(98 days)
Not Found
Not Found
No
The provided text does not contain any mention of AI, ML, or related terms, nor does it describe functionalities typically associated with these technologies in medical imaging (e.g., automated analysis, segmentation, or diagnosis).
No.
The intended use clearly states "Diagnostic ultrasound imaging," indicating its purpose is for diagnosis, not therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states "Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application: Wounds".
No
The intended use clearly describes high-resolution ultrasound imaging, which requires hardware (an ultrasound transducer and processing unit) to acquire the images. The summary does not mention the device being software that processes images from an existing ultrasound machine.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The intended use and description clearly indicate that this device uses ultrasound imaging to visualize wounds directly on the human body. This is an in vivo (within the living body) diagnostic method, not an in vitro method.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples, reagents, or laboratory procedures, which are hallmarks of IVD devices.
Therefore, this device falls under the category of medical imaging devices used for diagnostic purposes directly on the patient.
N/A
Intended Use / Indications for Use
High resolution ultrasound imaging of wounds such as:
- Clean, closed surgical incisions
- Skin graft donor sites
- Superficial wounds such as abrasion, skin tears, and blisters
- Chafed skin or irritated skin
- Ulcers-cavernous and non-cavernous
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Wounds. In conjunction with Wendelken / Pope scanning method and contact media.
Product codes
90 IYO, ITX
Device Description
Hudson 2020/2040/2060 Ultrasound Scanners with Transducer Model Numbers 402198 (Hudson 2020/2040 Scanners), 410054 (Hudson 2020/2040 Scanners), and 410503 (Hudson 2060 Scanners).
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Ultrasound
Anatomical Site
Wounds (Clean, closed surgical incisions; Skin graft donor sites; Superficial wounds such as abrasion, skin tears, and blisters; Chafed skin or irritated skin; Ulcers-cavernous and non-cavernous)
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Not Found
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1560 Ultrasonic pulsed echo imaging system.
(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird in flight, composed of three curved lines.
Public Health Service
DEC 11 2002
Food and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850
Dr. Martin E. Wendelken Hudson Diagnostic Imaging, LLC 610 Boulevard · ELMWOOD PARK NJ
Re: K022928
Trade Name: Hudson 2020/2040/2060 Ultrasound Scanners Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYO and ITX Dated: November 6, 2002 Received: November 12. 2002
Dear Dr. Wendelken:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Hudson 2020/2040/2060 Ultrasound Scanners, as described in your premarket notification:
Transducer Model Number
402198 (Hudson 2020/2040 Scanners) 410054 (Hudson 2020/2040 Scanners) 410503 (Hudson 2060 Scanners)
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device
1
Page 2 - Dr. Wendelken
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G. (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address **http://www.fda.gov/cdrh/dsmamain.html` .
2
Page 3 - Dr. Wendelken
.
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212. . . . . 11.11 .
Sincerely yours,
David C. Hegpison
for Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
3
Page 1
510(k) Number (if known):
Device Name: Hudson 2020 Ultrasound Scanner
Indications For Use:
High resolution ultrasound imaging of wounds such as:
- Clean, closed surgical incisions
- Skin graft donor sites
- Superficial wounds such as abrasion, skin tears, and blisters
- Chafed skin or irritated skin
- Ulcers-cavernous and non-cavernous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ewind C. Eagaton
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number Over-The-Counter Use
4 Prescription Use (Per 21 CFR 801.109)
OR
(Optional Format 1-2-96)
4
Page 2 of 3
510(k) Number (if known): K082998
· Device Name: Hudson 2040 Ultrasound Scanner
Indications For Use:
High resolution ultrasound imaging of wounds such as:
- Clean, closed surgical incisions
- Skin graft donor sites
- Superficial wounds such as abrasion, skin tears, and blisters
- Chafed skin or irritated skin
- Ulcers—cavernous and non-cavernous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Eavid h. Layson
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 22920 K) Number ver-The-Counter Use
Prescription Use (Per 21 CFR 801.109)
(Optional Format 1-2-96)
5
: Page
510(k) Number (if known): K072928
Device Name: Hudson 2060 Ultrasound Scanner
Indications For Use:
High resolution ultrasound imaging of wounds such as:
- Clean, closed surgical incisions
- Skin graft donor sites
- Superficial wounds such as abrasion, skin tears, and blisters
- Chafed skin or irritated skin
- Ulcers-cavernous and non-cavernous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Eygenson
(Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number -
Prescription Use (Per 21 CFR 801.109)
OR Over-The-Counter Use
(Optional Format 1-2-96)
6
Hudson 2020 Ultrasound System
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) | |
Ophthalmic | |||||||||||
Fetal | |||||||||||
Abdominal | |||||||||||
Intraoperative (specify) | |||||||||||
Intraoperative Neurological | |||||||||||
Pediatric | |||||||||||
Small Organ (specify) | |||||||||||
Neonatal Cephalic | |||||||||||
Adult Cephalic | |||||||||||
Cardiac | |||||||||||
Transesophageal | |||||||||||
Transrectal | |||||||||||
Transvaginal | |||||||||||
Transurethral | |||||||||||
Intravascular | |||||||||||
Peripheral Vascular | |||||||||||
Laparoscopic | |||||||||||
Musculo-skeletal | |||||||||||
Conventional | |||||||||||
Musculo-skeletal Superficial | |||||||||||
Other (specify) | |||||||||||
Wounds | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:_
In conjunction with Wendelken / Pope scanning method and contact media.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Ewind h. Sgrom
(Division Sign-Off) 4
Division of Reproductive, Abdominal,
46 and Radiological Devices
510(k) Number K022928
7
Hudson 2040 Ultrasound System
Diagnostic Ultrasound Indications for Use For
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Clinical Application | A | B | M | PWD | CWD | Color | ||||
Doppler | Amplitude | |||||||||
Doppler | Color | |||||||||
Velocity | ||||||||||
Imaging | Combined | |||||||||
(specify) | Other | |||||||||
(specify) | ||||||||||
Ophthalmic | ||||||||||
Fetal | ||||||||||
Abdominal | ||||||||||
Intraoperative (specify) | ||||||||||
Intraoperative Neurological | ||||||||||
Pediatric | ||||||||||
Small Organ (specify) | ||||||||||
Neonatal Cephalic | ||||||||||
Adult Cephalic | ||||||||||
Cardiac | ||||||||||
Transesophageal | ||||||||||
Transrectal | ||||||||||
Transvaginal | ||||||||||
Transurethral | ||||||||||
Intravascular | ||||||||||
Peripheral Vascular | ||||||||||
Laparoscopic | ||||||||||
Musculo-skeletal | ||||||||||
Conventional | ||||||||||
Musculo-skeletal Superficial | ||||||||||
Wounds | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
In conjunction with Wendelken / Pope scanning method and contact media.
(PLEASE DO NOT WRITE BELOW THIS LINE . CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluat (ODE
Gavid G. Ligarson
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
4
8
Hudson 2060 Ultrasound System
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Application | Mode of Operation | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
A | B | M | PWD | CWD | Color | |||||
Doppler | Amplitude | |||||||||
Doppler | Color | |||||||||
Velocity | ||||||||||
Imaging | Combined | |||||||||
(specify) | Other | |||||||||
(specify) | ||||||||||
Ophthalmic | ||||||||||
Fetal | ||||||||||
Abdominal | ||||||||||
Intraoperative (specify) | ||||||||||
Intraoperative Neurological | ||||||||||
Pediatric | ||||||||||
Small Organ (specify) | ||||||||||
Neonatal Cephalic | ||||||||||
Adult Cephalic | ||||||||||
Cardiac | ||||||||||
Transesophageal | ||||||||||
Transrectal | ||||||||||
Transvaginal | ||||||||||
Transurethral | ||||||||||
Intravascular | ||||||||||
Peripheral Vascular | ||||||||||
Laparoscopic | ||||||||||
Musculo-skeletal | ||||||||||
Conventional | ||||||||||
Musculo-skeletal Superficial | ||||||||||
Other (specify) | ||||||||||
Wounds | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
In conjunction with Wendelken / Pope scanning method and contact media.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE
(
Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K022928
48
9
Hudson 2020 # 402198 Transducer 8.0.Mhz Linear Array 40 mm
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Clinical Application | A | B | M | PWD | CWD | Color | ||||
Doppler | Amplitude | |||||||||
Doppler | Color | |||||||||
Velocity | ||||||||||
Imaging | Combined | |||||||||
(specify) | Other | |||||||||
(specify) | ||||||||||
Ophthalmic | ||||||||||
Fetal | ||||||||||
Abdominal | ||||||||||
Intraoperative (specify) | ||||||||||
Intraoperative Neurological | ||||||||||
Pediatric | ||||||||||
Small Organ (specify) | ||||||||||
Neonatal Cephalic | ||||||||||
Adult Cephalic | ||||||||||
Cardiac | ||||||||||
Transesophageal | ||||||||||
Transrectal | ||||||||||
Transvaginal | ||||||||||
Transurethral | ||||||||||
Intravascular | ||||||||||
Peripheral Vascular | ||||||||||
Laparoscopic | ||||||||||
Musculo-skeletal | ||||||||||
Conventional | ||||||||||
Musculo-skeletal Superficial | ||||||||||
Other (specify) | ||||||||||
Wounds | N |
N= new indication; P= previously deared by FDA; E= added under Appendix E
Additional Comments:_
In conjunction with Wendelken / Pope scanning method and contact media.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CORH, Office of Device Evaluation (ODE)
David A. Bryson
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _
49
10
Hudson 2040 # 402198 Transducer 8.0 Mhz Linear Array 40 mm
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound Imaging or fluid flow analysis of the human body as follows:
Mode of Operation | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Clinical Application | A | B | M | PWD | CWD | Color | ||||
Doppler | Amplitude | |||||||||
Doppler | Color | |||||||||
Velocity | ||||||||||
Imaging | Combined | |||||||||
(specify) | Other | |||||||||
(specify) | ||||||||||
Ophthalmic | ||||||||||
Fetal | ||||||||||
Abdominal | ||||||||||
Intraoperative (specify) | ||||||||||
Intraoperative Neurological | ||||||||||
Pediatric | ||||||||||
Small Organ (specify) | ||||||||||
Neonatal Cephalic | ||||||||||
Adult Cephalic | ||||||||||
Cardiac | ||||||||||
Transesophageal | ||||||||||
Transrectal | ||||||||||
Transvaginal | ||||||||||
Transurethral | ||||||||||
Intravascular | ||||||||||
Peripheral Vascular | ||||||||||
Laparoscopic | ||||||||||
Musculo-skeletal | ||||||||||
Conventional | ||||||||||
Musculo-skeletal Superficial | ||||||||||
Other (specify) | ||||||||||
Wounds | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
In conjunction with Wendelken / Pope scanning method and contact media.
(PLEASE DO NOT WRITE BELOW THIS LINE . CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
C
David A. Layman
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K022978
11
Hudson 2020 - #410054 Transduce 8.0 Mhz Linear Array 60 mm 6.0
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
Ophthalmic | ||||||||||
Fetal | ||||||||||
Abdominal | ||||||||||
Intraoperative (specify) | ||||||||||
Intraoperative Neurological | ||||||||||
Pediatric | ||||||||||
Small Organ (specify) | ||||||||||
Neonatal Cephalic | ||||||||||
Adult Cephalic | ||||||||||
Cardiac | ||||||||||
Transesophageal | ||||||||||
Transrectal | ||||||||||
Transvaginal | ||||||||||
Transurethral | ||||||||||
Intravascular | ||||||||||
Peripheral Vascular | ||||||||||
Laparoscopic | ||||||||||
Musculo-skeletal | ||||||||||
Conventional | ||||||||||
Musculo-skeletal Superficial | ||||||||||
Other (specify) | ||||||||||
Wounds | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:_
In conjunction with Wendelken / Pope scanning method and contact media.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David C. Dejarm
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K022928
51
12
Hudson 2040 - # 410054 Transducer 8.0 Mhz Linear Array 60 mm 6.0
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Clinical Application | A | B | M | PWD | CWD | Color | ||||
Doppler | Amplitude | |||||||||
Doppler | Color | |||||||||
Velocity | ||||||||||
Imaging | Combined | |||||||||
(specify) | Other | |||||||||
(specify) | ||||||||||
Ophthalmic | ||||||||||
Fetal | ||||||||||
Abdominal | ||||||||||
Intraoperative (specify) | ||||||||||
Intraoperative Neurological | ||||||||||
Pediatric | ||||||||||
Small Organ (specify) | ||||||||||
Neonatal Cephalic | ||||||||||
Adult Cephalic | ||||||||||
Cardiac | ||||||||||
Transesophageal | ||||||||||
Transrectal | ||||||||||
Transvaginal | ||||||||||
Transurethral | ||||||||||
Intravascular | ||||||||||
Peripheral Vascular | ||||||||||
Laparoscopic | ||||||||||
Musculo-skeletal | ||||||||||
Conventional | ||||||||||
Musculo-skeletal Superficial | ||||||||||
Other (specify) | ||||||||||
Wounds | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:__________________________________________________________________________________________________________________________________________________________
In conjunction with Wendelken / Pope scanning method and contact media.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David C. Lyson
(Division Sign-Off) /
Division of Reproductive, Abdominal,
and Radiological Devices K022928
510(k) Number
13
Hudson 2060 - # 410503 Transducer 10.0 Mhz Linear Array 40 mm 5.0
Diagnostic Ultrasound Indications for Use Form
Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Clinical Application | A | B | M | PWD | CWD | Color | ||||
Doppler | Amplitude | |||||||||
Doppler | Color | |||||||||
Velocity | ||||||||||
Imaging | Combined | |||||||||
(specify) | Other | |||||||||
(specify) | ||||||||||
Ophthalmic | ||||||||||
Fetal | ||||||||||
Abdominal | ||||||||||
Intraoperative (specify) | ||||||||||
Intraoperative Neurological | ||||||||||
Pediatric | ||||||||||
Small Organ (specify) | ||||||||||
Neonatal Cephalic | ||||||||||
Adult Cephalic | ||||||||||
Cardiac | ||||||||||
Transesophageal | ||||||||||
Transrectal | ||||||||||
Transvaginal | ||||||||||
Transurethral | ||||||||||
Intravascular | ||||||||||
Peripheral Vascular | ||||||||||
Laparoscopic | ||||||||||
Musculo-skeletal | ||||||||||
Conventional | ||||||||||
Musculo-skeletal Superficial | ||||||||||
Wounds | ||||||||||
Other (specify) | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:_
In conjunction with Wendelken / Pope scanning method and contact media.
(PLEASE DO NOT WRITE BELOW THIS LINE . CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Chil A. Dyson
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number 4022928