K Number
K024187
Date Cleared
2003-02-04

(47 days)

Product Code
Regulation Number
892.1000
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Spirit III TotalSENSE Cardiac Coil is a receive-only phased array RF coil, used for obtaining diagnostic images of the cardiovascular system in Magnetic Resonance Imaging Systems. The Spirit III TotalSENSE Cardiac Coil is designed for use with the Magnetom Trio 3.0T MRI system manufactured by Siemens Medical Solutions, Inc. The indications for use are the same as for standard MR Imaging.

The Spirit III TotalSENSE Cardiac Coil is designed to provide Magnetic Resonance Images of the cardiovascular system. The Spirit III TotalSENSE Cardiac Coil is designed for use with the Siemens Magnetom Trio 3.0T MRI scanner manufactured by Siemens Medical Solutions, Inc.

The indications for use are the same as for standard imaging:

The 3.0T MRI system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.

Device Description

The Spirit III TotalSENSE Cardiac Coil is an eight element receive only phased array coil. The coil is composed of a flexible top piece and a rigid bottom piece. The flexible top piece allows for imaging of patients of different sizes. The open, patient friendly design minimizes claustrophobic effects and maximizes patient comfort. The coil elements and accessory electronics are enclosed in flexible and rigid plastic housings, which are fire rated. All rigid plastic housing parts have a high impact and tensile strength.

AI/ML Overview

I am sorry, but the provided text does not contain information about the acceptance criteria or a study proving the device meets acceptance criteria for the Spirit III TotalSENSE Cardiac Coil. The document is a 510(k) premarket notification summary and a letter from the FDA determining substantial equivalence to predicate devices, focusing on the device's intended use, description, and comparison to existing technology. It does not include specific performance metrics, study designs, or data about the device's performance against predefined acceptance criteria.

{0}------------------------------------------------

KD24187

SUMMARY OF SAFETY AND EFFECTIVENESS

FEB 0 4 2003

1. Device Name:Magnetic Resonance Imaging Accessory
2. Proprietary Name:Spirit III TotalSENSE Cardiac Coil
3. Classification:Class II
4. Establishment Registration #:1529041
5. Manufacture Facility Location:USA Instruments, Inc.,1515 Danner DriveAurora, Ohio 44202, USATelephone: 330-562-1000; Fax: 330-562-1422.
6. Performance Standard:No applicable performance standards have beenissued under Section 514 of the Food, Drug andCosmetic Act.
7. Intended Use:The Spirit III TotalSENSE Cardiac Coil is a receive-only phased array RF coil, used for obtainingdiagnostic images of the cardiovascular system inMagnetic Resonance Imaging Systems. The Spirit IIITotalSENSE Cardiac Coil is designed for use withthe Magnetom Trio 3.0T MRI system manufacturedby Siemens Medical Solutions, Inc. The indicationsfor use are the same as for standard MR Imaging.
8. Device Description:The Spirit III TotalSENSE Cardiac Coil is an eightelement receive only phased array coil. The coil iscomposed of a flexible top piece and a rigid bottompiece. The flexible top piece allows for imaging ofpatients of different sizes. The open, patient friendlydesign minimizes claustrophobic effects andmaximizes patient comfort. The coil elements andaccessory electronics are enclosed in flexible andrigid plastic housings, which are fire rated. All rigidplastic housing parts have a high impact and tensilestrength.

Please turn over

. . . . . . .

{1}------------------------------------------------

Spirit III TotalSENSE Cardiac Coil ProductFeaturesComparison to predicate device or other 510(k)cleared products
Intended Use: imaging of thecardiovascular system-Similar to the Vision 5000 Torso Coil manufactured byUSA Instruments, Inc. (K013594).-Similar to the Insight Plus 9000 Phased Array Torsoand Pelvis Coil manufactured by USA Instruments, Inc.(K001209).
Indications for Use: Identical to routineMRI imaging.-Similar to the Vision 5000 Torso Coil manufactured byUSA Instruments, Inc. (K013594).-Similar to the Insight Plus 9000 Phased Array Torsoand Pelvis Coil manufactured by USA Instruments, Inc.(K001209).
Coil Enclosure Material:Flame Retardant PolyurethaneVinyl Coated EVA foamFlame Retardant Polycarbonate- Similar to the Mark 5000 Quadrature Shoulder coilmanufactured by USA Instruments, Inc. (K013854)
Coil Design: Receive-only phased arraycoil.-Similar to the Insight Plus 9000 Phased Array Torsoand Pelvis Coil manufactured by USA Instruments, Inc.(K001209).
Decoupling: Switching diode decoupling.Additional RF fuses incorporated.-Similar to the Insight Plus 9000 Phased Array Torsoand Pelvis Coil manufactured by USA Instruments, Inc.(K001209).
Prevention of RF Burns: Does not transmitRF power; decoupling isolates the coilelements from RF fields during RFtransmission; coil elements and circuitry areenclosed in a non-conductive housing.-Similar to the Insight Plus 9000 Phased Array Torsoand Pelvis Coil manufactured by USA Instruments, Inc.(K001209).
Radio Frequency Absorption: Coil is areceive only coil and does not transmit RFpower.-Similar to the Insight Plus 9000 Phased Array Torsoand Pelvis Coil manufactured by USA Instruments, Inc.(K001209).
Formation of Resonant Loop: Decouplingisolates the coil elements from RF fieldsduring RF transmission; length of cable andstiffness does not permit looping.-Similar to the Insight Plus 9000 Phased Array Torsoand Pelvis Coil manufactured by USA Instruments, Inc.(K001209).

9. Safety and Effectiveness

.

:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Christie Shumaker Manager, OA and Regulatory Affairs USA Instruments, Inc. 1515 Danner Drive AURORA OH 44202

FEB 0 4 2003

Re: K024187 Trade/Device Name: Spirit III TotalSENSE Cardiac Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: December 16, 2002 Received: December 19, 2002

Dear Ms. Shumaker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx. 1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): _ 长 ○ 2 41 87

Device Name: Spirit III TotalSENSE Cardiac Coil

Indications for Use: The Spirit III TotalSENSE Cardiac Coil is designed to provide Magnetic Resonance Images of the cardiovascular system. The Spirit III TotalSENSE Cardiac Coil is designed for use with the Siemens Magnetom Trio 3.0T MRI scanner manufactured by Siemens Medical Solutions, Inc.

Anatomic Regions: Cardiovascular system Nuclei Excited: Hydrogen

The indications for use are the same as for standard imaging:

The 3.0T MRI system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) OR Over-The-Counter Use Prescription Use (Per 21 CFR 801.109) (Optional Format 1-2-96)

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.