(302 days)
The Methadone Metabolite Immunoassay is intended to be used for the qualitative and semi-quantitative determination of the presence of Methadone Metabolite (2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine or EDDP) in human urine at cutoffs of 300 and 1000 ng/mL. The semi-quantitative range of the assay is 31-2000 ng/mL. The assay provides a simple and rapid analytical screening procedure to detect methadone metabolite in human urine.
The subject device, the DR1 Methadone Metabolite Enzyme Assay, is a ready-to-use, liquid homogeneous enzyme immunoassay. The assay uses specific antibodies that detect EDDP in human urine without cross-reactivity to the parent drug, methadone. The assay is based on the competition between a drug labeled with glucose-6-phosphate dehydrogenase (G6PDH), and free drug from the urine sample for a fixed amount of specific antibody binding sites. In the absence of free drug from the sample, the specific antibody binds the drug labeled with G6PDH and causes a decrease in enzyme activity. This phenomenon creates a direct relationship between drug concentration in urine and the enzyme activity. The enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH. The subject DRI® Methadone Metabolite Enzyme Assay utilizes a new monoclonal antibody and enzyme-conjugate, with increased reagent stability and calibrator and control separations.
Here's a breakdown of the acceptance criteria and the study information based on the provided text:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Accuracy: Agreement with predicate device | 100% agreement (Predicate device: 57 true positives, 54 true negatives) |
| Accuracy: Agreement with GC/MS | 95% agreement (69 true positives, 73 true negatives) |
| Total Imprecision (Predicate Device) | ≤ 0.8% CV across 4 levels (0, 300, 1000, 2000 ng/mL) |
| Total Imprecision (Subject Device) | ≤ 0.9% CV across 6 levels (150, 300, 500, 750, 1000, 1250 ng/mL) |
| Total Imprecision (Subject Device, dose mode) | ≤ 5.7% CV |
| Assay Sensitivity | Approximately 31 ng/mL |
| Intra- and Inter-assay Reproducibility | ≤ 5.7% |
| Linearity | Up to 2000 ng/mL |
| Adequate separation for 300 ng/mL cutoff | Demonstrated for specimens between 150 and 500 ng/mL |
| Adequate separation for 1000 ng/mL cutoff | Demonstrated for specimens between 750 and 1250 ng/mL |
| Cross-reactivity: Methadone | Does not cross-react |
| Interference: Common endogenous substances, urinary pH, unrelated compounds | No deleterious effect |
| Interference: Potentially cross-reacting compounds | No deleterious effect |
Study Details
-
Sample size used for the test set and the data provenance:
- Accuracy (comparison to predicate device): 111 samples (57 true positives, 54 true negatives). Data provenance is not specified (e.g., country of origin, retrospective/prospective).
- Accuracy (comparison to GC/MS): 142 samples (69 true positives, 73 true negatives). Data provenance is not specified.
- Imprecision studies: The number of unique samples is not explicitly stated. Instead, it refers to "4 levels of analyte concentration" for the predicate and "6 levels of analyte concentration" for the subject device, suggesting controlled spiked samples or pooled samples at specific concentrations rather than a large set of varied patient samples for this particular aspect.
- Sensitivity, reproducibility, linearity, clinical accuracy: Established through "a series of laboratory and clinical studies." The specific sample sizes for each of these broader categories are not detailed beyond the accuracy numbers mentioned above. The type of study (retrospective or prospective) is not explicitly stated, but "clinical studies" for accuracy suggest the use of patient samples.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not mentioned. The ground truth for accuracy was established using a "commercial EIA Assay" (for the predicate comparison) and the "GC/MS reference method" (for the subject device comparison). These are analytical reference methods, not expert human readers.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. The ground truth was established by laboratory reference methods (commercial EIA and GC/MS), not through human adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This device is an in vitro diagnostic (IVD) immunoassay, which is an automated or semi-automated laboratory test. It does not involve human readers interpreting images or data where AI assistance would be relevant for "improving" reader performance.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, the performance data presented (accuracy, imprecision, sensitivity, reproducibility, linearity, specificity) represent the standalone performance of the DRI® Methadone Metabolite Enzyme Assay. As an IVD, it is designed to operate primarily as an algorithm/assay without human interpretive input for the final result (though human oversight, calibration, and QC are essential).
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Analytical Reference Methods:
- For the predicate device comparison: a "commercial EIA Assay."
- For the subject device comparison: "GC/MS reference method" (Gas Chromatography/Mass Spectrometry), which is considered the gold standard for confirmatory drug testing.
- Analytical Reference Methods:
-
The sample size for the training set:
- Not explicitly stated. The document describes performance characteristics and studies (e.g., sensitivity, reproducibility, linearity, clinical accuracy), but it does not detail a separate "training set" as one would typically discuss for machine learning models. For an immunoassay, the "training" analogous process involves assay development, optimization, and validation using various samples to establish performance parameters, but not in the sense of a distinct, labeled training dataset for an AI algorithm.
-
How the ground truth for the training set was established:
- Not applicable in the context of a "training set" for an AI algorithm. The development and optimization of the immunoassay likely involved using known spiked samples, clinical samples characterized by reference methods (like GC/MS), and internal controls to optimize reagent concentrations, antibody specificity, and overall assay performance.
{0}------------------------------------------------
AUG 2 7 2003 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
K023617 510(k) Number:
Applicant Information:
| Date Prepared: | July 16, 2003 |
|---|---|
| Applicant: | Microgenics Corporation46360 Fremont BoulevardFremont, CA 94538 |
| Contact Person: | David C Casal, Ph.D.Vice-President, Clinical, Regulatory and Quality Affairs |
| Telephone: | (510) 979-5012 |
| Facsimile: | (510) 979-5212 |
| E-mail: | dcasal@microgenics.com |
Device Information:
| Classification/Name: | Clinical Toxicology Test System, Methadone Test SystemClass II per 21 CFR §862.3620, Product Code DJR |
|---|---|
| Trade Name: | DRI® Methadone Metabolite Enzyme Assay |
| Common Name: | Methadone Metabolite Assay |
Legally Marketed Predicate Device:
The subject DRI® Methadone Metabolite Enzyme Assay is substantially equivalent to the previously cleared (K931780) predicate DR10 Methadone Metabolite Enzyme Assay cleared. The subject DRT® Methadone Metabolite Enzyme Assay is identical or similar to its predicate in terms of intended use, method principle, device components, risk to the patient, and clinical performance.
Device Description:
The subject device, the DR1 Methadone Metabolite Enzyme Assay, is a ready-to-use, liquid homogeneous enzyme immunoassay. The assay uses specific antibodies that detect EDDP in human urine without cross-reactivity to the parent drug, methadone. The assay is based on the competition between a drug labeled with glucose-6-phosphate dehydrogenase (G6PDH), and free drug from the urine sample for a fixed amount of specific antibody binding sites. In the absence of free drug from the sample, the specific antibody binds the drug labeled with G6PDH and causes a decrease in enzyme activity. This phenomenon creates a direct relationship between drug concentration in urine and the enzyme activity. The enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH. The subject DRI® Methadone Metabolite Enzyme Assay utilizes a new monoclonal antibody and enzyme-conjugate, with increased reagent stability and calibrator
{1}------------------------------------------------
and control separations.
Intended Use:
The Methadone Metabolite Immunoassay is intended to be used for the qualitative and semi-quantitative determination of the presence of Methadone Metabolite (2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine or EDDP) in human urine at cutoffs of 300 and 1000 ng/mL. The semi-quantitative range of the assay is 31-2000 ng/mL. The assay provides a simple and rapid analytical screening procedure to detect methadone metabolite in human urine.
Comparison to Predicate Device:
The subject DRI® Methadone Metabolite Enzyme Assay is essentially identical and therefore substantially equivalent to the predicate device. A comparison of salient descriptive and performance characteristics for the devices amply demonstrates that the subject device is substantially equivalent to the predicate device in form and function.
| Device Name | Predicate Methadone Metabolite | Subject Methadone Metabolite |
|---|---|---|
| Assay (K931780) | Assay | |
| Indicationsfor Use | "This homogeneous methadonemetabolite enzyme immunoassay isintended for the qualitative andsemiquantitative determination of themethadone metabolite, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine(EDDP), in human urine." | The Methadone MetaboliteImmunoassay is intended to be usedfor the qualitative and semi-quantitative determination of thepresence of Methadone Metabolite (2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine or EDDP) inhuman urine at cutoffs of 300 and1000 ng/mL. The semi-quantitativerange of the assay is 31-2000 ng/mL.The assay provides a simple and rapidanalytical screening procedure todetect methadone metabolite inhuman urine. |
| MethodPrinciple | The assay is based on the competitionbetween a drug labeled with glucose-6-phosphate dehydrogenase (G6PDH),and free drug from the urine samplefor a fixed amount of specificantibody binding sites. In the absenceof free drug from the sample, thespecific antibody binds the druglabeled with G6PDH and causes adecrease in enzyme activity. Thisphenomenon creates a directrelationship between drugconcentration in urine and the enzyme | The assay is based on the competitionbetween a drug labeled with glucose-6-phosphate dehydrogenase (G6PDH),and free drug from the urine samplefor a fixed amount of specific antibodybinding sites. In the absence of freedrug from the sample, the specificantibody binds the drug labeled withG6PDH and causes a decrease inenzyme activity. This phenomenoncreates a direct relationship betweendrug concentration in urine and theenzyme activity. The enzyme activity |
Comparison of the Predicate (K931780) and Subject (K023617) Methadone Metabolite Assays
{2}------------------------------------------------
| Device Name | Predicate Methadone MetaboliteAssay (K931780) | Subject Methadone MetaboliteAssay |
|---|---|---|
| activity. The enzyme activity isdetermined spectrophotometrically at340 nm by measuring its ability toconvert nicotinamide adeninedinucleotide (NAD) to NADH. | is determined spectrophotometricallyat 340 nm by measuring its ability toconvert nicotinamide adeninedinucleotide (NAD) to NADH. | |
| Components | - Antibody/Substrate Reagent- Enzyme Conjugate Reagent | - Antibody/Substrate Reagent- Enzyme Conjugate Reagent |
| Risk topatient | In vitro device, positive results must beconfirmed by GC/MS. | In vitro device, positive results must beconfirmed by GC/MS. |
| ClinicalPerformance | Accuracy: %Agreement with acommercial EIA Assay was 100% (57true positives, 54 true negatives). | Accuracy: %Agreement with GC/MSwas 95% (69 true positives, 73 truenegatives). |
| Total Imprecision: Percent rate CVsacross 4 levels of analyte concentration(0 ng/mL, 300 ng/mL, 1000 ng/mL,2000 ng/mL) were ≤ 0.8%. | Total Imprecision: Percent rate CVsacross 6 levels of analyte concentration(150 ng/mL, 300 ng/mL, and 500ng/mL, 750 ng/mL, 1000 ng/mL, and1250 ng/mL) were ≤ 0.9%. In dosemode, the %CV ≤ 5.7% |
Summary of Non-Clinical and Clinical Data:
Performance characteristics for the subject DRI® Methadone Metabolite Enzyme Assay, including sensitivity, reproducibility, linearity, and clinical accuracy, were established in a series of laboratory and clinical studies. These studies demonstrated that the assay sensitivity is approximately 31 ng/mL, intra- and inter-assay reproducibility across the dynamic range of the assay is less than or equal to 5.7%, and that the assay is linear up to 2000 ng/mL. Additional studies also showed that there was adequate separation offered for specimens between 150 and 500 ng/mL using the 300 ng/mL control (cutoff) and between 750 and 1250 ng/mL for the 1000 ng/mL control (cutoff).
Method comparison studies show that methadone metabolite concentrations obtained using the subject DR10 Methadone Metabolite Enzyme Assay and the GC/MS reference method are 95% concordant. Specificity testing demonstrated that the DR10 Methadone Metabolite Enzyme Assay does not cross-react with methadone nor is the assay result deleteriously by common endogenous substances, variations in urinary pH levels, structurally unrelated pharmaceutical compounds, or potentially cross-reacting compounds.
Summary:
Based upon the product performance characteristics, intended use, and technical information provided in this premarket notification, the DR10 Methadone Metabolite Enzyme Assay has been shown to be substantially equivalent to the currently marketed predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized lines forming its body and wings.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Lakshmi Anne, Ph.D. Director, Product Development Microgenics Corporation 46360 Fremont Boulevard Fremont, California 94538
AUG 2 7 2003
Re: K023617
Trade/Device Name: DRI® Methadone Metabolite Enzyme Assay Regulation Number: 21 CFR & 862.3620 Regulation Name: Methadone Test System Regulatory Class: II Product Code: DJR, DJR, DLJ, LAS Dated: May 29, 2003 Received: May 30, 2003
Dear Dr. Anne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K023617
Device name: DRJ® Methadone Metabolite Enzyme Assay
Indications for Use:
4
The Methadone Metabolite Immunoassay is intended to be used for the qualitative and semi-quantitative determination of the presence of Methadone Metabolite (2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine or EDDP) in human urine at cutoffs of 300 and 1000 ng/mL. The semi-quantitative range of the assay is 31-2000 ng/mL. The assay provides a simple and rapid analytical screening procedure to detect methadone metabolite in human urine.
Ian Cooper
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K023617
PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use √
Over-the Counter Use
(per 21 CFR §801.109
(Optional Format 1-2-96)
б
2
§ 862.3620 Methadone test system.
(a)
Identification. A methadone test system is a device intended to measure methadone, an addictive narcotic pain-relieving drug, in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of methadone use or overdose and to determine compliance with regulations in methadone maintenance treatment.(b)
Classification. Class II (special controls). A methadone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).