K Number
K030051

Validate with FDA (Live)

Date Cleared
2003-04-04

(88 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VersaBoard patient positioning system developed and manufactured by Bionix Development Corporation, Toledo, Ohio, is intended to be used for the positioning and re-positioning of patients undergoing or receiving a course of external beam radiation therapy for the treatment of cancer and other diseases.

It is intended to be used by or under the direction of a licensed physician.

Device Description

The Bionix VersaBoard is substantially equivalent to the Med-Tec S-Type Baseplate in design, construction, and function. The VersaBoard is flat and has a similar, generally torso-shaped contour. with an area specifically for the head, shoulders, and back. The head portion has a central open area where a thin plate of carbon fiber/epoxy is placed. This thin plate may have a waffle-like cutout, or a more sophisticated design that allows for the prone as well as supine positioning of the patient. The Bionix VersaBoard is manufactured according to the FDA Good Manufacturing Practice guidelines using standard methods and practices. The VersaBoard is constructed in the same manner as the S-Type Baseplate from Med-Tec, having a carbon fiber/epoxy/foam core composite structure) that is an accepted standard in radiation therapy. The carbon fiber/epoxy again provides stiffness and strength, while the foam core allows for almost no attenuation of the radiation beam during the treatment process. The VersaBoard also has simple mechanical interlocks that allow the board to be secured to the tabletop of the therapy couch. Other interlocks or clamps allow low-melt thermoplastic to be attached to the VersaBoard during the patient positioning process.

AI/ML Overview

The provided document is a 510(k) Premarket Notification Summary for the VersaBoard Patient Positioning System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing detailed acceptance criteria and studies as might be found in a De Novo submission or a PMA.

Therefore, many of the requested elements regarding acceptance criteria, specific study design, sample sizes, expert involvement, and ground truth establishment are not typically included or required in a 510(k) summary focused on substantial equivalence for a device like a patient positioning system.

Here's an analysis based on the information available in the document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not define specific, quantifiable acceptance criteria or provide a table of reported device performance in the way a clinical study would for a novel therapeutic or diagnostic device. Instead, the submission relies on demonstrating substantial equivalence to predicate devices. The key "performance" metrics for patient positioning systems like the VersaBoard relate to their physical properties (strength, stiffness, minimal radiation attenuation) and their ability to reproducibly position patients without interfering with radiation therapy.

Acceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device Performance (as claimed in submission)
Similar Design and Construction"The Bionix VersaBoard is substantially equivalent to the Med-Tec S-Type Baseplate in design, construction, and function." "The VersaBoard is constructed in the same manner as the S-Type Baseplate from Med-Tec, having a carbon fiber/epoxy/foam core composite structure."
Similar Functionality for Patient Positioning and Repositioning"In clinical practice the VersaBoard again functions similarly to the Med-Tec S-Type Baseplate." "The patient is positioned on the VersaBoard... allowing him to be positioned and repositioned securely during the radiation therapy regimen."
Minimal Radiation Attenuation"The carbon fiber/epoxy/foam composite structure of the board has a minimal attenuation factor." "The foam core allows for almost no attenuation of the radiation beam during the treatment process." "Similar to that exhibited by the VersaBoard." (in comparison to other predicate devices)
Strength and Stiffness for Patient Support"The carbon fiber/epoxy skin provides strength and stiffness;" "It is reasonable to expect that the two devices will have similar properties as regards to stiffness, support strength..."
Compatibility with Radiation Therapy Couch & Thermoplastic Masks"The VersaBoard also has simple mechanical interlocks that allow the board to be secured to the tabletop of the therapy couch." "Other interlocks or clamps allow low-melt thermoplastic to be attached to the VersaBoard."

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: Not applicable. The submission does not describe a clinical "test set" in the context of a performance study with patient data. It is a comparison of design and function of the device itself to predicate devices.
  • Data Provenance: Not applicable. The "data" presented comprises descriptions of the device's design, construction materials, and intended use, along with comparisons to existing predicate devices. No patient or clinical data is provided or referenced from a specific country of origin, nor is it described as retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

  • Not applicable. No ground truth was established by experts for a test set in the context of this 510(k) submission. The submission relies on the established safety and effectiveness of the predicate devices. The submitter, James Huttner M.D., Ph.D., as Vice President, New Product Development, is an expert in the field.

4. Adjudication Method for the Test Set:

  • Not applicable. There was no test set requiring adjudication in the context of a clinical study.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices to assess reader performance. The VersaBoard is a patient positioning system, not a diagnostic device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. The VersaBoard is a physical medical device, not an algorithm. Therefore, "standalone" performance in the AI context does not apply. The device's function is inherently "with human-in-the-loop" as it is used by clinicians to position patients.

7. The Type of Ground Truth Used:

  • Not applicable in the conventional sense of clinical ground truth (e.g., pathology, outcomes data). The "ground truth" for this submission is effectively the established safety and effectiveness of the legally marketed predicate devices (Med-Tec S-Type Baseplate, Med-Tec Carbon Fiber Breast Board, Bionix Max 2 Deluxe TorsoBoard) as recognized by the FDA through prior clearances (K933227, K974703, K905007). The submission argues that since the VersaBoard is substantially equivalent in design, materials, and intended use, it inherently shares the "ground truth" of safety and effectiveness with these predicates.

8. The Sample Size for the Training Set:

  • Not applicable. There is no concept of a "training set" for this type of physical medical device in a 510(k) submission.

9. How the Ground Truth for the Training Set was Established:

  • Not applicable, as there was no training set.

{0}------------------------------------------------

KD30051

VersaBoard Patient Positioning System-Premarket Notification Submission

Premarket Notification (510K) Summary

APR 0 4 2003

Date Prepared: October 14, 2002

Bionix Development Corporation Submitter: 5154 Enterprise Blvd. Toledo, Ohio 43612 419.727.8421 (phone) 419.727.4430 (fax)

Contact Person: James Huttner M.D., Ph.D. [jjhuttner(@yahoo.com (email)]

Trade Name: VersaBoard Patient Positioning System

Common Name: Carbon Fiber Patient Immobilization System

Classification Name: Medical charged-particle radiation therapy system, accessory (per CFR section 892.5050)

Intended Use: The VersaBoard from Bionix Development Corporation is designed to be used for the positioning and re-positioning of patients for receiving radiation therapy.

Claim of Substantial Equivalence:

This product is similar in design and function to existing patient positioning devices currently in use as accessories to radiation therapy systems.

One such device is the S-Type Baseplate manufactured and legally marketed by Med-Tec. Inc. of Orange City, Iowa. This device has been classified as a Class II device by the FDA, and has been granted marketing clearance and has been assigned the document control number K933227.

The S-Type Baseplate from Med-Tec consists of a flat "board" comprised of a composite material with a carbon fiber/epoxy skin and a foam core. The device has a generally torso-shaped contour, with a specific area for the head, shoulders, and back. The head portion contains in its center a mesh-like cutout section, where the carbon fiber/epoxy skin is bonded to itself with no foam core, and then a waffle-like cutout pattern is cut into the carbon fiber to give a more open area roughly where the patient's head should be. Other important features of the device are simple mechanical interlocking systems for attaching the S-Type Baseplate to the top of the radiation therapy couch, and for the attachment of contoured low-melt thermoplastic sheets that are used to further position and hold the patient.

{1}------------------------------------------------

VersaBoard Patient Positioning System-Premarket Notification Submission

The carbon fiber/epoxy/foam composite structure of the board has a minimal attenuation factor. This is due primarily to the foam core of the composite, which being mostly air blocks little of the radiation. Standard dosimetry has been used to document this fact, and such results have been widely published in the medical literature. The carbon fiber/epoxy skin provides strength and stiffness; in aggregate such composite structures are ideal for producing devices that reproducibly position patients and yet do not interfere with the administration of the therapeutic radiation. Patient positioning devices with this type of composite structure are common in radiation therapy. They come in many varieties and are manufactured by several companies; examples include Med-Tec, Aktina, Arplay, and Bionix.

In practice, the Med-Tec S-Type Base-plate is secured to the therapy couch tabletop either by a lock-down mechanism, or by the patient's own weight. The patient is positioned supine on the board with his head resting on a cradle over the area of the waffle cut-out, and then a mask of his/her upper torso is made by stretching warm lowmelt thermoplastic over the patient, and then securing that mask to the board using the interlocking mechanism (in this case, custom panel-rivets) described earlier. As the lowmelt thermoplastic cools it becomes rigid, taking and holding the shape of the patient. In this fashion the patient is positioned reproducibly on the board. Radiation therapy is then administered in the usual fashion. (Copies and marketing materials from the Med-Tec, Inc. catalog and web-site are appended to this document to substantiate and clarify the above claims as to design and use of the Med-Tec S-Type Baseplate.)

The Bionix VersaBoard is substantially equivalent to the Med-Tec S-Type Baseplate in design, construction, and function. The VersaBoard is flat and has a similar, generally torso-shaped contour. with an area specifically for the head, shoulders, and back. The head portion has a central open area where a thin plate of carbon fiber/epoxy is placed. This thin plate may have a waffle-like cutout, or a more sophisticated design that allows for the prone as well as supine positioning of the patient.

The Bionix VersaBoard is manufactured according to the FDA Good Manufacturing Practice guidelines using standard methods and practices. The VersaBoard is constructed in the same manner as the S-Type Baseplate from Med-Tec, having a carbon fiber/epoxy/foam core composite structure) that is an accepted standard in radiation therapy. The carbon fiber/epoxy again provides stiffness and strength, while the foam core allows for almost no attenuation of the radiation beam during the treatment process. The VersaBoard also has simple mechanical interlocks that allow the board to be secured to the tabletop of the therapy couch. Other interlocks or clamps allow low-melt thermoplastic to be attached to the VersaBoard during the patient positioning process. (Engineering drawings and perspective views, as well as digital images of the prototype device are appended to this document to substantiate the above claims as to design and structure of the Bionix VersaBoard, as production models are not yet available.)

In clinical practice the VersaBoard again functions similarly to the Med-Tec S-Type Baseplate. The patient is positioned on the VersaBoard in either the prone or supine position, with his head resting on a foam cushion or support. Warm low-melt thermoplastic in its pliable state is then draped over the patient's head and shoulders

{2}------------------------------------------------

where it conforms to the patient's anatomy. It is then secured to the VersaBoard using clamps or other simple mechanical interlocks. When it cools, the low-melt thermoplastic becomes rigid and retains the shape of the patient, allowing him to be positioned and repositioned securely during the radiation therapy regimen.

Based on the almost identical design and construction of the Bionix VersaBoard to the S-Type Baseplate currently manufactured and sold by Med-Tec, Inc., it is reasonable to expect that the two devices will have similar properties as regards to stiffness, support strength, and minimal attenuation of the radiotherapy beam, and should function in a substantially equivalent fashion during the patient positioning and the radiation therapy process. Both the VersaBoard and the Med-Tec S-Type Baseplate are intended for use in positioning and re-positioning patients during radiation therapy procedures, and both boards are employed in clinically identical fashions. Therefore, it is reasonable to conclude that the VersaBoard manufactured by Bionix Development Corporation is substantially equivalent in all aspects to the S-Type Baseplate manufactured by Med-Tec, Inc.

The Bionix VersaBoard is also substantially equivalent to other similar patient positioning devices constructed from carbon fiber/epoxy/foam core composites. Two such devices are the Carbon Fiber Breast Board manufactured and legally sold by Med-Tec, Inc. of Orange City, Iowa (K974703), and the Max 2 Deluxe TorsoBoard manufactured and legally sold by Bionix Development Corp., Toledo, Ohio (K905007). Both the Carbon Fiber Breast Board from Med-Tec and the Max 2 Deluxe TorsoBoard from Bionix are intended to be used to accurately position and hold patients securely during a radiation therapy treatment regimen. Both of these devices are constructed from carbon fiber/epoxy/foam core composites that produce minimal attenuation to the radiotherapy beam, similar to that exhibited by the VersaBoard. These devices also contain or use simple mechanical interlock mechanisms to secure the boards to the treatment couch tabletop, and others that secure low-melt thermoplastic to the boards themselves, allowing the low-melt thermoplastic to be formed into a mask that contours to the patient and can be used to accurately position and hold the patient during the radiation therapy procedure. In like fashion, the VersaBoard is intended to accurately position and securely hold patients undergoing radiation therapy. The VersaBoard has similar mechanical interlock mechanisms to secure the board to the therapy couch tabletop and to low-melt thermoplastic, as described above.

The almost identical design, construction, materials, properties, performance and intended use of the VersaBoard to the Carbon Fiber Breast Board from Med-Tec and the Max 2 Deluxe TorsoBoard form Bionix Development Corp., both existing legally sold devices, prove the Bionix VersaBoard to be substantially equivalent to these devices as well.

Submitted by:

pho

James Huttner M.D., Ph.D. Vice President, New Product Development Bionix Development Corporation

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the logo is a stylized image of three birds in flight, stacked on top of each other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 0 4 2003

Re: K030051 James Huttner, M.D., Ph.D. Vice President, New Product Development Bionix Development Corporation 5154 Enterprise Blvd. TOLEDO OH 43612

Trade/Device Name: Versaboard Patient Positioning System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charge-particle radiation therapy system Regulatory Class: II Product Code: 90 IYE Dated: October 15, 2002 Received: January 6, 2003

Dear Dr. Huttner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

ices and Radiological Heal

CDRH Home | Search | A-Z Index | Feedback

Page of

510(k) Number (if known): {} 3 0 0 5

Device Name: VersaBoard Patient Positioning System

Indications for Use:

The VersaBoard patient positioning system developed and manufactured by Bionix Development Corporation, Toledo, Ohio, is intended to be used for the positioning and re-positioning of patients undergoing or receiving a course of external beam radiation therapy for the treatment of cancer and other diseases.

It is intended to be used by or under the direction of a licensed physician.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

(Posted July 1, 1998)

Back to the Indications for Use Page

(Optional Format 3-10-98)

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Dise.

Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K030

CDRH Home | Search | A-Z Index | Feedback | Accessib

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.