(70 days)
The Welch Allyn Vital Signs Monitor is intended for monitoring of blood pressure, pulse rate, temperature and oxygen saturation (SpO2) of adult and pediatric patients. The device is not designed, sold or intended for use except as indicated.
- The Welch Allyn Vital Signs Monitor is not designed for use with neonates. To ensure pediatric blood pressure accuracy and safety, note that the Welch Allyn Small Child cuffs are the smallest cuffs approved for use with young children and infants. The circumference of the child's arm must fit within the range markings on the cuff.
- The Welch Allyn Vital Signs Monitor should not be used on patients who are linked to heart/lung machines.
The Welch Allyn Vital Signs Monitor is designed to non-invasively and automatically measure systolic and diastolic blood pressure, pulse rate, temperature and oxygen saturation (SpO2) for adult and pediatric patients. THE WELCH ALLYN VITAL SIGNS MONITOR IS NOT INTENDED TO BE USED ON NEONATAL PATIENTS. All blood pressure, pulse, temperature and SpO2 values are displayed on large, easy-to-read displays, and may be printed via an integrated thermal printer, as desired. The rechargeable battery and wide variety of mounting accessories make the Welch Allyn Vital Signs Monitor convenient for many locations. The operator may choose any combination of simultaneous measurement modalities. This flexibility, combined with features such as programmable alarms and automatic blood pressure cycles, makes the Welch Allyn Vital Signs Monitor ideal for a wide variety of patient monitoring needs.
The provided document describes the Welch Allyn Vital Signs Monitor, an updated version of a previously cleared device, primarily focusing on the change of the SpO2 module.
Here's an analysis of the acceptance criteria and the study justifying the device's performance, based on the provided text:
1. Table of Acceptance Criteria and the Reported Device Performance:
The document mentions "Accuracy test" as a validation test. However, it does not explicitly state specific numerical acceptance criteria (e.g., within X mmHg for blood pressure, within Y% for SpO2) or the quantitative reported device performance against such criteria. The submission states that the device "passed all the safety and effectiveness testing," implying that any implicit acceptance criteria for these tests were met.
| Metric (Implied) | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| SpO2 Accuracy | (Specific numerical accuracy not provided, but implicitly met Nellcor standards) | "The full protocols were tested on the VSM with MP506 and these tested were used to show that Welch Allyn implemented correctly the new MP506 SpO2 module." and "The Vital Signs Monitor with MP506 Pulse Oximeter OEM module has passed all the safety and effectiveness testing." |
| Blood Pressure Accuracy | (Specific numerical accuracy not provided) | "The Vital Signs Monitor with MP506 Pulse Oximeter OEM module has passed all the safety and effectiveness testing." |
| Pulse Rate Accuracy | (Specific numerical accuracy not provided) | "The Vital Signs Monitor with MP506 Pulse Oximeter OEM module has passed all the safety and effectiveness testing." |
| Temperature Accuracy | (Specific numerical accuracy not provided) | "The Vital Signs Monitor with MP506 Pulse Oximeter OEM module has passed all the safety and effectiveness testing." |
| Software System Level Functional Tests | All functionalities perform as specified. | Passed |
| Pre-scan IEC60601-1-2 Testing | Compliance with electrical safety and electromagnetic compatibility standards. | Passed |
| Usage Test | Device functions as intended during typical use. | Passed |
| Basic Functional Tests | All fundamental device functions operate correctly. | Passed |
2. Sample Size Used for the Test Set and the Data Provenance:
The document does not explicitly state the sample sizes used for the "Accuracy test" or "Usage test" for the Welch Allyn Vital Signs Monitor with the new MP506 module. It also does not specify the provenance (e.g., country of origin, retrospective/prospective) of any data collected during these tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
The document does not mention the use of experts to establish ground truth for any of the described tests. The testing described appears to be focused on device functionality, safety, and compliance with manufacturing/module-specific standards rather than a clinical performance study requiring expert adjudication of patient data for ground truth.
4. Adjudication Method for the Test Set:
No adjudication method is mentioned, as there is no indication that expert review of patient data was part of the testing described.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:
No MRMC comparative effectiveness study is mentioned in the provided text. The submission focuses on demonstrating substantial equivalence to a predicate device through verification and validation of changes (specifically the SpO2 module).
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done:
The "Accuracy test" and "Nellcor compliance testing" performed on the SpO2 module can be considered standalone performance evaluations of that component. The overall device testing (software functional, basic functional) also assesses the device's inherent performance. There's no mention of a human-in-the-loop performance study in the context of its accuracy, but rather general "usage test" which might involve human interaction.
7. The Type of Ground Truth Used:
For the SpO2 module, the "Nellcor compliance testing" implies that the ground truth would be based on the established gold standards and methodologies used by Nellcor for validating pulse oximetry, which typically involves controlled desaturation studies against arterial blood gas measurements (co-oximetry). However, this is inferred, not explicitly stated regarding the data collected for this submission. For other metrics like blood pressure, temperature, and pulse rate, the ground truth for an "Accuracy test" would typically involve comparison against a reference standard device. The document does not explicitly detail these ground truth methods.
8. The Sample Size for the Training Set:
The document describes verification and validation testing for the updated device, not the development of an algorithm that would require a training set. The device is a measurement instrument, and the core technology for SpO2, blood pressure, etc., relies on established physical principles and algorithms, not a machine learning model that requires a distinct training/test split in the way AI/ML devices do. Therefore, a "training set" is not applicable in the context described.
9. How the Ground Truth for the Training Set Was Established:
As there is no mention of a training set for an AI/ML algorithm, this question is not applicable.
{0}------------------------------------------------
FEB 1 2 2003
KO2400
Section 1 - Revised 510(k) Summary
[As described in CFR 807.92]
Submitter's Name & Address David Klementowski 4341 State Street Road Skaneateles Falls, NY 13153 Phone: (315) 685-4100 ext. 4133 Fax: (315) 685-7889
B. Other Contact Persons
Chuck Setzer Quality Assurance and Regulatory Affairs Manager (Arden Facility), 95 Old Shoals Road Arden, NC 28704 Phone: (828) 684-4895 ext. 5262 Fax: (828) 687-1002
Brett Pierce Project Manager 95 Old Shoals Road Arden, NC 28704 Phone: (828) 684-4895 ext. 5248 Fax: (828) 687-1002
C. Device Name
Proprietary Name: Welch Allyn Vital Signs Monitor Common Name: Vital Signs Measurement Device Classification Name: Noninvasive Blood Pressure Measurement System
D. Device Classification Class II (CFR21 870.1130 Noninvasive Blood Pressure Measurement System).
E. Predicate Device Welch Allyn Vital Signs Monitor Welch Allyn, Inc. 510(K) Document Control Number K951193
F. Description of the Device
The Welch Allyn Vital Signs Monitor is designed to non-invasively and automaticallymeasure systolic and diastolic blood pressure, pulse rate, temperature and oxygen saturation (SpO2) for adult and pediatric patients. THE
{1}------------------------------------------------
WELCH ALLYN VITAL SIGNS MONITOR IS NOT INTENDED TO BE USED ON NEONATAL PATIENTS. All blood pressure, pulse, temperature and SpO2 values are displayed on large, easy-to-read displays, and may be printed via an integrated thermal printer, as desired.
The rechargeable battery and wide variety of mounting accessories make the Welch Allyn Vital Signs Monitor convenient for many locations. The operator may choose any combination of simultaneous measurement modalities. This flexibility, combined with features such as programmable alarms and automatic blood pressure cycles, makes the Welch Allyn Vital Signs Monitor ideal for a wide variety of patient monitoring needs.
The Welch Allyn Vital Signs Monitor is intended for use in a wide variety of health care settings. This includes hospital departments, as well as patient transport within the hospital environment. The Welch Allyn Vital Signs Monitor is also intended for use in alternate care settings, such as physician offices, freestanding ambulatory care and surgery centers, health clinics and nursing homes. The Welch Allyn Vital Signs Monitor may also be used during patient transport within any of these alternate care environments.
The Welch Allyn Vital Signs Monitor is not intended for the monitoring of patients during transport external to the health care environment (eg. ambulance, helicopter transport). The Welch Allyn Vital Signs Monitor is not intended for use in environments, which are not supervised by a health care practitioner.
H. Intended Use
The Welch Allyn Vital Signs Monitor is intended for monitoring of blood pressure, pulse rate, temperature and oxygen saturation (SpO2) of adult and pediatric patients. The device is not designed, sold or intended for use except as indicated.
The Welch Allyn Spot Check Device is not designed for use with neonates. To ensure pediatric blood pressure accuracy and safety, note that the Welch Allyn Small Child cuffs are the smallest cuffs approved for use with young children and infants. The circumference of the child's arm must fit within the range markings on the cuff.
The Welch Allyn Spot Check Device should not be used on patients who are linked to heart/lung machines.
I. Technological Similarities and Differences
The Welch Allyn Vital Signs Monitor and the Welch Allyn Vital Signs Monitor with MP506 Pulse Oximeter OEM Module are essentially the same device. Welch Allyn has added the new SpO2 module from Nellcor (The MP506 (same as in the N-595 Pulse Oximeter mad ey Nellcor)) to the Vital Signs Monitor. This is an upgraded SpO2 module that was cleared by the FDA 510(k) K012891. The Vital
{2}------------------------------------------------
Signs Monitor was cleared by the FDA 510(k) K002530.
In regards to the change in the SpO2 module for to the Vital Signs Monitor, the basic fundamental technology is the same. A sensor is placed on the finger, toe, ear lobe. or forehead and a microprocessor interprets the amount of red and infrared light that is passing through the tissue and calculates Specific Oxygen (SpO2) of the blood.
J. Testing for Equivalence
Equivalence to the legally market predicate device was determine through the means of verification and validation testing. The following testing was conducted on the VSM with MP506:
Verification: 3 test were conducted on the VSM 506
-
- Software system level functional tests
-
- Pre-scan IEC60601-1-2 testing at a third party test house
-
- Nellcor compliance testing (whereby Nellcor tested our VSM with their MP506 to their standards).
The full protocols were tested on the VSM with MP506 and these tested were used to show that Welch Allyn implemented correctly the new MP506 SpO2 module.
Validation: 4 test were conducted on the VSM with MP506
-
- Software system level functional tests
-
- Accuracy test
-
- Usage test
-
- Basic functional tests
The full protocols were tested on the VSM with MP506 and these tests were used to show that the Welch Allyn VSM meet the customer requirements.
K. 510(k) Summary Conclusions
Based on the following, it is our conclusion that the new Welch Allyn Vital Signs Monitor with MP506 Pulse Oximeter OEM module is substantially equivalent to the Welch Allyn Vital Signs Monitor.
- The features and intended use has not changed ●
- The Vital Signs Monitor with MP506 Pulse Oximeter OEM module has . passed all the safety and effectiveness testing.
- The fundamental technology of the Welch Allyn Vital Signs Monitor with MP506 Pulse Oximeter OEM module has not changed from that of the Welch Allyn Vital Signs Monitor cleared under K951193.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or movement. The profiles are stacked on top of each other, creating a sense of unity. Encircling the image is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
FEB 1 2 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David Klementowski Corporate Regulatory Affairs Manager Welch Allyn, Incorporated 4341 State Street Road Skaneatels Falls, New York 13153
Re: K024005
Trade/Device Name: Welch Allyn Vital Signs Monitor Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQN, DQA Dated: January 29, 2003 Received: January 30, 2003
Dear Mr. Klementowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Klementowski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Runner
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Indications for Use Statement.
| 510(k) Number: | Unknown |
|---|---|
| Device Name: | Welch Allyn Vital Signs Monitor |
| Indications for use: | The Welch Allyn Vital Signs Monitor is intended formonitoring of blood pressure, pulse rate, temperature andoxygen saturation (SpO2) of adult and pediatric patients.The device is not designed, sold or intended for use exceptas indicated.* The Welch Allyn Vital Signs Monitor is not designed foruse with neonates. To ensure pediatric blood pressureaccuracy and safety, note that the Welch Allyn Small Childcuffs are the smallest cuffs approved for use with youngchildren and infants. The circumference of the child's armmust fit within the range markings on the cuff.* The Welch Allyn Vital Signs Monitor should not be usedon patients who are linked to heart/lung machines. |
(Please Do Not Write Below This Line - Continue On Another Page If Needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_
Or Over-The-Counter Use
(Per 21 CFR 801.109)

y, General Hospital,
510(k) Number:
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).