a sandwich enzyme immunoassay using two specific monoclonal antibodies. The assay system utilizes monoclonal anti-Lo-PLA2 antibody (2C10) for solid phase immobilization on the microtiter stripwells. The test sample is first diluted with the sample diluent and incubated at 2-8°C for 60 minutes. The diluted test sample is then allowed to react with the immobilized monoclonal antibody at 2-8°C for 90 minutes. The wells are washed with distilled water to remove any unbound antigen. A second monoclonal anti-Lo-PLA2 antibody (4B4) labeled with the enzyme horseradish peroxidase (HRP) is then added and reacted with the immobilized antigen at 2-8℃ for 60 minutes, resulting in the Lp-PLA2 molecules being captured between the solid phase and the enzyme-labeled antibodies. The wells are washed with distilled water to remove unbound labeled antibodies. The substrate, tetramethylbenzidine (TMB). is then added and incubated at 2-8°C for 20 minutes resulting in the development of a blue color. Color development is stopped with the addition of Stop Solution (1N HCl), changing the color to yellow. The absorbance of the enzymatic turnover of the substrate is determined spectrophotometrically at 450 nm using a standard microplate reader and is directly proportional to the concentration of Lp-PLA2 present. A set of Lp-PLA2 calibrators is used to plot a standard curve of absorbance (v-axis) versus Lp-PLA2 concentration in ng/mL (x-axis) from which the Lp-PLA2 concentration in the test sample can be interpolated. The standard curve is constructed using a point-to-point curve fit manually or by using appropriate calibration curve fitting software.
Here's an analysis of the diaDexus PLAC™ Test based on the provided text, addressing your specific questions:
Acceptance Criteria and Device Performance Study for diaDexus PLAC™ Test
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria for the analytical performance characteristics. Instead, it presents the results of these performance tests as factual observations. For the clinical performance, the acceptance criteria are implicitly defined by demonstrating a statistically significant association between Lp-PLA2 and CHD risk across different models.
| Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Analytical Performance | ||
| Analytical Sensitivity (Detection Limit) | A low detection limit suitable for intended use. | 1.2 ng/mL |
| Linearity (Dilution) | Percent recovery close to 100%. | Average recovery of 104% (range 133-1310 ng/mL) |
| Linearity (Dilution) | Percent recovery close to 100%. | Average recovery of 104% (range 79-982 ng/mL) |
| Antigen Recovery | Mean recovery close to 100% (with an acceptable range). | Mean recovery of 109% (range 97-119%) |
| Interfering Substances | No appreciable interference from common endogenous substances. | No appreciable interference for hemoglobin, triglycerides, cholesterol, bilirubin, and albumin at specified levels. |
| Precision (Intra-assay) | Low Coefficient of Variation (CV). | < 7% CV (n=80) |
| Precision (Inter-assay) | Low Coefficient of Variation (CV). | < 9% CV (n=20) |
| Clinical Performance | ||
| Prediction of CHD Risk | Statistically significant predictor of CHD risk. | Hazard Ratios (95% CI) demonstrated significance (e.g., Model 3: 1.71 (p=0.029) and 2.12 (p=0.003) for intermediate and highest levels respectively). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: 1348 banked EDTA-plasma samples.
- 608 from CHD cases
- 740 from participants free of CHD (controls)
- Data Provenance:
- Country of Origin: Not explicitly stated, but the "Atherosclerosis Risk In Communities (ARIC) Study" sponsored by the National Institutes of Health (NIH) strongly suggests USA.
- Retrospective or Prospective: The study is described as a "case-cohort study" where participants "free of CHD were enrolled and followed for the development of CHD for up to nine years." This indicates a prospective design for patient enrollment and follow-up data collection, with the Lp-PLA2 levels being measured retrospectively on banked samples from this cohort.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number of experts or their qualifications for establishing the ground truth (CHD diagnosis) for the test set. It states that the clinical data was collected "following a well-designed protocol conducted by qualified experts in a CLIA-certified laboratory" within the ARIC Study. This implies that the diagnosis of CHD and the determination of "free of CHD" were established through the standard clinical protocols and expert medical judgment within the large epidemiological study, rather than a specific expert panel convened solely for this device's ground truth adjudication.
4. Adjudication Method for the Test Set
The document describes the origin of the "ground truth" (CHD cases vs. controls) as being derived from the ARIC Study's follow-up data. It does not mention a specific adjudication method (like 2+1 or 3+1) for the diagnostic outcomes used as ground truth for this device study. The implication is that the diagnoses from the ARIC study are considered the established ground truth.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) test measuring a biomarker (Lp-PLA2) in plasma, not an imaging device or AI algorithm that human readers would interact with. Therefore, there is no "human readers improve with AI vs without AI assistance" component.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, the study presents the standalone performance of the diaDexus PLAC™ Test. The device measures Lp-PLA2 levels in plasma samples, and these levels are then used in statistical models (Cox regression) to predict the risk of CHD. This is an algorithm-only performance in the sense that the test results are generated by the assay (the device) independently, and then interpreted statistically. There is no human-in-the-loop interaction with the device's measurement process.
7. The Type of Ground Truth Used
The ground truth used was outcomes data, specifically the development of Coronary Heart Disease (CHD) over a nine-year follow-up period in participants from the ARIC Study.
8. The Sample Size for the Training Set
The document does not explicitly state a separate training set size for the device's development or the clinical study. The 1348 samples from the ARIC study are used for the clinical validation. It is possible that the assay itself (e.g., antibody selection, assay parameters) was developed using internal samples, but details on a formal training set for the prediction model are not provided. The cut-points (310 and 420 ng/mL) used in the Cox regression analysis were "generated from the ARIC data set," which implies they were derived from the same 1348 samples, suggesting this dataset served as both a development and validation set, or that these cut-points were post-hoc derived from the same studied population based on disease prevalence.
9. How the Ground Truth for the Training Set Was Established
As noted above, a distinct "training set" with separate ground truth establishment is not described. If the ARIC data was used for both deriving cut-points and demonstrating performance, then the ground truth (CHD outcomes) for this dataset was established through the follow-up of participants in the ARIC study for up to nine years, where clinical diagnosis of CHD was determined by "qualified experts in a CLIA-certified laboratory" as part of that large epidemiology study.
{0}------------------------------------------------
510(k) Summary diaDexus PLAC™ Test
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K030477.
General Information
| Name and Address of Applicant: | diaDexus, Inc.343 Oyster Point Blvd.South San Francisco, CA 94080650-246-6400Robert Wolfert, Ph.D. |
|---|---|
| Date Prepared | July 16, 2003 |
| Device Trade Name: | diaDexus PLAC™ Test |
| Generic Name: | Enzyme Immunoassay forthe Quantitative Determination of LpPLA2 (Lipoprotein-AssociatedPhospholipase A2) in HumanPlasma |
Identification of Legally Marketed Device
The Lp(a) Test - K013128 (N Latex Lp(a) Reagent, Dade Behring Inc., Newark, DE)
Intended Use
The diaDexus PLAC™ test is an enzyme immunoassay for the quantitative determination of Lp-PLA2 (lipoprotein-associated phospholipase A2) in human plasma, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predicting risk for coronary heart disease.
{1}------------------------------------------------
Device Description
The diaDexus PLAC™ test is based on the principle of a sandwich enzyme immunoassay using two specific monoclonal antibodies.
The assay system utilizes monoclonal anti-Lo-PLA2 antibody (2C10) for solid phase immobilization on the microtiter stripwells. The test sample is first diluted with the sample diluent and incubated at 2-8°C for 60 minutes. The diluted test sample is then allowed to react with the immobilized monoclonal antibody at 2-8°C for 90 minutes. The wells are washed with distilled water to remove any unbound antigen. A second monoclonal anti-Lo-PLA2 antibody (4B4) labeled with the enzyme horseradish peroxidase (HRP) is then added and reacted with the immobilized antigen at 2-8℃ for 60 minutes, resulting in the Lp-PLA2 molecules being captured between the solid phase and the enzyme-labeled antibodies. The wells are washed with distilled water to remove unbound labeled antibodies. The substrate, tetramethylbenzidine (TMB). is then added and incubated at 2-8°C for 20 minutes resulting in the development of a blue color. Color development is stopped with the addition of Stop Solution (1N HCl), changing the color to yellow. The absorbance of the enzymatic turnover of the substrate is determined spectrophotometrically at 450 nm using a standard microplate reader and is directly proportional to the concentration of Lp-PLA2 present. A set of Lp-PLA2 calibrators is used to plot a standard curve of absorbance (v-axis) versus Lp-PLA2 concentration in ng/mL (x-axis) from which the Lp-PLA2 concentration in the test sample can be interpolated. The standard curve is constructed using a point-to-point curve fit manually or by using appropriate calibration curve fitting software.
Characterization of Rare Reagents
Antigen
The antigen used in the diaDexus enzyme immunoassay PLAC™ test is purified recombinant Lo-PLA2 (DDX-RA). Antigen preparations were characterized using SDSpolyacrylamide gels under reducing and non-reducing conditions and Western blot analysis using an anti-Lp-PLA2 antibody, to demonstrate consistency with the molecular weight of the antigen reported in the literature.
Antibodies
The monoclonal anti-Lp-PLA2 antibodies used in the preparation of the coated microtiter stripwells (2010) and conjugate (4B4) were characterized for purity and reactivity in a series of procedures including Paragon gel electrophoresis, SDS-PAGE, size exclusion chromatography, isotyping and enzyme immunoassay. These results demonstrated that the monoclonal antibodies bind to the Lp-PLA2 antigen quantitatively and specifically.
{2}------------------------------------------------
Comparison of New Device to Predicate Device
The chart below identifies the similarities and differences between PLAC™ test and the predicate device, the Lp(a) test (N Latex Lp(a) Reagent, Dade Behring Inc., Newark, DE).
| Characteristic | PLAC™ Test | N Latex Lp(a) Reagent(K013128) |
|---|---|---|
| Intended Use | The diaDexus PLAC™ test is anenzyme immunoassay for thequantitative determination of Lp-PLA2 (lipoprotein-associatedphospholipase A2) in humanplasma, to be used in conjunctionwith clinical evaluation and patientrisk assessment as an aid inpredicting risk for coronary heartdisease. | Measurement of Lp(a) aids in theidentification of individuals at riskfrom cardiovascular disease inspecific populations when used inconjunction with clinical evaluation. |
| Analyte | Lp-PLA2 (lipoprotein-associatedphospholipase A2) | Lipoprotein (a) [Lp(a)] |
| Sample | Human EDTA-plasma | Human serum and heparin-plasma |
| Methodology | Microplate enzyme immunoassay | Particle enhancedimmunonephelometry |
| DetectionMethod | Spectrophotometer at 450 nm | BN Systems Nephelometer |
| Risk to Patients | Minimal risk | Minimal risk |
| LaboratoryEnvironment | Professional laboratory | Professional laboratory |
Performance Characteristics
Analytical Sensitivity (Detection Limit)
The minimum detection limit, as calculated by interpolation of the mean plus two standard deviations of 24 replicates of the 0 ng/mL Lp-PLA2 Calibrator is 1.2 ng/mL.
Linearity
Four EDTA-plasma samples containing different levels of endogenous Lp-PLA2 were diluted with Sample Diluent and assayed (dilution range: 1:5 to 1:20). Percent recovery was determined as the observed value divided by the expected value, multiplied by 100. The average recovery, demonstrating linearity of diluted samples over a range of 133 to 1310 ng/mL Lp-PLA2, was 104%.
{3}------------------------------------------------
Dilution Linearity
Six EDTA-plasma samples with known high Lp-PLA2 levels were intermixed with six plasma samples with known low Lp-PLA2 levels. Percent recovery was determined as the measured value divided by the expected value, multiplied by 100. The average recovery, demonstrating linearity of diluted samples over a range of 79 to 982 ng/mL Lb-PLA2, was 104%.
Antigen Recovery
Known amounts of recombinant Lp-PLA2 (80, 179, and 436 ng/mL) were added to 15 human plasma samples and fetal bovine serum (control). Recovery was determined as the observed spiked concentration in the plasma samples, calculated as a percent of the observed control concentration (with no endogenous Lp-PLA2). Results demonstrated a mean antigen recovery from 97-119% across the assay range, with an overall mean recovery of 109%.
Interfering Substances
Five endogenous substances found in blood were evaluated for interference in the assay. Five individual plasma samples with Lp-PLA2 values ranging from 92 to 664 ng/ml_ were spiked with potential interferents endogenous to blood. No appreciable interference was observed at spiked levels of 500 mg/dL hemoglobin, 3000 mg/dL trialycerides, 500 mg/dL cholesterol, 20 mg/dL bilirubin and up to 6.2 g/dL albumin.
Precision
Intra-assay and inter-assay variability were determined by testing three human plasma pools with Lp-PLA2 concentrations distributed throughout the calibration range of the assay. The three plasma pools were assayed, using a single lot of reagents, in duplicate. on two separate stripwells per day, for twenty days. The intra-assay precision (n = 80) was < 7% CV and the inter-assay precision (n = 20) was < 9% CV.
Reference Interval
EDTA-plasma samples obtained from apparently healthy males (n=251) and apparently healthy females (n=174), in the clinically relevant age range of 40-70 years, were evaluated with the diaDexus PLAC™ test. The reference interval calculated from the samples (central 90%) was determined to be 120-342 ng/mL for females and 131-376 ng/mL for males.
Summary of Clinical Study
To determine the efficacy of the diaDexus PLAC™ test as a predictor of risk for coronary heart disease (CHD). Lp-PLA2 levels were measured in 1348 banked EDTA-plasma samples from a large, multi-center, epidemiology study, the Atherosclerosis Risk In Communities (ARIC) Study, sponsored by the National Institutes of Health's, National Heart, Lung, and Blood Institute. In this case-cohort study, participants, age 47-69, free of CHD were enrolled and followed for the development of CHD for up to nine vears. Of the 1348
{4}------------------------------------------------
samples selected for the Lp-PLA2 study, 608 were from CHD cases, and 740 were from participants who were free of CHD at the time of censor (controls).
Cox regression models were used to evaluate the association of Lp-PLA2 and CHD in a univariate analysis (Model 1), a univariate analysis adjusted for demographics (Model 2), and a multivariate model adjusted for demographics and other prognostics factors (Model 3). Using high and low cutpoints of Lp-PLA2, generated from the ARIC data set (420 and 310 ng/mL, the 67th and 330 percentiles, respectively), the hazard ratios of the Cox regression analyses demonstrated that Lp-PLA2 is a significant predictor of risk for CHD, for the highest and intermediate levels when compared to the lowest level of Lp-PLA2, for all participants (see Table 1). It should be noted that different cutpoints may be appropriate for different clinical populations.
| Lp-PLA2 Risk Ratio (95% Cl, p value)* | |||
|---|---|---|---|
| Lp-PLA2 (ng/mL) | <310 | 310-420 | >420 |
| Model 1 | 1.0 | 1.49(1.11-1.99, p=0.008) | 2.50(1.89-3.31, p<0.001) |
| Model 2 | 1.0 | 1.24(0.92-1.66, p=0.154) | 1.76(1.32-2.36, p<0.001) |
| Model 3 | 1.0 | 1.71(1.06-2.75, p=0.029) | 2.12(1.29-3.48, p=0.003) |
Table 1. Risk Ratios of CHD for Subjects Across All LDL Levels
*The lowest tertile with Lp-PLA2 values <310 ng/mL is used as the reference group
Model 1: univariate analysis
Model 2: adjusted for age, race, gender
- Model 3: adjusted for age, race, gender, current smoking status, blood pressure, diabetes, HDL, LDL, CRP and Lp-PLA2 - LDL interaction
Conclusions
The PLAC™ test demonstrated acceptable analytical performance, i.e., the test is reproducible, linear and accurate, and is not subject to appreciable crossreactivity or interference. The detection limit of 1.2 na/mL is acceptable for the intended use of this assay. Results were consistent across several manufactured reagent lots evaluated over a period of time. Therefore, this assay should provide reliable and reproducible results when used by clinical laboratories.
Clinical data were collected from a large, multi-center epidemiology study sponsored by the National Institutes of Health, following a well-designed protocol conducted by qualified experts in a CLIA-certified laboratory. The results of this clinical study and performance testing demonstrate that the levels of Lp-PLA2 are associated with the risk of CHD, and support the safety and effectiveness of the PLAC™ test for use as a predictor of risk for coronary heart disease.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of a human figure, represented by three curved lines that suggest a profile view of a person's head and shoulders.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUL 1 8 2003
Robert L. Wolfert, Ph.D. Vice President of Diagnostics diaDexus, Inc. 343 Oyster Point Blvd. South San Francisco, CA 94080
K030477 Re:
Trade/Device Name: diaDexus PLACTM Test Regulation Number: 21 CFR 866.5600 Regulation Name: Low-density lipoprotein immunological test system Regulatory Class: Class II Product Code: NOE; JJX Dated: May 21, 2003 Received: May 22, 2003
Dear Dr. Wolfert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
INTENDED USE
510(k) Number: K030477
Device Name: diaDexus PLAC™ test
Indications for Use:
The diaDexus PLAC™ test is an enzyme immunoassay for the quantitative determination of Lp-PLA2 (lipoprotein-associated phospholipase A2) in human plasma, to be used in conjunction with clinical evaluation and patient risk assessment as an aid in predicting risk for coronary heart disease.

Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K030477
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE AS NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________
§ 866.5600 Low-density lipoprotein immunological test system.
(a)
Identification. A low-density lipoprotein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the low-density lipoprotein in serum and other body fluids. Measurement of low-density lipoprotein in serum may aid in the diagnosis of disorders of lipid (fat) metabolism and help to identify young persons at risk from cardiovascular diseases.(b)
Classification. Class II (performance standards).