K Number
K024067

Validate with FDA (Live)

Date Cleared
2003-01-15

(36 days)

Product Code
Regulation Number
866.5775
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Rheumatoid Factor assay is used for the quantitation of rheumatoid factor in human serum. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.

Device Description

Rheumatoid Factor is an in vitro diagnostic assay for the quantitative determination of rheumatoid factor in human serum. The RF assay is a latex enhanced immunoturbidimetric assay that involves an antigen-antibody reaction between the rheumatoid factor in the sample and the denatured human IgG, which has been adsorbed to latex particles. The resulting agglutination is detected as an absorbance change (572 nm), with the magnitude of the change being proportional to the quantity of RF in the sample. The actual concentration is then determined by interpolation from a calibration curve prepared from calibrators of known concentration.

AI/ML Overview

The Abbott Laboratories Rheumatoid Factor assay is an in vitro diagnostic device for the quantitative determination of rheumatoid factor in human serum. The study described aimed to demonstrate the substantial equivalence of this device to a legally marketed predicate device, the Roche Diagnostics Corp. Tina-quant RF II assay.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly derived from the comparison to the predicate device and the expected performance of such assays. The study primarily focused on correlation and precision, as well as the assay range and sensitivity.

Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Abbott Rheumatoid Factor assay on AEROSET® System)
Correlation with Predicate Device (Roche Diagnostics Corp. Tina-quant RF II assay on Hitachi 717 Analyzer)Strong positive correlation (e.g., correlation coefficient close to 1)Correlation Coefficient: 0.996 Slope: 1.14 Y-intercept: 2.14 IU/mL
Precision (Total %CV)Acceptable variability for clinical use across different levels (e.g., within-run, between-run, between-day)Level 1: Total %CV ranged from 4.6% to 7.5% Level 2: Total %CV ranged from 3.6% to 4.3% Level 3: Total %CV ranged from 1.2% to 2.0%
Assay RangeComparable to predicate device or suitable for clinical use4.8 to 200.0 IU/mL
Limit of Quantitation (Sensitivity)Clinically relevant and comparable to predicate device7.71 IU/mL

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the comparative performance studies (test set). It mentions "comparative performance studies were conducted" and "On the AEROSET System," but specific numbers of patient samples are not provided. The data provenance is not explicitly stated (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable and not provided in the document. The study involves comparing the analytical performance of two in vitro diagnostic (IVD) assays by measuring quantitative results. The "ground truth" in this context refers to the actual concentration of rheumatoid factor, which is determined by the assay itself, rather than expert interpretation of images or clinical outcomes that require expert consensus.

4. Adjudication Method for the Test Set

This information is not applicable and not provided in the document. As mentioned above, this study measures quantitative concentrations, not subjective interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study design is typically used for diagnostic imaging devices where human readers interpret images, sometimes with the assistance of AI, to determine diagnostic accuracy. This document describes an in vitro diagnostic assay that produces quantitative results, not an imaging device or a system requiring human interpretation.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, this study inherently represents a standalone performance. The Abbott Rheumatoid Factor assay, like other IVD assays, operates as an algorithm/instrument to produce a quantitative result. Its performance is measured independently based on its analytical capabilities (correlation, precision, range, sensitivity) with the input sample. There is no human-in-the-loop component for the measurement process itself.

7. The Type of Ground Truth Used

The ground truth for the comparison was the results obtained from the predicate device (Roche Diagnostics Corp. Tina-quant RF II assay on the Hitachi 717 Analyzer). For precision studies, the ground truth was derived from the known concentrations of control materials.

8. The Sample Size for the Training Set

This information is not applicable and not provided. The Abbott Rheumatoid Factor assay is an immunoassay, not a machine learning or AI-driven algorithm that requires a "training set" in the conventional sense for model development. Its calibration curve is established using calibrators, which are distinct from a training set for an AI model.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable. As explained in point 8, there is no "training set" in the context of an AI/ML model for this type of immunoassay. The assay uses calibrators of known concentrations to establish its calibration curve, which allows for the interpolation of actual concentrations from measured absorbance changes. The ground truth for these calibrators would typically be established through highly accurate and traceable reference methods.

{0}------------------------------------------------

510(k) Summary

Submitter's Name/Address Abbott Laboratories 1920 Hurd Drive Irving, TX 75038

Contact Person Linda Morris Senior Regulatory Specialist MS 1-8 Regulatory Affairs (972) 518-6711 Fax (972) 753-3367

Date of Preparation of this Summary:December 5, 2002
Device Trade or Proprietary Name:Rheumatoid Factor
Device Common/Usual Name or Classification Name:Rheumatoid Factor
Classification Number/Class:DHR, Class II

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K024067

Test Description:

Rheumatoid Factor is an in vitro diagnostic assay for the quantitative determination of rheumatoid factor in human serum. The RF assay is a latex enhanced immunoturbidimetric assay that involves an antigen-antibody reaction between the rheumatoid factor in the sample and the denatured human IgG, which has been adsorbed to latex particles. The resulting agglutination is detected as an absorbance change (572 nm), with the magnitude of the change being proportional to the quantity of RF in the sample. The actual concentration is then determined by interpolation from a calibration curve prepared from calibrators of known concentration.

Rheumatoid Factor 510(k) 1/9/2003 RF Summary R2.doc

Section III Page 1

7

{1}------------------------------------------------

Substantial Equivalence:

The Rheumatoid Factor assay is substantially equivalent to the Roche Diagnostics Corp. Tina-quant RF II assay (K002609) on the Hitachi® 717 Analyzer.

Both assays yield similar Performance Characteristics.

Similarities:

  • . Both assays are in vitro immunoassays.
  • Both assays can be used for the quantitative determination of rheumatoid factor. .
  • . Both assays yield similar clinical results.
  • Both assays are based on the measurement of the agglutination following . antigen-antibody reaction.
  • . Human serum is a suitable specimen for both assays.

Differences:

  • . There is a difference between the assay range.
  • . Suitable specimens for the Roche Diagnostics Corp. Tina-quant RF II assay include plasma specimens.

Intended Use:

The Rheumatoid Factor assay is used for the quantitation of rheumatoid factor in human serum.

Section II Page 2

Image /page/1/Picture/16 description: The image shows the number 000013 in a bold, sans-serif font. The numbers are black against a white background. The numbers are evenly spaced and aligned horizontally.

{2}------------------------------------------------

Performance Characteristics:

Comparative performance studies were conducted using the AEROSET® System. The Rheumatoid Factor assay method comparison yielded acceptable correlation with the Roche Diagnostics Corp. Tina-quant RF II assay on the Hitachi 717 Analyzer. On the AEROSET System, the correlation coefficient = 0.996, slope = 1.14, and the Y-intercept = 2.14 IU/mL. Precision studies were conducted using the Rheumatoid Factor assay. Within-run, between-run, and between-day studies were performed using three levels of control material. On the AEROSET System, the total %CV for Level 1 ranged from 4.6% to 7.5%, Level 2 ranged from 3.6% to 4.3%, and Level 3 ranged from 1.2% to 2.0%. The Rheumatoid Factor assay range is 4.8 to 200.0 IU/mL. The limit of quantitation (sensitivity) of the Rheumatoid Factor assay is 7.71 IU/mL on the AEROSET System. These data demonstrate that the performance of the Rheumatoid Factor assay is substantially equivalent to the performance of the Roche Diagnostics Corp. Tina-quant RF II assay on the Hitachi 717 Analyzer.

Conclusion:

The Rheumatoid Factor assay is substantially equivalent to the Roche Diagnostics Corp. Tina-quant RF II assay on the Hitachi 717 Analyzer as demonstrated by results obtained in the studies.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 1 5 2003

Ms. Linda Morris Senior Regulatory Specialist ADD Regulatory Affairs Abbott Laboratories 1920 Hurd Drive Irving, Texas 75038

Re: K024067

Trade/Device Name: Rheumatoid Factor Regulation Number: 21 CFR § 866.5775 Regulation Name: System, Test, Rheumatoid Factor Regulatory Class: II Product Code: DHR Dated: December 5, 2002 Received: December 10, 2002

Dear Ms. Morris:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): HOAYOLe7

Device Name: Rheumatoid Factor ------------------------------------------------------------------------------------------------------------------------------------------------------------

Indications For Use:

The Rheumatoid Factor assay is used for the quantitation of rheumatoid factor in human serum. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use_ OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)

(Optional Format 1-2-96)

Af. Reeves for J. Bautista

linical Laborato

Rheumatoid Factor 510(k) 1 2/5/2002 RF_5_R1.doc

§ 866.5775 Rheumatoid factor immunological test system.

(a)
Identification. A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.(b)
Classification. Class II (performance standards).