(185 days)
The Opus™ Spinal System is intended for posterior, noncervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture and dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and failed previous fusion.
The Opus™ Spinal System is also intended to be used in conjunction with the titanium hooks from the OSS/Diapason Spinal System and the Xia Spinal System. The Opus™ Spinal System is also intended to be used in conjunction with the Multi-Axis Cross Connectors.
The Opus Spinal System is made up of a range of screws, which are compatible with both the rod and plate components of the system. The components of the system are manufactured from ISO 5832/3 Titanium Alloy (Ti-6Al-4V).
Here's an analysis of the acceptance criteria and study information for the Opus™ Spinal System based on the provided document:
This document is a 510(k) summary for an additional indication for the Opus™ Spinal System. It primarily focuses on demonstrating substantial equivalence to a predicate device through mechanical testing. It does not contain information about AI/ML device performance, human reader studies, ground truth establishment, or training data for such applications.
Here's a breakdown of the requested information based on the document:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Mechanical properties comparable to predicate device | Static and fatigue test results show the constructs of the Opus™ Spinal System demonstrated comparable mechanical properties to the predicate device. |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not specify a "test set" in the context of clinical data or image-based diagnostics. The testing involved mechanical performance on device constructs. The sample size for these mechanical tests is not explicitly stated in the provided text. The provenance is the testing facility, not patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable. This device is a spinal fixation system, not a diagnostic AI/ML device. Ground truth, in this context, would relate to mechanical engineering standards and material properties, not expert interpretations of medical images or clinical outcomes in the usual AI/ML sense.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like "2+1" or "3+1" are typically used for establishing ground truth in clinical or image-based studies where expert consensus is needed. This document describes mechanical testing, which relies on engineering standards and measurements, not expert adjudication in the same way.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This document is for a spinal implant, not an AI-assisted diagnostic tool. No MRMC studies were performed related to AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This device is a physical spinal implant; there is no algorithm or standalone performance in the context of AI/ML.
7. The Type of Ground Truth Used
The ground truth for this device's performance evaluation is based on mechanical engineering standards and benchmark performance of the predicate device. The goal was to demonstrate "comparable mechanical properties" to the legally marketed predicate device (OSS/Diapason and Xia® Titanium Hooks).
8. The Sample Size for the Training Set
Not applicable. This device is not an AI/ML system that utilizes a "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable. There is no training set for this device in the AI/ML context.
Summary of Device and Study Focus:
The Opus™ Spinal System's 510(k) summary focuses on demonstrating substantial equivalence to existing predicate devices (OSS/Diapason and Xia® Titanium Hooks) for additional indications. This equivalence is primarily supported by mechanical performance data (static and fatigue tests) showing "comparable mechanical properties" between the Opus™ Spinal System constructs and the predicate device. The document does not describe any clinical trials, human-in-the-loop studies, or AI/ML performance evaluations.
§ 888.3070 Thoracolumbosacral pedicle screw system.
(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.